← Product Code [NBW](/productcode/NBW) · K191657

# Bioland Blood Glucose Monitoring System (K191657)

_Bioland Technology, Ltd. · NBW · Mar 20, 2020 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K191657

## Device Facts

- **Applicant:** Bioland Technology, Ltd.
- **Product Code:** [NBW](/productcode/NBW.md)
- **Decision Date:** Mar 20, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Pediatric

## Indications for Use

The Bioland Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The Bioland Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. It is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Bioland Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes. The Bioland Blood Glucose Monitoring System is not for use in neonates.

## Device Story

System comprises blood glucose meter, test strips, lancing device, and control solution. Operates as electrochemical biosensor; carbon electrodes detect current generated by reaction of capillary blood glucose with glucose oxidase on test strip. Meter measures current strength proportional to glucose concentration; displays result. Used at home by patients with diabetes for self-monitoring. Data can be transferred via Bluetooth to connected equipment. Provides quantitative glucose measurement to aid diabetes management; does not diagnose or screen for diabetes. Benefits patient by enabling convenient, rapid home monitoring of blood glucose levels.

## Clinical Evidence

Lay-user study with 351 participants demonstrated system accuracy. 100% of results were within ±20% of YSI 2300 comparator across the measuring range. Accuracy at extreme glucose values (<80 mg/dL and >250 mg/dL) confirmed via spiked/glycolyzed samples (n=444), with 99.48% and 100% of results within ±20% respectively. Bench testing included precision (CV 1.9-6.7%), linearity (R² > 0.999), interference testing (24 substances), hematocrit (20-60%), altitude (up to 10,744 ft), and environmental robustness.

## Technological Characteristics

Electrochemical biosensor with carbon electrodes; glucose oxidase enzyme. Meter dimensions 90x54x13 mm; weight 48g. Power: 2x AAA alkaline batteries (DC 3V). Connectivity: Bluetooth. Materials: ABS (housing), PMMA (test slot), PC (LCD lens). Automatic temperature compensation via built-in thermistor. Measuring range 40-600 mg/dL.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Bioland G-423 Blood Glucose Monitoring System ([K113077](/device/K113077.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT

## I Background Information:

A 510(k) Number
K191657

B Applicant
Bioland Technology Ltd.

C Proprietary and Established Names
Bioland Blood Glucose Monitoring System

D Regulatory Information
|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  NBW | Class II | 21 CFR 862.1345 - Glucose Test System | CH - Clinical Chemistry  |

## II Submission/Device Overview:

A Purpose for Submission:
New device

B Measurand:
Glucose in fresh capillary whole blood samples obtained from the fingertip

C Type of Test:
Quantitative amperometric assay (glucose oxidase)

{1}

III Intended Use/Indications for Use:

A Intended Use(s):

See Indications for Use below.

B Indication(s) for Use:

The Bioland Blood Glucose Monitoring System is comprised of Bioland Blood Glucose Meter and Bioland Blood Glucose Test Strips.

The Bioland Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The Bioland Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. It is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Bioland Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes. The Bioland Blood Glucose Monitoring System is not for use in neonates.

C Special Conditions for Use Statement(s):

- OTC - Over The Counter
- This device is not intended for use in healthcare or assisted-use settings such as hospitals, physician's offices, or long-term care facilities because it has not been cleared by FDA for use in these settings, including for routine assisted testing or as part of glycemic control procedures. Use of this device on multiple patients may lead to transmission of Human Immunodeficiency Virus (HIV), Hepatitis C Virus (HCV), Hepatitis B Virus (HBV), or other bloodborne pathogens.
- The blood monitoring system is not intended for use on neonates.
- Altitudes above 10744 feet (3275 meters) may cause inaccurate results.
- Not for use for patients in a hyperglycemic-hyperosmolar state, with or without ketosis.
- Not for use on critically ill patients.
- Not to be used for patients who are dehydrated, hypertensive, hypotensive or in shock.
- Very low (less than 20%) or very high (more than 60%) red blood cell count (hematocrit) can lead to incorrect test results. If you do not know your hematocrit level, please consult your health care provider.
- High temperature (more than 104°F) and low temperature (less than 50°F) may lead to incorrect test results.
- High humidity (more than 85%) and low humidity (less than 10%) may lead to incorrect test results.

