K191632 · Microport Orthopedics, Inc. · LZO · Oct 7, 2019 · Orthopedic
Device Facts
Record ID
K191632
Device Name
PROFEMUR TL2 Stems
Applicant
Microport Orthopedics, Inc.
Product Code
LZO · Orthopedic
Decision Date
Oct 7, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PROFEMUR® TL2 stems are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients. 1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2) inflammatory degenerative joint disease such as rheumatoid arthritis; 3) correction of functional deformity; and, 4) revision procedures where other treatments or devices have failed Titanium plasma spray coatings applied to implant surfaces are intended for uncemented arthroplasty.
Device Story
Monolithic wedge-taper femoral hip stem; used in total hip arthroplasty; implanted by orthopedic surgeons in clinical settings. Device provides structural support for hip joint; restores biomechanics, leg length, and head center offset. Features vertical distal groove for rotational stability; proximal titanium plasma spray coating for uncemented fixation. Offered in three neck variants (Standard, High Offset, Ultra High Offset) to accommodate patient anatomy. Single-use; sterile; gamma-irradiated.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified via Finite Element Analysis (FEA), proximal and distal fatigue evaluation, and range of motion (ROM) testing per ISO 7206-4/6 and ASTM F2996. Bacterial endotoxin testing confirmed compliance with ANSI/AAMI ST72 and USP <161>.
Technological Characteristics
Monolithic forged titanium alloy (ASTM F620) wedge-taper stem. Proximal titanium plasma spray coating (ASTM F1580). Glass-beaded finish on non-coated/non-trunnion surfaces. Vertical distal groove. Sizes 0-12 (Standard/High Offset) and 1-5, 11, 12 (Ultra High Offset). Sterilized via Gamma Irradiation.
Indications for Use
Indicated for skeletally mature patients requiring total hip arthroplasty for pain relief or improved function due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, painful hip dysplasia), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity, or revision procedures.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
October 7, 2019
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
MicroPort Orthopedics Inc. Sneh Pingle Project Manager, Regulatory Affairs 5677 Airline Road Memphis, Tennessee 38002
### Re: K191632
Trade/Device Name: PROFEMUR TL2 Stems Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO Dated: September 9, 2019 Received: September 10, 2019
Dear Sneh Pingle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
FOR Vesa Vuniqi Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K191632
Device Name PROFEMUR® TL2 Stem System
#### Indications for Use (Describe)
PROFEMUR® TL2 stems are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.
1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
2) inflammatory degenerative joint disease such as rheumatoid arthritis;
3) correction of functional deformity; and,
4) revision procedures where other treatments or devices have failed
Titanium plasma spray coatings applied to implant surfaces are intended for uncemented arthroplasty.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Submitted By: | MicroPort Orthopedics Inc.<br>5677 Airline Road<br>Arlington, TN 38002 |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| | Telephone Number: (901) 290-5175 |
| Date of Submission: | June 18, 2019 |
| Contact Person: | Sneh Pingle<br>Project Manager, Regulatory Affairs<br>Telephone number: (901) 867-4656<br>Email: sneh.pingle@ortho.microport.com |
| Name of Device: | PROFEMUR® TL2 Stem |
| Common Name: | Femoral Hip Stem |
| Device Classification Name | 21 CFR 888.3353- Hip joint metal/ceramic/polymer |
| And Reference: | semi-constrained cemented or nonporous uncemented<br>prosthesis |
| Device Class: | Class II |
| Panel Code: | Orthopedics/87 |
| Product Code: | LZO |
| Predicate Device: | PROFEMUR® TL CLASSIC HIP STEM (K123688, SE<br>02/08/2013)<br>This predicate has not been subject to a design-related<br>recall. |
{4}------------------------------------------------
#### Device Description:
The PROFEMUR® TL2 stems are monolithic wedge taper style stems manufactured from forged titanium alloy conforming to ASTM F620. The proximal portion (top third) of the TL2 stem possesses a titanium plasma spray coating, manufactured from pure titanium powder conforming to ASTM F1580. The stems will have a glass-beaded finish everywhere except for the trunnion and plasma spray coating region. The PROFEMUR® TL2 stems also have a vertical groove along most of the distal length of the stem to assist in rotational stability in the canal. The PROFEMUR® TL2 stems are designed for use in uncemented total hip arthroplasty. The PROFEMUR TL2 Stem is offered in 3 different neck variants, Standard (STD), High Offset (HO), and Ultra High Offset (UHO). The PROFEMUR TL2 Stems will be offered in sizes 0-12 for Standard (STD) and High Offset (HO), and 1-5, 11 & 12 in Ultra High Offset (UHO) with M/L Width ranging from 27-40mm and A/P width ranging from 13-17mm. All stem variants: standard, high offset, and ultra high offset were designed to accommodate a wide range of patient anatomy and to provide an optimized fit to adequately restore the biomechanics of the respective hip (including restoration of the natural leg length and head center offset). The PROFEMUR® TL2 stems are single use devices which will be provided sterile via Gamma Irradiation.
