K191785 · Compumedics Limited · OLY · Feb 14, 2020 · Neurology
Device Facts
Record ID
K191785
Device Name
Orion LifeSpan MEG
Applicant
Compumedics Limited
Product Code
OLY · Neurology
Decision Date
Feb 14, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Orion LifeSpan MEG non-invasively measures the magnetoencephalographic (MEG) (and, optionally, electroencephalographic (EEG) signals) produced by electrically active tissue of the brain. These signals are recorded by a computerized data acquisition system, displayed, and may then be interpreted by trained physicians to help localize these active areas. The locations may then be correlated with anatomical information of the brain. MEG is routinely used to identify the locations of visual, auditory, and somatosensory in the brain when used in conjunction with evoked response averaging devices. MEG is also used to non-invasively locate regions of epileptic activity within the brain. The localization information provided by the device may be used, in conjunction with other diagnostic data, as an aid in neurosurgical planning. It is assumed that the users of the Orion LifeSpan™ MEG are physicians or neurology laboratory technicians who have received training in the following areas: Hospital procedures; Physiological monitoring of patients; Training relevant to the specific discipline or disorder under investigation. Note: This indication for use specifically excludes use of the Orion LifeSpan™ MEG as life support equipment, for example vital signs monitoring in intensive care units.
Device Story
Orion LifeSpan™ MEG records magnetic signals from human brain using superconducting magnetometry; optionally records EEG via integrated SynAmps2 hardware. System uses CURRY software platform for acquisition, processing, and display. Operated by trained physicians/technicians in clinical settings. Input signals transformed into localized active brain areas; output displayed for physician interpretation. Localization data correlated with anatomical brain information to aid neurosurgical planning. Benefits include non-invasive identification of epileptic regions and sensory cortex locations, supporting surgical decision-making.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including empty-room noise performance, phantom dipole localization accuracy, device compatibility, and software verification/validation. Phantom testing showed average localization error of 2.3mm (Orion) vs 1.9mm (Elekta).
Technological Characteristics
Superconducting magnetometry; DROS SQUID architecture; flux-locked loop readout. 186 (adult) or 138 (pediatric) axial gradiometer pickup coils. Fiberglass composite dewar/helmet. Mains power. Connectivity via CURRY software. Standards: IEC 60601-1, IEC 60601-1-2, IEC 80601-2-26, IEC 62304, ISO 14971.
Indications for Use
Indicated for non-invasive measurement of MEG and optional EEG signals from electrically active brain tissue to localize active areas for neurosurgical planning. Used to identify visual, auditory, and somatosensory locations and epileptic activity. For use by trained physicians or neurology technicians. Excludes life support/vital signs monitoring.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
February 14, 2020
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Compumedics Limited William Alam Quality Engineering Manager 30-40 Flockhart Street Abbotsford, 3067 Au
### Re: K191785
Trade/Device Name: Orion LifeSpan™ MEG Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLY, GWQ, OLX Dated: January 10, 2020 Received: January 13, 2020
#### Dear William Alam:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jay Gupta, M.S.E. Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K191785
Device Name Orion LifeSpan MEG
#### Indications for Use (Describe)
The Orion LifeSpan MEG non-invasively measures the magnetoencephalographic (MEG) (and, optionally, electroencephalographic (EEG) signals) produced by electrically active tissue of the brain. These signals are recorded by a computerized data acquisition system, displayed, and may then be interpreted by trained physicians to help localize these active areas. The locations may then be correlated with anatomical information of the brain. MEG is routinely used to identify the locations of visual, auditory, and somatosensory in the brain when used in conjunction with evoked response averaging devices. MEG is also used to non-invasively locate regions of epileptic activity within the brain. The localization information provided by the device may be used, in conjunction with other diagnostic data, as an aid in neurosurgical planning.
It is assumed that the users of the Orion LifeSpan™ MEG are physicians or neurology laboratory technicians who have received training in the following areas:
#### • Hospital procedures
- · Physiological monitoring of patients
- · Training relevant to the specific discipline or disorder under investigation
Note: This indication for use specifically excludes use of the Orion LifeSpan™ MEG as life support equipment, for example vital signs monitoring in intensive care units.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <b> \[X] Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | <div> <span> \[] Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
The following 510(k) summary is being submitted in accordance with 21CFR807.92.
