Encision AEM Monopolar Laparoscopic Instruments and Accessories
K191612 · Encision Incorporated · GEI · Mar 20, 2020 · General, Plastic Surgery
Device Facts
Record ID
K191612
Device Name
Encision AEM Monopolar Laparoscopic Instruments and Accessories
Applicant
Encision Incorporated
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 20, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
These AEM Instruments incorporate the use of AEM technology and are intended for use in delivering monopolar electrosurgical energy during laparoscopic procedures only. AEM Instruments are intended for use with the AEM Monitoring System and electrosurgical generators having compatibility with the AEM Monitor. Scissors Inserts are intended for use on soft tissue only.
Device Story
Encision AEM Monopolar Laparoscopic Instruments are surgical accessories used in laparoscopic procedures to deliver monopolar electrosurgical energy. The system includes reusable handles, reusable/disposable insert electrodes, and disposable sheaths. The device operates in conjunction with an AEM Monitor connected to an electrosurgical generator to continuously monitor and dynamically manage stray energy, specifically insulation failure and capacitive coupling, which may occur outside the surgeon's field of view. The instruments are used by surgeons in an operating room setting. The output is the delivery of electrosurgical energy for cutting or coagulation, with the monitor providing safety management of stray energy. This helps prevent unintended tissue damage, enhancing patient safety during minimally invasive surgery.
Clinical Evidence
Bench testing only. No clinical or animal testing was performed. Performance analysis demonstrated that the subject devices perform as intended and are equivalent to the predicate devices under similar conditions.
Technological Characteristics
Materials: Biocompatible per ISO 10993-1. Sensing: Active Electrode Monitoring (AEM) for stray energy management. Energy: Monopolar electrosurgical. Form factor: Laparoscopic instruments (handles, inserts, sheaths) in 35cm and 45cm lengths. Sterilization: E-beam (disposables) per ISO 11137-1; Steam (reusables) per ISO 17665-1. Connectivity: Compatible with AEM Monitoring System and electrosurgical generators. Standards: IEC 60601-2-2, IEC 60601-1-2.
Indications for Use
Indicated for use in delivering monopolar electrosurgical energy during laparoscopic procedures. Scissors inserts are indicated for soft tissue use only. Intended for use with AEM Monitoring System and compatible electrosurgical generators.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 20, 2020
Encision Incorporated % Charles Hart Principle Consultant HART Consulting LLC 615 Reid Place Castle Rock. Colorado 80108
Re: K191612
Trade/Device Name: Encision AEM Monopolar Laparoscopic Instruments and Accessories Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting & Coagulation Devices and Accessories Regulatory Class: Class II Product Code: GEI Dated: February 18, 2020 Received: February 21, 2020
Dear Charles Hart:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K191612
Device Name
Encision AEM® (Active Electrode Monitoring) Monopolar Laparoscopic Instruments and Accessories
Indications for Use (Describe)
These AEM Instruments incorporate the use of AEM technology and are intended for use in delivering monopolar electrosurgical energy during laparoscopic procedures only. AEM Instruments are intended for use with the AEM Monitoring System and electrosurgical generators having compatibility with the AEM Monitor. Scissors Inserts are intended for use on soft tissue only.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "K191612" in a simple, sans-serif font. The text is horizontally oriented and appears to be a code or identifier. The characters are uniformly sized and spaced, making the text easily readable.
Image /page/3/Picture/1 description: The image shows the logo for Encision. The logo features the word "ENCISION" in green, with a stylized orange circle to the right of the word. Below the word "ENCISION" is the phrase "DELIVERING ENERGY SAFELY" in a smaller font size.
### K191612 510(k) Summary
Device: Encision AEM® (Active Electrode Monitoring) Monopolar Laparoscopic Instruments and Accessories
Owner: Encision Inc.
