Direct Applications: - Light cured composite restorations - Light cured compomers - Cavity sealing prior to placement of indirect restorations - Composite repair of porcelain and hybrid ceramic restorations - Sealing exposed root surfaces which are causing hypersensitivity - Core build ups using light cure/dual cure composites - Protective varnish for glass ionomer (GI) filling Indirect Applications: - Cementation indirect restorations
Device Story
Zipbond is a single-component, (meth)acrylate-based dental adhesive used by dentists in clinical settings. It functions as a bonding agent for direct and indirect dental restorations. The device is applied to enamel, dentin, composites, or cements and requires activation via an external LED light curing device. It supports three pre-treatment techniques: self-etch, selective-etch, and total-etch. By creating a bond between the tooth structure and restorative materials, it facilitates the placement of fillings, core build-ups, and indirect restorations, and helps treat hypersensitivity by sealing exposed root surfaces. The inclusion of sodium fluoride provides a fluoride source. The clinician applies the liquid adhesive, then light-cures it before placing the restorative material. This process ensures stable adhesion, improving the longevity and success of dental restorations.
Clinical Evidence
Bench testing only. Performance verified via in-vitro testing including pH measurement and shear bond strength on enamel and dentin per ISO 29022:2013. Biocompatibility evaluated per ISO 10993-1 and ISO 7405. Results demonstrate performance equivalent to the predicate.
Technological Characteristics
Single-component (meth)acrylate-based liquid adhesive. Composition: functional monomers, photoinitiators, stabilizers, silica, ethanol, water, sodium fluoride. Energy source: external LED light curing device. Application: self-etch, selective-etch, or total-etch techniques. Sterilization: not applicable (non-sterile dental material).
Indications for Use
Indicated for direct restorations (composite/compomer), cavity sealing, repair of porcelain/hybrid ceramic restorations, root surface desensitization, core build-ups, protective varnish for glass ionomer, and cementation of indirect restorations.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 2, 2019
SDI Limited Ray Cahill Chief Quality and Compliance Officer 3-15 Brunsdon Street Bayswater, 3153 AU
Re: K191656
Trade/Device Name: ZipbondTM Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: September 2, 2019 Received: September 4, 2019
Dear Ray Cahill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
#### K191656
Device Name
ZIPBOND™
Indications for Use (Describe)
Direct Applications:
- · Light cured composite restorations
- · Light cured compomers
- · Cavity sealing prior to placement of indirect restorations
- · Composite repair of porcelain and hybrid ceramic restorations
- · Sealing exposed root surfaces which are causing hypersensitivity
- · Core build ups using light cure/dual cure composites
- · Protective varnish for glass ionomer (GI) filling
Indirect Applications:
- · Cementation indirect restorations
| Type of Use ( <i>Select one or both, as applicable</i> ) |
|------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
PSC Publishing Services (301) 443-6740 EF
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary
# For
# Zipbond™
#### 1. Submitter Information:
SDI Limited 3-15 Brunsdon Street Bayswater Victoria 3153 Australia
| Contact Person: | Ray Cahill |
|-----------------|-----------------------|
| Phone: | +61 3 8727 7111 |
| Fax: | +61 3 8727 7222 |
| Email: | ray.cahill@sdi.com.au |
Date Prepared: 26 November 2019
### 2. Device Details
| Proprietary Name: | Zipbond™ |
|--------------------|---------------------------|
| Common name: | Dental bonding agent |
| Regulation Name: | Resin tooth bonding agent |
| Regulation Number: | 21 CFR 872.3200 |
| Product Code: | KLE |
| Regulatory Class: | II |
### 3. Predicate device:
| Predicate Device name | 510(k) | Company name |
|----------------------------------------------------------------------|---------|--------------|
| Scotchbond Universal Adhesive<br>(named as Adhesive EXL 759 in 510k) | K110302 | 3M Espe AG |
### 4. Device Description:
Zipbond is a single-component, (meth)acrylate-based material which functions as a combined etch-and-rinse (total-etch), self-etch and selective-etch dental adhesive. It offers an adhesive application technique for both direct indications and bonds to enamel, dentin, composites, cements.
