Arcadia Balloon Catheter

K191596 · Merit Medical Systems, Inc. · HRX · Oct 9, 2019 · Orthopedic

Device Facts

Record IDK191596
Device NameArcadia Balloon Catheter
ApplicantMerit Medical Systems, Inc.
Product CodeHRX · Orthopedic
Decision DateOct 9, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.1100
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Arcadia™ Steerable and Straight Balloon Catheters are intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. These balloon catheters are to be used with cleared spinal polymethylmethacrylate (PMMA) bone cement for use during percutaneous vertebral augmentation, such as kyphoplasty.

Device Story

Arcadia Balloon Catheter is an inflatable bone tamp for spinal procedures; used for fracture reduction and void creation in cancellous bone. Device consists of Y-adapter, dual-lumen catheter shaft, and inflatable balloon. Inner lumen houses straight or articulating stylet; outer lumen serves as inflation conduit. Radiopaque marker bands on shaft and balloon aid placement and depth verification. Exterior surface lubricated for introducer access. Operated by physician in clinical setting; connected to external inflation device using contrast solution. Inflation creates void in bone for subsequent PMMA cement delivery. Benefits include controlled fracture reduction and stabilization. Single-use, sterile device.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), insertion/withdrawal force, inflation dimensions, articulation profile, deflation time, constrained pressure testing, burst volume, and tensile strength.

Technological Characteristics

Materials: Polyurethane balloon/catheter, Platinum/Iridium marker bands. Principle: Inflatable balloon tamp for bone void creation. Energy: Manual inflation via external device. Connectivity: None. Sterilization: Radiation (ISO 11137-2). Standards: ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-11, USP <151>, ASTM D4169, ASTM F88, ASTM F1140, ASTM F1929, ASTM F1980, ASTM F2096, ISO 11607.

Indications for Use

Indicated for reduction and fixation of spinal fractures and creation of a void in cancellous bone during percutaneous vertebral augmentation (e.g., kyphoplasty) using cleared spinal PMMA bone cement.

