# pORTAL Steerable Hydrophilic Guidewire and pORTAL EXT Extension Wire (K191687)

_Phenox Limited · DQX · Oct 9, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K191687

## Device Facts

- **Applicant:** Phenox Limited
- **Product Code:** [DQX](/productcode/DQX.md)
- **Decision Date:** Oct 9, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The pORTAL Steerable Hydrophilic Guidewire and pORTAL EXT Extension Wire is intended for use in the neuro vasculature. The wire can be steered to facilitate the selective placement of diagnostic or therapeutic catheters. This device is not intended for use in peripheral or coronary arteries.

## Device Story

Disposable, single-use steerable hydrophilic guidewire and extension wire system; used in neurovasculature to navigate tortuous vessels; facilitates selective placement of diagnostic/therapeutic catheters. Device features push/torque capability; distal tip flexibility; hydrophilic coating for lubricity. Includes torquer and introducer needle accessories. Extension wire allows length increase for over-the-wire exchanges. Used in sterile operating theatre by physicians; provides tactile feedback for navigation; enables precise catheter positioning; benefits patient by facilitating minimally invasive neurointerventional procedures.

## Clinical Evidence

Bench testing only. Performance testing included tensile strength, torque strength, tip flexibility, coating adherence/integrity, catheter compatibility, radiopacity, kink resistance, and particulate residue. Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemocompatibility, pyrogenicity) confirmed safety.

## Technological Characteristics

Guidewire: 35NLT Cobalt Chrome alloy core, Platinum/Tungsten distal tip coil, hydrophilic distal coating, PTFE proximal coating. Extension wire: Stainless steel core, Nitinol distal tube. 0.014" diameter. EO sterilization. Single-use. Mechanical steering via push/torque.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Traxcess 14EX Guidewire and Traxcess Docking Wire ([K093397](/device/K093397.md))
- Transend EX Guidewire ([K934122](/device/K934122.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 9, 2019

phenox Limited % Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW Buffalo, Minnesota 55313

Re: K191687

Trade/Device Name: pORTAL Steerable Hydrophilic Guidewire and pORTAL EXT Extension Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: MOF Dated: September 9, 2019 Received: September 10, 2019

Dear Mark Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Xiaolin Zheng, Ph.D., M.S. Director (Acting) DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K191687

Device Name

pORTAL Steerable Hydrophilic Guidewire and pORTAL EXT Extension Wire

Indications for Use (Describe)

The pORTAL Steerable Hydrophilic Guidewire and pORTAL EXT Extension Wire is intended for use in the neuro vasculature. The wire can be steered to facilitate the selective placement of diagnostic or therapeutic catheters. This device is not intended for use in peripheral or coronary arteries.

| Type of Use (Select one or both, as applicable)                                                                  |
|------------------------------------------------------------------------------------------------------------------|
| <span> <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)   </span> |
| <span>    □ Over-The-Counter Use (21 CFR 801 Subpart C)   </span>                                                |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary (21 CFR 807.92)

### I. SUBMITTER

phenox Limited, Ballybrit Business Park, Galway, Ireland

Contact Person: Garv Brogan Director of Regulatory Affairs, Phone: +353 91 740103

Date Prepared: May 31st, 2019

#### II. DEVICE

Name of Device: pORTAL® Steerable Hydrophilic Guidewire and pORTAL®EXT Extension Wire

Common or Usual Name: Catheter Guidewire

Classification: Class II, 21 CFR 870.1330, Catheter Guide wire

Product Code: MOF

#### III. PREDICATE DEVICE(S)

Traxcess 14EX Guidewire and Traxcess Docking Wire, K093397, Cleared Dec 2009. Transend EX Guidewire, K934122, cleared October 1993.

#### IV. DEVICE DESCRIPTION

The pORTAL Steerable Hydrophilic Guidewire is a disposable medical device designed for single use only. It is designed to selectively introduce and position catheters and other interventional devices within target vessels. In order to achieve this purpose, the guidewire must be capable of being steered through blood vessels. This necessitates pushing and torqueing capability within the product. The design of the distal section of the guides steerability, while maintaining the flexibility necessary to negotiate the tortuous path of the blood vessel network. Coatings are placed on the device to improve the lubricity and ease in its advancement through the guide catheter and the blood vessels. Two accessories are provided within the package. These are a Torquer device and an introducer Needle. Both can be used to aid in the use of the device.

The pORTAL EXT Extension Wire is intended to interface with pORTAL Steerable Hydrophilic Guidewire. It is provided sterile and is sold separately in its own packaging configuration. The 115cm extension wire provides a facility to extend the overall length of the pORTAL quidewire to 315cm (see table 1 for dimensional specifications). This enable length to be extended during use to aid with over-the wire exchange. This facilitates introduction and positioning of catheters and other interventional devices within the target anatomy, while the guidewire retains its working position at the location of intervention. The stiffness of the extension wire is comparable to the proximal, unground section of the pORTAL Guidewire.

