The Electric Breast Pumps are intended to express and collect milk from the breast of a lactating woman.
Device Story
Electric Breast Pump (BLA8015-02) is a single-user, software-controlled, battery-powered device for expressing and collecting breast milk. Used in hospital or home settings by lactating women. Device features four modes (massage, continuous express, frequency conversion, normal express) with adjustable suction levels (6 or 9 levels depending on mode) ranging from 20-240 mmHg. Operation involves a vacuum pump exhausting air from a cavity; once target suction is reached, the pump stops, allowing milk to flow into a collection bottle via pressure differential. User interface includes an LCD display and mode/level keys. Device includes backflow protection to prevent milk from entering the motor. Healthcare providers or patients use the device to facilitate milk expression; output is collected milk for infant feeding.
Clinical Evidence
Bench testing only. Testing included vacuum pressure verification, backflow prevention, electrical safety (IEC 60601-1, IEC 60601-1-11), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-5, ISO 10993-10). Software verification and validation were performed for a minor level of concern.
Technological Characteristics
Powered by 3.7V 2200mAh Li-ion battery. Materials: polypropylene and silicone gel (food contact compliant). Reciprocating diaphragm pump mechanism. Microprocessor-controlled suction and cycle speed (6.5-92.3 cycles/min). Suction range 20-240 mmHg. Non-sterile, reusable. Complies with IEC 60601-1, IEC 60601-1-11, and IEC 60601-1-2.
Indications for Use
Indicated for lactating women to express and collect breast milk. No specific contraindications listed.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 16, 2020
Ningbo Huiyoo Baby Products Co. Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O Box 120-119 Shanghai, 200120 CHINA
Re: K191802
> Trade/Device Name: Electric Breast Pump Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: December 19, 2019 Received: December 23, 2019
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Sharon M. Andrews Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K191802
Device Name Electric Breast Pump
Indications for Use (Describe)
The Electric Breast Pumps are intended to express and collect milk from the breast of a lactating woman.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K191802
- 1. Sponsor Identification
Ningbo Huiyoo Baby Products Co. Ltd. 523# Cizhang Road, Zhangqi Industrial Area, Cixi, 315313, Zhejiang, China
Contact Person: Xiaoping Liu Position: Electrical Engineer Tel: +86-0574-63679665 Email: liuxiaoping@foxmail.com
- 2. Date of Preparation: January 15, 2020
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)
### Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
- 4. Identification of Proposed Device
Trade Name: Electric Breast Pump; Common Name: Breast Pump Models: BLA8015-02
Regulatory Information Classification Name: Pump, Breast, Powered; Classification: II Product Code: HGX; Regulation Number: 21 CFR 884.5160; Review Panel: Obstetrics/Gynecology;
- 5. Identification of Predicate Device
510(k) Number: K151284 Product Name: Electric Double Breast Pumps Manufacturer: Wuxi Xinzhongrui Baby Supplies Co., Ltd.
The predicate device has not been subject to a design-related recall.
#### б. Device Description
The Electric Breast Pump BLA8015-02 is designed and manufactured to express and collect milk from a lactating woman's breast. The subject device is electrically powered, software-controlled, digital single-user pumps.
The pump is available in 4 different modes, which are massage mode, continuous express mode, frequency conversion mode and normal express mode. The different modes can be distinguished by different keys displayed on the LCD. There are 6 expression levels for massage mode and continuous expression mode, and 9 expression levels for frequency conversion mode and normal expression mode. The suction strengths are different for different working modes. The suction strength can be adjusted within 20~240mmHg by the user. When powered, the subject device starts with massage mode at level 1. The mode switch and level adjustment can be achieved by mode keys.
A suction cycle consists of an inhale process and an exhaust process. When the breast pump is powered on, the vacuum pump exhausts the air inside the cavity. When the suction strength reaches the setting value, the vacuum pump stops working and the expressed milk opens the valve and then flows into the bottle due to pressure difference.