| Device Information:<br>Submitter: | Compumedics Limited |
|-----------------------------------|---------------------------------------------------------------------------------|
| Address: | 30-40 Flockhart Street, Abbotsford 3067,<br>Victoria, Australia |
| Phone number: | +61 (0) 3 8420 7300 |
| Fax number: | +61 (0) 3 8420 7399 |
| Contact person: | William Alam<br>30-40 Flockhart Street, Abbotsford 3067,<br>Victoria, Australia |
| Phone number: | +61 (0) 3 8420 7300 |
| Fax number: | +61 (0) 3 8420 7399 |
| Date prepared: | June 28, 2019 |
| Trade name: | Orion LifeSpan™ MEG |
| Common name: | Magnetoencephalograph (MEG) |
| Primary product code: | OLY |
| Secondary product code: | GWQ, OLX |
| Device class: | II |
| Regulation numbers: | 21 CFR 882.1400 |
# Predicate Device Information:
| Trade Name: | Elekta Neuromag |
|---------------------|--------------------|
| Manufacturer: | Elekta Neuromag Oy |
| 510(k) number: | K041264 |
| Product code: | OLY, GWQ, OLX |
| Device class: | II |
| Regulation numbers: | 21 CFR 882.1400 |
{4}------------------------------------------------
### Device Description
The Orion LifeSpan™ MEG is a magnetoencephalograph (MEG) which records magnetic signals from the human brain. The Orion LifeSpan™ MEG uses the CURRY software platform (K001781) to acquire, process and display these signals. Optionally EEG can be recorded simultaneously with the MEG using an integrated SynAmps2 (K023771) hardware system.
The Orion LifeSpan™ MEG can optionally be provided with adult, child or both sized helmets.
The Orion LifeSpan™ MEG is used by skilled operators or physicians trained in the acquisition and interpretation of such signals. It is used as an adjunct as part of a range of measurements performed, such as imaging and other studies, to form a more complete picture of the patient's pathology alongside a standard clinical workup.
### Indications for Use
The Orion LifeSpan™ MEG non-invasively measures the magnetoencephalographic (MEG) signals (and, optionally, electroencephalographic (EEG) signals) produced by electrically active tissue of the brain. These signals are recorded by a computerized data acquisition system, displayed, and may then be interpreted by trained physicians to help localize these active areas. The locations may then be correlated with anatomical information of the brain. MEG is routinely used to identify the locations of visual, auditory, and somatosensory in the brain when used in conjunction with evoked response stimulators. MEG is also used to non-invasively locate regions of epileptic activity within the brain. The localization information provided by MEG may be used, in conjunction with other diagnostic data, as an aid in neurosurgical planning.
It is assumed that the users of the Orion LifeSpan™ MEG are physicians or neurology laboratory technicians who have received training in the following areas:
*Hospital procedures
*Physiological monitoring of patients
*Training relevant to the specific discipline or disorder under investigation
Note: This indication for use specifically excludes use of the Orion LifeSpan™ MEG as life support equipment, for example vital signs monitoring in intensive care units.
{5}------------------------------------------------
# Substantial Equivalence Discussion
The following technological differences exist between the Orion LifeSpan™ MEG and Elekta Neuromag.
| Orion<br>LifeSpan™ MEG (K191785) | Elekta Neuromag<br>(K041264) | Differences<br>(if any) | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | | | |
| The Orion LifeSpan™ MEG non-invasively<br>measures the magnetoencephalographic<br>(MEG) signals (and, optionally,<br>electroencephalographic (EEG) signals)<br>produced by electrically active tissue of the<br>brain. These signals are recorded by a<br>computerized data acquisition system,<br>displayed, and may then be interpreted by<br>trained physicians to help localize these<br>active areas. The locations may then be<br>correlated with anatomical information of the<br>brain. MEG is routinely used to identify the<br>locations of visual, auditory, and<br>somatosensory in the brain when used in<br>conjunction with evoked response<br>stimulators. MEG is also used to non-<br>invasively locate regions of epileptic activity<br>within the brain. The localization information<br>provided by MEG may be used, in<br>conjunction with other diagnostic data, as an<br>aid in neurosurgical planning.