Regulatory Classification

Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Merit Medical Systems, Inc. Shamsa Karimi Principal Regulatory Affairs Specialist 1600 West Merit Parkway South Jordan, Utah 84095 Re: K191596 Trade/Device Name: Arcadia™ Balloon Catheter Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX, HXG, NDN Dated: September 11, 2019 Received: September 12, 2019 Dear Ms. Karimi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Laurence D. Coyne, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K191596 Device Name Arcadia TM Balloon Catheter #### Indications for Use (Describe) The Arcadia ™ Steerable and Straight Balloon Catheters are intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. These balloon catheters are to be used with cleared spinal polymethylmethacrylate (PMMA) bone cement for use during percutaneous vertebral augmentation, such as kyphoplasty. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Contact Person:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(408) 770-1115<br>Shamsa Karimi<br>1721504 | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Subject Device | Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code:<br>21 CFR §:<br>Review Panel: | Arcadia™ Balloon Catheter<br>Inflatable Balloon Tamp<br>Arthroscope, Orthopedic Manual Surgical<br>Instrument, Polymethylmethacrylate (PMMA)<br>bone cement<br>II<br>HRX, HXG, NDN<br>888.1100, and 888.4540, 888.3027<br>Orthopedics | | Predicate<br>Device | Predicate #1:<br>Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer:<br>Predicate#2:<br>Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | Kyphon® Express II™ Inflatable Bone Tamp<br>Arthroscope, Orthopedic Manual Surgical<br>Instrument, Polymethylmethacrylate (PMMA)<br>Bone Cement<br>K123771<br>Medtronic Sofamor Danek USA<br>Osseoflex® SB<br>Arthroscope, Polymethylmethacrylate (PMMA)<br>Bone Cement<br>K141930<br>Osseon LLC<br>This predicate device has not been subject to a design-related recall. | | Device<br>Description | | The Arcadia Balloon Catheters consist of a Y-Adapter with a dual-lumen catheter shaft and a balloon. The inner shaft lumen contains a<br>stylet (either straight or articulating). The outer lumen is an inflation conduit for the balloon. A valved luer port on the Y-adapter allows for | | marker bands are printed on the shaft and serve as indicators of when the distal tip of the balloon catheter has reached the distal end of the working cannula of the StabiliT® Introducer. Radiopaque markers are located on the proximal and distal end of the balloon to aid in balloon placement prior to inflation. The exterior surface of balloon is covered by a lubricant which ease access through the introducer cannula. | | | | The Arcadia Steerable Balloon Catheter features a mechanism that enables steering of the device. The Arcadia Straight Balloon Catheter has a removable stylet. | | | | Indications for<br>Use | The ArcadiaTM Steerable and Straight Balloon Catheters are intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. These balloon catheters are to be used with cleared spinal polymethylmethacrylate (PMMA) bone cement for use during percutaneous vertebral augmentation, such as kyphoplasty. | | connecting an inflation device to inflate and deflate the balloon. Two # 510(k) Summarv {4}------------------------------------------------ {5}------------------------------------------------ The design and technological characteristics of the subject Arcadia Balloon Catheter are substantially equivalent to the predicate devices. Kyphon® Express II™ Inflatable Bone Tamp and Osseoflex® SB. The comparison between the subject and the predicate devices were based on the following: ● Intended Use Indications for Use ● ● Labeling Basic Design and Material ● Principle of Operation/Fundamental Technology ● Performance ● Sterilization ● Biocompatibility ● Comparison to Predicate Devices Summary of Technological Characteristics of the Subject Device to the Predicate Devices Attribute Predicate Device Predicate Device Subject Device Osseoflex® SB Kyphon® Express II™ Arcadia Balloon Catheter (Steerable) Inflatable Bone Tamp (Steerable and Straight) (Straight) Balloon Inflation Medium 60% Contrast 60% Contrast 60% Contrast Polyurethane Device Balloon Polyurethane Polyurethane Materials Catheter Polyurethane Polyurethane Polyurethane Marker Platinum/Iridium Platinum/Iridium Platinum/Iridium Bands Balloon Length (pre-10mm/(2mL), 10mm/(3mL), 15mm/(4mL), 10mm/(3mL), 15mm/(4mL), (inflation)/Maximum 20mm/(5mL) 20mm/(5mL), 25mm/(7mL), 20mm/(4mL) 30mm/(8mL) Balloon Volume Max Inflation Pressure 400 psi 700 psi 700 psi {6}------------------------------------------------ | Summary of Non-Clinical Tests Conducted for Determination of Substantial Equivalence | | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------| | Performance Test Summary | | | Test Performed | Acceptance Criteria | | Biocompatibility Testing | Device meets applicable ISO 10993-1:2018 requirements | | Insertion and Withdrawal Force | Insertion and withdrawal of the device through the<br>introducer do not create forces that would damage the<br>balloon catheter. | | Inflated Balloon Dimensions<br>(Unconstrained) | Inflated balloon dimensions are similar to similarly labeled<br>predicate devices. | | Articulated