When used with a LED light curing device, Zipbond can be used with self-cure, dual-cure and light cure resin cements, composites and core-build up materials. Zipbond provides the operator with three choices of enamel and dentin pre-treatment:
- . Self-etch mode: no phosphoric acid etching required prior to application
{4}------------------------------------------------
- Selective-etch mode: selective phosphoric acid etching to enamel
- Total-etch mode: phosphoric acid etching of both enamel and dentin
#### 5. Indications of use:
Direct Applications
- Light cured composite restorations
- Light cured compomers
- Cavity sealing prior to placement of indirect restorations
- Composite repair of porcelain and hybrid ceramic restorations
- Sealing exposed root surfaces which are causing hypersensitivity
- · Core build ups using light cure/dual cure composites
- · Protective varnish for glass ionomer (GI) filling
Indirect Applications
- Cementation indirect restorations
#### 6. Substantial equivalence:
#### Technical characteristics:
| Proposed device:<br>Zipbond, 510K - K191656 | Predicate device:<br>Scotchbond Universal<br>(named as Adhesive EXL<br>759, in 510K - K110302) | Differences: |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation number: | Regulation number: | Differences: |
| 21 CFR 872.3200 | 21 CFR 872.3200 | No differences |
| Product code: | Product code: | Differences: |
| KLE | KLE | No differences |
| Regulatory class: | Regulatory class: | Differences: |
| II | II | No differences |
| Regulation name: | Regulation name: | Differences: |
| Resin tooth bonding agent | Resin tooth bonding agent | No differences |
| Common name: | Common name: | Differences: |
| Dental bonding agent | Dental bonding agent | No differences |
| Intended use: | Intended use: | Differences: |
| Universal dental adhesive | Universal dental adhesive | No differences |
| Composition: | Composition: | Differences: |
| Functional monomers<br>(methacrylates),<br>photoinitiators, stabilisers,<br>silica, ethanol, water,<br>sodium fluoride | Functional monomers<br>(methacrylates),<br>photoinitiators, stabilisers,<br>silica, ethanol, water | Minor – Zipbond Universal<br>Adhesive contains sodium<br>fluoride. Scotchbond<br>Universal does not mention<br>in their Instructions of Use<br>that it contains fluoride. |
| Application: | Application: | Differences: |
| Liquid | Liquid | No differences |
| Rx / OTC: | Rx / OTC: | Differences: |
| Rx | Rx | No differences |
| Indications for Use<br>statement: | Indications for Use<br>statement: | Differences: |
| Direct Applications:<br>• Light cured composite<br>restorations<br>• Light cured compomers<br>• Cavity sealing prior to<br>placement of indirect<br>restorations<br>• Composite repair of<br>porcelain and hybrid<br>ceramic restorations<br>• Sealing exposed root<br>surfaces which are causing<br>hypersensitivity<br>• Core build ups using light<br>cure/dual cure composites<br>• Protective varnish for<br>glass ionomer (GI) filling<br><br>Indirect Applications:<br>• Cementation indirect<br>restorations | • All classes of fillings<br>(according to Black) with<br>light-curing composite or<br>compomer filling materials<br>• Cementation of indirect<br>restorations when combined<br>with RelyX Ultimate<br>Adhesive Resin Cement,<br>manufactured by 3M ESPE<br>• Cementation of veneers<br>when combined with RelyX<br>Veneer Cement,<br>manufactured<br>by 3M ESPE<br>• Bonding of core build-ups<br>made of light-curing<br>composite or core build-up<br>materials<br>• Bonding of dual-cure<br>cements and core build-up<br>materials and self-cure<br>composites when combined<br>with Scotchbond Universal<br>DCA<br>• Repair of composite or<br>compomer fillings<br>• Intraoral repair of<br>composite restorations,<br>porcelain fused to metal, and<br>all-ceramic restorations<br>without extra primer<br>• Root surface<br>desensitization<br>• Sealing of cavities prior to<br>cementation of amalgam<br>restorations<br>• Sealing of cavities and<br>preparation of tooth stumps<br>prior to temporary<br>cementation of indirect<br>restorations<br>• Bonding of fissure sealants<br>• Protective varnish for glass<br>ionomer fillings | Minor – Zipbond is not used<br>with a dual-cure activator,<br>and requires light curing<br>before placement of dual-<br>cure cements, core build up<br>materials and self-cure<br>composites. |
| Indications for Use: | Indications for Use: | Differences: |
| Light cured composite<br>restorations | All classes of fillings<br>(according to Black) with<br>light-curing composite or<br>compomer filling materials | No differences – classes of<br>fillings according to Black<br>include direct restorations |
| Light cured compomers | Cementation of veneers<br>when combined with RelyX<br>Veneer Cement,<br>manufactured<br>by 3M ESPE | |
| | | |
| Cavity sealing prior to<br>placement of indirect<br>restorations | Sealing of cavities prior to<br>cementation of amalgam<br>restorations | No differences – sealing of<br>cavities prior to placement<br>of indirect restorations |
| | Sealing of cavities and<br>preparation of tooth stumps<br>prior to temporary<br>cementation of indirect<br>restorations | |
| Composite repair of<br>porcelain and hybrid<br>ceramic restorations | Intraoral repair of composite<br>restorations, porcelain fused<br>to metal, and<br>all-ceramic restorations<br>without extra primer<br>Repair of composite or<br>compomer fillings | No differences- Zipbond<br>hybrid ceramic restorations<br>include porcelain fused to<br>metal. |
| Sealing exposed root<br>surfaces which are causing<br>hypersensitivity | Root surface desensitization | No differences – sealing<br>root surfaces results in<br>desensitization |
| Core build ups using light<br>cure/dual cure composites | Bonding of core build-ups<br>made of light-curing<br>composite or core build-up<br>materials | Minor indication – predicate<br>device Scotchbond<br>Universal is used with a<br>dual cure activator when<br>bonding with dual-cure<br>cements, core build-up |
| | Bonding of dual-cure<br>cements and core build-up<br>materials and self-cure<br>composites when combined<br>with Scotchbond Universal<br>DCA | materials and self-cure<br>composites, whereas<br>subject device Zipbond is<br>not used with a dual cure<br>activator and is to be prior<br>light cured when bonding to<br>these materials. |
| Protective varnish for glass<br>ionomer (GI) filling | Protective varnish for glass<br>ionomer fillings | No differences |
| Cementation indirect<br>restorations | Cementation of indirect<br>restorations when combined<br>with RelyX Ultimate<br>Adhesive Resin Cement,<br>manufactured by 3M ESPE | No differences – predicate<br>device Scotchbond<br>Universal indication<br>describes the compatibility<br>of the dental adhesive with<br>a specific resin cement,<br>RelyX Ultimate Adhesive<br>Resin Cement.<br>The adhesive bonds the<br>indirect substrate to the<br>prepared tooth cavity. The<br>tooth cavity may<br>incorporate a luting cement,<br>based on the substrate<br>manufacturer's Instructions<br>for Use. |
| Light cured composite | Bonding of fissure sealants…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.