Enlighten BB Base, developed by ENLIGHTEN MATERIALS Co., Ltd., is a light-cured resin indicated for the fabrication of full removable denture bases in dental laboratories. It is an alternative to traditional heat-cured, auto-cured or light-cured denture base resins. Enlighten BB Base should be used exclusively by dental professionals. Fabrication of denture bases with Enlighten BB Base requires a computer-aided and manufacturing (CAD/CAM) system including the following; optical impression system, design software (CAD software), stereolithographic additive printer, and post-cure unit.
Device Story
BB Base is a photosensitive, methacrylate-based resin used for additive manufacturing of full removable denture bases. The workflow involves capturing patient anatomy via an optical impression system, designing the denture base using CAD software, and printing the structure using a stereolithographic additive printer (385-405 nm light). Following printing, the device undergoes final polymerization in a UV-light curing box. The device is intended for use by dental professionals in laboratory settings. The resulting denture base provides a functional alternative to traditional heat- or auto-cured acrylic resins. Clinical benefit is derived from the precision and efficiency of the CAD/CAM fabrication process for edentulous patients.
Clinical Evidence
No clinical data. Evidence consists of bench testing for mechanical properties and biocompatibility. Mechanical testing (flexural strength, flexural modulus, residual monomer, water sorption, water solubility) confirmed compliance with ISO 20795-1. Biocompatibility testing followed ISO 10993-1, -3, -5, -10, and -11 standards. Risk assessment conducted per ISO 14971.
Technological Characteristics
Photosensitive, methacrylate-based resin containing photoinitiators (385-405 nm activation) and pigments. Classified as ISO 20795-1 Type 4 light-activated acrylic resin. Form factor: 1 kg liquid bottle. Requires external CAD/CAM system (optical scanner, CAD software, stereolithographic printer, UV post-cure unit).
Indications for Use
Indicated for the fabrication of full removable denture bases for edentulous patients in dental laboratories. For use exclusively by dental professionals.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
August 18, 2020
Enlighten Materials Co., Ltd % Chiao-Min Chang Regulatory Affairs Voler Biotech Consulting Co., Ltd. No. 3-1, Ln 58, Hejiang St., Zhongshan Dist., Taipei City, 10480 TAIWAN
Re: K191591
Trade/Device Name: BB Base 3D Printing Resin for Denture Base Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: July 9, 2020 Received: July 20, 2020
Dear Chiao-Min Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
## 510(k) Number
Device Name: BB Base 3D printing resin for denture base Indications for Use:
Enlighten BB Base, developed by ENLIGHTEN MATERIALS Co., Ltd., is a light-cured resin indicated for the fabrication of full removable denture bases in dental laboratories. It is an alternative to traditional heat-cured, auto-cured or light-cured denture base resins. Enlighten BB Base should be used exclusively by dental professionals. Fabrication of denture bases with Enlighten BB Base requires a computer-aided and manufacturing (CAD/CAM) system including the following; optical impression system, design software (CAD software), stereolithographic additive printer, and post-cure unit.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
{3}------------------------------------------------
## 510(k) SUMMARY
# BB Base 3D printing resin for denture base
## 1. Contact Person
Name: Jiun Ming Su Title: Chief Financial Officer E-mail: jimmy.su.js@gmail.com Office address: 8F., NO.138, SEC. 2, JINSHAN S. RD., DA-AN DIST., TAIPEI CITY 10642, TAIWAN (R.O.C.) Tel: (O) +886956973958
| Date prepared: | 2020.07.07 |
|-----------------------|-------------------------------------------------|
| Product Name: | BB Base 3D printing resin for denture base |
| Classification Name: | Denture relining, repairing, or rebasing resin. |
| Common or Usual Name: | Dental Material |
| Classification Panel: | 21 CFR 872.3760 |
| Regulation Number: | 872.3760 |
| Device Class: | Class 2 |
| Product Code: | EBI |
#### で Device Name and Classification
#### 3. Predicate Device(s)
Dentca Denture Base II K162044
## 4. Device Description
BB Base is a photosensitive resin intended to fabricate removable dentures in a CAD/CAM additive printing process. The BB Base 3D printing resin for denture base is a viscous solution of the following compounds: methacrylate-based resins, a photoinitiator that activates between 385nm to 405 nm light, and pigments. It comes in one size, one kilogram per bottle. It is a Type 4 (light-activated) acrylic resin as classified ISO20795-1. The material is used in a 3D printer, which prints the shape determined by a 3D stereolithographic drawing. After printing, the printed product is placed in a UV-light curing box for final polymerization. 3D
{4}------------------------------------------------
printer is not included with the device.