Angle and Radius | Articulation profile of steerable devices matches that of<br>tools used to create channels in the bone for balloon<br>placement. | | Deflation Time with Contrast Solution | Deflation time is similar to predicate devices of the same<br>size and volume profile. | | Constrained Max Pressure Testing | Device is capable of sustaining its maximum rated pressure<br>when the balloon is constrained. | | Unconstrained Burst Volume | Device exceeds its maximum volume rating before burst<br>failure | | Tensile Force Testing | Device material and bond strengths exceed forces that may<br>be applied during normal clinical use. | | Summary of Clinical Tests for Determination of Substantial Equivalence | | | N/A- No clinical test was conducted for this submission. | | | Substantial Equivalence Conclusion | | | Based on the indications for use, intended use, design, safety and performance testing, the subject<br>Arcadia Balloon Catheter meets the requirements that are considered essential for its intended use and is<br>substantially equivalent to the Predicate Devices, Kyphon® Express II™ Inflatable Bone Tamp [K123771]<br>and Osseoflex® SB [K141930]. | | {7}------------------------------------------------ | | No performance standards have been established under Section 514 of<br>the Food, Drug and Cosmetic Act for these devices. Performance<br>testing of the subject device, Arcadia Balloon Catheter was conducted<br>based on the risk assessment and based on the requirements of the<br>following international standards: | |---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Performance<br>Data | FDA Guidance Document: Use of International Standard ISO<br>10993-1, Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management process. ISO 10993-1:2018, Biological evaluation of medical devices -<br>Part 1: Evaluation and testing within a risk management process ISO 10993-5:2009: Biological evaluation of medical devices-<br>Tests for in vitro cytotoxicity. ISO 10993-10:2010: Biological evaluation of medical devices-<br>Tests for irritation and skin sensitization ISO 10993-11:2017: Biological evaluation of medical devices-<br>Tests for systemic toxicity. USP 41- NF 36: 2018, <151>, Pyrogen Test (USP Rabbit Test) ASTM D 4169-16: Standard practice for performance testing of<br>shipping and systems. ASTM F88M/F88M-15: Standard test method for seal strength<br>of flexible barrier materials. ASTM F1140/F1140M-13: Standard test method for internal<br>pressurization failure resistance of unrestrained packages. ASTM F1929-15: Standard test method for detecting seal leaks<br>in porous medical packaging by dye penetration. ASTM F1980-16: Standard Guide for accelerated aging of<br>sterile barrier systems for medical devices. ASTM F2096-11: Standard test method for detecting gross<br>leaks in packaging by internal pressurization (bubble test). EN ISO 2233:2000: Packaging—Complete filled transport<br>packages and unit loads—Conditioning for testing. ISO 11607-1:2006: Packaging for terminally sterilized medical<br>devices-Part 1: Requirements for materials, sterile barrier<br>systems and packaging systems [Including: Amendment 1<br>(2014)]. ISO 11607-2: 2006: Packaging for terminally sterilized medical<br>devices - Part 2: Validation requirements for forming, sealing<br>and assembly processes [Including: Amendment 1 (2014)] ISO 11137-2:2013: Sterilization of health care products –<br>Radiation - Part 2 Establishing the sterilization dose. | | Performance<br>Data cont. | Pre-Inflation Dimensional Analysis Insertion and Withdrawal Force Inflated Balloon Dimensions (Unconstrained) Articulated Angle and Radius Deflation Time with Contrast Solution Constrained Max Pressure Testing Unconstrained Burst Volume Tensile Force Testing | | Design validation testing confirmed that the subject de | | {8}------------------------------------------------ The following performance data were provided in support of the substantial equivalence determination. ## Biocompatibility testing The biocompatibility evaluation and testing for the subject Arcadia Balloon Catheter was conducted in accordance with ISO 10993-1:2018, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and FDA Guidance Document: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (2016)". The battery of testing included the following tests: - Cytotoxicity ● - Sensitization ● - Irritation . - Acute Systemic Toxicity - Pyrogenicity The Arcadia Balloon Catheter is categorized as an externally communicating device with tissue/bone/dentin contact for a limited (< 24 hours) duration. All biocompatibility testing met the requirements of the respective test methods, thus supporting the biocompatibility of the subject device, Arcadia Balloon Catheter. ## Performance Testing Design verification and validation testing were completed to establish the performance of the subject device, Arcadia Balloon Catheter and substantial equivalence to the predicate device. ## Design Verification Testing: ice, Arcadia Balloon Catheter conforms to user needs and intended uses. {9}------------------------------------------------ | | Performance testing demonstrates equivalence between the subject<br>device and the predicate device. A risk management assessment was<br>performed and found that the subject device does