<br>It is assumed that the users of the Orion<br>LifeSpan™ MEG are physicians or<br>neurology laboratory technicians who have<br>received training in the following areas:<br>*Hospital procedures<br>*Physiological monitoring of patients<br>*Training relevant to the specific discipline<br>or disorder under investigation<br>Note: This indication for use specifically<br>excludes use of the Orion LifeSpan™ MEG<br>as life support equipment, for example vital<br>signs monitoring in intensive care units. | The Elekta Neuromag non-invasively measures the<br>magnetoencephalographic (meg)<br>signals (and, optionally,<br>electroencephalographic eeg<br>signals) produced by electrically<br>active tissue of the brain. These<br>signals are recorded by a<br>computerized data acquisition<br>system, displayed, and may then<br>be interpreted by trained<br>physicians to help localize these<br>active areas. The locations may<br>then be correlated with<br>anatomical information of the<br>brain. Meg is routinely used to<br>identify the locations of visual,<br>auditory, somatosensory, and<br>motor cortex in the brain. Meg is<br>also used to non-invasively<br>locate regions of epileptic<br>activity within the brain. The<br>localization information<br>provided by meg may be used, in<br>conjunction with other<br>diagnostic data, in neurosurgical<br>planning. | No significant<br>difference, except<br>the Orion<br>LifeSpan™ MEG<br>is not<br>recommended for<br>use with motor<br>evoked response<br>stimulators.<br><br>Assumptions<br>regarding evoked<br>response<br>stimulators, the<br>types of users<br>permitted, and use<br>case exclusions are<br>made more<br>explicitly clear to<br>further ensure safe<br>usage. | |
| Features | Orion<br>LifeSpan™<br>MEG<br>(K191785) | Elekta<br>Neuromag<br>(K041264) | Differences (if any) |
| Technology | | | |
| Site of<br>patient | Head and Scalp | Head and Scalp | No difference. |
| Patient<br>Position | Supine | Supine and<br>upright | No significant difference.<br>Upright position is available<br>mostly for research purposes.<br>Supine position is common<br>for MEG study. |
| Underlying<br>Technology | Superconducting<br>magnetometry | Superconducting<br>magnetometry | No difference. |
| Overall<br>Sensitivity | 10fT/ $√$ Hz | 10fT/ $√$ Hz | No difference. See Non-<br>Clinical Performance Data. |
| MEG<br>Sensing<br>Locations | 186 (adult)<br>138 (pediatric) | 102 | No significant difference. |
| SQUID<br>Readout | Flux locked loop | Flux locked loop | No difference. |
| Features | Orion<br>LifeSpan™<br>MEG<br>(K191785) | Elekta<br>Neuromag<br>(K041264) | Differences (if any) |
| Pickup Coil<br>Design | Axial<br>gradiometer | Mix of planar<br>gradiometers<br>and<br>magnetometers | Despite difference in<br>orientation, each pickup coil<br>type measures components of<br>the magnetic field generated<br>from the same brain sources.<br>During source estimation<br>these signals are converted to<br>(for example) equivalent<br>current dipole locations,<br>orientations and strengths.<br>Therefore, there is no<br>significant difference arising<br>from the pickup coil designs. |
| Average<br>Coil-to-Coil<br>Spacing | 27.3 mm (Adult),<br>25 mm<br>(Pediatric)<br>average distance<br>b/w centers of<br>each sensing loc. | 34 mm average<br>distance<br>between centers<br>of each sensing<br>location. | No difference. See Non-<br>Clinical Performance Data<br>for discussion of<br>localization accuracy<br>testing. |
| Detector<br>Architecture | DROS SQUID | dc SQUID | No significant difference.<br>DROS SQUID is a type of dc<br>SQUID. |
| Interference<br>Elimination<br>(i.e., SSS,<br>SPS) | SSP, DSSP | SSP, tSSS (which<br>has superseded<br>SSS) | Note that DSSP is<br>conceptually equivalent to<br>tSSS in that it uses<br>temporal information to<br>separate interference from<br>signal.1 Therefore, no<br>significant difference. |
| Head<br>Position<br>Indicator<br>(Y/N) | Y | Y | No difference. |
| Features | Orion<br>LifeSpan™<br>MEG<br>(K191785) | Elekta<br>Neuromag<br>(K041264) | Differences (if any) |
| Offline<br>Average<br>Function to<br>Process<br>Raw Data | Y | Y | No difference. |
| Number of<br>SQUID<br>Sensor<br>Elements | 186 (adult) 138<br>(pediatric) Each<br>at a unique<br>location | 306 at 102<br>unique locations | Phantom dipole localization<br>accuracy has shown that<br>sampling density is sufficient<br>in the Orion. Therefore, no<br>functionally significant<br>difference is noted due to the<br>number of SQUID elements. |
| Number of<br>Auxiliary<br>Channels<br>for Other<br>Types of<br>Data (i.e.,<br>EEG) | Up to 256<br>unipolar and<br>16 bipolar<br>EEG channels<br><br>Up to 8 ADC<br>channels. | Up to 124<br>unipolar and<br>4 bipolar<br>EEG<br>channels.<br><br>Up to 8<br>ADC<br>channels. | No significant difference,<br>provided that 128 EEG<br>channels are included with<br>the Orion, instead of the<br>maximum (256). |
| Method of<br>Calculation<br>(i.e.,<br>idealized v.<br>individual<br>head model)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.