Gabriel Feeding Tube with Balloon, and EnFit Connector
K191784 · Syncro Medical Innovations, Inc. · PIF · Nov 15, 2019 · Gastroenterology, Urology
Device Facts
Record ID
K191784
Device Name
Gabriel Feeding Tube with Balloon, and EnFit Connector
Applicant
Syncro Medical Innovations, Inc.
Product Code
PIF · Gastroenterology, Urology
Decision Date
Nov 15, 2019
Decision
SEKD
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Gabriel® Feeding Tube with Balloon functions as a conduit to facilitate enteral feeding, and may be used in adult or elderly patients who cannot consume an adequate diet orally. Small bowel feeding may be indicated for patients with functioning gut who require short to moderate term feeding support, such as post-trauma patients, general trauma patients, high-risk patients prone to tube misplacement complications, and patients in whom malnutrition exist, or may result, secondary to an underlying disease or condition.
Device Story
Nasoenteral feeding tube with distal balloon; facilitates enteral feeding into small bowel. Input: manual insertion by clinician; peristalsis-driven distal advancement. Features: DEHP-free PVC shaft; spiral 316 stainless steel wire reinforcement; central stainless steel stiffening stylet. Procedure: clinician applies lubricant/numbing gel; inserts tube; uses optional silk thread for deflection; inflates distal balloon at 30 cm mark to detect airway misplacement via pulse oximetry and CO2 sampling line. Output: enteral feeding conduit. Used in clinical/bedside settings. Benefits: reduces misconnection risk via EnFit connector; provides MR conditional safety; facilitates bedside placement with included convenience kit (stopcock, CO2 line).
Clinical Evidence
No clinical testing performed. Substantial equivalence supported by bench testing, including flow rate, tensile strength, leakage, stress cracking, and MR safety evaluations.
Indicated for adult or elderly patients requiring enteral feeding support due to inability to consume adequate oral diet; includes post-trauma, burn, and high-risk patients prone to tube misplacement or malnutrition.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 15, 2019
Syncro Medical Innovations, Inc. Sabry Gabriel, MD President 515 Mulberry St., Suite 200 Macon, GA 31201-6308
Re: K191784
Trade/Device Name: Gabriel Feeding Tube with Balloon, and EnFit Connector Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: PIF Dated: October 4, 2019 Received: October 8, 2019
Dear Sabry Gabriel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the Gabriel Feeding Tube with Balloon, and enfit connector have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
In addition, we have determined that your device kit contains Benzocaine Gel 20% which is subject to regulation as a drug.
Our substantially equivalent determination does not apply to the drug component of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug component. For information on applicable Agency requirements for marketing this drug, we suggest you contact:
Director, Division of Drug Labeling Compliance Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Shani P. Haugen, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K191784
#### Device Name
Gabriel Feeding® Tube with Balloon, and ENFit Connnector
Indications for Use (Describe)
The Gabriel® Feeding Tube with Balloon functions as a conduit to facilitate enteral feeding, and may be used in adult or elderly patients who cannot consume an adequate diet orally. Small bowel feeding may be indicated for patients with functioning gut who require short to moderate term feeding support, such as post-trauma patients, general trauma patients, high-risk patients prone to tube misplacement complications, and patients in whom malnutrition exist, or may result, secondary to an underlying disease or condition.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|------------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <input checked="true" type="checkbox"/> </div> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <input type="checkbox"/> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 5. 510(k) Summary
The following information is provided in accordance with 21 CFR 807.92 for the Premarket Notification 510(k) Summary:
### 5.1 Submitter Information
| Company: | Sabry Gabriel<br>MD, President<br>Syncro Medical Innovations, Inc.<br>515 Mulberry St Suite 200<br>Macon, GA 31201-6308 USA<br>Telephone: 478-335-8311<br>Fax: 716-809-3504<br>sabrygabriel@syncromedical.com |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Sabry Gabriel<br>MD, President<br>Syncro Medical Innovations, Inc.<br>515 Mulberry St Suite 200<br>Macon, GA 31201-6308 USA<br>Telephone: 478-335-8311<br>Fax: 716-809-3504<br>sabrygabriel@syncromedical.com |
| Date Summary Prepared: | November 14, 2019 |
#### 5.2 Name of the Device
| Trade Name: | Gabriel® Feeding Tube with Balloon, and EnFit Connector |
|----------------------|---------------------------------------------------------|
| Common Name: | Nasoenteral Feeding Tube with Balloon |
| Classification Name: | Gastrointestinal tubes and accessories |
| Review Panel: | Gastroenterology & Urology (GU) |
| Regulation: | 876.5980 |
| Class: | Class II |
| Product Code: | PIF |
# 5.3 Equivalence Claimed to Predicate Device
The Gabriel® Feeding Tube with Balloon, and EnFit Connector is equivalent to the Gabriel® Feeding Tube with Balloon (K160787), manufactured by SYNCRO MEDICAL INNOVATIONS, INC.
