K191722 · Zimmer Biomet Spine, Inc. · OLO · Dec 2, 2019 · Neurology
Device Facts
Record ID
K191722
Device Name
Vital Navigation System
Applicant
Zimmer Biomet Spine, Inc.
Product Code
OLO · Neurology
Decision Date
Dec 2, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
Vital Navigation instruments are to be used during the preparation and placement of Vital and Vitality screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. Vital Navigation instruments are specifically designed for use with the Medtronic StealthStation System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.
Device Story
System consists of nonsterile, reusable manual surgical instruments (awls, probes, taps, drivers); designed for use with Medtronic StealthStation System. Instruments interface with navigation system to provide real-time, multi-plane 3D/2D visualization of anatomical structures during open or minimally invasive spinal screw placement. Surgeon uses visual feedback from navigation system to guide instrument positioning relative to patient anatomy (CT/MR models, fluoroscopy, or digitized landmarks). System assists in precise hardware placement; benefits include improved accuracy for complex/MIS procedures.
Clinical Evidence
Bench testing only. Accuracy testing performed in accordance with ASTM F2554-18 (Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems) to validate instrument performance for Vital and Vitality screw implantation.
Technological Characteristics
Reusable manual surgical instruments (awls, probes, taps, drivers). Designed for integration with Medtronic StealthStation stereotactic navigation system. Accuracy validated per ASTM F2554-18. Materials and manufacturing methods consistent with predicate navigated instrument systems.
Indications for Use
Indicated for patients undergoing spinal surgery requiring preparation and placement of Vital and Vitality screws where stereotactic surgical navigation is appropriate for locating anatomical structures.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
December 2, 2019
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is the Department of Health & Human Services logo. The right side of the logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Zimmer Biomet Spine, Inc. Alex Pawlowski Regulatory Affairs Senior Specialist 10225 Westminster Drive Westminster, Colorado 80021
Re: K191722
Trade/Device Name: Vital Navigation System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: October 31, 2019 Received: November 1, 2019
Dear Alex Pawlowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191722
Device Name Vital Navigation System
#### Indications for Use (Describe)
Vital Navigation instruments are to be used during the preparation and placement of Vital and Vitality screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. Vital Navigation instruments are specifically designed for use with the Medtronic StealthStation System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> <b> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for Zimmer Biomet. The logo consists of a blue circle with a stylized "Z" inside, followed by the text "ZIMMER BIOMET" in a sans-serif font. The text is in a dark gray color.
### K191722
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92.
| Preparation Date | June 26, 2019 |
|---------------------|----------------------------------------------------------------------------------------------------|
| Applicant/Sponsor | Zimmer Biomet Spine, Inc.<br>10225 Westmoor Dr.<br>Westminster, CO 80021 |
| Contact Person | Alex Pawlowski<br>Regulatory Affairs Senior Specialist<br>Phone: 303-533-1062<br>Fax: 303-501-8444 |
| Alternate Contact | Ted Kuhn<br>Regulatory Affairs Associate Director<br>Phone: 720-469-5055<br>Fax: 303-501-8444 |
| Trade Name | Vital Navigation System |
| Common Name | Stereotaxic Instrument |
| Device Class | Class II |
| Classification Name | OLO - Orthopedic Stereotaxic Instrument (21 CFR 882.4560) |
# Device Description & Technological Characteristics:
Orthopedic
The Vital Navigation System is comprised of nonsterile, reusable instruments including awls, probes, taps, and drivers that can be operated manually. These instruments are intended to be used with the Medtronic Synergy™ Experience StealthStation® System to assist surgeons in precisely locating anatomical structures in either open or minimally invasive procedures for preparation and placement of Vital and Vitality screws. This surgical imaging technology provides surgeons visualization for complex and MIS procedures and confirms the accuracy of advanced surgical procedures. Use of these navigation systems provides the surgeon access to real-time, multi-plane 3D images (and 2D images) providing confirmation of hardware placement.
Device Panel
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Zimmer Biomet. The logo consists of a blue circle with a stylized "Z" inside, followed by the text "ZIMMER BIOMET" in a sans-serif font. The text is in a dark gray color, contrasting with the white background.
### Intended Use / Indications for Use:
Vital Navigation System Instruments are to be used during the preparation and placement of Vital and Vitality screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. Vital Navigation System Instruments are specifically designed for use with the Medtronic StealthStation System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.
### Summary of Technological Characteristics:
The technological characteristics of the subject Vital Navigation System components remain the same as, or similar to, the predicate Orthofix Navigated Instrument System (K172115) in regards to intended use, indications for use, design, manufacturing methods, fundamental technology, and operational principles. The purpose of this submission is to seek clearance for the release of stereotactic instruments for the implantation of Vital and Vitality screws, which utilize the existing Medtronic StealthStation system.
### Summary of Performance Data:
To support substantial equivalence, accuracy testing of the stereotactic instruments of the subject Vital Navigation System were assessed and tested appropriately in accordance with ASTM standards. The assessment concluded that the use of the proposed stereotactic instruments for the implantation of Vital and Vitality screws are safe and effective for use. Performance testing included tests per ASTM F2554-18 Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems. In all instances, the proposed instrument system functioned as intended and demonstrated substantial equivalence to the predicate device(s).
# Predicate Devices:
Primary Predicate: Orthofix Navigated Instrument System (K172115)
Additional Predicates: Medtronic StealthStation (K140454)
# Substantial Equivalence Conclusion:
The Vital Navigation System is substantially equivalent to the predicate system as a spinal fixation device in regards to intended use, indications for use, fundamental technology including design, materials, manufacturing methods, sterility, and operational principles. Furthermore, accuracy testing and other supporting information sufficiently demonstrate the substantial equivalence of the subject components to the Navigated Instrument System which has been cleared for stereotactic guidance during orthopedic surgery procedures. Based on this information, the subject device does not raise any new issues regarding the safety or efficacy when compared to its predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.