← Product Code [MAX](/productcode/MAX) · K191796

# Neo Cage System (K191796)

_Neo Medical SA · MAX · Dec 13, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K191796

## Device Facts

- **Applicant:** Neo Medical SA
- **Product Code:** [MAX](/productcode/MAX.md)
- **Decision Date:** Dec 13, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

NEO Cage System™ is an intervertebral body fusion device intended to be used with autogenous bone graft to facilitate interbody fusion and to be used with supplemental spinal fixation systems that have been cleared for the use in the lumbosacral spine. The cage is to be implanted in open surgery via a posterior or transforaminal approach. The indication for use is Degenerative Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have grade 1 Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and six months of non-operative treatment.

## Device Story

NEO Cage System is an intervertebral body fusion device used in open posterior or transforaminal spinal surgery. It consists of titanium alloy cages of varying sizes and lordotic angles to accommodate patient anatomy. The device is implanted by orthopedic surgeons to stabilize the spine and promote fusion when used with autogenous bone graft and supplemental spinal fixation. It functions as a mechanical spacer to maintain disc height and facilitate bone growth. The system includes both single-use and reusable instruments for implantation. The device is MRI conditional. It benefits patients by providing structural support during the healing process following surgical correction of spinal disorders.

## Clinical Evidence

No clinical data. Evidence consists of non-clinical bench testing, including mechanical testing (static/dynamic compression, shear, torsion, subsidence, expulsion, wear), biocompatibility (ISO 10993), sterilization validation (ISO 11137-2), and MRI compatibility testing (ASTM F2052, F2182, F2119). Usability was confirmed via a WetLab study with orthopedic surgeons.

## Technological Characteristics

Materials: Titanium alloy (Ti6Al4V) per ISO 5832-3 or ASTM F3001. Form factor: Intervertebral cages of varying lengths, sizes, and lordotic angles. Energy source: None (mechanical). Connectivity: None. Sterilization: Validated 25kGy (VDmax25 method) for SAL 10-6. Software: None.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- NEO Cage System ([K181048](/device/K181048.md))
- Stryker Tritanium Cage ([K152304](/device/K152304.md))
- Spineart Juliet Ti ([K153621](/device/K153621.md))
- Medtronic Capstone ([K073291](/device/K073291.md), [K082342](/device/K082342.md))
- SpineVision SpaceVision System ([K153783](/device/K153783.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 13, 2019

Neo Medical S.A. % Mr. Charles Cathlin President Confinis Corporation 15807 Glacier Ct North Potomac, Maryland 20878

Re: K191796

Trade/Device Name: NEO Cage System™ Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: November 4, 2019 Received: November 12, 2019

Dear Mr. Cathlin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent L. Showalter, PhD Assistant Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K191796

Device Name NEO Cage System™

#### Indications for Use (Describe)

NEO Cage System™ is an intervertebral body fusion device intended to be used with autogenous bone graft to facilitate interbody fusion and to be used with supplemental spinal fixation systems that have been cleared for the use in the lumbosacral spine. The cage is to be implanted in open surgery via a posterior or transforaminal approach.

The indication for use is Degenerative Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have grade 1 Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and six months of non-operative treatment.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Date prepared | October 28, 2019, according to 21 CFR 807.92 |
|---------------|----------------------------------------------|
|---------------|----------------------------------------------|

#### Purpose of submission

| 510(k) type           | Traditional 510(k)                                               |
|-----------------------|------------------------------------------------------------------|
| Purpose of submission | Line extension to currently cleared NEO Cage<br>SystemTM K181048 |

## Submitter information

| Submitter                                | Neo Medical S.A.<br>Route de Lausanne 157 A<br>1096 Villette (Lavaux), Switzerland    |
|------------------------------------------|---------------------------------------------------------------------------------------|
| Contact US agent<br>confinis corporation | Mr. Charles Cathlin<br>Mailto: charles.cathlin@confinis.com<br>Phone: +1 240 252 0891 |

#### Device name and classification

| Trade name          | NEO Cage System™                  |
|---------------------|-----------------------------------|
| Common name         | Intervertebral body fusion device |
| Device panel        | Orthopedic                        |
| Classification name | Intervertebral body fusion device |
| Class               | II                                |
| Product code        | MAX                               |
| CFR section         | 888.3080                          |

## Predicate devices

The NEO Cage System™ is substantially equivalent to the primary predicate device NEO Cage System™ (K181048) as well as the following additional predicate devices: Stryker Tritanium Cage (K152304), Spineart Juliet Ti (K153621), Medtronic Capstone (K073291 (initial), K082342 (PEEK introduced)) and SpineVision SpaceVision System (K153783).

