← Product Code [MUD](/productcode/MUD) · K191676

# Intra.Ox 2.0 Handheld Tissue Oximeter (K191676)

_Vioptix, Inc. · MUD · Sep 16, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K191676

## Device Facts

- **Applicant:** Vioptix, Inc.
- **Product Code:** [MUD](/productcode/MUD.md)
- **Decision Date:** Sep 16, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Intra.Ox 2.0 Handheld Tissue Oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue. The Intra.Ox 2.0 Handheld Tissue Oximeter is indicated for use in monitoring patients during circulatory or perfusion examinations. The Intra.Ox 2.0 Handheld Tissue Oximeter is intended to be used by physicians, surgeons, nurses, or other skilled users in a medical environment. The Intra.Ox 2.0 Handheld Tissue Oximeter should only be used on adult patients.

## Device Story

Intra.Ox 2.0 is a handheld, battery-powered tissue oximeter used by clinicians in medical environments to estimate StO2. The device consists of a reusable main unit, a sterile single-use sheath, a single-use battery pack, and a single-use quality control target. It operates via spectrophotometric oximetry, using five near-infrared wavelengths (730nm, 760nm, 810nm, 845nm, 895nm) to perform spatially resolved optical measurements through direct tissue contact. The main unit processes reflected light to calculate StO2, which is displayed on an integrated screen. Clinicians use these real-time measurements to assess tissue perfusion and circulatory status, aiding in clinical decision-making regarding patient monitoring. The device provides a non-invasive method to evaluate tissue oxygenation levels.

## Clinical Evidence

Clinical performance was evaluated in a non-significant risk (NSR) study with 18 healthy volunteers. The study compared StO2 measurements between the subject and predicate devices during transient ischemic events. Results demonstrated excellent agreement in deoxygenation curve shapes and similar StO2 ranges for both healthy and compromised tissue, supporting substantial equivalence.

## Technological Characteristics

Handheld, battery-powered (4 Lithium AA, 6V). Polycarbonate housing. Spectrophotometric oximetry using 5 LED wavelengths: 730nm, 760nm, 810nm, 845nm, 895nm. Dimensions: 7.3" x 2.7" x 3.1". Standards: ASTM F2096:2011, ISO 10993-1:2009, ISO 10993-2:2006, ISO 10993-5:2009, ISO 10993-10:2010, ISO 10993-12:2012, ISO 13485:2003, BS EN ISO 14971:2012, ISTA 2A:2011.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- Intra.Ox Handheld Tissue Oximeter ([K163472](/device/K163472.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

ViOptix, Inc. % Valerie Defiesta-Ng Vice President, Regulatory Affairs Experien Group, LLC 224 Airport Parkway, Suite 250 San Jose, California 95110

September 16, 2019

Re: K191676

Trade/Device Name: Intra.Ox 2.0 Handheld Tissue Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: June 21, 2019 Received: June 24, 2019

Dear Valerie Defiesta-Ng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Neil R.P. Ogden, MS Assistant Director, THT4A4 DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K191676

Device Name Intra.OxTM 2.0 Handheld Tissue Oximeter

Indications for Use (Describe)

The Intra. Oxim 2.0 Handheld Tissue Oximeter is intended to non-invasively estimate the percent oxygen saturation (SKO2) in a volume of tissue.

The Intra. OxTM 2.0 Handheld Tissue Oximeter is indicated for use in monitoring patients during circulatory or perfusion examinations.

The Intra. Oxim 2.0 Handheld Tissue Oximeter is intended to be used by physicians, surgeons, nurses, or other skilled users in a medical environment.