{4}------------------------------------------------

| Part #           | Content of<br>Package | Description on label                         | Total<br>Length | Outer<br>Diameter  | Length of<br>Hydrophilic Coating |
|------------------|-----------------------|----------------------------------------------|-----------------|--------------------|----------------------------------|
| PORT-14-200-1    | 1                     | pORTAL<br>Steerable Hydrophilic<br>Guidewire | 205 cm          | 0.014"<br>(0.36mm) | 35.5 cm<br>(nominal)             |
| PORT-14-200-5    | 5                     | pORTAL<br>Steerable Hydrophilic<br>Guidewire | 205 cm          | 0.014"<br>(0.36mm) | 35.5 cm<br>(nominal)             |
| PORT-14-115-1-EX | 1                     | pORTAL<br>EXT Extension Wire                 | 115 cm          | 0.014"<br>(0.36mm) | N/A                              |
| PORT-14-115-5-EX | 5                     | pORTAL<br>EXT Extension Wire                 | 115 cm          | 0.014"<br>(0.36mm) | N/A                              |

#### Table 1: Product details of the pORTAL components

#### V. INDICATIONS FOR USE

The pORTAL Steerable Hydrophilic Guidewire and pORTAL EXT Extension Wire is intended for use in the neuro vasculature. The wire can be steered to facilitate the selective placement of diagnostic or therapeutic catheters. This device is not intended for use in peripheral or coronary arteries.

#### IV. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES

| Parameter                 | Predicate Device<br>(Transend EX)                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Device<br>(Traxcess 14EX)                                                                                                                                                                                                                                                                                                    | Subject Device                                                                                                                                                                                                                                                                                                                        | Assessment of<br>differences                                                                                                                                                                                                                                                                                                                                             |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model                     | Transend EX Guidewire                                                                                                                                                                                                                                                                                                                                                                                                   | Traxcess 14EX Guidewire<br>and Docking wire                                                                                                                                                                                                                                                                                            | pORTAL Guidewire and<br>Extension System                                                                                                                                                                                                                                                                                              | Equivalent                                                                                                                                                                                                                                                                                                                                                               |
| 510(k) number             | K934122                                                                                                                                                                                                                                                                                                                                                                                                                 | K093397                                                                                                                                                                                                                                                                                                                                | K191687                                                                                                                                                                                                                                                                                                                               | Equivalent                                                                                                                                                                                                                                                                                                                                                               |
| Product<br>Classification | 21 CFR 870.1330,<br>Class II                                                                                                                                                                                                                                                                                                                                                                                            | 21 CFR 870.1330,<br>Class II                                                                                                                                                                                                                                                                                                           | 21 CFR 870.1330,<br>Class II                                                                                                                                                                                                                                                                                                          | Equivalent                                                                                                                                                                                                                                                                                                                                                               |
| Device Design             | Steerable hydrophilic<br>guidewire                                                                                                                                                                                                                                                                                                                                                                                      | Steerable hydrophilic<br>guidewire and docking<br>wire                                                                                                                                                                                                                                                                                 | Steerable hydrophilic<br>guidewire and extension<br>system                                                                                                                                                                                                                                                                            | Equivalent                                                                                                                                                                                                                                                                                                                                                               |
| Indications for<br>Use    | The Transend EX<br>guidewire is intended<br>for general intravascular<br>use, including<br>neurovascular and<br>peripheral vasculatures.<br>The guidewire can be<br>torqued to facilitate the<br>selective placement of<br>diagnostic or<br>therapeutic catheters.<br>The device is not<br>intended for use in<br>coronary arteries. The<br>torque device (pin vase)<br>is included with the<br>guidewire to facilitate | The Traxcess Guidewire<br>and Docking Wire is<br>intended for general<br>intravascular use,<br>including the neuro and<br>peripheral vasculature.<br>The wire can be steered<br>to facilitate the selective<br>placement of diagnostic<br>or therapeutic catheters.<br>This device is not<br>intended for use in<br>coronary arteries. | The pORTAL Steerable<br>Hydrophilic Guidewire and<br>pORTAL EXT Extension Wire is<br>intended for use in the neuro<br>vasculature. The wire can be<br>steered to facilitate the<br>selective placement of<br>diagnostic or therapeutic<br>catheters.<br>This device is not intended<br>for use in peripheral or<br>coronary arteries. | The Indications for Use<br>statement is similar to the<br>predicates with the<br>exception that the<br>treatment indication is<br>reduced to only include<br>neuro as opposed to a<br>general intravascular use<br>including peripheral. This<br>tightening of the<br>indication statement does<br>not raise any additional<br>questions on safety and<br>effectiveness. |
| Mode of Action            | The guidewire can be<br>steered to facilitate the<br>selective placement of<br>diagnostic or<br>therapeutic catheters.                                                                                                                                                                                                                                                                                                  | The guidewire can be<br>steered to facilitate the<br>selective placement of<br>diagnostic or therapeutic<br>catheters.                                                                                                                                                                                                                 | The guidewire can be<br>steered to facilitate the<br>selective placement of<br>diagnostic or therapeutic<br>catheters.                                                                                                                                                                                                                | Equivalent                                                                                                                                                                                                                                                                                                                                                               |