### 5. Intended Use / Indications for Use
BB Base 3D printing resin for denture base, developed by ENLIGHTEN MATERIALS Co., Ltd., is a light-cured resin indicated for the fabrication of full removable denture bases in dental laboratories. It is an alternative to traditional heat-cured, light-cured or light-cured denture base resins. BB Base 3D printing resin for denture base should be used exclusively by dental professionals. Fabrication of denture bases with BB Base 3D printing resin for denture base requires a computer-aided and manufacturing (CAD/CAM) system including the following; optical impression system, design software (CAD software), stereolithographic additive printer, and post-cure unit.
#### Scanner and CAD software:
| | Brand | Model |
|-----------------|------------|--------|
| Scanner | Carestream | 3600 |
| Design software | Exocad | Exocad |
Printing system:385~405nm 3D printers
| Printer Brand | Model | Software |
|---------------|-------|------------|
| Miicraft | Ultra | MiiUtility |
#### Post-Curing: 385~405nm UV box
| Post-cure unit brand | Type |
|----------------------|-----------|
| Formlabs | Form Cure |
{5}------------------------------------------------
## 6. Comparison of technology
| | Subject Device | Predicate Device |
|---------------------------|------------------------------------------------------------|------------------------------------------------------------|
| 510(k) No. | K191591 | Dentca Denture Base II<br>(K162044) |
| Item | BB Base 3D printing resin for<br>denture base | e-dent Temporary Resin and<br>Extra-Oral |
| Classification | 2 | 2 |
| Product Code | EBI | EBI |
| Intended Use | Fabrication and repair of<br>removable dentures | Fabrication and repair of<br>removable dentures |
| Target population | Edentulus patients | Edentulus patients |
| Standard and<br>Approvals | ISO20795-1 type 4<br>(light-cure denture base<br>material) | ISO20795-1 type 4<br>(light-cure denture base<br>material) |
## Device Description
| | K191591 | K162044 | Differences |
|-----------------------------------|-----------------------------------------|-----------------------------------------|-------------|
| | Subject Device | Predicate Device | |
| Acrylic Resin | Light- Cure Resin | Light- Cure Resin | Identical |
| Chemical<br>Characterization | Methacrylate-<br>based resin | Methacrylate-<br>based resin | Identical |
| Polymerization<br>(curing) Method | Visible light | Visible light | Identical |
| Product State | Pre-mix resin | Pre-mix resin | Identical |
| Fabrication of<br>denture Base | CAD/CAM<br>additive printing<br>process | CAD/CAM<br>additive printing<br>process | Identical |
| Teeth Assemble | Bonding | Bonding | Identical |
{6}------------------------------------------------
## Physical Properties
| | K191591<br>Subject Device | K162044<br>Predicate Device | Comparison |
|-------------------|---------------------------|----------------------------------------------------------------------------|------------------------------------|
| Flexural Strength | >85.2 MPa | >65 MPa | Met the criteria of<br>ISO 20975-1 |
| Flexural modulus | >2049 MPa | >2000 MPa | Met the criteria of<br>ISO 20975-1 |
| Viscosity | ~1180 cps | 1000 <x<2000< td=""><td>Met the criteria of<br/>ISO 20975-1</td></x<2000<> | Met the criteria of<br>ISO 20975-1 |
The physical properties of BB base are complied with ISO 20975-1, and these differences do not raise concerns of safety and effectiveness for this submission.
## 7. Performance Data
- Non-clinical performance testing
BB base has been tested for mechanical properties as part of the product specification. The most applicable standard for mechanical characteristics determination of denture base polymers and copolymers is the ISO 20975-1 Dentistry - Base polymers - Part 1: Denture base polymers.
BB base is considered a surface device, in contact with the mucosal membrane, for > 30 days. The ISO 10993-1 standard was followed, and the following biological safety aspects have been addressed:
- Cytotoxicity ISO10993-5 >
- Sensitization ISO10993-10 A
- A Irritation or intracutaneous reactivity - ISO10993-10
- Acute systemic toxicity ISO10993-11 A
- Genotoxicity ISO 10993-3 A
We also conducted a risk assessment following ISO 14971 to conform the biocompatibility.
BB base has been tested for conformity with the industry standard ISO 20795-1. BB base is compliant to the requirements defined in ISO 20975-1 for Type 4 materials.
{7}------------------------------------------------
The following bench tests are conducted on BB base using all the compatible CAD/CAM systems, including the post-curing process:
- Flexural strength A
- > Flexural modulus
- > Residual monomer
- Water sorption A
- > Water solubility
Clinical Performance Testing ●
We did not conduct clinical performance testing for this submission device.
## 8. Conclusions:
Based on similar technology and Indications for Use as well as results of performance testing, we believe that BB Base 3D printing resin for denture base is substantially equivalent to the predicate, Dentca Denture Base II.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.