K191554 · Fotona D.O.O. · GEX · Aug 30, 2019 · General, Plastic Surgery
Device Facts
Record ID
K191554
Device Name
LightWalker Laser System Family
Applicant
Fotona D.O.O.
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 30, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fotona LightWalker Laser System Family and its accessories will be marketed for the following indication added in this submission: Nd:YAG laser (1064 nm wavelength) therapy: Temporary relief of muscle and joint pain and stiffness, arthritis pain or muscle spasm, temporary increase in local blood circulation and/or promoting relaxation of muscle. Fotona LightWalker Laser System Family and its accessories will also be marketed for the following indications for use, already previously cleared under K172819: Er: YAG laser (2940 nm wavelength) in dentistry: Intra-oral soft tissue surgery (incision, excision, ablation, coagulation); Leukoplakia; Pulpotomy as adjunct to root canal retreatment; Pulp extirpation; Removal of fibromae; Removal of granulated tissue; Caries removal, cavity preparation, enamel roughening; Sulcular debridement; Tooth preparation to obtain access to root canal debridement and cleaning, root canal preparation including enlargement; Cutting, shaving, contouring and resection of oral osseous tissue (bone); Osteotomy, osseous crown lengthening, osteoplasty; Apicectomy surgery; Removal of subgingival calculi in periodontal pockets with periodontitis by closed or open curetage. Er: YAG laser (2940 nm wavelength) in dermatology and other surgical areas: Dermatology and Plastic Surgery Indications: Epidermal nevi, actinic cheilitis, verrucae, skin tags, keratoses and skin resurfacing; ENT Surgery Indications: ENT lesions, cysts, polyps, hyperkeratosis, oral leukoplakia; Oral/Maxillofacial Indications: Oral and glossal lesions, gingivectomy; General Surgery Indications: Surgical incision/excision, vaporization and coagulation of soft tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation; Podiatry Indications: Warts, plantar verrucae, large mosaic verrucae, matrixectomy; Ophtalmology Indications: Soft tissue surrounding the eye; Gynecology Indications: Herpes simplex, endometrial adhesion, CIN (Cervical intraepithelial neoplasia), cysts, condiloma; Nd:YAG laser (1064 nm wavelength) in dentistry: Excisional and incisional biopsies; Excision and vaporization of herpes simplex I and II; Exposure of unerupted teeth; Fibroma removal; Frenectomy and frenotomy; Gingival troughing for crown impressions; Gingivectomy; Gingivoplastv; Gingival incision and excision; Hemostasis; Implant recovery; Incision and drainage of abscess; Laser assisted uvulopaletoplasty (LAUP); Operculectomy; Oral papillectomies; Pulpotomy and pulpotomy as an adjunct to root canal therapy; Reduction of denture hyperplasia; Reduction of gingival hypertrophy; Removal of filling material such as gutta percha or resin as adjunct treatment during root canal therapy; Removal of post-surgical granulations; Soft tissue crown lengthening; Sulcular debridement or soft tissue curettage (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss, and tooth mobility); Tissue retraction for impression; Treatment of aphtous ulcers; Vestibuloplasty; Laser assisted new attachment procedure (cementum-mediated periodontal ligament new-attachment to the root surface in the absence of long junctional epithelium); Periodontal regeneration - true regeneration of the attachment apparatus (new cementum, new periodontal ligament, and new alveolar bone) on a previously diseased root surface. Nd:YAG laser (1064 nm wavelength) in dermatology and other surgical areas: Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin; Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to, port wine stains, hemangiomae, warts, telangiectasiae, rosacea, venus lake, leg veins and spider veins; Treatment of wrinkles; Treatment of mild to moderate inflammatory acne vulgaris; General surgery indications: surgical incision, excision, vaporization and coagulation of soft tissue. All soft tissue is included, striated and smooth tissue, muscle, cartillage, meniscus, mucous membrane, lymph vessels and nodes, organs and glands, fibroma removal; Podiatry (ablation, vaporization, excision, and coagulation of soft tissue) including: Matrixectomy, Periungual and subungual warts, Plantar warts, Radical nail excision, Neuromas; Temporary increase of clear nail in patients with onychomycosis (e.g. dermatophytes Trichophyton rubrum and T mentagrophytes and/or yeasts Candida albicans, etc.).