not introduce any<br>new risk related to its safety or effectiveness. | |------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Summary of<br>Substantial<br>Equivalence | Based on the indications for use, intended use, design, safety and<br>performance testing, the subject Arcadia Balloon Catheter meets the<br>requirements that are considered essential for its intended use and is<br>substantially equivalent to the Predicate Devices, Kyphon® Express<br>IIT™ Inflatable Bone Tamp [K123771] and Osseoflex® SB [K141930]. | | Performance<br>Data | No performance standards have been established under Section 514 of<br>the Food, Drug and Cosmetic Act for these devices. Performance<br>testing of the subject device, Arcadia Balloon Catheter was conducted<br>based on the risk assessment and based on the requirements of the<br>following international standards:<br>FDA Guidance Document: Use of International Standard ISO<br>10993-1, Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management process. ISO 10993-1:2018, Biological evaluation of medical devices -<br>Part 1: Evaluation and testing within a risk management process ISO 10993-5:2009: Biological evaluation of medical devices—<br>Tests for in vitro cytotoxicity. ISO 10993-10:2010: Biological evaluation of medical devices—<br>Tests for irritation and skin sensitization ISO 10993-11:2017: Biological evaluation of medical devices—<br>Tests for systemic toxicity. USP 41- NF 36: 2018, <151>, Pyrogen Test (USP Rabbit Test) ASTM D 4169-16: Standard practice for performance testing of<br>shipping and systems. ASTM F88M/F88M-15: Standard test method for seal strength<br>of flexible barrier materials. ASTM F1140/F1140M-13: Standard test method for internal<br>pressurization failure resistance of unrestrained packages. ASTM F1929-15: Standard test method for detecting seal leaks<br>in porous medical packaging by dye penetration. ASTM F1980-16: Standard Guide for accelerated aging of<br>sterile barrier systems for medical devices. ASTM F2096-11: Standard test method for detecting gross<br>leaks in packaging by internal pressurization (bubble test). EN ISO 2233:2000: Packaging—Complete filled transport<br>packages and unit loads—Conditioning for testing. ISO 11607-1:2006: Packaging for terminally sterilized medical<br>devices—Part 1: Requirements for materials, sterile barrier<br>systems and packaging systems [Including: Amendment 1<br>(2014)]. ISO 11607-2: 2006: Packaging for terminally sterilized medical<br>devices - Part 2: Validation requirements for forming, sealing<br>and assembly processes [Including: Amendment 1 (2014)] ISO 11137-2:2013: Sterilization of health care products –<br>Radiation - Part 2 Establishing the sterilization dose. | | Performance<br>Data cont. | • Pre-Inflation Dimensional Analysis<br>• Insertion and Withdrawal Force<br>• Inflated Balloon Dimensions (Unconstrained)<br>• Articulated Angle and Radius<br>• Deflation Time with Contrast Solution<br>• Constrained Max Pressure Testing<br>• Unconstrained Burst Volume<br>• Tensile Force Testing | | | Design validation testing confirmed that the subject de | {10}------------------------------------------------ {11}------------------------------------------------ The following performance data were provided in support of the substantial equivalence determination. ## Biocompatibility testing The biocompatibility evaluation and testing for the subject Arcadia Balloon Catheter was conducted in accordance with ISO 10993-1:2018, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and FDA Guidance Document: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (2016)". The battery of testing included the following tests: - Cytotoxicity ● - Sensitization ● - Irritation . - Acute Systemic Toxicity - Pyrogenicity The Arcadia Balloon Catheter is categorized as an externally communicating device with tissue/bone/dentin contact for a limited (< 24 hours) duration. All biocompatibility testing met the requirements of the respective test methods, thus supporting the biocompatibility of the subject device, Arcadia Balloon Catheter. ## Performance Testing Design verification and validation testing were completed to establish the performance of the subject device, Arcadia Balloon Catheter and substantial equivalence to the predicate device. ## Design Verification Testing: vice, Arcadia Balloon Catheter conforms to user needs and intended uses. {12}------------------------------------------------ | | Performance testing demonstrates equivalence between the subject device and the predicate device. A risk management assessment was performed and found that the subject device does not introduce any new risk related to its safety or effectiveness. | |------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The Arcadia Balloon Catheter is qualified and labeled as a sterile, single-use device with the 6-month shelf-life at this time. | | Summary of<br>Substantial<br>Equivalence | Based on the indications for use, intended use, design, safety and performance testing, the subject Arcadia Balloon Catheter meets the requirements that are considered essential for its intended use and is substantially equivalent to the Predicate Device, Kyphon® Express II™ Inflatable Bone Tamp [K123771]. |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...