{5}------------------------------------------------
### 5.4 Basis for substantial Equivalence
The device has identical indications for use, intended population, materials used and functionality as the predicate device "Gabriel Feeding Tube with Balloon" (K160787), cleared on August 9, 2016.
The device label is changed from MR unsafe to MR conditional based on supporting test results.
The device proximal Y shaped catheter tip connector is replaced with a rigid male EnFit connector to mitigate misconnection risk without changing principal function, indication for use, intended use or intended patient population.
Three previously cleared items are added to the convenience kit to facilitate bedside insertion procedure. None of the changes or added items to the convenience kit negatively impact the device intended function or raise a new safety concern.
The three items added to the convenience kit are: (A) A 3-way EnFit stopcock valve (K915171), (B) Silk thread (K931271) that can be optionally used in the updated insertion technique, (C) CO2 sampling line (K181981) that can facilitate quick detection of misplacement in the airway in addition to observing pulse oximetry as in the predicate device. The feeding tube is a single use, finished device and sold for use by prescription only. It is provided non-sterile in three different sizes; 12 Fr, 10 Fr and 8 Fr. All three sizes are preassembled with nonmagnetic stylets.
# 5.5 Legally Marketed Device To Which The Device is Substantially Equivalent (Predicate Device):
Gabriel Feeding Tube with Balloon (K160787)
#### 5.6 Description Of The Device and Placement Procedure:
The Gabriel Feeding Tube with Balloon serves as conduit through which enteral feeding solutions are directly infused into patients small bowel. The tube has two distal end openings, one distal end balloon, one proximal end pilot balloon that indicates the status of the inflation of the distal end balloon and a proximal end EnFit connector that replaced our Y shaped catheter tip connector of the predicate device. The tube shaft is lined by a monofilament spiral wire to prevent occlusion from kinking. A braided 7 strands stainless steel stiffening central stylet prevents coiling in the mouth during insertion and provides column strength to facilitate placement into the stomach.The tube patient contacting material is made from DEHP-free PVC. Size 12 Fr tube is 130 cm long. During insertion of the tube, a lubricant and or a numbing gel is applied to one nostril. A silk thread is looped through the tube two distal openings. It can be optionally used to deflect the tube downwards towards the oropharynx, then discarded. A CO2 sampling line is connected to the tube proximal end to faciltate detection of feeding tube misplacement into the trachea, then discarded at the insertion procedure. At 18 cm mark patient, if awake, is asked to swallow. At 30 cm depth mark, the feeding tube balloon are inflated with 6 ml of air. Observing fluctuation in CO2 sampling wave or a drop in pulse oximetry by 5 or more points is an indication of tube misplacement into the trachea. This can be corrected by deflating the balloon immediately, withdrawal to 18 cm depth mark then reinsertion. Once esophageal placement is established, the tube with balloon inflated is advanced to the stomach. At 70 cm mark the stylet is partially removed as the tube is advanced to the 100 cm mark and taped at the nose. The stylet and CO2 sampling lines are removed and x-ray is obtained to confirm gastric or duodenal placement. The inflated tube distal end balloon allows peristalsis to advance the feeding tube distally. The balloon is deflated after 2 days, or before removal of the tube.
{6}------------------------------------------------
# 5.7 Indication For Use
The Gabriel Feeding Tube with Balloon functions as a conduit to facilitate enteral feeding, and may be used in adult or elderly patients who cannot consume an adequate diet orally. Small bowel feeding may be indicated for patients with functioning gut who require short to moderate term feeding support, such as post-trauma patients, burn patients, general trauma patients, high-risk patients prone to tube misplacement complications, and patients in whom malnutrition exist, or may result, secondary to an underlying disease or condition.