The Intra. OxTM 2.0 Handheld Tissue Oximeter should only be used on adult patients.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Notification K_191676

### GENERAL INFORMATION [807.92(a)(1)]

## Applicant:

ViOptix, Inc. 39655 Eureka Drive Newark, CA 94560-4806 USA Phone: 510-226-5860 Fax: 510-226-5864

#### Contact Person:

Valerie Defiesta-Ng Vice President, Regulatory Affairs Experien Group, LLC 224 Airport Parkway, Suite 250 San Jose, CA, 95110 USA

Date Prepared: June 21, 2019

# DEVICE INFORMATION [807.92(a)(2)]

Trade Name:

Intra.Ox™ 2.0 Handheld Tissue Oximeter

## Generic/Common Name:

Oximeter, Tissue Saturation

#### Classification:

21 CFR§870.2700

Product Code:

MUD

{4}------------------------------------------------

# PREDICATE DEVICE(S) [807.92(a)(3)]

Intra.Ox™ Handheld Tissue Oximeter (K163472)

### DEVICE DESCRIPTION [807.92(a)(4)]

The ViOptix Intra.Ox 2.0 Handheld Tissue Oximeter is a sterile, cordless, battery-powered device that non-invasively estimates the percent oxygen saturation (StO2) in a volume of tissue. The device includes three components and an accessory:

- . Main Unit: a re-usable module consists of light sources, detectors, and processing electronics to convert measurements of reflected light into an estimate of StO2;
- . Sheath: a single-use, sterile Sheath placed around the Main Unit during device use (provided in the Disposable Kit);
- Battery Pack: a single-use battery (provided in the Disposable Kit) that is paired with the . Main Unit to provide power; and
- . Quality Control (QC) Target: a single-use accessory (provided in the Disposable Kit) that provides an optical check once the Sheath is placed around the Main Unit.

The device uses spatially resolved optical measurements at five wavelengths. The device performs measurements on the patient by direct physical contact to the patient's tissue and displays the StO2 estimate on the built-in screen.

The device shares the same Indications for Use and the same technological characteristics as the predicate device, including Principle of Operation, StO2 measured parameters, accuracy and range, energy delivered and power source.

#### INDICATIONS FOR USE [807.92(a)(5)]

The Intra.Ox™ 2.0 Handheld Tissue Oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue.

The Intra.Ox" 2.0 Handheld Tissue Oximeter is indicated for use in monitoring patients during circulatory or perfusion examinations.

The Intra.Ox" 2.0 Handheld Tissue Oximeter is intended to be used by physicians, surgeons, nurses, or other skilled users in a medical environment.

The Intra.Ox™ 2.0 Handheld Tissue Oximeter should only be used on adult patients.

#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(a)(6)]

The technological characteristics of the Intra.Ox 2.0 Handheld Tissue Oximeter are substantially equivalent to the predicate device, the Intra.Ox Handheld Tissue Oximeter (K163472). The table below lists the technological characteristics of the proposed device and predicate device and provides rationale to support a determination of substantial equivalence. Any differences between the devices do not raise any different questions of safety or efficacy.