D Special Instrument Requirements:

Bioland Blood Glucose Meter

{2}

IV Device/System Characteristics:

A Device Description:

The Bioland Blood Glucose Monitoring System consists of the Bioland Blood Glucose Meter, Bioland Blood Glucose Test Strips, lancing device, sterile lancets and Bioland Glucose Control Solutions (Levels 1, 2 and 3).

Bioland Blood Glucose Test Strips and Bioland Glucose Control Solutions (Levels 1, 2 and 3) are not included in the kit package and should be purchased separately.

B Principle of Operation:

The Bioland Blood Glucose Monitoring System quantitatively measures the amount of glucose in whole blood from the fingertip using amperometric technology. The test is based on the measurement of electrical current generated by the reaction of capillary whole blood glucose with glucose oxidase and a mediator on the test strip. The detected current signal is proportional to the glucose concentration in the sample, which is then calculated and displayed on the meter.

C Instrument Description Information:

|  Modes of Operation | Yes | No  |
| --- | --- | --- |
|  Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? | ☑ | ☐  |
|  Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? | ☑ | ☐  |
|  Software |  |   |
|  FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types. | ☑ | ☐  |

1. Instrument Name:

Bioland Blood Glucose Meter.

2. Specimen Identification:

There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.

3. Specimen Sampling and Handling:

The glucose test is intended to be used with capillary whole blood from the finger. The whole blood sample is applied directly to the test strip by capillary action.

4. Calibration:

The meter does not require calibration or coding by the user.

{3}

5. Quality Control:

Bioland Glucose Control Solutions are aqueous solutions containing glucose and are available at three levels (Level 1, 2 and 3). Instructions on how to order control solutions, and when to perform a control solution test are included in the user guide. The control solution readings are not included in the average of the patient results when the measurements are performed in the 'CTL' measurement mode. The user is cautioned not to use the meter and to contact the customer support if the control result falls outside these ranges the range printed on the test strip vial label.

V Substantial Equivalence Information:

A Predicate Device Name(s):

Bioland G-423 Blood Glucose Monitoring System

B Predicate 510(k) Number(s):

K113077

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K191657 | K113077  |
| --- | --- | --- |
|  Device Trade Name | Bioland Blood Glucose Monitoring System | Bioland G-423 Blood Glucose Monitoring System  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | Quantitative measurement of glucose in capillary whole blood from the fingertip. It is intended for use by people with diabetes mellitus at home (over-the counter) as an aid in monitoring the effectiveness of diabetes control program. | Same  |
|  Test Principle | Amperometric glucose biosensor | Same  |
|  Enzyme | Glucose Oxidase | Same  |
|  Specimen Type | Capillary whole blood from fingertip | Same  |

{4}

|  Device & Predicate Device(s): | K191657 | K113077  |
| --- | --- | --- |
|  Detecting Range | 40 ~ 600 mg/dL | Same  |
|  General Device Characteristic Differences |  |   |
|  Test Strip Calibration | No Code function, no need to calibrate the meter | Use Code test strip to calibrate the meter  |
|  Sample volume | ≥ 0.7 μL | ≥ 1.8 μL  |
|  Measuring Time | 6 s | 10 s  |

# VI Standards/Guidance Documents Referenced:

FDA Guidance Document; Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use, issued on October 11, 2016

IEC 61010-1:2010, Safety requirements for electrical equipment for measurement, control and laboratory use – Part 1: General Requirements.

IEC 60601-1-2: 2007, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.

CLSI, EP06-A: 2003, Evaluation of the linearity of quantitative analytical Methods: A Statistical Approach.

CLSI EP07-A2: 2005, Interference Testing in Clinical Chemistry, Approved Guideline, 2nd Edition.