#### Table 2: Comparison of pORTAL technological characteristics with predicate device

{5}------------------------------------------------

| Parameter                        | Predicate Device<br>(Transend EX)                                   | Predicate Device<br>(Traxcess 14EX)                                 | Subject Device                                                      | Assessment of<br>differences                                                                                                                                                                                                                                                                              |
|----------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Sizes                     | Two lengths 182cm and<br>205cm guidewire                            | 200cm guidewire<br>115cm docking wire                               | 205cm guidewire<br>115cm extension wire                             | The guidewire devices<br>have similar overall<br>lengths. Traxcess may<br>be extended using a<br>docking wire, resulting<br>in comparable length<br>to the pORTAL<br>Extension wire. These<br>minor differences in<br>lengths do not raise<br>any additional<br>questions on safety<br>and effectiveness. |
| Use                              | Single Use Device                                                   | Single Use Device                                                   | Single Use Device                                                   | Equivalent                                                                                                                                                                                                                                                                                                |
| Sterilization                    | EO Sterilization                                                    | EO Sterilization                                                    | EO Sterilization                                                    | Equivalent                                                                                                                                                                                                                                                                                                |
| Method of supply                 | Stored with a dispenser<br>coil, Tyvek Pouch and<br>shipping carton | Stored with a dispenser coil,<br>Tyvek Pouch and shipping<br>carton | Stored with a dispenser coil,<br>Tyvek Pouch and shipping<br>carton | Equivalent                                                                                                                                                                                                                                                                                                |
| Intended Clinical<br>Environment | Sterile operating theatre                                           | Sterile operating theatre                                           | Sterile operating theatre                                           | Equivalent                                                                                                                                                                                                                                                                                                |
| Biocompatible                    | Yes                                                                 | Yes                                                                 | Yes                                                                 | Equivalent                                                                                                                                                                                                                                                                                                |

# V. COMPARISON OF MATERIALS USED WITH THE PREDICATE DEVICES

| Parameter        | Predicate Device<br>(Transend EX)                          | Predicate Device<br>(Traxcess 14EX)                                                                                                             | Subject Device                                                                                                                       | Assessment of<br>differences                                                                                                                                                                                                                                                                                                             |
|------------------|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Wire<br>Material | Guidewire:<br>Scitanium (proprietary alloy)<br>core wire   | Guidewire:<br>Stainless-steel core with<br>Nitinol tapered wire<br><br>Docking Wire:<br>Stainless-steel core with<br>Nitinol tube on distal end | Guidewire:<br>35NLT Cobalt Chrome alloy<br><br>Extension Wire:<br>Stainless Steel core wire with<br>a Nitinol tube on the distal end | While there are differences in<br>materials, pORTAL uses<br>similar materials which are<br>commonly used in medical<br>devices. DV testing on the<br>pORTAL confirmed it will<br>perform as intended and the<br>difference in materials do not<br>raise questions of safety and<br>effectiveness                                         |
| Coating          | Distal:<br>Hydrophilic coating "ICE"<br><br>Proximal: PTFE | Distal: Hydrophilic coating<br><br>Proximal: PTFE                                                                                               | Distal: Hydrophilic coating<br><br>Proximal: PTFE                                                                                    | The function of each device<br>coating is equivalent. Whilst<br>having different materials,<br>each of the device coatings<br>functions are used equivalent<br>and incorporated in the<br>device to increase the<br>lubricity of the guidewires.<br>The difference in materials do<br>not raise questions of safety<br>and effectiveness |
| Tip<br>Material  | Tapered ribbon Scitanium<br>(proprietary alloy)            | Platinum / Nitinol coil                                                                                                                         | Platinum/Tungsten coil                                                                                                               | The function of each device tip<br>is equivalent. The difference in<br>materials do not raise<br>questions of safety and<br>effectiveness                                                                                                                                                                                                |

Table 3: Comparison of pORTAL materials with predicate device

{6}------------------------------------------------

# VI. BIOCOMPATIBILITY

Biocompatibility testing for the pORTAL Steerable Hydrophilic Guidewire and pORTAL EXT Extension Wire was conducted to conform with the FDA consensus standard, recognition number 2-156, AAMI/ANSI/ISO 10993-1:2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, along with the FDA Blue Book Memorandum# G95. Table 4 and 5 below summarizes the biocompatibility testing performed on the pORTAL Steerable Hydrophilic Guidewire and pORTAL EXT Extension Wire.