Device Story
LightWalker Laser System Family combines Er:YAG (2940 nm) and Nd:YAG (1064 nm) flashlamp-pumped laser sources in one housing. Er:YAG beam delivered via articulated arm to optical handpiece; Nd:YAG beam delivered via fiber optic cable to handpiece or bare fiber. System includes diode aiming beam. Operated by clinicians in dental, dermatological, and surgical settings. Microprocessor-controlled interface allows parameter selection (pulse duration, frequency, power). Output used for tissue incision, excision, ablation, coagulation, and therapeutic pain relief. Internal closed-loop water-air heat exchanger manages thermal output. Clinical benefit includes precise tissue management and non-invasive therapeutic pain relief.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and performance specifications identical to predicate devices.
Technological Characteristics
Flashlamp-pumped Er:YAG (2940 nm) and Nd:YAG (1064 nm) lasers. Microprocessor-controlled. Internal closed-loop water-air heat exchanger. Articulated arm (Er:YAG) and fiber optic (Nd:YAG) delivery. Touch screen interface. Class II aiming beam (635/650 nm or 520-532 nm).
Indications for Use
Indicated for dental, dermatological, surgical, and podiatric procedures including soft/hard tissue surgery, hair reduction, lesion removal, and pain relief. Applicable to all skin types (Fitzpatrick I-VI). Used for temporary relief of muscle/joint pain, stiffness, arthritis, and muscle spasms, and to increase local blood circulation.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 15, 2019
Fotona d.o.o. Lina Rupnik R&D Quality Department Stegne 7 Ljubljana, 1000 Si
Re: K191554
Trade/Device Name: LightWalker Laser System Family Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: June 7, 2019 Received: June 12, 2019
Dear Lina Rupnik:
This letter corrects our substantially equivalent letter of August 30, 2019.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmr/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or post-marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Purva U. Pandya -S
Purva Pandva Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K191554
#### Device Name
LightWalker Laser System Family
#### Indications for Use (Describe)
Er: Y A G laser (2940 nm wavelength) in dentistry:
- · Intra-oral soft tissue surgery (incision, excision, ablation, coagulation)
- · Leukoplakia
- · Pulpotomy as adjunct to root canal retreatment
- Pulp extirpation
- · Removal of fibromae
- · Removal of granulated tissue
- · Caries removal, cavity preparation, enamel roughening
- · Sulcular debridement
- · Tooth preparation to obtain access to root canal debridement and cleaning, root canal preparation including enlargement
- · Cutting, shaving, contouring and resection of oral osseous tissue (bone)
- · Osteotomy, osseous crown lengthening, osteoplasty
- · Apicectomy surgery
- · Removal of subgingival calculi in periodontal pockets with periodontitis by closed or open curetage
Er: Y AG laser (2940 nm wavelength) in dermatology and other surgical areas:
· Dermatology and Plastic Surgery Indications: Epidermal nevi, actinic cheilitis, verrucae, skin tags, keratoses and skin resurfacing
- ENT Surgery Indications: ENT lesions, cysts, polyps, hyperkeratosis, oral leukoplakia
- · Oral/Maxillofacial Indications: Oral and glossal lesions, gingivectomy
- · General Surgery Indications: Surgical incision, vaporization and coagulation of soft tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation
- · Podiatry Indications: Warts, plantar verrucae, large mosaic verrucae, matrixectomy
- · Ophtalmology Indications: Soft tissue surrounding the eye
· Gynecology Indications: Herpes simplex, endometrial adhesion, CIN (Cervical intraepithelial neoplasia), cysts, condiloma
Nd:YAG laser (1064 nm wavelength) in dentistry:
- · Excisional and incisional biopsies
- · Excision and vaporization of herpes simplex I and II
- · Exposure of unerupted teeth
- Fibroma removal
- Frenectomy and frenotomy
- · Gingival troughing for crown impressions
- · Gingivectomy
- Gingivoplasty
- · Gingival incision and excision
- · Hemostasis
- · Implant recovery
- Incision and drainage of abscess
- · Laser assisted uvulopaletoplasty (LAUP)
{3}------------------------------------------------
- · Operculectomy
- Oral papillectomies
- Pulpotomy and pulpotomy as an adjunct to root canal therapy
- Reduction of denture hyperplasia
- · Reduction of gingival hypertrophy
- Removal of filling material such as gutta percha or resin as adjunct treatment during root canal therapy
- Removal of post-surgical granulations
- · Soft tissue crown lengthening
· Sulcular debridement or soft tissue curettage (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss, and tooth mobility)
- Tissue retraction for impression
- Treatment of aphtous ulcers
- · Vestibuloplasty
· Laser assisted new attachment procedure (cementum-mediated periodontal ligament new-attachment to the root surface in the absence of long junctional epithelium)
• Periodontal regeneration - true regeneration of the attachment apparatus (new cementum, new periodontal ligament, and new alveolar bone) on a previously diseased root surface
Nd:YAG laser (1064 nm wavelength) in dermatology and other surgical areas:
· Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin*
- · Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to, port wine stains,
hemangiomae, warts, telangiectasiae, rosacea, venus lake, leg veins and spider veins
- Treatment of wrinkles
- · Treatment of mild to moderate inflammatory acne vulgaris
· General surgery indications: surgical incision, vaporization and coagulation of soft tissue. All soft tissue is included, striated and smooth tissue, muscle, cartillage, meniscus, mucous membrane, lymph vessels and nodes, organs and glands, fibroma removal.