#### Indications for Use:
PROFEMUR® TL2 stems are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.
#### Indications for Use
- 1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2) inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3) correction of functional deformity; and,
- 4) revision procedures where other treatments or devices have failed
Titanium plasma spray coatings applied to implant surfaces are intended for uncemented arthroplasty.
### Comparison of Technological Characteristics with the Predicate devices:
Device comparison described in this premarket notification demonstrate that the subject PROFEMUR® TL2 stems are substantially equivalent to the identified predicate PROFEMUR® TL Classic Hip Stems cleared by the FDA for commercial distribution in the United States. The subject devices were shown to be substantially equivalent and have equivalent technological characteristics to their predicate devices in areas including design, intended use, material composition, operational principles and function.
{5}------------------------------------------------
## Discussion of the Non-clinical Testing/Performance Data:
### Mechanical Testing:
MicroPort has evaluated the subject PROFEMUR® TL2 Stems to demonstrate substantial equivalence to the identified predicate PROFEMUR® TL Classic Hip Stems. Design verification testing for the subject implants was completed in accordance with
- . ASTM-F2996 (2013): Standard Practice for Fine Element Analysis (FEA) of Non-Modular Metallic Orthopedic Hip Femoral Stems
- ISO 7206-4 (2010): Implants for surgery Partial and total hip joint prostheses -● part 4 Determination of endurance properties and performance of stemmed femoral components.
- ISO 7206-6 (2013): Implants for surgery -- Partial and total hip joint prostheses Part ● 6: Endurance properties testing and performance requirements of neck region of stemmed femoral components.
- ASTM F2068 (2015): Standard Specification for Femoral Prostheses—Metallic . Implants.
- ISO 21535 (2007): Non-active surgical implants -- Joint replacement implants --● Specific requirements for hip-joint replacement implants.
The testing performed for the subject PROFEMUR® TL2 Stems were:
- A. Finite Element Analysis, (FEA),
- B. Proximal Fatigue Evaluation,
- C. Distal Fatigue Evaluation,
- D. Range of Motion (ROM),
The mechanical testing verifies that the subject PROFEMUR® TL2 Stems are substantially equivalent to the predicate PROFEMUR® TL Classic Hip Stems currently on the market and has met all mechanical testing requirements based on the worst case construct testing.
### Non-Pyrogenicity Endotoxin Testing:
The bacterial endotoxin test, also known as Limulus amebocyte lysate (LAL) test, was performed utilizing worst case implants to verify that the subject implants meet the 20 endotoxin units (EU)/device pyrogen limit specification. The subject PROFEMUR® TL2 Stems were adopted into the current worst case Hip Stem and Neck Endotoxin Family based on the product process operations. Testing was successfully performed and it was confirmed that the worst case implants meet the 20 EU/device testing limit for general medical devices that are implanted as outlined in ANSI/AAMI ST72, Bacterial endotoxins – Test methods, routine monitoring, and
{6}------------------------------------------------
alternatives to batch testing and USP <161>, Medical Devices – Bacterial Endotoxin and Pyrogen Tests.
# Conclusion:
Based on the design features, the use of established well-known materials, feature comparisons, indications for use, principle of operations and results of the mechanical testing, the subject PROFEMUR® TL2 Stem have shown to be substantially equivalent to the legally marketed predicate devices cited in this premarket notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.