{5}------------------------------------------------

# 510(k) SUMMARY

|                                   | Intra.Ox Handheld Tissue<br>Oximeter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Intra.Ox 2.0 Handheld issue<br>Oximeter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Substantial<br>Equivalence Rationale                                                                                                                                                                                            |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature                           | Primary Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                 |
| 510(k) Number                     | K163472                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | K191676                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | --                                                                                                                                                                                                                              |
| Indications for Use               | The Intra.Ox Handheld Tissue<br>Oximeter is intended to<br>non-invasively estimate the percent<br>oxygen saturation (StO2) in a<br>volume of tissue.<br>The Intra.Ox Handheld Tissue<br>Oximeter is indicated for use in<br>monitoring patients during<br>circulatory or perfusion<br>examinations.<br>The Intra.Ox Handheld Tissue<br>Oximeter is intended to be used by<br>physicians, surgeons, nurses, or<br>other skilled users in a medical<br>environment.<br>The Intra.Ox Handheld Tissue<br>Oximeter should only be used on<br>adult patients. | The Intra.OxTM 2.0 Handheld Tissue<br>Oximeter is intended to<br>non-invasively estimate the percent<br>oxygen saturation (StO2) in a<br>volume of tissue.<br>The Intra.OxTM 2.0 Handheld Tissue<br>Oximeter is indicated for use in<br>monitoring patients during<br>circulatory or perfusion<br>examinations.<br>The Intra.OxTM 2.0 Handheld Tissue<br>Oximeter is intended to be used by<br>physicians, surgeons, nurses, or<br>other skilled users in a medical<br>environment.<br>The Intra.OxTM 2.0 Handheld Tissue<br>Oximeter should only be used on<br>adult patients. | Same                                                                                                                                                                                                                            |
| Measured<br>Parameters            | Tissue oxygen saturation<br>(% StO2)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Tissue oxygen saturation<br>(% StO2)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                            |
| Operating Principle               | Spectrophotometric oximetry                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Spectrophotometric oximetry                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                            |
| Energy Delivered                  | Near-infrared light<br>Source: LED chips<br>Wavelengths: 730nm, 760mm,<br>810mm, 845mm, and 895mm                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Near-infrared light<br>Source: LED chips<br>Wavelengths: 730nm, 760mm,<br>810mm, 845mm, and 895mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                            |
| Single Patient Use<br>or Reusable | Main Unit - Single Patient Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Main Unit - Reusable<br>Sheath - Single Patient Use<br>Battery Pack - Single Patient Use<br>Quality Control Target – Single<br>Patient Use                                                                                                                                                                                                                                                                                                                                                                                                                                      | Similar<br>The difference in use of<br>the Main Unit of the<br>Subject Device does not<br>raise different questions<br>related to safety and<br>effectiveness, as<br>demonstrated by<br>verification and<br>validation testing. |
| Power Source                      | Battery powered<br>Battery type: 4 Lithium AA<br>Battery voltage 6V total                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Battery powered<br>Battery type: 4 Lithium AA<br>Battery voltage 6V total                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                            |
| Measurement Range                 | 1-99% StO2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 1-99% StO2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                            |
| Feature                           | Intra.Ox Handheld Tissue<br>Oximeter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Intra.Ox 2.0 Handheld issue<br>Oximeter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Substantial<br>Equivalence Rationale                                                                                                                                                                                            |
| Physical Design                   | Ergonomic hand-held design for<br>range of hand sizes, includes thumb<br>rest, ambidextrous use                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Ergonomic hand-held design for<br>range of hand sizes, includes thumb<br>rest, ambidextrous use<br>Device contains a disposable kit that<br>contains a single-use Battery Pack<br>and sterile single-use disposable<br>sheath.                                                                                                                                                                                                                                                                                                                                                  | Not the same.<br>The difference in<br>physical design does<br>not raise different<br>questions related to<br>safety and<br>effectiveness, as<br>demonstrated by<br>verification and<br>validation testing.                      |
| Dimensions and<br>Weight          | 7in. x 2.9in x 3in.<br>L x W x H                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 7.3" x 2.7" x 3.1"<br>L x W x H                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Not the same<br>The differences in<br>physical dimensions do<br>not raise different<br>questions related to<br>safety and<br>effectiveness, as<br>demonstrated by<br>verification and<br>validation testing.                    |
| Housing Materials                 | Polycarbonate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Polycarbonate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                            |

{6}------------------------------------------------

#### SUBSTANTIAL EQUIVALENCE

The Intra.Ox 2.0 Handheld Tissue Oximeter is substantially equivalent to the predicate device with regard to intended use, Indications for Use, principle of operation and fundamental scientific technology. Any differences in the technological characteristics between the devices do not raise any different questions of safety or effectiveness. Thus, the Intra.Ox 2.0 Handheld Tissue Oximeter is substantially equivalent to the predicate device.

#### PERFORMANCE DATA [807.92(b)]

All necessary bench and clinical testing was conducted on the Intra.Ox 2.0 Handheld Tissue Oximeter to support a determination of substantial equivalence to the predicate device. The tests performed include:

- Biocompatibility .
	- o Cytotoxicity MEM Elution
	- Guinea Pig Maximization Sensitization O
	- Intracutaneous Reactivity Irritation O
	- System Toxicity O
	- Material Mediated Pyrogenicity o
- Mechanical Testing ●
- Sterilization Validation
- Packaging, Shelf-Life and Transportation Testing
- Software Verification and Validation ●
- Heterogeneous Blood Phantom Study ●

{7}------------------------------------------------

- Human Factors Usability Study ●
- Non-Significant Risk Clinical Study

The collective performance testing demonstrates that the Intra.Ox 2.0 Handheld Tissue Oximeter does not raise any different questions of safety or effectiveness when compared to the predicate device. The results of performance testing demonstrate that the Intra. Ox 2.0 Handheld Tissue Oximeter performs as intended.