- Contact: Greg Trudel President & CEO 6797 Winchester Circle Boulder, CO, 80301, USA gtrudel@encision.com 303-444-2600
Date: 20 Mar 2020
### Subject Device:
| Trade Name | See table below |
|-------------------|--------------------------------------------------------|
| Common Name | Monopolar laparoscopic accessory |
| Regulation Name | Electrosurgical, Cutting & Coagulation,<br>Accessories |
| Regulation Number | 21 CFR 878.4400 |
| Product Code | GEI |
| Device Class | 2 |
| Review Panel | General and Plastic Surgery |
| Prior Submissions | No prior submissions for the subject device |
### Primary Predicate Devices:
| Trade Name | See table below |
|-----------------|--------------------------------------------------------|
| Regulation Name | Electrosurgical, Cutting & Coagulation,<br>Accessories |
| Manufacturer | Encision Inc. |
| 510(k) Number | K912780 |
### Secondary Predicate Devices:
| Trade Name | ES03XX Series |
|-----------------|--------------------------------------------------------|
| Regulation Name | Electrosurgical, Cutting & Coagulation,<br>Accessories |
| Manufacturer | Encision Inc. |
| 510(k) Number | K091074 |
{4}------------------------------------------------
## ENCISIGN
### Device Description:
Encision AEM® Monopolar Laparoscopic Instruments and Accessories incorporate the use of AEM technology and are intended for use in delivering monopolar electrosurgical energy during laparoscopic procedures only. Encision AEM® Monopolar Laparoscopic Instruments and Accessories are intended for use with the AEM Monitoring System and electrosurgical generators having compatibility with the AEM Monitor. These Encision AEM® Monopolar Laparoscopic Instruments and Accessories include reusable handles and reusable & disposable insert electrodes plus disposable sheath. Encision AEM® Monopolar Laparoscopic Instruments and Accessories, in conjunction with an AEM Monitor properly connected to the electrosurgical generator (ESU), continuously monitor and dynamically manage "stray energy" (insulation failure and capacitive coupling), which are likely out of the surgeon's field of view.
## Model Numbers and Trade Names:
| Model Number<br>(subject device) | Trade Name | Product Family | Predicate Device<br>(K912780) |
|----------------------------------|--------------------------------------------------------------|----------------------------|-------------------------------|
| ES0101 | AEM e·Edge® Curved<br>Scissors, 1/2"Insert | Disposable Scissors | ES0101 |
| ES0102 | AEM e·Edge® Curved<br>Scissors, 3/4" Insert | Disposable Scissors | ES0102 |
| ES0110 | AEM e-Edge® Hook Scissors<br>Insert | Disposable Scissors | ES0110 |
| ES0120 | AEM e·Edge® Scissors, 3/4",<br>Low Profile Insert | Disposable Scissors | ES0120 |
| ES8000 Series | AEM enTouch® Handles | AEM® Handles<br>(Reusable) | ES5700 |
| ES8200 Series | AEM enTouch® Handles with<br>Indexing and Locking | AEM® Handles<br>(Reusable) | ES5700 |
| ES0004 | AEM® Right Angle Dissector<br>Insert | Reusable Inserts | ES0004 |
| ES0008 | AEM® Tapered Right Angle<br>Dissector Insert | Reusable Inserts | ES0008 |
| ES0009 | AEM® Blunt Nose Grasper<br>Insert | Reusable Inserts | ES0009 |
| ES0011 | AEM® Short Right Angle<br>Dissector Insert | Reusable Inserts | ES0011 |
| ES0012 | AEM® Bottle Nose Grasper<br>Insert | Reusable Inserts | ES0012 |
| ES0013 | AEM® Bullet Nose Grasper<br>Insert | Reusable Inserts | ES0013 |
