← Product Code [GEI](/productcode/GEI) · K191542

# Apyx Plasma/RF Handpiece (K191542)

_Bovie Medical Corporation Dba Apyx Medical Corporation · GEI · Oct 11, 2019 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K191542

## Device Facts

- **Applicant:** Bovie Medical Corporation Dba Apyx Medical Corporation
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Oct 11, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Apyx Plasma/RF Handpiece is intended to be used with compatible electrosurgical generators for the delivery of radiofrequency energy and/or helium plasma for cutting, coagulation and ablation of soft tissue during open surgical procedures. The Apyx Plasma/RF Handpiece is compatible with the Electrosurgical Generators, BVX-200H and BVX-200P, owned by Apyx Medical.

## Device Story

The Apyx Plasma/RF Handpiece is a sterile, single-use monopolar electrosurgical device used in open surgery. It connects to Apyx Medical's BVX-200H/P generators, which supply radiofrequency (RF) energy and helium gas. The device transforms these inputs into helium plasma at the electrode tip, which is used by surgeons for cutting, coagulation, and ablation of soft tissue. The handpiece features a non-extendable electrode, a 3.0 mm shaft (15cm or 27cm length), and a side-port configuration for plasma exit. It is activated via a hand button or footswitch. By delivering controlled thermal energy, the device enables precise tissue management. Bench and animal studies confirm that the device maintains safe tissue temperature profiles and provides equivalent thermal effects to predicate devices, benefiting patients through effective soft tissue handling without requiring skin temperature monitoring.

## Clinical Evidence

No human clinical data. Evidence consists of bench testing (mechanical functionality, energy/power output, electrical safety/EMC per IEC 60601-1/2-2), biocompatibility (ISO 10993), and pre-clinical/ex-vivo studies. Ex-vivo studies on muscle, liver, and kidney tissue demonstrated equivalent thermal effects (depth/lateral spread) to predicates. Live porcine model testing confirmed safe epidermal temperatures (<41°C) during worst-case scenarios.

## Technological Characteristics

Monopolar electrosurgical handpiece; 15cm or 27cm shaft length; 3.0mm shaft diameter; non-extendable electrode; side-port plasma exit; helium gas/RF energy source; compatible with BVX-200H/P generators; sterile, single-use; complies with ANSI/AAMI/IEC 60601-1 and 60601-2-2; biocompatible per ISO 10993.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Renuvion®/J-Plasma® Precise Open Handpiece ([K183610](/device/K183610.md) & [K170188](/device/K170188.md))
- Renuvion/J-Plasma® Precise Handpiece ([K183610](/device/K183610.md))

## Reference Devices

- Thermi Temperature Controlled Radiofrequency (RF) System ([K173582](/device/K173582.md))
- Thermi Injectable RF Electrodes ([K170116](/device/K170116.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 11, 2019

Apyx Medical Corporation (formerly Bovie Medical) Topaz Kirlew Vice President QA & RA 5115 Ulmerton Road Clearwater, Florida 33760

Re: K191542

Trade/Device Name: Apyx Plasma/RF Handpiece Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: June 10, 2019 Received: June 11, 2019

Dear Topaz Kirlew:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K191542

#### Device Name

Apyx Plasma/RF Handpiece (APYX-15-SP, APYX-15-TP and APXY-27-TP)

#### Indications for Use (Describe)

The Apyx Plasma/RF Handpiece is intended to be used with compatible electrosurgical generators for the delivery of radiofrequency energy and/or helium plasma for cutting, coagulation of soft tissue during open surgical procedures. The Apyx Plasma/RF Handpiece is compatible with the Electrosurgical Generators, BVX-200H and BVX-200P, owned by Apyx Medical.

| Type of Use (Select one or both, as applicable) |                                                                                                    |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------|
|                                                 | <span> <span style="font-size:12px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)  </span> |
|                                                 | <span> <span style="font-size:12px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  </span>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for Aryx Medical. The logo features the word "Aryx" in orange, with a blue triangle replacing the "A". Below "Aryx" is the word "MEDICAL" in blue. The logo is simple and modern, with a focus on the company name.