{4}------------------------------------------------

#### Indications for use

Neo Cage System™ is an intervertebral body fusion device intended to be used with autogenous bone graft to facilitate interbody fusion and to be used with supplemental spinal fixation systems that have been cleared for the lumbosacral spine. The cage is to be implanted in open surgery via a posterior or transforaminal approach.

The indication for use is Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have grade 1 Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

#### Device description

The NEO Cage System™ is an intervertebral fusion device with bone graft intended for stabilization and to promote bone fusion during the normal healing process following surgical correction of disorders of the spine. The cages are intended to be used with supplemental spinal fixation systems that have been cleared for the use in the lumbosacral spine. The cage implants may be implanted via an open posterior approach.

The Neo Cage System™ consists of cages of different lengths, sizes and lordotic angles to adapt to a variety of patient anatomies. The size and form of the devices is adjusted to the morphology of the body and the surgical technique. The cages of the Neo Cage system are made of titanium alloy (Ti6Al4V) according to ISO 5832-3 or ASTM F3001 and are supplied sterile. The system includes the relevant instruments which are either single use only and delivered sterile or reusable. The Instruments are to be used for the implantation of the above mentioned medical devices.

## Summary of testing

Sterilization validation: The minimal dose of 25kGy is validated using VDmax25 method as described in ISO 11137-2 and confirmed a Sterility Assurance Level SAL of 10-6.

Pyrogenicity: The pyrogenicity is tested with the LAL-Test (Limulustest).

Packaging validation: The sterile barrier system complies with ISO 11607-1. Real time and accelerated ageing studies as well as transport studies were performed to demonstrate 5-year shelf life and packaging integrity.

Biocompatibility: Biological evaluation has been performed in accordance with ISO 10993-1. Chemical Analysis according to ISO 10993-18 and Cytotoxicity testing according to ISO 10993-5 were performed on worst case components of the NEO Cage System™ for K181048. Cytotoxicity testing and testing of inorganic and organic residues was repeated to confirm biocompatibility of the line extension.

{5}------------------------------------------------

Mechanical testing: A new worst case cage is introduced with the line extension. Mechanical testing has therefore been repeated and completed successfully. The following tests were performed in accordance with ASTM F2077

- Static compression test ●
- Static shear compression test ●
- Static torsion test ●
- Dynamic compression test ●
- . Dynamic shear compression test
- . Dynamic torsion test

Additional testing included

- Subsidence test according to ASTM F2267
- Wear testing (debris) according to ISO 17853 ●
- Expulsion test ●

Trabecular structure: The trabecular structure has been characterized by tomography examination for the currently cleared cage NEO Cage System™ K181048 and is not affected by the line extension.

MRI Compatibility: The following tests were performed with the new worst-case regarding MRI related issues

- . Magnetic field interactions ASTM F2052-15
- MRI-related heating ASTM F2182-11a and .
- Artifacts ASTM F2119-07 (R2013) .

The NEO Cage System™ is MRI conditional and information is provided in the labeling.

Usability: Suitability of the instruments and use of the NEO Caqe System™ including its labeling was confirmed in a WetLab study with orthopedic surgeons.

Clinical Evaluation: Clinical studies were not required, based upon review an evaluation of scientific literature, comparison with similar devices and non-clinical performance data.

## Conclusion

The NEO Cage System™ is substantially equivalent to its predicate devices in terms of intended use, material, design, mechanical properties and function.

Non-clinical performance testing demonstrates that the NEO Cage System™ meets the requirements according to FDA Guidance for Intervertebral Body Fusion Device, issued on June 12, 2007, and is as safe and effective as its predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K191796](https://fda-staging.innolitics.com/device/K191796)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com