# [807.92(b)(1)]Non-clinical Testing Summary:

In a heterogenous phantom study, measurements were made using the Intra.Ox 2.0 Handheld Tissue Oximeter in a heterogeneous phantom prepared with swine whole blood using an Intralipid solution to mimic tissue scattering and compared to a "gold standard" blood cooximeter. The results show that the Intra.Ox 2.0 Handheld Tissue Oximeter is substantially equivalent to the predicate device in limits of agreement to the gold standard, as well as 95% confidence intervals in slope and intercept. These data support the conclusion that the Intra.Ox 2.0 Handheld Tissue Oximeter provides as good or better agreement to the gold standard over the clinically relevant range than the predicate Intra.Ox Handheld Tissue Oximeter.

## [807.92(b)(2)|Clinical Testing Summary:

Clinical performance was determined by measuring tissue oxygen saturation (StO2) with the Intra.Ox 2.0 Handheld Tissue Oximeter and the predicate Intra.Ox Handheld Tissue Oximeter during transient ischemic events on healthy human volunteers, which temporarily mimics compromised tissue via a non-significant risk (NSR) IRB-approved study. A total of 18 data sets from 18 subjects, who were near-evenly distributed over age, gender, and skin color as determined by the Fitzpatrick skin type, were analyzed.

There was excellent agreement in shape of the deoxygenation curve observed during ischemic events between the Intra.Ox 2.0 Handheld Tissue Oximeter and the predicate device for each analyzed subject. A direct comparison with paired data showed good agreement considering the physiological variances inherent between measurement sites.

Figure 1 shows the similarities in values for healthy tissue (65% [Intra.Ox 2.0 Handheld Tissue Oximeter] and 61% [predicate Intra.Ox]) and mean desaturation dynamic range of 27 percentage points (both devices) agrees well with literature-reported values of skin and muscle transient ischemia.

Importantly, the Intra.Ox 2.0 Handheld Tissue Oximeter and predicate Intra.Ox devices measure similar ranges of StO2 values for both healthy and compromised tissue, thus demonstrating substantial equivalence.

{8}------------------------------------------------

Image /page/8/Figure/1 description: This image is a boxplot comparing Intra.Ox 2.0 and Predicate. The y-axis is labeled "StO2%" and ranges from 0 to 100. The x-axis shows two groups, "Healthy" and "Compromised", for each of the two methods, Intra.Ox 2.0 and Predicate. The boxplot shows the distribution of StO2 values for each group and method, with the median, quartiles, and outliers displayed.

Predicate Device for Healthy and Compromised Tissue.

# Performance Standards

Compliance to the following performance standards were followed to conduct the performance tests as listed above.

- . ASTM F2096:2011, Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
- . ISO 10993-1:2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- ISO 10993-2:2006, Biological evaluation of medical devices - Part 2: Animal welfare requirements
- ISO 10993-5:2009, Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- ISO 10993-12:2012, Biological evaluation of medical devices - - Part 12: Sample preparation and reference materials
- ISO 13485:2003, Medical devices - Quality management systems - Requirements for regulatory purposes
- BSI BS EN ISO 14971:2012, Medical devices - Application of risk management to medical devices
- ISTA 2A:2011, Partial Simulation Performance Tests

{9}------------------------------------------------

# CONCLUSIONS [807.92(b)(3)]

The Intra.Ox 2.0 Handheld Tissue Oximeter and the predicate Intra.Ox Handheld Tissue Oximeter have the same Indications for Use and utilize the same technological characteristics which do not raise different questions of safety or effectiveness. The performance data show that the Intra.Ox 2.0 Handheld Tissue Oximeter is safe and performs as intended and as such demonstrates substantial equivalence compared to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K191676](https://fda-staging.innolitics.com/device/K191676)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).