# VII Performance Characteristics (if/when applicable):

# A Analytical Performance:

# 1. Precision/Reproducibility:

Within-run precision studies were performed using venous whole blood samples adjusted to five glucose concentration levels: 40.3, 89.9, 131.5, 199.3 and  $323.8\mathrm{mg / dL}$ . Each glucose level was analyzed in replicates of 10, using 3 lots of test strips and 10 meters for a total of 300 measurements per each glucose level. Results are summarized below:

|  Glucose Level (mg/dL) | Lot | N | Mean (mg/dL) | SD (mg/dL) | CV (%)  |
| --- | --- | --- | --- | --- | --- |
|  Level I (30-50) | 1 | 100 | 40.0 | 2.7 | 6.6  |
|   |  2 | 100 | 40.2 | 2.7 | 6.7  |
|   |  3 | 100 | 40.0 | 2.5 | 6.3  |

{5}

Intermediate (day-to-day) precision was evaluated using 5 levels of glucose control solutions with concentrations of 41.9, 94.0, 131.1, 199.2 and 325.3 mg/dL. For each concentration, 10 measurements were taken over 10 days using 10 meters and 3 lots of test strips, for a total of 300 measurements per sample. Results are summarized below:

|  Glucose Level (mg/dL) | Lot | N | Mean (mg/dL) | SD (mg/dL) | CV (%)  |
| --- | --- | --- | --- | --- | --- |
|  Level I (30-50) | 1 | 100 | 40.1 | 2.0 | 5.1  |
|   |  2 | 100 | 40.1 | 2.0 | 5.1  |
|   |  3 | 100 | 40.0 | 1.9 | 4.9  |
|   |  Combined | 300 | 40.1 | 2.0 | 5.0  |
|  Level II (51-110) | 1 | 100 | 130.8 | 4.6 | 3.5  |
|   |  2 | 100 | 130.0 | 4.7 | 3.6  |
|   |  3 | 100 | 130.7 | 4.6 | 3.5  |
|   |  Combined | 300 | 130.5 | 4.6 | 3.6  |
|  Level V (251-400) | 1 | 100 | 200.0 | 7.2 | 3.6  |
|   |  2 | 100 | 199.9 | 6.5 | 3.3  |
|   |  3 | 100 | 201.3 | 7.0 | 3.5  |
|   |  Combined | 300 | 200.4 | 6.9 | 3.4  |
|  Level VI (151-250) | 1 | 100 | 322.1 | 11.1 | 3.4  |
|   |  2 | 100 | 323.2 | 10.9 | 3.4  |
|   |  3 | 100 | 320.9 | 10.1 | 3.1  |
|   |  Combined | 300 | 322.1 | 10.7 | 3.3  |

Intermediate (day-to-day) precision was evaluated using 5 levels of glucose control solutions with concentrations of 41.9, 94.0, 131.1, 199.2 and 325.3 mg/dL. For each concentration, 10 measurements were taken over 10 days using 10 meters and 3 lots of test strips, for a total of 300 measurements per sample. Results are summarized below:

|  Glucose Level (mg/dL) | Lot | N | Mean (mg/dL) | SD (mg/dL) | CV (%)  |
| --- | --- | --- | --- | --- | --- |
|  Level I (30-50) | 1 | 100 | 40.1 | 2.0 | 5.1  |
|   |  2 | 100 | 40.1 | 2.0 | 5.1  |
|   |  3 | 100 | 40.0 | 1.9 | 4.9  |
|   |  Combined | 300 | 40.1 | 2.0 | 5.0  |
|  Level II (51-110) | 1 | 100 | 90.5 | 2.5 | 2.8  |
|   |  2 | 100 | 90.3 | 2.7 | 3.0  |
|   |  3 | 100 | 89.8 | 2.5 | 2.8  |
|   |  Combined | 300 | 90.2 | 2.6 | 2.9  |
|  Level III (111-150) | 1 | 100 | 116.8 | 3.2 | 2.8  |
|   |  2 | 100 | 116.9 | 3.2 | 2.8  |
|   |  3 | 100 | 117.3 | 2.9 | 2.5  |
|   |  Combined | 300 | 117.0 | 3.1 | 2.7  |
|  Level IV (151-250) | 1 | 100 | 193.0 | 4.8 | 2.5  |
|   |  2 | 100 | 192.4 | 4.6 | 2.4  |
|   |  3 | 100 | 192.1 | 4.0 | 2.1  |
|   |  Combined | 300 | 192.5 | 4.5 | 2.3  |
|  Level V (251-400) | 1 | 100 | 315.6 | 6.3 | 2.0  |
|   |  2 | 100 | 314.2 | 6.2 | 2.0  |
|   |  3 | 100 | 315.0 | 6.0 | 1.9  |
|   |  Combined | 300 | 314.9 | 6.2 | 2.0  |