| Test                                                              | Test Description                                                                                                                                           | Result Summary                                                                                                                                                                                                                                                                                                                                                                              | Conclusion |
|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Chemical<br>Characterization                                      | Exhaustive Extraction in<br>Purified Water (PW), 50%<br>Isopropyl<br>Alcohol (IPA), and Hexane (Hex) for<br>Total Non-Volatile Residue (NVR)               | The exaggerated extractions of the Phenox/LRM<br>0.014" Neurovascular Guidewire yielded low<br>(<0.24mg) amounts of residue from the PW<br>extraction vehicle; the EtOH and Hex extraction<br>vehicles yielded a slightly higher amount of NVR<br>than the PW.                                                                                                                              | Acceptable |
|                                                                   | Fourier Transform Infrared (FTIR)<br>Spectroscopy of residues obtained<br>from the test article extracts                                                   | Purified Water: Most closely matches cellulose<br>acetate ester EtOH: Most closely matches<br>polyurethane Hexane: Most closely matches<br>phenolic antioxidant                                                                                                                                                                                                                             | Acceptable |
|                                                                   | Determination of Trace<br>Metals/Elements Using Inductively<br>Coupled Plasma (ICP) Spectroscopy<br>(PW extract only)                                      | The PW extract of the PORTAL Guidewire contained<br>trace amounts of metallics. The most significant<br>amount of a metal detected was tungsten which was<br>present at a level of 6.13 x $10^4$ mg/device.                                                                                                                                                                                 | Acceptable |
|                                                                   | Gas Chromatography with Mass<br>Spectrometry (GC/MS) for detection<br>of volatile and semi-volatile organic<br>compounds from the test article<br>extracts | No semi-volatile compounds were detected in the<br>PW, EtOH And Hex extracts of the pORTAL<br>Guidewire at a level greater than 1.0 ppm, the<br>detection limit of the method.                                                                                                                                                                                                              | Acceptable |
|                                                                   | Ultra-performance Liquid<br>Chromatography with Mass<br>Spectrometry (LC/MS) for detection of<br>non-volatile compounds from the test<br>article extracts  | The LC/MS analyses of the EtOH and Hex extracts<br>of the pORTAL Guidewire indicated the presence of<br>the antioxidant Irganox 1010                                                                                                                                                                                                                                                        | Acceptable |
| Cytotoxicity                                                      | Cytotoxicity Study Using the ISO<br>Elution Method                                                                                                         | The test article extract showed no evidence of<br>causing cell lysis or toxicity and was a grade 0. The<br>test article extract met the requirements of the test<br>since the grade was less than a grade 2 (mild<br>reactivity).                                                                                                                                                           | Pass       |
| Sensitization                                                     | ISO Guinea Pig Maximisation<br>Sensitisation Test                                                                                                          | The test article extracts showed no evidence of<br>causing delayed dermal contact sensitization in the<br>guinea pig. The test article was not considered a<br>sensitizer in the guinea pig maximization test.                                                                                                                                                                              | Pass       |
| Intracutaneous<br>Reactivity/<br>irritation                       | ISO Intracutaneous Study in Rabbits                                                                                                                        | The test article met the requirements of the test<br>since the difference between each test article<br>extract overall mean score and corresponding<br>control extract overall mean score was 0.0 and 0.0<br>for the SC and SO test article extracts, respectively.                                                                                                                         | Pass       |
| Acute Systemic<br>Toxicity                                        | ISO Systemic Toxicity Study in Mice                                                                                                                        | There was no mortality or evidence of systemic<br>toxicity from the extracts injected into mice. Each<br>test article extract met the requirements of the study.                                                                                                                                                                                                                            | Pass       |
| pORTAL Steerable Hydrophilic Guidewire                            |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                             |            |
| Test                                                              | Test Description                                                                                                                                           | Result Summary                                                                                                                                                                                                                                                                                                                                                                              | Conclusion |
| Hemocompatibility                                                 | ASTM Hemolysis Study                                                                                                                                       | The hemolytic index for the test article in direct<br>contact with blood was 0.0% and the hemolytic<br>index for the test article extract was 0.3%. Both the<br>test article in direct contact with blood and the test<br>article extract were non-hemolytic.                                                                                                                               | Pass       |
| Hemocompatibility                                                 | In Vivo Thromboresistance<br>Study in Dogs Jugular Vein                                                                                                    | Under the conditions of the study, minimal to slight<br>thrombus formation was noted with the test article<br>and control article. The test and control article<br>appeared equivalent in this exaggerated model with<br>neither demonstrating thromboresistance.<br>Consequently, these results should be considered<br>with respect to clinical use when assessing clinical<br>relevance. | Pass       |
| Acute Systemic<br>Toxicity<br>(Material Mediated<br>Pyrogenicity) | USP Rabbit Pyrogen Study, Material<br>Mediated                                                                                                             | The total rise of rabbit temperatures during the 3-<br>hour observation period was within acceptable<br>USP limits. The test article was judged as<br>nonpyrogenic.                                                                                                                                                                                                                         | Pass       |