{5}------------------------------------------------
The pump is provided in non-sterile and can be re-used by a single user. The device has embedded software and the level of concern for the software is minor. The device is powered by Li-ion battery.
The patient contact component and milk contact component are massage cushion and milk bottle. The contact duration and contact level is limited skin contact. The massage cushion and milk bottle are respectively made of polypropylene and silicone gel. The materials meet the requirements of FDA regulations concerning food contact.
The device can be used in both hospital and home environment.
#### 7. Indication for Use
The Electric Breast Pumps are intended to express and collect milk from the breast of a lactating woman.
| Subject Device | Predicate Device |
|----------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The Electric Breast Pumps are intended to<br>express and collect milk from the breast of a<br>lactating woman. | The powered Electric Double Breast Pumps are<br>intended to express and collect milk from the<br>breast of a lactating woman. They are double<br>pumps with a single pumping option and intended<br>for single user. |
The indications for use of the subject device are similar to that of the predicate. The subject device is indicated for single pumping and for a single user, while the predicate is indicated for single and double pumping in a single user. The differences in the indications statements do not alter the intended use of the subject device as compared to the predicate.
- 8. Comparison of Technological Characteristics
| ITEM | Proposed Device | Predicate Device, K151284 |
|--------------------|----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | HGX | HGX |
| Regulation No. | 21 CFR 884.5160 | 21 CFR 884.5160 |
| Class | II | II |
| Indication for Use | The Electric Breast Pumps are intended to express and collect milk from the breast of a lactating woman. | The powered Electric Double Breast Pumps are intended to express and collect milk from the breast of a lactating woman. They are double pumps with a single |
Table 1 Comparison of Technological Characteristics
{6}------------------------------------------------
| | | pumping option and intended for |
|-----------------------------------------|-------------------------------------------------|---------------------------------------------------------------|
| | | single user. |
| Patient | Breastfeeding women | Breastfeeding women |
| Population | | |
| Anatomical Sites | Breast | Breast |
| Pump Type | Reciprocating Diaphragm | Reciprocating Diaphragm |
| Pumping Options | Single | Single or Double |
| Cycling/Suction<br>Control<br>Mechanism | Microprocessor | Microprocessor |
| Suction Levels | 6 and 9 | 7 and 9 |
| Suction Strength | 20~240 mmHg | 60 -240 mmHg |
| Cycle Speed | 6.5~92.3 cycle/min | Single: 36 – 103 cycles / min<br>Double: 25 – 76 cycles / min |
| Power Supply | 3.7V 2200mAh Li-ion | 6V DC Adaptor or batteries |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 |
| Back Flow<br>Protection | Yes | Yes |
| Patient Contact<br>Material | Silicone gel | Silicone gel |
| Biocompatibility | No cytotoxicity, irritation or<br>sensitization | No cytotoxicity, irritation or<br>sensitization |
The differences between proposed device and predicate include suction levels, suction strength, cycle speed and power supply. These differences do not raise any question its safety and effectiveness. The differences in technological characteristics may be evaluated through performance testing.
- 9. Summary of Non-clinical Performance Testing
Non clinical tests were conducted to verify that the proposed device met all design specifications and that the subject device is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
IEC 60601-1:2005+CORR.1:2006+CORR.2:2007+AM1:2012, Medical electrical equipment – Part 1: General requirements for basic safety, and essential performance.
IEC 60601-1-11:2010, Medical electrical equipment – Part 1-11: General requirements for basic safety, and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
{7}------------------------------------------------
IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic Compatibility
ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
The software was verified and validated. Software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, published on May 11, 2005 for a minor level of concern.
Additionally, the following non-clinical tests were conducted:
Vacuum pressure of proposed devices was tested. All the tests results complied with the design specifications of the proposed devices.
Backflow testing was conducted to ensure that even if the bottle is over-filled, no liquid will backflow into the tubing, and therefore no liquid can backflow into the pump motor. The test results showed that there was no backflow during the test.
### 10. Conclusion
The performance testing demonstrate that the subject device is as safe and effective as the predicate. Therefore, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.