AOCup Lens Case with AODisc

K191795 · Alcon Laboratories, Inc. · LPN · Jul 31, 2019 · Ophthalmic

Device Facts

Record IDK191795
Device NameAOCup Lens Case with AODisc
ApplicantAlcon Laboratories, Inc.
Product CodeLPN · Ophthalmic
Decision DateJul 31, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 886.5928
Device ClassClass 2

Indications for Use

For storage of soft (hydrophilic), and rigid gas permeable (RGP) hard contact lenses during disinfection with buffered 3% hydrogen peroxide contact lens solution.

Device Story

AOCup Lens Case with AODisc is a specialized contact lens storage container used for disinfection. The device consists of a transparent cup, a screw cap assembly, lens baskets, and a platinum-coated neutralization disc. The user places contact lenses into the baskets and fills the cup with 3% hydrogen peroxide solution. The platinum catalyst on the disc neutralizes the hydrogen peroxide into water and oxygen gas during the disinfection process. The device is intended for patient self-use at home. The current modification involves a change in the silicone material used for the O-ring gasket within the screw cap assembly to ensure proper sealing.

Clinical Evidence

Bench testing only. The device underwent biocompatibility, physicochemical, and design functionality testing to evaluate the modified silicone gasket material. Results demonstrated that the modified device performs equivalently to the predicate.

Technological Characteristics

The device is a lens case featuring a screw cap with a silicone O-ring gasket, lens baskets, and a platinum-coated neutralization disc. The platinum catalyst facilitates the chemical neutralization of hydrogen peroxide. The device is a passive mechanical system; no electronics or software are involved.

Indications for Use

Indicated for storage of soft (hydrophilic) and rigid gas permeable (RGP) hard contact lenses during disinfection with buffered 3% hydrogen peroxide solution.

Regulatory Classification

Identification

A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.

Special Controls

*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 31, 2019 Alcon Laboratories, Inc. Dr. Ralf Finke Senior Regulatory Specialist Alcon Vision Care Industriering 1 Grosswallstadt, DE 63868 Bavaria Re: K191795 Trade/Device Name: AOCup Lens Case with AODisc Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (Hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LRX, LPN Dated: July 2, 2019 Received: July 3, 2019 Dear Dr. Finke: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketingsafety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) K191795 Device Name AOCup Lens Case with AODisc Indications for Use (Describe) For storage of soft (hydrophilic), and rigid gas permeable (RGP) hard contact lenses during disinfection with buffered 3% hydrogen peroxide contact lens solution. | Type of Use (Select one or both, as applicable) | | | |-------------------------------------------------|---------------------------------------------|--| | ☐ Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED | | | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary This 510(k) summary document has been prepared in accordance with section 21 CFR 807.92. # I. Submitter of the 510(k) | Company: | Alcon Laboratories, Inc.<br>6201 South Freeway<br>Fort Worth, TX 76134-2099, USA | |-------------------------|----------------------------------------------------------------------------------| | Primary Contact Person: | Ralf Finke | | Phone: | (Germany) +49 (6022) 240.515 | | Fax: | (Germany) +49 (6022) 240.512 | | Email: | ralf.finke@alcon.com | | Back up Contact Person: | Katryna Warren | | Phone: | (USA, Fort Worth, TX) 817-615-5076 | | Email: | katryna.warren@alcon.com | | Date Prepared: | June 24, 2019 | # II. Devices Subject to this 510(k) | Trade Names: | AOCup Lens Case with AODisc<br>(part of the Clear Care Cleaning & Disinfecting Solution system<br>and the Clear Care Plus Cleaning & Disinfecting Solution system) | |------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Contact lens case | | Classification Name: | Ophthalmic | | Device Classification: | Class II (21 CFR 886.5928) Soft (hydrophilic) contact lens care<br>products; (21 CFR 886.5918) Rigid gas permeable contact lens<br>care products | | Product Code: | LRX (lens case)<br>LPN (Solution) | {4}------------------------------------------------ # III. Predicate Device The legally marketed device(s) to which we are claiming substantial equivalence are: | 510(k) Number | Device | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------| | Previously approved in PMA P820040, Supplement S032, 23-Oct-1995<br>(Reclassified to Class II in 1997)<br>Special 510k K162597, 19-Oct-2016<br>Special 510k K173538, 04-Oct-2017 | AOCup Lens Case with AODisc | | K003345, 26-Mar-2001;<br>K013512, 20-Dec-2001;<br>K022687, 19-Nov-2002;<br>K023455, 28-Feb-2003;<br>K030522, 12-Sep-2003;<br>K031521, 27-Jun-2003 | Clear Care Cleaning & Disinfecting Solution (contains the AOCup Lens Case with AODisc as part of the system) | | K142284, 26-Jan-2015 | Clear Care Plus Cleaning & Disinfecting Solution (contains the AOCup Lens Case with AODisc as part of the system) | {5}------------------------------------------------ # IV. Device Description The AOCup Lens Case with AODisc is equivalent to the predicate device except change to the silicone material of a gasket. The gasket has O-ring form and is included in the screw cap assembly of the lens case. The device is a specialized lens case consisting of a transparent cup with a connected unit of screw cap, lens holders (baskets) and platinum-coated neutralization disc. The platinum catalyst coating and shape of the plastic disc are designed to effectively neutralize hydrogen peroxide in the lens care system to harmless water and oxygen gas. The modifications of gasket material does not change any indications for use nor the basic technical principle of the device functions. #### Indications for Use V. For storage of soft (hydrophilic), and rigid gas permeable (RGP) hard contact lenses during disinfection with buffered 3% hydrogen peroxide contact lens solution. ### VI. Comparison to Technological Characteristics with the Predicate Device The silicone material of an o-ring gasket in the screw cap of predicate device AOCup Lens Case cleared in 510k K162597 and K173538 has been changed to a different silicone material. Dimensions and technical function of the gasket are unchanged. {6}------------------------------------------------ # VII. Performance Data The AOCup Lens Case with AODisc and the changed gasket was evaluated in biocompatibility, physicochemical and design functionality tests. ## VIII. Conclusions Successful results of all nonclinical testing supported the substantial equivalence and therefore safety and efficacy of the modified AOCup Lens Case with AODisc to the existing product for its intended use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...