### 5.8 Technological Characteristics
The proposed device(K191784) has the same technological characteristics as the predicate device (K160787). Specifically, both feeding tubes function as conduit for enteral feeding.
Both the proposed and predicate feeding tubes are propelled distally by effect of peristalsis on the tube distal end balloon.
Both the proposed and predicate feeding tubes utilize inflation of the distal end balloon at 30 cm mark to detect airway misplacement by observing changes in pulse oximetry. In addition, the proposed device utilizes a previously cleared CO2 sampling line for additional detection of airway misplacement. The CO2 sampling line is connected to the proximal end of the feeding tube outside the patient and is discarded at the end of the insertion procedure.
Both the proposed and predicate feeding tubes have a stainless steel wire-enforced wall to prevent occlusion by kinking.
Both the proposed and predicate feeding tubes have a central stainless steel non patient contacting stiffening stylet.
# 5.9 Non-clinical testing
#### Tests were performed to demonstrate substantial equivalence with the predicate device in the following areas:
Feeding formula flow rate test using model with EnFit connector and compared to predicate device with catheter tip connector (pass)
Tube shaft to male EnFit connector tensile test (pass)
Stylet to stylet hub tensile test (pass)
48 Months shelf life test on the device with catheter tip connector (pass)
#### Tests conducted on the EnFit connector:
Leakage by pressure decay test (pass)
{7}------------------------------------------------
Positive pressure liquid leakage test (pass)
Stress cracking test (pass)
Resistance to separation from Axial load test (pass)
Resistance to separation from unscrewing test (pass)
Thread overriding resistance test (pass)
Disconnection by unscrewing test (pass)
Dimensional verification of EnFit connector (pass)
EnFit connector material elasticity modulus above 700 MPa (pass)
MR safety test (MR conditional)
### 5.10 Biocompatibility
Materials were tested for cytotoxicity, sensitization, acute and sub-acute toxicity. The materials were confirmed to be biocompatible.
# 5.11 Clinical Testing
No clinical testing was performed in association with this submission.
| 5.1: Device Comparison Table | | |
|---------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Proposed Device: Gabriel® Feeding<br>Tube with Balloon, and EnFit<br>Connector (K191784). | Predicate Device: Gabriel Feeding<br>Tube with Balloon (K160787). |
| Intended Use | A nasoenteric conduit used to facilitate<br>enteral feeding. The device is intended to<br>be inserted during bedside procedures. | A nasoenteric conduit used to facilitate<br>enteral feeding. The device is intended to<br>be inserted during bedside procedures. |
| Indication for Use | The Gabriel® Feeding Tube with<br>Balloon functions as a conduit to<br>facilitate enteral feeding, and may be<br>used in adult or elderly patients who<br>cannot consume an adequate diet orally.<br>Small bowel feeding may be indicated<br>for patients with functioning gut who<br>require short to moderate term feeding<br>support, such as post-trauma patients,<br>burn patients, general trauma patients, | The Gabriel® Feeding Tube with<br>Balloon functions as a conduit to<br>facilitate enteral feeding, and may be<br>used in adult or elderly patients who<br>cannot consume an adequate diet orally.<br>Small bowel feeding may be indicated<br>for patients with functioning gut who<br>require short to moderate term feeding<br>support, such as post-trauma patients,<br>burn patients, general trauma patients, |
| | | K19178<br>Page 5 |
| | high-risk patients prone to tube<br>misplacement complications, and<br>patients in whom malnutrition exist, or<br>may result, secondary to an underlying<br>disease or condition. | high-risk patients prone to tube<br>misplacement complications, and<br>patients in whom malnutrition exist, or<br>may result, secondary to an underlying<br>disease or condition. |
