← Product Code [PHX](/productcode/PHX) · K191561

# Exactech Equinoxe Humeral Augmented Trays (K191561)

_Exactech, Inc. · PHX · Aug 20, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K191561

## Device Facts

- **Applicant:** Exactech, Inc.
- **Product Code:** [PHX](/productcode/PHX.md)
- **Decision Date:** Aug 20, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3660
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Equinoxe Humeral Augmented Tray is indicated for proximal humeral replacement in conjunction with reverse shoulder arthroplasty in skeletally mature individuals with an intact medial calcar for whom the rotator cuff is irreparable or grossly deficient, a functional deltoid muscle is present, and proximal humeral bone loss is present in cases of: - Rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic degenerative problems - Congenital abnormalities in the skeletally mature - Primary and secondary necrosis of the humeral head - Humeral head fracture with displacement of the tuberosities - Pathologies where arthrodesis or resectional arthroplasty of the humeral head are not acceptable - Revisions of humeral prostheses when other treatments or devices have failed (where adequate humeral stem fixation can be achieved) - Displaced three- and four-part upper humeral fractures - Failed ORIF requiring rTSA with an absent or deficient greater tuberosity - Revision of failed previous reconstructions when distal anchorage is required - Treatment of unplanned proximal bone loss during a revision - To restore mobility from previous procedures (e.g. previous fusion) The Equinoxe Humeral Augmented Tray can be used in either primary or revision arthroplasty procedures.

## Device Story

Modular orthopedic implant system for proximal humeral bone loss; used in reverse shoulder arthroplasty. Components include humeral tray, modular extension (augment), and universal locking screw. Surgeon selects components based on patient anatomy; modular extension attaches to tray via locking screw to address severe bone loss. Implanted by orthopedic surgeons in clinical/OR settings. Device restores joint mechanics, improves deltoid wrapping, and provides soft tissue tensioning. Titanium alloy construction. Benefits include improved stability and mobility in revision or primary cases with significant bone deficiency.

## Clinical Evidence

Bench testing only. Included fatigue strength testing under worst-case lateral and medial loading, range of motion analysis, and cadaveric evaluation. Pyrogen testing performed per USP <161>, USP <85>, and ANSI/AAMI ST72.

## Technological Characteristics

Materials: Titanium Alloy (Ti-6Al-4V). Components: Humeral tray, modular extension, universal locking screw. Design: Anatomical, modular, multiple offsets/thicknesses/lateralizations. Sterilization: Compliant with FDA guidance for sterile devices.

## Regulatory Identification

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

## Predicate Devices

- Equinoxe Reverse Shoulder System (Humeral Adapter Trays) ([K063569](/device/K063569.md))
- Equinoxe Reverse Shoulder System + 15mm Humeral Adapter Tray ([K082702](/device/K082702.md))

## Reference Devices

- Equinoxe Humeral Reconstruction Prosthesis ([K143659](/device/K143659.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 20, 2019

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

Exactech, Inc Shing Jen Tai Principal Regulatory Affairs Specialist 2320 NW 66th Court Gainesville, Florida 32653

Re: K191561

Trade/Device Name: Exactech® Equinoxe® Humeral Augmented Trays Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: PHX, KWT, KWS Dated: June 11, 2019 Received: June 13, 2019

Dear Shing Jen Tai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

FOR Vesa Vuniqi Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

#### 510(k) Number (if known) K191561

Device Name

Exactech® Equinoxe® Humeral Augmented Trays

#### Indications for Use (Describe)

The Equinoxe Humeral Augmented Tray is indicated for proximal humeral replacement in conjunction with reverse shoulder arthroplasty in skeletally mature individuals with an intact medial calcar for whom the rotator cuff is irreparable or grossly deficient, a functional deltoid muscle is present, and proximal humeral bone loss is present in cases of:

- · Rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic degenerative problems
- · Congenital abnormalities in the skeletally mature
- · Primary and secondary necrosis of the humeral head
- · Humeral head fracture with displacement of the tuberosities
- · Pathologies where arthrodesis or resectional arthroplasty of the humeral head are not acceptable
- Revisions of humeral prostheses when other treatments or devices have failed (where adequate humeral stem fixation can be achieved)
- · Displaced three- and four-part upper humeral fractures
- · Failed ORIF requiring rTSA with an absent or deficient greater tuberosity
- Revision of failed previous reconstructions when distal anchorage is required
- · Treatment of unplanned proximal bone loss during a revision
- · To restore mobility from previous procedures (e.g. previous fusion)

The Equinoxe Humeral Augmented Tray can be used in either primary or revision arthroplasty procedures.

Type of Use (Select one or both, as applicable)

| <div> <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> </span>Prescription Use (Part 21 CFR 801 Subpart D)   </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> </span>Over-The-Counter Use (21 CFR 801 Subpart C)   </div>                                                                                                                                                                                                       |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Exactech®Equinoxe® Humeral Augmented Trays Traditional 510(k) - 510(k) Summary

| Sponsor:              | Exactech®, Inc<br>2320 N.W. 66th Court<br>Gainesville, FL 32653                                                                             |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
|                       | Phone: (352) 377-1140<br>Fax: (352) 378-2617                                                                                                |
|                       | FDA Establishment Number 1038671                                                                                                            |
| Contact Person:       | Shing Jen Tai, PhD<br>Principal Regulatory Affairs Specialist                                                                               |
| Date:                 | August 8, 2019                                                                                                                              |
| Proprietary Name:     | Exactech® Equinoxe® Humeral Augmented Trays                                                                                                 |
| Common Name:          | Reverse shoulder prosthesis                                                                                                                 |
| Regulation Number:    | 21 CFR 888.3650<br>21 CFR 888.3660                                                                                                          |
| Classification Name:  | Shoulder joint metal/polymer non-constrained cemented<br>prosthesis<br>Shoulder joint metal/polymer semi-constrained cemented<br>prosthesis |
| Device Class:         | Class II                                                                                                                                    |
| Product Code:         | PHX, KWT, KWS                                                                                                                               |
| Classification Panel: | Orthopedic                                                                                                                                  |