#### Table 4: Biocompatibility Summary for pORTAL Guidewire

{7}------------------------------------------------

# Table 5: Biocompatibility Summary for pORTAL EXT Extension Wire

| pORTAL EXT Extension Wire    |                                                                                                                                            |                                                                                                                                                                                                                                                                                              |            |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Test                         | Test Description                                                                                                                           | Result Summary                                                                                                                                                                                                                                                                               | Conclusion |
| Chemical<br>Characterization | Exhaustive Extraction in Purified<br>Water (PW), 50% Isopropyl Alcohol,<br>and Hexane for Total Non-Volatile<br>Residue                    | The exaggerated extractions of the pORTAL<br>Extension Wire yielded low (0.04 mg) amounts of<br>residue from the PW extraction vehicle; the EtOH<br>(0.071) and Hex (0.026) extraction vehicles yielded<br>a slightly higher and lower amounts of NVR<br>respectively than the PW.           | Acceptable |
|                              | Fourier Transform Infrared Spec. of<br>residues obtained from the test article<br>extracts                                                 | Purified Water: No major bands of interest were<br>observed. EtOH: Most closely matches trace<br>hydrocarbons, Hexane: Most closely matches trace<br>hydrocarbons                                                                                                                            | Acceptable |
|                              | Determination of Trace Metal<br>Elements Using Inductively Coupled<br>Plasma Spec.                                                         | The PW extract of the pORTAL Extension Wire<br>contained trace amounts of metallics. The most<br>significant amount of a metal detected was<br>aluminium which was present at a level of 5.06 x 10<br>mg/device.                                                                             | Acceptable |
|                              | Gas Chromatography with Mass Spec.<br>for detection of volatile & semi-volatile<br>organic compounds from the test<br>article extracts     | No semi-volatile compounds were detected in the<br>PW, EtOH And Hex extracts of the pORTAL<br>Extension Wire at a level greater than 1.0 ppm, the<br>detection limit of the method.                                                                                                          | Acceptable |
|                              | Ultra-Performance<br>Liquid Chromatography with Mass<br>Spec. for detection of non-volatile<br>compounds from the test article<br>extracts | The LC/MS data indicate the presence of numerous<br>incompletely identified compounds in the PW, EtOH,<br>and Hex extracts of the pORTAL Extension Wire.<br>Analysis of the types of compounds detected<br>indicates that many of the chemical species<br>detected are structurally related. | Acceptable |
| Cytotoxicity                 | Cytotoxicity Study<br>Using the ISO<br>Elution Method                                                                                      | The test article extract showed no evidence of<br>causing cell lysis or toxicity and was a grade 0. The<br>test article extract met the requirements of the test<br>since the grade was less than a grade 2 (mild<br>reactivity).                                                            | Pass       |
| Hemocompatibility            | ASTM Hemolysis Study                                                                                                                       | The hemolytic index for the test article in direct<br>contact with blood was 0.3% and the hemolytic<br>index for the test article extract was 0.6%. Both the<br>test article in direct contact with blood and the test<br>article extract were non-hemolytic.                                | Pass       |

{8}------------------------------------------------

#### VII. NON-CLINICAL PERFORMANCE DATA

The following non-clinical performance tests were executed to provide support for the substantial equivalence determination. These are summarized below.

### Performance Testing - Bench

A full suite of performance testing on the bench was carried out on the pORTAL Steerable Hydrophilic Guidewire and pORTAL EXT Extension Wire. This testing included, Tensile Strength, Torque Strength, Tip Flexibility, Coating Adherence/Integrity & Catheter Compatibility under simulated conditions. The results of this testing demonstrated compliance to all the design attributes as well as demonstrating an acceptable performance outcome. The results of these tests, in conjunction with the substantial equivalence claims as outlined in the premarket notification, effectively demonstrate substantial equivalence to the cited predicates.