{6}

2. Linearity:

Linearity was evaluated using 10 levels of venous whole blood glucose concentrations (24.9, 40.3, 80.3, 149.8, 249.0, 340.0, 409.5, 508.8, 580.3 and 620.8 mg/dL) as determined by the comparator method YSI 2300. Each glucose level was measured in replicates of 10 using 10 glucose meters and 3 lots of test strips, and the values from the Bioland Blood Glucose Monitoring System were compared to the values obtained from the comparator method, YSI 2300. Linear regression analysis of the results yielded the following results:

Test strip lot 1:  $y = 0.9934x + 2.1362$  $R^2 = 0.9997$

Test strip lot 2:  $y = 1.0042x + 0.0776$  $R^2 = 0.9996$

Test strip lot 3:  $y = 1.0007x - 0.0366$  $R^2 = 0.9999$

The results of the linearity study support the sponsor's claimed glucose measurement range of  $40 - 600\mathrm{mg / dL}$ . If a sample is less than  $40\mathrm{mg / dL}$  glucose, the result is flagged by the meter as "Lo". If a sample exceeds  $600\mathrm{mg / dL}$  glucose, the result is flagged by the meter as "Hi". The "Lo" and "Hi" functions were validated and demonstrated to function as intended.

3. Analytical Specificity/Interference:

To assess potential interferences, a study was conducted using venous whole blood samples adjusted to three glucose levels, as measured by the comparator method, YSI 2300: 50-70, 110-130 and  $225 - 270\mathrm{mg / dL}$ . Each of these samples was divided into a test pool and a control pool, with each of the potential endogenous and exogenous interfering substances added to the test pool. The sponsor tested 24 exogenous and endogenous substances. The study was conducted using 3 lots of test strips and 10 meters, for a total of 30 replicates per test sample. The difference between test sample and control sample meter results was calculated. The table below shows the highest concentration of substance tested at which no significant interference (as defined by the sponsor as  $\pm 10\mathrm{mg / dL}$  for glucose concentrations  $&lt; 100\mathrm{mg / dL}$  and  $\pm 10\%$  for glucose concentrations  $&gt;100\mathrm{mg / dL}$ ) was observed:

|  Potential Interfering Substance | Highest Concentration Without Significant Interference (mg/dL)  |
| --- | --- |
|  Acetaminophen | 8.0  |
|  Ascorbic Acid | 5.0  |
|  Unconjugated Bilirubin | 90  |
|  Cholesterol | 500  |
|  Creatinine | 5.0  |
|  Dopamine | 2.0  |
|  EDTA | 360  |
|  Galactose | 900  |
|  Gentisic Acid | 5.0  |
|  Reduced Glutathione | 53  |
|  Hemoglobin | 500  |
|  Heparin | 8000 U/dL  |
|  Ibuprofen | 50  |

{7}

|  Potential Interfering Substance | Highest Concentration Without Significant Interference (mg/dL)  |
| --- | --- |
|  Icodextrin | 13  |
|  L-Dopa | 10  |
|  Maltose | 900  |
|  Mannitol | 3.0  |
|  Methyldopa | 25  |
|  Salicylic Acid | 60  |
|  Tolazamide | 100  |
|  Tolbutamide | 400  |
|  Triglycerides | 2000  |
|  Uric Acid | 8.0  |
|  Xylose | 100  |

The sponsor included the following statements in the labeling:

- Acetaminophen in your blood (&gt; 8.0 mg/dL) might affect the reliability of your blood glucose results. If you are taking Tylenol, or other acetaminophen containing drugs, your glucose results may not be reliable. If you are unsure, then ask your healthcare professional.
- If you have a disease or condition that elevates your blood uric acid level (&gt; 8.0 mg/dL), such as gout, your blood glucose results may not be reliable. If you are unsure, then ask your healthcare professional.
- Xylose: Do not test blood glucose during or soon after a xylose absorption test. Xylose in the blood can give falsely elevated results.

4. Assay Reportable Range:

The reportable range is 40 - 600 mg/dL.

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

The Bioland Blood Glucose Monitoring System is traceable to the NIST (SRM) 917A glucose reference material. A method comparison was performed using the candidate device and YSI 2300 as the comparator method.