| Patient Contacting<br>Materials | -Tube shaft outside: PVC Tekni-Plex<br>Technologies (Suzhou) Co., Ltd - Grade<br>or Model 7811<br>- Tube balloon: PVC Tekni-Plex<br>Technologies (Suzhou) Co., Ltd Grade or<br>Model 6511<br>- Tube inner surface: PVC Tekni-Plex<br>Technologies (Suzhou) Co., Ltd - Grade<br>or Model 7811 with a spiral 316 stainless<br>steel coil | -Tube shaft outside: PVC Tekni-Plex<br>Technologies (Suzhou) Co., Ltd - Grade<br>or Model 7811<br>- Tube balloon: PVC Tekni-Plex<br>Technologies (Suzhou) Co., Ltd Grade or<br>Model 6511<br>- Tube inner surface: PVC Tekni-Plex<br>Technologies (Suzhou) Co., Ltd - Grade<br>or Model 7811 with a spiral 316 stainless<br>steel coil |
| Dimensions/Sizes | GFTB 512 with EnFit connector, 12 Fr x<br>130 cm, Non-Magnetic Stylet Design,<br>1.2 mm Stylet Diameter, 3 mL Balloon<br>GFTB 510 with EnFit connector, 10 Fr X<br>100 cm, Non-Magnetic Stylet Design,<br>1.2 mm Stylet Diameter, 3 mL Balloon;<br>GFTB 508 with EnFit Connector, 8 Fr x<br>80 cm, Non-Magnetic Stylet Design, 1.0<br>mm Stylet Diameter, 1 mL Balloon; | GFBT 412, 12 Fr x 130 cm, Magnetic<br>and Non-Magnetic Stylet Design, 1.2<br>mm Stylet Diameter, 3 mL Balloon<br>GFTB 410, 10 Fr x 130 cm, Magnetic<br>and Non-Magnetic Stylet Design, 1.2<br>mm Stylet Diameter, 3 mL Balloon<br>GFTB 408, 8 Fr x 100 cm, Magnetic and<br>Non-Magnetic Stylet Design, 1.0 mm<br>Stylet Diameter, 1 mL Balloon |
| Provided<br>Accessories (all<br>received premarket<br>notification (K)number,<br>or exempt) | Numbing gel, lubricant gel, cotton swab,<br>luer lock syringe, skin adhesive, securing<br>tape, silk suture thread, EnFit 3-way<br>stopcock valve (k)915171 and CO2<br>sampling line (k)181981 | Numbing gel, lubricant gel, pH paper,<br>cotton swab, luer lock syringe, skin<br>adhesive, and securing tape |
| Shelf Life/Expiration | Shelf life increased to 48 months after<br>appropriate tests met acceptance criteria. | 18 months |
| MR safety | MR Conditional based on test result | MR Unsafe due to lack of testing |
| Proximal end connector | Rigid male EnFit with elastic modulus<br>above 700 MPa | Flexible Y shaped catheter tip port |
| Connectivity | Directly to EnFit feeding line or through 3-way EnFit stopcock valves (k)915171 | Requires adapter with EnFit feeding line or Lopez 3-way valve with catheter tip port |
| Flushing and administration of medicine | Using EnFit syringe | Using catheter tip syringe |
| FDA Product Code | PIF | KNT |
| Sterilization | Non-sterile | Non-sterile |
| Silk thread | Utilized for patient comfort and proposed to be included in the kit | Utilized in clinical practice for patient comfort but not included in the kit |
| Detection of airway misplacement | Utilizes inflation of the tube distal end balloon at 30 cm depth mark and observing pulse oximetry and CO2 sampling | Utilizes inflation of the tube distal end balloon at 30 cm depth mark and observing pulse oximetry |
{8}------------------------------------------------
{9}------------------------------------------------
# 5.12 Conclusions Drawn from Non-Clinical Testing
The results of the comparison of design, materials, intended use and technological characteristics demonstrate that the proposed device (K191784) is as safe, as effective, and performs as well as the identified predicate legally marketed device (K160787) and support a determination of substantial equivalence.
#### 5.13 Conclusion
The Gabriel Feeding Tube with Balloon and ENFit connector (K191784) is substantially equivalent to our predicate device Gabriel Feeding tube with Balloon (K160787). Based on the intended use, principal of operation, performance characteristics, and technological characteristics, the proposed Gabriel Feeding Tube with Balloon and EnFit connector is substantially equivalent to and as safe, as effective, and performs as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.