| pORTAL Steerable Hydrophilic Guidewire                          |                                                                                                                                                                                                                            |                                                                                                                                                                                    |
|-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test                                                            | Test Method Summary                                                                                                                                                                                                        | Results                                                                                                                                                                            |
| Dimensional                                                     | The overall guidewire length, polymer jacket length<br>and overall guidewire diameter were measured.                                                                                                                       | Dimensional results meet acceptance<br>criteria.                                                                                                                                   |
| Tip Tensile Strength                                            | The distal section of the wire was pulled until it<br>failed.                                                                                                                                                              | Results show the tensile strength meets<br>specification.                                                                                                                          |
| Catheter Compatibility                                          | The guidewire was tested with relevant interfacing<br>devices such as over the wire micro catheters under<br>simulated use.                                                                                                | The guidewires tested performed<br>acceptably and were shown to be<br>compatible with tested microcatheters.                                                                       |
| Coating Adherence/Integrity                                     | The guidewire was tracked multiple times through a<br>catheter in a simulated anatomy. The coatings were<br>inspected for signs of damage or degradation.                                                                  | No coating was rubbed off during or after<br>testing of guidewire in any of the<br>microcatheters.                                                                                 |
| Tip Stiffness (Gram Weight)                                     | The distal tip of the guidewire was compressed. The<br>peak load was electronically recorded and reported.                                                                                                                 | Device stiffness is equivalent or less<br>than predicate device. Device tip<br>stiffness is within specification.                                                                  |
| Tip Shape Retention                                             | A J-Shape was formed on the distal end of the wire,<br>then recorded using an image measuring system.                                                                                                                      | The guidewire distal tip retention is<br>better than or equivalent to predicate<br>device. Tip shape retention is within<br>specification.                                         |
| Lubricity and Durability Test                                   | The guidewire, within a microcatheter was tracked<br>multiple times through a 3-Dimensional Neuro<br>Silicone anatomy block. Resulting lubricity was<br>measured.                                                          | Lubricity and durability of the guidewire<br>is better than or equivalent to the<br>predicate device. The device meets<br>acceptance criteria on both lubricity and<br>durability. |
| PTFE Coating Adherence (Saline<br>Soak & Wipe Test)             | The entire length was pulled through a fixed clamp.<br>The wire was inspected for PTFE degradation or<br>damage.                                                                                                           | All devices passed saline soak & wipe<br>test.                                                                                                                                     |
| PTFE Coating Mandrel Adhesion<br>Test (Saline Soak & Wrap Test) | The wire was wrapped around a mandrel of a fixed<br>diameter (as defined in ISO11070 standard). The<br>wrapped PTFE coated wire was inspected under a<br>microscope for any signs of flaking, cracking or<br>delamination. | All devices passed saline soak & wrap<br>test.                                                                                                                                     |
| PTFE Adhesion / Durability<br>Testing (Eraser Test)             | The wire was placed in a fixture and subject to a<br>number of cycles underweight. The wire was<br>removed and inspected for failure.                                                                                      | All devices passed eraser test.                                                                                                                                                    |
| Radiopacity                                                     | The radiopaque distal end of the guidewire was<br>imaged using standard fluoroscopy equipment. An<br>image was taken and reviewed to ensure visibility                                                                     | Device radiopacity was seen to be<br>comparable to predicate device. Device<br>meets specification.                                                                                |
| Torque Strength (Combined<br>Load/Turns to Failure)             | The guidewire was loaded into the guide catheter<br>within a simulated neurovascular anatomy. Force is<br>applied and the specimen is rotated to failure.                                                                  | Device torqueability was demonstrated<br>within specification.                                                                                                                     |
| ISO Flex Resistance Test<br>(Bending Durability)                | Test method defined in ISO 11070 Annex G. The<br>test region of the wire was wrapped around<br>mandrels in opposing directions a number of times.<br>The wire was removed and examined.                                    | Flex resistance was comparable to<br>predicate device.                                                                                                                             |

#### Table 6: DV Summary for pORTAL Guidewire

{9}------------------------------------------------

| pORTAL Steerable Hydrophilic Guidewire |                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                      |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test                                   | Test Method Summary                                                                                                                                                                                                                                                                               | Results                                                                                                                                                                              |
| Corrosion Resistance                   | Test method defined in ISO 11070 Annex B. The<br>products were placed in a beaker of salt water at<br>room temp for 5 hours, and then placed in boiling<br>distilled water for 30 minutes, followed by 48 hours<br>at 37°C. The products were removed, and examined<br>for evidence of corrosion. | Device meets acceptance criteria, no<br>evidence of corrosion on device.                                                                                                             |
| FDA Tip Flexibility                    | This test measures the load required to deflect the<br>distal tip of the guidewire to angles of 45° and 90. A<br>force curve was generated for each specimen at the<br>specified deflection points and were electronically<br>collected.                                                          | Tip stiffness properties were shown to<br>be comparable or less atraumatic than<br>predicate device. Peak load was within<br>specification.                                          |
| Kink Resistance / Flexibility          | The tip of each device was wrapped around<br>mandrels, until kinking occurs.                                                                                                                                                                                                                      | Kink resistance was equivalent or better<br>than predicate device.                                                                                                                   |
| Particle Residue                       | Following simulated use, particulates generated are<br>collected, and counted by size for each wire.                                                                                                                                                                                              | All parts & calculated tolerance interval<br>are below max specification for ≥10μm<br>particles/device and ≥25um particles/<br>device and compared to a predicate<br>device control. |
| Torque Response                        | The test measured the reaction of the distal end of<br>the guidewire when the wire was placed within a<br>simulated anatomy.                                                                                                                                                                      | Equivalent torque response when<br>compared to predicate device. Both<br>fresh and fatigued devices are within<br>specifications.                                                    |