Test Strip Stability

Test strip stability was assessed using real-time stability studies. Testing protocols and acceptance criteria were reviewed and found to be acceptable. The sponsor claims a shelf life stability of 24 months and an open vial stability of 90 days at the recommended storage conditions of 39.2-104 °F (4-40°C) and relative humidity (RH) of 10%-85%.

6. Detection Limit:

Not applicable.

{8}

7. Assay Cut-Off:
Not applicable.

8. Accuracy (Instrument):
Not applicable.

9. Carry-Over:
Not applicable.

## B Comparison Studies:

1. Method Comparison with Predicate Device:
See lay user study below in section C3.

2. Matrix Comparison:
Not applicable. The device is only intended for use with fresh capillary whole blood from a fingerstick.

## C Clinical Studies:

1. Clinical Sensitivity:
Not applicable.

2. Clinical Specificity:
Not applicable.

3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
To assess the performance of Bioland Blood Glucose Monitoring System in the hands of the intended users, the sponsor performed a study with 351 lay-user participants. The users were responsible for obtaining their own fingertip capillary sample and performing a blood glucose test according to the instructions and training materials routinely provided with the system. A total of 100 Bioland meters and 3 lots of Bioland test strips were used. Results were analyzed by comparing the blood glucose results obtained by the lay users with the Bioland Test System against results obtained with the laboratory-based comparator method (YSI 2300 glucose analyzer). The glucose concentrations in the samples ranged between 50.7 - 384 mg/dL, as measured by the YSI 2300. The set included 57 native samples with glucose concentration &lt; 80 mg/dL and 16 samples with glucose concentration &gt; 250 mg/dL. The results obtained with the candidate device were compared to the results obtained with the comparator method, YSI 2300, and are summarized below:

{9}

Glucose concentration &lt; 75 mg/dL

|  Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL | Within ± 20 mg/dL  |
| --- | --- | --- | --- |
|  26/40
(65.00%) | 38/40
(95.00%) | 40/40
(100.00%) | 40/40
(100.00%)  |

Glucose concentration ≥ 75 mg/dL

|  Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20%  |
| --- | --- | --- | --- |
|  154/311
(44.52%) | 260/311
(83.60%) | 308/311
(99.04%) | 311/311
(100.00%)  |

Combined glucose concentrations across the entire measuring range:

|  Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20%  |
| --- | --- | --- | --- |
|  173/351
(49.29%) | 293/351
(83.48%) | 346/351
(98.58%) | 351/351
(100.0%)  |

Results of the linear regression analysis:

|  Slope | Y-intercept | R²  |
| --- | --- | --- |
|  1.0289 | -1.7316 | 0.9716  |

A Flesch-Kincaid readability assessment was conducted, and the results demonstrate that the Meter Manual and Test Strip Insert were written at an 8th grade level or less.

The study participants answered questions in two questionnaires for the investigator to evaluate their understanding of the system after participating in the lay-user study. The results demonstrated that the participants were able to understand the labeling and conduct the testing on their own.

## Accuracy at extreme glucose values:

To further assess the performance of the Bioland Blood Glucose Monitoring System at the extreme upper and lower ends of the claimed measuring range, the sponsor altered 444 capillary whole blood samples, by spiking or allowing samples to glycolyze, to achieve glucose concentrations below 80 mg/dL (216 samples; 32.1 – 79.9 mg/dL according to YSI 2300) and above 250 mg/dL (250 samples; 254.5 – 598.0 mg/dL according to YSI 2300). Results obtained with the candidate device using three test strip lots and were compared the results obtained with the comparator method, YSI 2300. The results are summarized below:

Glucose concentrations &lt; 80 mg/dL:

|  Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20%  |
| --- | --- | --- | --- |
|  94/194
(48.45%) | 159/194
(81.96%) | 187/194
(96.39%) | 193/194
(99.48%)  |

{10}

Glucose concentrations &gt; 250 mg/dL:

|  Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20%  |
| --- | --- | --- | --- |
|  141/250
(56.40%) | 212/250
(84.80%) | 249/250
(99.60%) | 250/250
(100.00%)  |

## D Clinical Cut-Off:

Not applicable.