· Podiatry (ablation, vaporization, incision, and coagulation of soft tissue) including: Matrixectomy, Periungual and subungual warts, Plantar warts, Radical nail excision, Neuromas.
· Temporary increase of clear nail in patients with onychomycosis (e.g. dermatophytes Trichophyton rubrum and T mentagrophytes and/or yeasts Candida albicans, etc.)
Nd:YAG laser (1064 nm wavelength) therapy:
· Temporary relief of muscle and joint pain and stiffhess, arthritis pain or muscle spassn. temporary increase in local blood circulation and/or promoting relaxation of muscle.
*Note: Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
### 510(k) Summary
### SUBMITTER'S INFORMATION
| Submitter: | Fotona d.o.o.<br>Stegne 7<br>1000 Ljubljana, Slovenia<br>Phone: +386 1 5009 100<br>Fax: +386 1 5009 200 |
|-----------------|---------------------------------------------------------------------------------------------------------|
| Contact Person: | Lina Rupnik, R&D Quality Department<br>Phone: +386 1 5009 314<br>E-mail: lina.rupnik@fotona.com |
| Date: | August 21, 2019 |
#### DEVICE INFORMATION
| Device Trade Name: | LightWalker Laser System Family |
|----------------------|-------------------------------------------------------------------------------------------------|
| Common name: | Medical Laser System |
| Classification name: | GEX-Powered Laser Surgical Instrument, General and Plastic Surgery<br>21 CFR 878.4810, Class II |
| Product Code: | GEX |
#### PREDICATE DEVICES
- LightWalker Laser System Family (K172819) .
- . Hilthera 4.0 Therapeutical Laser System (K141861)
## DEVICE DESCRIPTION
The device is Fotona LightWalker laser system family (same as K172819). It is based on Er:YAG (2940 nm) and Nd:YAG (1064 nm) laser technology. It combines two flashlamp-pumped laser sources in one housing, with optical cavities containing the Er: YAG and Nd: YAG crystals. A diode aiming beam is combined with both therapeutic laser beams. The combined therapeutic and aiming beams are guided through an articulated arm to an optical handpiece (in the case of the Er.YAG laser), or, in the case of the Nd:YAG laser, through an optical fiber delivery system to an optical handpiece or to the bare fiber distal end.
The Er: YAG laser is capable of delivering up to 1.5 J of laser energy in pulses with durations of 50-1000 us and frequencies (repetition rates) of up to 50 Hz. The maximum average output power is 20
{6}------------------------------------------------
#### K191554
W. The laser is intended to be used for incision/excision, cutting, ablation, vaporization and coagulation of soft and hard tissue in dentistry, dermatology and other surgical areas.
The Nd:YAG laser is capable of delivering up to 0.5 J of energy with pulse durations from 0.1-25 ms and frequencies of up to 100 Hz. For dental indications it is capable of delivering laser pulses with durations of up to 650 us, frequencies (repetition rates) of up to 100 Hz and a maximum output power of 15 W. The laser is intended to be used for various intraoral treatments in dentistry, for various surgical and aesthetic applications in dermatology and other surgical areas, and for pain reduction indication stated below.
#### INTENDED USE
a) The Fotona LightWalker Laser System Family and its accessories will be marketed for the following indication added in this submission:
Nd:YAG laser (1064 nm wavelength) therapy:
- Temporary relief of muscle and joint pain and stiffness, arthritis pain or muscle spasm, temporary increase in local blood circulation and/or promoting relaxation of muscle.
b) Fotona LightWalker Laser System Family and its accessories will also be marketed for the following indications for use, already previously cleared under K172819:
Er: YAG laser (2940 nm wavelength) in dentistry:
- Intra-oral soft tissue surgery (incision, excision, ablation, coagulation) .