Table 7: DV Summary for pORTAL EXT Extension Wire

| pORTAL EXT Extension Wire                                       |                                                                                                                                                                                                                                                     |                                                                                                                                                                                      |
|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test                                                            | Test Method Summary                                                                                                                                                                                                                                 | Results                                                                                                                                                                              |
| Catheter Compatibility Test                                     | The guidewire was tested with relevant interfacing<br>devices under simulated use.                                                                                                                                                                  | The extension systems performed<br>acceptably and were shown to be<br>compatible with tested microcatheters.                                                                         |
| PTFE Coating Mandrel Adhesion<br>Test (Saline Soak & Wrap Test) | The wire was wrapped around a mandrel of a fixed<br>diameter (as defined in ISO11070 standard). The<br>wrapped PTFE coated wire was inspected under a<br>microscope for any signs of flaking, cracking or<br>delamination.                          | No damage was observed post soak &<br>wrap test. All devices passed.                                                                                                                 |
| PTFE Coating Adherence (Saline<br>Soak & Wipe Test)             | The entire length was pulled through a fixed<br>clamp. The wire was inspected for PTFE<br>degradation or damage.                                                                                                                                    | No damage was observed post soak &<br>wipe test. All devices passed.                                                                                                                 |
| PTFE Adhesion / Durability Testing<br>(Eraser Test)             | The wire was placed in a fixture and subject to a<br>number of cycles underweight. The wire was<br>removed and inspected for failure.                                                                                                               | No damage was observed post eraser<br>test. All devices passed.                                                                                                                      |
| Dimensional Measurements                                        | The overall length and maximum diameter of the<br>extension wire were measured.                                                                                                                                                                     | All dimensional measurements were<br>within specifications.                                                                                                                          |
| Pull Test (Glue Joint Strength)                                 | The joint strength of the extension wire is<br>measured using a calibrated force tester                                                                                                                                                             | Joint strength measurements were<br>within specifications.                                                                                                                           |
| Pull Test (Strength of Union)                                   | The peak force to separate the guidewire from the<br>extension wire is measured using a calibrated<br>force tester                                                                                                                                  | Strength of union measurements were<br>comparable to predicate device and<br>within specifications.                                                                                  |
| Hypotube Joint Flex Test                                        | This test involves flex testing the integrity of the<br>connecting joint between the extension wire and<br>the guidewire. A force curve is generated for each<br>specimen, and the peak load is electronically<br>calculated and put into a report. | Results have shown comparable joint<br>flex properties between the subject<br>device & predicate device. Device meets<br>specification.                                              |
| Particulate Residue Test                                        | Following simulated use, particulates generated<br>are collected, and counted by size for each wire.                                                                                                                                                | All parts & calculated tolerance interval<br>are below max specification for ≥10µm<br>particles/device and ≥25µm particles/<br>device and compared to a predicate<br>device control. |

{10}------------------------------------------------

| pORTAL Guidewire and Extension System                  |                                                                                          |                                                                    |
|--------------------------------------------------------|------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Test                                                   | Test Method Summary                                                                      | Results                                                            |
| Residual Study for guidewire after<br>1x sterilisation | Wires were sterilized, then tested for EtO and ECH residual levels.                      | All test articles passed acceptance criteria.                      |
| LAL Testing                                            | LAL enhancement and inhibition validation was performed followed by endotoxin testing.   | All test articles passed acceptance criteria for endotoxin limits. |
| Product Bioburden Testing                              | Bioburden validation followed by bioburden testing was carried out on each test article. | All test articles passed bioburden acceptance criteria.            |

### Table 8: Sterility Summary for pORTAL Guidewire and EXT Extension Wire

#### Table 9: Packaging Integrity Summary for pORTAL Guidewire and EXT Extension Wire ORTAL Guidewir od Evto nsion Syste

| 000 TAL Gülüsmile allu Excelsion SVSCHI |                                                                                                                                        |                                                                  |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| Test                                    | Test Method Summary                                                                                                                    | Results                                                          |
| Visual Inspection                       | All samples are visually inspected for any damage<br>or defects.                                                                       | All samples passed visual inspection.                            |
| Seal Integrity: Dye Penetration         | Packaging is exposed to a dye solution and is<br>observed and inspected for leakage. As per ASTM<br>F1929.                             | All samples passed dye penetration<br>testing.                   |
| Pouch Peel Test                         | The force required to separate the two sealed<br>materials is recorded.                                                                | All samples passed pouch peel testing.                           |
| Seal Tensile Strength                   | The maximum tensile force required to separate<br>the seals is recorded. Test Method as per ASTM<br>F88.                               | All samples were within seal tensile<br>strength specifications. |
| Sterilisation                           | Completed products were sterilized, then tested for<br>EtO and ECH residual levels.                                                    | All samples passed sterilisation tests.                          |
| Seal integrity: Bubble leak test        | Packaging is pressurized and submerged under<br>water while visually inspecting for streams of<br>escaping bubbles. As per ASTM F2096. | All samples passed the bubble leak test.                         |
| Cytotoxicity                            | ISO MEM elution -<br>L929 fibroblast cultures - ISO10993-5                                                                             | All samples were within cytotoxicity<br>specifications.          |
| Physiochemical                          | The methods for these tests are outlined in The<br>United States Pharmacopeia Convention <661>                                         | All samples were within physiochemical<br>specifications.        |
| Age Testing                             | All samples are visually inspected for any damage<br>or defects.                                                                       | All samples met age testing criteria.                            |