| ES0014 | AEM® Fine Tooth Fenestrated<br>Grasper Insert | Reusable Inserts | ES0014 |
| ES0501 | AEM® Curved Maryland<br>Dissector Insert | Reusable Inserts | ES0501 |
| ES0506 | AEM® 90° Grasper Insert | Reusable Inserts | ES0506 |
| ES0507 | AEM® Fenestrated Grasping<br>Forceps Insert | Reusable Inserts | ES0507 |
| ES0508 | AEM® Round Nose Grasper<br>Insert | Reusable Inserts | ES0508 |
| Model Number<br>(subject device) | Trade Name | Product Family | Predicate Device<br>(K912780) |
| ES0509 | AEM® Pointed Nose Grasper Insert | Reusable Inserts | ES0509 |
| ES0510 | AEM® Petelin Dissector Insert | Reusable Inserts | ES0510 |
| ES0511 | AEM® Dolphin Nose Grasper Insert | Reusable Inserts | ES0511 |
| ES0512 | AEM® Long Dolphin Nose Grasper Insert | Reusable Inserts | ES0512 |
| ES0513 | AEM® Straight Dissector Insert | Reusable Inserts | ES0513 |
| ES0514 | AEM® Standard Grasper Insert | Reusable Inserts | ES0514 |
| ES0521 | AEM® Bowel Grasper Insert | Reusable Inserts | ES0521 |
| ES0522 | AEM® Fenestrated Bowel Grasper Insert | Reusable Inserts | ES0522 |
| ES0526 | AEM® Tapered Maryland Dissector Insert | Reusable Inserts | ES0526 |
| ES0533 | AEM® Strong Curved Maryland, 7-8mm Insert | Reusable Inserts | ES0533 |
| ES0535 | AEM® Endo Cinch Extreme Atraumatic Serrated Insert | Reusable Inserts | ES0535 |
| ES0537 | AEM® Wave Grasper Insert | Reusable Inserts | ES0537 |
| ES0538 | AEM® Mixter Clamp, 90° Long Insert | Reusable Inserts | ES0538 |
| ES0541 | AEM® Kelly Forceps Insert | Reusable Inserts | ES0541 |
| ES0543 | AEM® Maxi Grasper Insert | Reusable Inserts | ES0543 |
| ES0547 | AEM® Maryland Dissector, Diamond Serrations Insert | Reusable Inserts | ES0547 |
| ES0548 | AEM® Dissecting Forceps, Right Angled Insert | Reusable Inserts | ES0548 |
| ES0549 | AEM® Maryland Dissector, Aggressive Insert | Reusable Inserts | ES0549 |
| ES0552 | AEM® Dissecting Forceps, Right Angled, Cross Serrated Insert | Reusable Inserts | ES0552 |
| ES0553 | AEM® Straight Micro-Grasper Insert | Reusable Inserts | ES0553 |
| ES0557 | AEM® Micro-Fenestrated Grasper Insert | Reusable Inserts | ES0557 |
| ES0558 | AEM® Tapered Micro-Fenestrated Grasper Insert | Reusable Inserts | ES0558 |
| ES0559 | AEM® Right Angle Dissector, 7mm, Diagonal Serrations Insert | Reusable Inserts | ES0559 |
| ES0565 | AEM® Maryland Dissector with 90° Tooth Insert | Reusable Inserts | ES0565 |
| ES0566 | AEM® Beveled Maryland Dissector, Tapered Insert | Reusable Inserts | ES0566 |
| ES0570 | AEM® Insert, Atraumatic Grasper (Single Action) | Reusable Inserts | ES0570 |
| Model Number<br>(subject device) | Trade Name | Product Family | Predicate Device<br>(K912780) |
| ES0571 | AEM® Insert, Traumatic Grasper | Reusable Inserts | ES0571 |
| ES0573 | AEM® Insert, Traumatic Grasper (Single Action) | Reusable Inserts | ES0573 |
| ES0574 | AEM® Insert, Fenestrated Bowel Grasper | Reusable Inserts | ES0574 |
| ES0001 | AEM® Curved Scissors, 1/2" Insert | Reusable Inserts | ES0001 |
| ES0002 | AEM® Curved Scissors, 3/4" Insert | Reusable Inserts | ES0002 |
| ES0010 | AEM® Hook Scissors Insert | Reusable Inserts | ES0010 |
| ES0150 Series | AEM® Disposable Sheath | Disposable Sheaths | ES0150 |
{5}------------------------------------------------
# ENCISI&N
{6}------------------------------------------------
### Indications for Use:
These AEM Instruments incorporate the use of AEM technology and are intended for use in delivering monopolar electrosurgical energy during laparoscopic procedures only.