- Leukoplakia
- Pulpotomy as adjunct to root canal retreatment .
- Pulp extirpation •
- Removal of fibromae •
- Removal of granulated tissue .
- Caries removal, cavity preparation, enamel roughening •
- Sulcular debridement
- Tooth preparation to obtain access to root canal debridement and cleaning, root canal preparation including enlargement
- Cutting, shaving, contouring and resection of oral osseous tissue (bone)
- Osteotomy, osseous crown lengthening, osteoplasty .
- Apicectomy surgery
- Removal of subgingival calculi in periodontal pockets with periodontitis by closed or open . curetage
Er: YAG laser (2940 nm wavelength) in dermatology and other surgical areas:
- Dermatology and Plastic Surgery Indications: Epidermal nevi, actinic cheilitis, verrucae, skin tags, keratoses and skin resurfacing;
- ENT Surgery Indications: ENT lesions, cysts, polyps, hyperkeratosis, oral leukoplakia; .
- Oral/Maxillofacial Indications: Oral and glossal lesions, gingivectomy;
- General Surgery Indications: Surgical incision/excision, vaporization and coagulation of soft • tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation;
{7}------------------------------------------------
# K191554
- Podiatry Indications: Warts, plantar verrucae, large mosaic verrucae, matrixectomy; .
- Ophtalmology Indications: Soft tissue surrounding the eye; •
- Gynecology Indications: Herpes simplex, endometrial adhesion, CIN (Cervical . intraepithelial neoplasia), cysts, condiloma;
Nd:YAG laser (1064 nm wavelength) in dentistry:
- Excisional and incisional biopsies
- Excision and vaporization of herpes simplex I and II •
- Exposure of unerupted teeth .
- . Fibroma removal
- Frenectomy and frenotomy •
- Gingival troughing for crown impressions .
- . Gingivectomy
- Gingivoplastv .
- Gingival incision and excision
- Hemostasis •
- Implant recovery •
- . Incision and drainage of abscess
- Laser assisted uvulopaletoplasty (LAUP)
- Operculectomy •
- Oral papillectomies •
- Pulpotomy and pulpotomy as an adjunct to root canal therapy •
- Reduction of denture hyperplasia .
- Reduction of gingival hypertrophy
- . Removal of filling material such as gutta percha or resin as adjunct treatment during root canal therapy
- Removal of post-surgical granulations .
- Soft tissue crown lengthening •
- Sulcular debridement or soft tissue curettage (removal of diseased or inflamed soft tissue in . the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss, and tooth mobility)
- Tissue retraction for impression .
- Treatment of aphtous ulcers •
- Vestibuloplasty .
- Laser assisted new attachment procedure (cementum-mediated periodontal ligament new-. attachment to the root surface in the absence of long junctional epithelium)
- . Periodontal regeneration - true regeneration of the attachment apparatus (new cementum, new periodontal ligament, and new alveolar bone) on a previously diseased root surface
Nd:YAG laser (1064 nm wavelength) in dermatology and other surgical areas:
- Removal of unwanted hair, for stable long term or permanent hair reduction and for ● treatment of PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin*
- Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to, port wine stains, hemangiomae, warts, telangiectasiae, rosacea, venus lake, leg veins and spider veins
- Treatment of wrinkles
- Treatment of mild to moderate inflammatory acne vulgaris •
- General surgery indications: surgical incision, excision, vaporization and coagulation of soft . tissue. All soft tissue is included, striated and smooth tissue, muscle, cartillage, meniscus, mucous membrane, lymph vessels and nodes, organs and glands, fibroma removal.
- Podiatry (ablation, vaporization, excision, and coagulation of soft tissue) including: .
{8}------------------------------------------------
Matrixectomy, Periungual and subungual warts, Plantar warts, Radical nail excision, Neuromas.
- Temporary increase of clear nail in patients with onychomycosis (e.g. dermatophytes Trichophyton rubrum and T mentagrophytes and/or yeasts Candida albicans, etc.).
*Note: Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The Fotona LightWalker Laser System Family has the same technological and design characteristics (design, chemical composition, energy source; wavelength, active medium, cooling system, power supply, beam delivery, controls, housing) as the previously cleared Fotona's LightWalker Laser System family Device (K172819) and the second predicate device Hilthera 4.0 Therapeutical Laser System (K141861). The output characteristics are for the intended use the same as those of the predicate devices. All lasers utilize Class II aiming beams which pose no hazard to the user. All systems are microprocessor controlled devices. The microprocessor control regulates normal operation, permits parameter selection and avoids hazard incidence. All systems utilize an internal closed loop water-air heat exchanger circuit for optimal thermal control of the risk and benefits for the Fotona LightWalker Laser System Family are identical to the predicate devices when used for similar clinical applications.