## E Expected Values/Reference Range:

The labeling includes the following expected glucose values for people without diabetes:

|  Time of day | People without diabetes  |
| --- | --- |
|  Before breakfast (fasting) | < 100 mg/dL  |
|  Two hours after meals | < 140 mg/dL  |

Source: American Diabetes Association, Standard of Medical Care in Diabetes 2020, Vol. 43.

## F Other Supportive Instrument Performance Characteristics Data:

1) Hematocrit Study:

The Bioland Blood Glucose Meter System was assessed at 20%, 25%, 30%, 35%, 40%, 45%, 50%, 55% and 60% hematocrit levels, and at 5 glucose concentrations (30-50, 51-110, 111-150, 151-250 and 251-400 mg/dL). Each sample was tested in replicates of 10 using 10 meters and 3 lots of test strips, for a total of 30 replicates per sample. The values were compared with the glucose measurements obtained from YSI 2300 comparator method. The % bias of the Bioland Blood Glucose Monitoring System relative to YSI 2300 demonstrated adequate performance to support the claimed hematocrit range of 20-60%.

2) Altitude study:

To assess the effect of altitude, venous whole blood samples adjusted to 5 glucose concentrations covering the measuring range (30-50, 51-110, 111-150, 151-250 and 251-400 mg/dL) were tested at elevations of 298, 4790 and 10744 feet above sea level (0 feet) in replicates of 10, using 10 meters and 3 lot of test strips, for a total of 30 replicates per samples). Values measured by the candidate device were compared to the comparator method YSI 2300. The results support the claim that the Bioland Blood Glucose Monitoring System can be operated at altitudes of up to 10,744 ft.

3) Operating conditions (temperature, humidity):

The Bioland Blood Glucose Monitoring System was tested at different temperature and humidity conditions to assess the effect of operation environment on the meter's performance. Temperatures ranging from 45-110°F (7.2-43.3°C) and relative humidity from 10-90% were tested. Meter results were compared to the YSI 2300 comparator

{11}

method. Four temperature and humidity combinations were tested including low temperature/low humidity, low temperature/high humidity, high temperature/low humidity and high temperature/high humidity. Each of 3 venous whole blood glucose levels (30-50; 111-150 and 251-400 mg/dL) were tested by 10 meters, using 3 lots of test strips, for a total of 10 replicates per sample. Values measured by the Bioland Blood Glucose Monitoring System were compared to the comparator method YSI 2300 method. The study results support the labeled operating conditions claim of 50-104°F (10-40°C) and 15-85% RH.

4) Sample volume study:
To verify the test strip minimum sample volume requirement (0.7 µL), venous whole blood samples with the following glucose concentrations levels were used: 30-50 mg/dL, 111-150 mg/dL and 251-400 mg/dL. Sample volumes of 0.6, 0.7 and 0.8 µL were tested using 10 meters and 3 lots of test strips for a total of 30 measurements per sample volume per glucose level. Values obtained with the candidate device were compared to values obtained using the comparator method (YSI 2300). When the sample volume is lower than 0.7 µL the measurement procedure will not start and the meter gives an error message. This feature was validated and was shown to function as intended. Results support the claimed minimum sample volume of 0.7 µL.

5) Flex Studies
Intermittent sampling, sample perturbation, testing with used test strips and a variety of mechanical/durability testing (i.e., vibration test, drop test) was completed by the sponsor. The testing performed demonstrated that the Bioland Blood Glucose Monitoring System is robust to these expected use scenarios.

6) EMC
The sponsor provided documentation certifying that acceptable electrical safety and electromagnetic compatibility (EMC) testing had been performed and the system was found to be compliant.

7) Cleaning and disinfection robustness evaluation (Infection control studies)
The device is intended for a single-patient use only. Disinfection efficacy studies were performed on the materials comprising the meter by an outside commercial testing laboratory, demonstrating removal of the HBsAg antigen with the chosen disinfectant, Clorox Germicidal Wipes (EPA Registration Number 67619-12). Robustness studies were also performed by the sponsor demonstrating that there was no change in performance or external materials of the meter after 3,650 cycles of cleaning and disinfection using the chosen disinfectant. The robustness studies were designed to simulate cleaning and disinfection over the 5-year single-patient use life of the meter. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.

8) Test strip lot release criteria
The test strip lot release protocols and criteria were reviewed and found to be acceptable.

{12}

VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K191657](https://fda-staging.innolitics.com/device/K191657)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