AEM Instruments are intended for use with the AEM Monitoring System and electrosurgical generators having compatibility with the AEM Monitor.
Scissors Inserts are intended for use on soft tissue only.
### Sterility/Packaging:
The Disposable Scissors and Disposable Sheaths are supplied sterile per ANSVAAMI/ISO 11137-1:2006 and ANSI/AAMI/ISO 11607-1:2006. The products are sterilized using E-beam radiation sterilization to achieve 106 SAL. Product is packaged in a Tyvek mylar pouch (10 pouches in a box) with a 5 year shelf life in the sterile packaging.
The AEM® Handles and Reusable Inserts are reusable and steam sterilized for repeated use per ANSI/AAMI/ISO 17664:2017 and ANSI/AAMI/ISO 17665-1:2006. These instruments are cleaned using manual, automatic or combined manual/automatic cleaning methods. They are sterilized through Prevac steam sterilization using FDA-cleared sterile wraps to achieve 10th SAL. These products are shipped non-sterile and require sterilization before use.
### Biocompatibility:
All patient contact materials are biocompatible. Material testing demonstrates conformance with ANSI/AAMI/ISO 10993-1:2009, 4th edition. Gap analysis 05927 demonstrates that testing to a previous revision of the standard is consistent with the current recognized revision of ISO 10993-1.
### Standards:
The AEM® Monopolar Laparoscopic Instruments and Accessories have been tested to show conformity to the following FDA recognized standards:
{7}------------------------------------------------
# ENCISI&N®
| Topic | Standard Used | Application |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| E-beam sterilization | ANSI/AAMI/ISO 11137-1:2006<br>Sterilization of Health Care Products – Radiation – Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices | Disposable Scissors and Disposable Sheaths are sterilized to 10-6 SAL using E-beam sterilization in accordance with this standard. |
| Sterile packaging | ANSI/AAMI/ISO 11607-1:2006<br>Packaging for Terminally Sterilized Medical Devices – Part 1: Requirements for Materials, Sterile Barrier Systems and Packaging Systems | Disposable Scissors and Disposable Sheaths are packaged with a sterile barrier system in accordance with this standard. |
| Reprocessing – cleaning and sterilization | ANSI/AAMI/ISO 17664:2017<br>Processing of Health Care Products – Information to be Provided by the Medical Device Manufacturer for the Processing of Medical Devices | AEM® Handles and Reusable Inserts are reprocessed for reuse in accordance with this standard |
| Reprocessing – steam sterilization | ANSI/AAMI/ISO 17665-1:2006<br>Sterilization of Health Care Products – Moist Heat - Part 1: Requirements for the Development, Validation and Routing Control of a Sterilization Process for Medical Devices | AEM® Handles and Reusable Inserts are reprocessed to 10-6 SAL using steam sterilization in accordance with this standard. |
| Biocompatibility | ANSI/AAMI/ISO 10993-1:2009<br>Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process | All patient-contact materials for the AEM® Monopolar Laparoscopic Instruments and Accessories have been tested for biocompatibility in accordance with this standard. |
| Electromagnetic compatibility | ANSI/AAMI/IEC 60601-1-2:2014<br>Medical Electrical Equipment – Part 1-6: General Requirements for Basic Safety and Essential Performance. Collateral Standard. Electromagnetic Compatibility. Requirements and Tests | The AEM® Monitoring System in which these electrodes are an accessory has been tested for electromagnetic compatibility in accordance with this standard. |
| Electrical safety | ANSI/AAMI/IEC 60601-2-2:2017<br>Medical Electrical Equipment – Part 2-2: Particular Requirements for the Basic Safety and Essential Performance of High Frequency Surgical Equipment and High Frequency Surgical Accessories | The AEM® Monitoring System in which these electrodes are an accessory has been tested for electrical safety in accordance with this standard. |
| Labeling | ANSI/AAMI/ISO 15223-1:2016<br>Medical Devices – Symbols to be Used with Medical Device Labels, Labelling and Information to be Supplied – Part 1: General Requirements | The AEM® Monopolar Laparoscopic Instruments and Accessories have associated labeling in accordance with this standard. |