# Information on devices to which substantial equivalence is claimed:

| 510(k) Number | Trade or Proprietary Model Name                                 | Manufacturer  |
|---------------|-----------------------------------------------------------------|---------------|
| K063569       | Equinoxe Reverse Shoulder System (Humeral<br>Adapter Trays)     | Exactech, Inc |
| K082702       | Equinoxe Reverse Shoulder System + 15mm<br>Humeral Adapter Tray | Exactech, Inc |

{4}------------------------------------------------

## Exactech® Equinoxe® Humeral Augmented Trays Traditional 510(k) - 510(k) Summary

#### Information on reference device:

| 510(k) Number | Trade or Proprietary Model Name            | Manufacturer  |
|---------------|--------------------------------------------|---------------|
| K143659       | Equinoxe Humeral Reconstruction Prosthesis | Exactech, Inc |

### Indications for Use

The Equinoxe Humeral Augmented Tray is indicated for proximal humeral replacement in conjunction with reverse shoulder arthroplasty in skeletally mature individuals with an intact medial calcar for whom the rotator cuff is irreparable or grossly deficient, a functional deltoid muscle is present, and proximal humeral bone loss is present in cases of:

- Rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic degenerative ● problems
- Congenital abnormalities in the skeletally mature ●
- Primary and secondary necrosis of the humeral head ●
- Humeral head fracture with displacement of the tuberosities ●
- Pathologies where arthrodesis or resectional arthroplasty of the humeral head are ● not acceptable
- Revisions of humeral prostheses when other treatments or devices have failed (where adequate humeral stem fixation can be achieved)
- Displaced three- and four-part upper humeral fractures
- Failed ORIF requiring TTSA with an absent or deficient greater tuberosity ●
- Revision of failed previous reconstructions when distal anchorage is required ●
- Treatment of unplanned proximal bone loss during a revision
- To restore mobility from previous procedures (e.g. previous fusion) ●

The Equinoxe Humeral Augmented Tray can be used in either primary or revision arthroplasty procedures.

## De vice De scription

The Exactech Equinoxe Humeral Augmented Trays are intended for use in total shoulder arthroplasty in skeletally mature patients with proximal humeral bone loss. Specifically, Equinoxe Humeral Augmented Trays are used in combination with Equinoxe Reverse Shoulder System components to compensate for proximal humeral bone loss in presence of an intact medial calcar, irreparable or grossly deficient rotator cuff and a functional deltoid muscle. One of the overall goals of the anatomically designed Equinoxe Humeral Augmented Trays is to improve joint mechanics and stability by providing increased deltoid muscle wrapping and adequate soft tissue tensioning in patients with proximal humeral bone loss.

The Equinoxe Humeral Augmented Trays are consisted of a humeral tray component, a modular extension (augment) and a universal locking screw for use in primary and revision cases. To accommodate patients' various anatomical needs. Equinoxe Humeral Augmented Trays are anatomically designed with two offsets/thicknesses and multiple lateralizations of left and right humeral trays and left and right modular extensions. Per

{5}------------------------------------------------

## Exactech® Equinoxe® Humeral Augmented Trays Traditional 510(k) - 510(k) Summary

surgeon evaluation, the modular extension secured with a locking screw is used to replace greater and more severe proximal humeral bone loss than can be adequately addressed with use of humeral tray component alone. When the modular extension is used, the locking screw must be used to secure it to the humeral tray component. All three components of the Equinoxe Humeral Augmented Trays are manufactured from Titanium Alloy (Ti-6Al-4V).

## Non-Clinical Testing

The following mechanical testing, engineering analysis, and cadaveric evaluation were performed to demonstrate that the Exactech Equinoxe Humeral Augmented Trays perform as intended and are substantially equivalent to the identified predicate Equinoxe Humeral Adapter Trays previously cleared in K063569 and K082702:

- Fatigue strength testing under worst-case lateral loading
- Fatigue strength testing under worst-case medial loading ●
- Range of Motion Analysis ●
- Cadaveric evaluation

Pyrogen testing was conducted in accordance with USP <161>, USP <85>, and ANSI/AAMI ST72 to ensure the proposed Equinoxe Humeral Augmented Trays meet recommended limits per FDA's Guidance Document Submission and Review of Sterility Information in Premarket (510(k)) Submission for Devices Labeled as Sterile.

## Substantial Equivalence Conclusion

Results of the non-clinical testing including mechanical testing, engineering analysis and cadaveric evaluation demonstrate the proposed Exactech Equinoxe Humeral Augmented Trays are substantially equivalent to the previously cleared Humeral Adapter Trays in the Exactech Equinoxe Reverse Shoulder System.

---

**Source:** [https://fda-staging.innolitics.com/device/K191561](https://fda-staging.innolitics.com/device/K191561)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