The indications for use required no changes to the device or software previously cleared to market, thus we believe LightWalker Laser System Family is substantially equivalent in terms of indications for use and technology based on technical characteristics to previously predicate devices.
A comparison of the technical specifications for the intended use of the LightWalker Laser System Family with the previously cleared devices is provided in Table 1 below.
{9}------------------------------------------------
## K191554
Table 1 : Comparison table of the technical specifications for Er:YAG and Nd:YAG laser of Fotona LightWalker Laser System Family, the primary predicate LightWalker Laser System Family (K172819) and the secondary predicate Hilthera 4.0 Therapeutical Laser System (K141861), which is the predicate for the newly submitted indication for use.
| | Fotona<br>LightWalker Family | | Jeysis Medical, Inc.<br>Hilthera 4.0 | Fotona<br>LightWalker Family | |
|----------------------|--------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | K172819 | | K141861 | K191554 | |
| Energy source | Er:YAG laser | Nd:YAG laser | Nd:YAG laser | Er:YAG laser | Nd:YAG laser |
| Intended use | See Chapter<br>5.5 b) | See Chapter<br>5.5 b) | To provide topical<br>heating for the<br>purpose of elevating<br>tissue temperature for<br>temporary relief of<br>muscle & joint pain<br>& stiffness, arthritis<br>pain or muscle<br>spasm; the temporary<br>increase in local<br>blood circulation<br>and/or promoting<br>relaxation of muscle. | See Chapter<br>5.5 b) | Temporary<br>relief of<br>muscle and<br>joint pain<br>and stiffness,<br>arthritis pain<br>or muscle<br>spasm,<br>temporary<br>increase in<br>local blood<br>circulation<br>and/or<br>promoting<br>relaxation of<br>muscle. |
| Laser mode | Pulsed | Pulsed | Pulsed | Pulsed | Pulsed |
| Wavelength | 2940 nm | 1064 nm | 1064 nm | 2940 nm | 1064 nm |
| Operating<br>voltage | 100-240 VAC | | 230 VAC | 100-240 VAC | |
| Current<br>frequency | 50/60 Hz | | 50/60 Hz | 50/60 Hz | |
| Power range | Up to 20 W | Up to 15 W | Up to 10 W | Up to 20 W | Up to 15 W |
| Fluence | Up to 48<br>J/cm² | Up to 300<br>J/cm² | 0.15-1.2 J/cm² | Up to 48<br>J/cm² | Up to 300<br>J/cm² |
| Pulse width | 50-1000 μs | 0.1-25 ms | 100-150 μs | 50-1000 μs | 0.1-25 ms |
| Repetition<br>rate | Up to 50 Hz | Up to 100<br>Hz | 10-30 Hz | Up to 50 Hz | Up to 100<br>Hz |
| Delivery<br>system | Contact and<br>non-contact<br>handpieces<br>connected to<br>the system<br>via an<br>articulated<br>arm | Contact and<br>non-contact<br>handpieces<br>connected to<br>the system<br>via a fiber<br>optic cable | Contact handpieces<br>connected to the<br>system via a fiber<br>optic cable | Contact and<br>non-contact<br>handpieces<br>connected to<br>the system via<br>an articulated<br>arm | Contact and non-contact<br>handpieces connected to the<br>system via a fiber optic<br>cable |
| Aiming beam | Laser diode 635 nm/650 nm<br>(red) or 520-532 nm<br>(green); < 1 mW | | Laser diode 655 nm<br>(red); 1 mW | Laser diode 635 nm/650 nm (red) or 520-<br>532 nm (green); < 1 mW | |
| User interface | Touch screen control | | Touch screen control | Touch screen control | |
{10}------------------------------------------------
### STATEMENT OF SUBSTANTIAL EQUIVALENCE
The LightWalker Laser System Family shares the same indication for use, similar design and functional features with the predicate devices, and therefore Fotona believes that LightWalker Laser System Family is substantially equivalent in terms of the indications for use and technology based on technical characteristics to LightWalker Laser System Family (K172819) and to Hilthera 4.0 Therapeutical Laser System (K141861).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.