| Topic | Standard Used | Application |
| Risk management | ANSI/AAMI/ISO 14971:2007<br>Medical Devices – Application of Risk<br>Management to Medical Devices | Risk management activities for the<br>AEM® Monopolar Laparoscopic<br>Instruments and Accessories have<br>been conducted in accordance with<br>this standard. |
{8}------------------------------------------------
### Technological Characteristics:
The technological characteristics of the subject device are identical to the predicate devices with regards to fundamental design, materials, operating principle, use, compatibility, packaging and sterilization. The accessories have both 35cm and 45cm configurations. The changes over time to the predicate devices do not introduce new concerns of safety or effectiveness. The differences between the subject and predicate device are:
- . Use of Metrilube Instrument Lubricant for Disposable and Reusable Scissors for effective cutting performance. The lubricant is a patient-contact material and is biocompatible.
- . Change from a removable trigger to a permanent trigger on the AEM® Handles for improved handle performance. The fundamental technology and operating principle is identical to the predicate devices.
- . Shelf life of Disposable Scissors and Disposable Sheaths has increased from 1 year to 5 years in the sterile packaging.
### Substantial Equivalence:
The subject device is substantially equivalent to the predicate device, as there is no change to the Intended Use, Operating Principle, Patient Contact Materials, Fundamental Technology, Sterilization and Performance.
| Topic | Subject Device (this submission) compared to Predicate Devices | |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | (K912780) | |
| Indications for Use | No change to Intended Use. | |
| Operating Principle | No change to Operating Principle. | |
| Materials | Substantially equivalent. Material changes are equivalent to predicate<br>devices and do not add new biocompatibility concerns. | |
| Use | No changes to Use. The AEM Instruments are prescription use only. | |
| Technology | Only minor changes to Technology. The design changes made to the<br>products have not affected the fundamental technology of the AEM<br>Instruments. | |
| Sterilization | No changes to Sterilization. Disposable Scissors and Disposable Sheaths are<br>E-beam sterilized and supplied sterile. AEM Handles and Reusable Inserts<br>are reusable and supplied non-sterile. | |
| Storage &<br>Transportation | No change to Storage and Transportation. | |
| Performance | No change to Performance. | |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "ENCISION" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. To the right of the letters is a stylized graphic that resembles a crescent shape with a curved line running through it.
### Bench Testing:
All testing required per design control procedures was conducted using bench testing. Performance analysis has shown the Encision AEM® Monopolar Laparoscopic reusable handpieces and Encision AEM® Monopolar Laparoscopic reusable & disposable insert electrodes plus disposable sheath to perform as intended and as well as the predicate devices under the same and/or similar conditions. These devices have been tested with settings set equal to those seen in typical procedures where these devices would be indicated for use. The requirements for electrodes function in the AEM Monitoring System have not changed from the original 510(k) submissions. The tests conducted show that the changes since the original 510(k) submissions demonstrate conformity to the original internal and regulatory requirements. No differences have been observed between the performance of the predicate devices and the device subject of this 510(k) Notification submission. No animal or clinical testing is required.
### Conclusion:
The AEM® Monopolar Laparoscopic Instruments and Accessories are substantially equivalent to the predicate devices. The design changes do not introduce new safety or effectiveness concerns and do not alter the fundamental technology or operating principle of the devices. The Intended Use is identical. Bench testing data has confirmed substantial equivalence to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.