#### Table 10: Shelf Life Testing Summary for pORTAL Guidewire for 2 years and 3 months

| pORTAL Steerable Hydrophilic Guidewire              |                                                                                                                                                                   |                                                                                                                                                                                    |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test                                                | Test Method Summary                                                                                                                                               | Results                                                                                                                                                                            |
| Dimensional                                         | The overall guidewire length, polymer jacket length<br>and overall guidewire diameter were measured.                                                              | Dimensional results meet acceptance<br>criteria.                                                                                                                                   |
| Tip Tensile Strength                                | The distal section of the wire was pulled until it<br>failed.                                                                                                     | Results show the tensile strength meets<br>specification.                                                                                                                          |
| Catheter Compatibility                              | The guidewire was tested with relevant interfacing<br>devices such as over the wire micro catheters under<br>simulated use.                                       | The guidewires tested performed<br>acceptably and were shown to be<br>compatible with tested microcatheters.                                                                       |
| Coating Adherence/Integrity                         | The guidewire was tracked multiple times through a<br>catheter in a simulated anatomy. The coatings were<br>inspected for signs of damage or degradation.         | No coating was rubbed off during or after<br>testing of guidewire in any of the<br>microcatheters.                                                                                 |
| Tip Stiffness (Gram Weight)                         | The distal tip of the guidewire was compressed. The<br>peak load was electronically recorded and reported.                                                        | Device stiffness is equivalent or less<br>than predicate device. Device tip<br>stiffness is within specification.                                                                  |
| Tip Shape Retention                                 | A J-Shape was formed on the distal end of the wire,<br>then recorded using an image measuring system.                                                             | The guidewire distal tip retention is<br>better than or equivalent to predicate<br>device. Tip shape retention is within<br>specification.                                         |
| Lubricity and Durability Test                       | The guidewire, within a microcatheter was tracked<br>multiple times through a 3-Dimensional Neuro<br>Silicone anatomy block. Resulting lubricity was<br>measured. | Lubricity and durability of the guidewire<br>is better than or equivalent to the<br>predicate device. The device meets<br>acceptance criteria on both lubricity and<br>durability. |
| PTFE Coating Adherence (Saline<br>Soak & Wipe Test) | The entire length was pulled through a fixed clamp.<br>The wire was inspected for PTFE degradation or<br>damage.                                                  | All devices passed saline soak & wipe<br>test.                                                                                                                                     |

{11}------------------------------------------------

| pORTAL Steerable Hydrophilic Guidewire                          |                                                                                                                                                                                                                                                                                                   |                                                                                                                                             |
|-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Test                                                            | Test Method Summary                                                                                                                                                                                                                                                                               | Results                                                                                                                                     |
| PTFE Coating Mandrel Adhesion<br>Test (Saline Soak & Wrap Test) | The wire was wrapped around a mandrel of a fixed<br>diameter (as defined in ISO11070 standard). The<br>wrapped PTFE coated wire was inspected under a<br>microscope for any signs of flaking, cracking or<br>delamination.                                                                        | All devices passed saline soak & wrap<br>test.                                                                                              |
| PTFE Adhesion / Durability<br>Testing (Eraser Test)             | The wire was placed in a fixture and subject to a<br>number of cycles underweight. The wire was<br>removed and inspected for failure.                                                                                                                                                             | All devices passed eraser test.                                                                                                             |
| Radiopacity                                                     | The radiopaque distal end of the guidewire was<br>imaged using standard fluoroscopy equipment. An<br>image was taken and reviewed to ensure visibility                                                                                                                                            | Device radiopacity was seen to be<br>comparable to predicate device. Device<br>meets specification.                                         |
| Torque Strength (Combined<br>Load/Turns to Failure)             | The guidewire was loaded into the guide catheter<br>within a simulated neurovascular anatomy. Force is<br>applied and the specimen is rotated to failure.                                                                                                                                         | Device torqueability was demonstrated<br>within specification.                                                                              |
| ISO Flex Resistance Test<br>(Bending Durability)                | Test method defined in ISO 11070 Annex G. The<br>test region of the wire was wrapped around<br>mandrels in opposing directions a number of times.<br>The wire was removed and examined.                                                                                                           | Flex resistance was comparable to<br>predicate device.                                                                                      |
| Corrosion Resistance                                            | Test method defined in ISO 11070 Annex B. The<br>products were placed in a beaker of salt water at<br>room temp for 5 hours, and then placed in boiling<br>distilled water for 30 minutes, followed by 48 hours<br>at 37°C. The products were removed, and examined<br>for evidence of corrosion. | Device meets acceptance criteria, no<br>evidence of corrosion on device.                                                                    |
| FDA Tip Flexibility                                             | This test measures the load required to deflect the<br>distal tip of the guidewire to angles of 45° and 90. A<br>force curve was generated for each specimen at the<br>specified deflection points and were electronically<br>collected.                                                          | Tip stiffness properties were shown to<br>be comparable or less atraumatic than<br>predicate device. Peak load was within<br>specification. |
| Kink Resistance / Flexibility                                   | The tip of each device was wrapped around<br>mandrels, until kinking occurs.                                                                                                                                                                                                                      | Kink resistance was equivalent or better<br>than predicate device.                                                                          |
| Particle Residue                                                | Following simulated use, particulates generated are<br>collected, and counted by size for each wire. T                                                                                                                                                                                            | All parts & calculated tolerance interval<br>are below max specification for ≥10µm<br>particles/device and ≥25µm<br>particles/device.       |
| Torque Response                                                 | The test measured the reaction of the distal end of<br>the guidewire when the wire was plac…

---

**Source:** [https://fda-staging.innolitics.com/device/K191687](https://fda-staging.innolitics.com/device/K191687)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).