K191826 · Beijing Keyi Medical Device Technology Co., Ltd. · LPH · May 14, 2020 · Orthopedic
Device Facts
Record ID
K191826
Device Name
KeYi Total Hip System
Applicant
Beijing Keyi Medical Device Technology Co., Ltd.
Product Code
LPH · Orthopedic
Decision Date
May 14, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The KeYi Total Hip System is designed for total hip arthroplasty. The indications for use are: a. Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis. b. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures. c. Proximal femoral fractures. d. Avascular necrosis of the femoral head. e. Non-union of proximal femoral neck fractures. f. Other indications such as congenital dysplasia, arthrodesis conversion, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteoporosis, pseudarthrosis conversion, and structural abnormalities. The Ke Yi femoral stems are indicated for cementless use. The KeYi acetabular shells are indicated for cemented or cementless use.
Device Story
KeYi Total Hip System is a semi-constrained, porous-coated hip prosthesis for primary or revision total hip arthroplasty. System components include forged titanium alloy femoral stems with proximal CP titanium plasma porous coating; cobalt chromium alloy femoral heads; titanium alloy acetabular shells with CP titanium plasma porous coating; and ultrahigh molecular weight polyethylene (UHMWPE) acetabular liners. Optional components include titanium alloy acetabular dome hole plugs and cancellous bone screws. The device is used by orthopedic surgeons in a clinical/surgical setting. Femoral heads attach to stems via 12/14 taper-locking. Acetabular shells support cemented or cementless fixation; femoral stems are for cementless use. The system replaces damaged hip joint structures to restore joint function and mobility, potentially reducing pain and improving patient quality of life.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing including stem and neck fatigue testing (ISO 7206-4/6), disassembly testing (ASTM F1820, ASTM F2009), wear simulation (ISO 14242-1), range of motion simulation, porous coating characterization, material property characterization, biocompatibility (ISO 10993), and pyrogen testing.
Indicated for total hip arthroplasty in patients with significantly impaired joints due to rheumatoid, osteoarthritis, or post-traumatic arthritis; revision of failed hip procedures; proximal femoral fractures; avascular necrosis; non-union of femoral neck fractures; or conditions including congenital dysplasia, arthrodesis conversion, coxa plana/vara/valga, metabolic/tumorous conditions, osteoporosis, pseudarthrosis, and structural abnormalities.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Depuy Duraloc Acetabular Cup System-Enduron Liner (K983491)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.
Beijing Ke Yi Medical Device Technology Co., Ltd. Jenny Jiang Regulatory Affairs Supervisor Building 1, 30 Yongchang South Road Beijing, 100176 China
May 14, 2020
### Re: K191826
Trade/Device Name: KeYi Total Hip System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH Dated: April 8, 2020 Received: April 14, 2020
Dear Jenny Jiang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vesa Vuniqi, MS Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K191826
Device Name KeYi Total Hip System
Indications for Use (Describe)
The KeYi Total Hip System is designed for total hip arthroplasty. The indications for use are:
a. Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
- b. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
- c. Proximal femoral fractures.
- d. Avascular necrosis of the femoral head.
- e. Non-union of proximal femoral neck fractures.
f. Other indications such as congenital dysplasia, arthrodesis conversion, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteoporosis, pseudarthrosis conversion, and structural abnormalities.
The Ke Yi femoral stems are indicated for cementless use. The KeYi acetabular shells are indicated for cemented or cementless use.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
{3}------------------------------------------------
### 510(k) SUMMARY KeYi Total Hip System
510(k) Owner Information Name: Beijing KeYi Medical Device Technology Co., Ltd. Address: Building 1, 30 Yongchang South Road, Beijing Economic Technological Development Area, 100176 Beijing, China Telephone Number: +86-10-67853877 Fax Number: +86-10-67853877 ext. 8117 Email: jiangli@keyibangen.com Contact Person: Jenny Jiang Regulatory Affairs Supervisor Date Prepared: July 1st , 2019 Name of Device KeYi Total Hip System Trade Name / Proprietary Name: Hip Joint Prosthesis Common Name: Product Code: LPH Regulatory Classification: Class II - 21 CFR § 888.3358 - Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Orthopedic Device Panel Review Panel: K111472 – HELICON Hip System Predicate Devices: Additional Predicate Device: K980513 - Depuy Articul/Eze femoral Heads K961186 - Depuy Duraloc Cementless Acetabular cup svstem K001991 - DePuy Titan Porocoat Hip Stem K983491 - Depuy Duraloc Acetabular Cup System-Enduron Liner
Reason for 510(k) Submission: New Devices
# Device Description
The KeYi Total Hip System is a porous-coated, semi-constrained, hip prosthesis designed for either primary or revision hip arthroplasty. The system is comprised of femoral stems and mating metal femoral heads; acetabular shells and mating acetabular liners; optional acetabular cancellous bone screws; and optional acetabular dome hole plug.
The KeYi Femoral Stems are forged titanium alloy, feature a proximal CP Titanium plasma porous coating, and are offered in a range of sizes. The KeYi Femoral Heads are polished
Premarket Notification Confidential Information
{4}------------------------------------------------
cobalt chromium alloy and designed to mate with 12/14 taper of the femoral stem through taperlocking arrangement and to articulate with the acetabular liner. The heads are available in different sizes with various offsets.
The KeYi Acetabular Shells are manufactured from titanium alloy and feature a porous CP Titanium plasma porous coating. The shells feature a dome hole, are available with a two or three-hole pattern for supplemental bone screw fixation, and come in a range of outer diameter sizes. The KeYi Acetabular Liners are manufactured from ultrahigh molecular weight polyethylene (standard/non-crosslinked UHMWPE).
Optional components include threaded acetabular dome hole plugs and acetabular cancellous bone screws, all manufactured from titanium alloy.
### Intended Use / Indications for Use
The KeYi Total Hip System is designed for total hip arthroplasty. The indications for use are:
- a. Significantly impaired ioints resulting from rheumatoid, osteo, and post-traumatic arthritis,
- Revision of failed femoral head replacement, cup arthroplasty or other hip procedures. ڪ
- Proximal femoral fractures. C.
- d. Avascular necrosis of the femoral head.
- e. Non-union of proximal femoral neck fractures.
- Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, f. coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities.
The KeYi femoral stems are indicated for cementless use. The KeYi acetabular shells are indicated for cemented or cementless use.
### Summary of the Technological Characteristics of the Device Compared to the Selected Predicate Devices
The technological characteristics of the KeYi Total Hip System are similar to predicate devices.
### PERFORMANCE DATA – Summary of Non-Clinical Test Conducted for Determination of Substantial Equivalence
The following mechanical tests of the KeYi Total Hip System were performed to support the determination of substantial equivalence:
- Stem fatigue test for KeYi femoral stems per ISO 7206-4 .
- Neck fatique test for KeYi femoral stems per ISO 7206-6 .
- Disassembly test of KeYi acetabular shells and UHMWPE liners per ASTM F1820 ●
- Taper disassembly test of KeYi femoral head and taper per ASTM F2009 ●
- . Wear Simulation Test of KeYi acetabular liners per ISO 14242-1
- Range of motion simulation for KeYi total hip system 0
- Porous coating characterization tests
- . Material properties characterization of non-crosslinked UHMWPE
- Biocompatibility test for non-crosslinked UHMWPE per ISO 10993
- . Pyrogen test
#### Conclusion
KeYi Total Hip System is substantially equivalent to the predicate devices.
Premarket Notification Confidential Information
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.