K191692 · Imedrix Inc. (Formerly Piitech Inc.) · DPS · Jan 10, 2020 · Cardiovascular
Device Facts
Record ID
K191692
Device Name
KardioScreen
Applicant
Imedrix Inc. (Formerly Piitech Inc.)
Product Code
DPS · Cardiovascular
Decision Date
Jan 10, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
KardioScreen is a portable and mobile device intended to acquire a resting, diagnostic ECG using surface electrodes from adult patients and pediatric patients age 14-21 years. The device is intended to be used by doctors and/or trained healthcare professionals in a hospital, medical professional's facility, clinic, physician's office or any medical outreach center. The device can acquire, display, analyze and print either 6-lead or 12-lead electrocardiograms which can be exported for external use. The recorded ECG signals can help users to analyze and diagnose heart disease, but the device is not meant as a sole means of diagnosis.
Device Story
KardioScreen is a portable, battery-operated electrocardiograph for clinical use by physicians or trained healthcare professionals. It acquires resting 6-lead or 12-lead ECG signals via surface electrodes connected to a KardioScreen Converter. The converter processes signals and transmits data via Bluetooth Low Energy (BLE) to an Android tablet running the KardioScreen application. The application displays, analyzes, and prints ECG waveforms for clinical review. The device aids in heart disease diagnosis but is not a sole diagnostic tool. It is intended for use in hospitals, clinics, and medical outreach centers.
Clinical Evidence
No clinical data were submitted. Substantial equivalence is supported by non-clinical testing, including electrocardiograph safety and performance, electrical safety, EMC, wireless co-existence, software verification and validation, usability validation, and biocompatibility.
Technological Characteristics
Portable, battery-operated electrocardiograph. Uses surface electrodes (disposable or optional reusable clamp/bulb). Connectivity via Bluetooth Low Energy (BLE) to Android tablet. Defibrillation-proof. Complies with electrical safety and EMC standards.
Indications for Use
Indicated for adult and pediatric patients (ages 14-21) for the acquisition of resting, diagnostic 6-lead or 12-lead ECGs to assist in the analysis and diagnosis of heart disease.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
January 10, 2020
iMedrix Inc. (formerly Piitech Inc.) Srikanth Jadcherla Chief Executive Officer 1525 McCarthy Blvd., Suite 1059 Milpitas, California 95035
Re: K191692
Trade/Device Name: KardioScreen Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: December 11, 2019 Received: December 12, 2019
Dear Srikanth Jadcherla:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191692
Device Name KardioScreen
#### Indications for Use (Describe)
KardioScreen is a portable and mobile device intended to acquire a resting, diagnostic ECG using surface electrodes from adult patients and pediatric patients age 14-21 years. The device is intended to be used by doctors and/or trained healthcare professionals in a hospital, medical professional's facility, clinic, physician's office or any medical outreach center. The device can acquire, display, analyze and print either 6-lead or 12-lead electrocardiograms which can be exported for external use. The recorded ECG signals can help users to analyze and diagnose heart disease, but the device is not meant as a sole means of diagnosis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <div> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Imedrix. The logo consists of a square made up of smaller squares in different shades of blue and pink. To the right of the square is the word "imedrix" in a blue sans-serif font. The "i" is lowercase, and the rest of the letters are lowercase.
## I. SUBMITTER
iMedrix Inc. (formerly Piitech Inc.) 1525 McCarthy Blvd., Suite 1059 Milpitas, CA 95035
Phone: 866-463-3749 Fax: 866-463-3749
Contact Person: Srikanth Jadcherla, Chief Executive Officer
Date Summary Prepared: December 11, 2019
## II. DEVICE
| Proprietary Name: | KardioScreen |
|----------------------|--------------------|
| Classification Name: | Electrocardiograph |
| Product Code: | DPS |
| Regulation: | 21 CFR 870.2340 |
| Regulatory Class: | II |
## III. PREDICATE DEVICE
The primary predicate device is Edan Instruments, Inc.'s PADECG, which was cleared under K161302.
This predicate device has not been to subject of a recall.
# IV. DEVICE DESCRIPTION
KardioScreen is a portable battery-operated electrocardiograph designed to acquire either a 6lead or 12-lead resting, diagnostic electrocardiogram (ECG) using surface electrodes from adult patients and pediatric patients aged 14-21 years. The device consists of the following three (3) parts:
- 1. Electrodes and patient cable with KardioScreen Converter
- 2. KardioScreen device
- 3. KardioScreen application running on Android tablet
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Imedrix. The logo consists of a square made up of smaller squares in different shades of blue and pink, followed by the word "imedrix" in blue lowercase letters. The logo is simple and modern, and the colors are calming and professional.
The device is supplied with patient cables with lead wires that connect to any commercially available disposable ECG electrode. Optional reusable clamp electrodes can be used for the limb leads and bulb electrodes for the chest leads.
Once the electrodes have been connected to the patient, the ECG cable with KardioScreen Converter is then connected to the "ECG" port on the KardioScreen device. This module receives the electrical ECG signals, performs processing on the signals and then wirelessly transmits the ECG data through a Bluetooth Low Energy (BLE) interface to a mobile application running on an Android tablet. The ECG waveforms are displayed on the KardioScreen application for viewing and analysis. From the Android tablet, the ECG report can be printed or exported for external use.
## V. INTENDED USE / INDICATIONS FOR USE
KardioScreen is a portable and mobile device intended to acquire a resting, diagnostic ECG using surface electrodes from adult patients and pediatric patients age 14-21 years. The device is intended to be used by doctors and/or trained healthcare professionals in a hospital, medical professional's facility, clinic, physician's office or any medical outreach center. The device can acquire, display, analyze and print either 6-lead or 12-lead electrocardiograms which can be exported for external use. The recorded ECG signals can help users to analyze and diagnose heart disease, but the device is not meant as a sole means of diagnosis.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
iMedrix's KardioScreen and Edan Instruments, Inc.'s PADECG have the same intended use, same clinical setting, same target user, same ECG measurements, and same pace detection. Both devices comply with the electrical safety, EMC and electrocardiograph safety and performance standards and are defibrillation-proof.
Both devices transmit wirelessly to a mobile device, though KardioScreen is designed to work with Android tablets while PADECG works with iOS devices. KardioScreen can also capture a 6lead ECG, in addition to the standard 12-lead ECG that PADECG captures. Both devices are designed for adult and pediatric populations, though KardioScreen restricts pediatric use to ages 14 and older.
These differences in technological characteristics do not raise new issues of safety or effectiveness.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Imedrix. The logo consists of a square made up of smaller squares in different shades of blue and pink. To the right of the square is the word "imedrix" in blue lowercase letters. The "i" is the same shade of blue as the darkest square in the logo, while the "x" is a lighter shade of blue.
### VII. PERFORMANCE DATA - NON-CLINICAL TESTING
No animal testing was submitted to support the substantial equivalence of KardioScreen to PADECG. The following non-clinical testing was performed to support the substantial equivalence of KardioScreen to the predicate device:
- Electrocardiograph safety and performance
- Electrical safety
- . Electromagnetic compatibility (EMC)
- . Wireless co-existence testing
- Software verification and validation ●
- . Usability validation
- Biocompatibility .
#### VIII. PERFORMANCE DATA - CLINICAL TESTING
No clinical data were submitted to support the substantial equivalence of KardioScreen to PADECG.
#### IX. CONCLUSIONS
Based on the indications for use, technological characteristics, results of non-clinical testing, and comparison to the predicate device, KardioScreen has been shown to be substantially equivalent to a legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.