## K191542

### 1. General Information:

| Submitted by:                 | Apyx Medical Corporation (formerly Bovie Medical)<br>5115 Ulmerton Road Clearwater, Florida 33760-4004<br>United States of America                                                                                                                                                                       |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration #: | 3007593903                                                                                                                                                                                                                                                                                               |
| Submitter FAX Number:         | (727) 322-4465                                                                                                                                                                                                                                                                                           |
| Contact Person:               | Dr. Topaz Kirlew, Vice President, QA & RA<br>5115 Ulmerton Road<br>Clearwater, Florida 33760 -4004<br>United States of America<br>Phone: (727) 803-8617<br>Email: topaz.kirlew@apyxmedical.com                                                                                                           |
| Date Prepared:                | June 10, 2019                                                                                                                                                                                                                                                                                            |
| Trade Names (Model Numbers):  | Apyx Plasma/RF Handpiece<br>(APYX-15-SP, APYX-15-TP, APYX-27-TP)                                                                                                                                                                                                                                         |
| Common Name:                  | Electrosurgical Handpiece                                                                                                                                                                                                                                                                                |
| Classification:               | Class II per 21CFR 878.4400 - Electrosurgical<br>Cutting and Coagulation Device and Accessories<br>Product Code GEI                                                                                                                                                                                      |
| Predicate Devices:            | Primary Predicate Devices<br>Renuvion®/J-Plasma® Precise Open Handpiece<br>(K183610 & K170188)<br><br>Renuvion/J-Plasma® Precise Handpiece<br>(K183610)<br><br>Reference Devices<br>Thermi Temperature Controlled Radiofrequency (RF)<br>System (K173582) & Thermi Injectable RF Electrodes<br>(K170116) |

#### 2. Indications for Use:

The Apyx Plasma/RF Handpiece (HP) is intended to be used with compatible electrosurgical generators for the delivery of radiofrequency energy and/or helium plasma for cutting, coagulation and ablation of soft tissue during open surgical procedures. The Apyx Plasma/RF

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for Apyx Medical. The logo is primarily orange, with the word "Apyx" in a stylized font. A small blue triangle is present to the left of the word. Below "Apyx" is the word "MEDICAL" in blue, with a horizontal blue line underneath.

Handpiece is compatible with the Electrosurgical Generators BVX-200H and BVX-200P owned by Apyx Medical.

### 3. Device Description:

The Apyx Plasma/RF Handpiece is a sterile, single use electrosurgical (monopolar) device intended to be used in conjunction with compatible generators for the percutaneous delivery of radiofrequency energy and/or helium plasma for cutting, coagulation and ablation of soft tissue. The compatible Generators operate at an adjustable power of up to 40W (expressed as 0-100% where 100% is 40W) and provide an adjustable helium gas flow of 1-5 LPM. Radiofrequency energy is delivered to the handpiece by the generator and used to energize the electrode. When helium gas is passed over the energized electrode, a helium plasma is generated for soft tissue cutting, coagulation or ablation.

The Apyx Plasma/RF Handpiece has a non-extendable electrode to generate helium plasma. The handpiece is available in two different lengths: 15cm and 27cm. The 15cm length is available in a single port (APYX-15-SP) as well as a twin port configuration (APYX-15-TP). The 27cm length is only available with a twin port configuration (APYX-27-TP). The Apyx Plasma/RF System with all components and accessories is depicted in the picture below.

Image /page/4/Figure/9 description: The image shows a medical device setup. A red helium tank is connected to a regulator with a pressure gauge. A generator (ESU) is connected to a handpiece and cable, a foot switch, and a grounding pad.

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the logo for Apyx Medical. The logo features the word "Apyx" in orange, with a blue triangle replacing the left side of the "A". Below "Apyx" is the word "MEDICAL" in blue. The logo is simple and modern, with a focus on the company name.

## K191542

### 4. Technological Characteristics:

The Apyx Plasma/RF Handpiece is a modification to the primary predicate device that was cleared in K183610 & K170188 (Renuvion®/J-Plasma® Precise Open Handpiece). There are no changes to indications for use, principle of operation, mechanism of action or sterilization methods.

The Apyx Plasma/RF Handpiece as compared to the primary predicate device (K183610 & K170188) has only the following differences in technological characteristics:

- V All models will have a side port configuration to allow the plasma to exit the handpiece through the side vs. from the point tip.
- > All 3 configurations will have a 'bullet shape' point tip and the electrode is not extendable.
- V The handle is thicker, and the shaft diameter is smaller (3.0 mm shaft vs. 5.0 mm shaft) and more flexible with tip distance indicators at the end.
- > Shaft length of one model is 27cm long (the primary predicate device used for clearance of the Renuvion/J-Plasma® Precise Open Handpiece is the Renuvion/J-Plasma® Handpiece (K151325) that has a shaft length of 27cm and was also cleared in K183610).

The Apyx Plasma/RF Handpiece with its components is pictured below.

Image /page/5/Figure/13 description: The image shows a medical device with several labeled components. The device has a long, thin flexible shaft with energy exit portals and tip distance indicators at the end. The device also has a handle with an activation button, a plug connector, and a cable.

Results of bench and animal testing demonstrate that the Apyx Plasma/RF Handpiece is at least as safe and effective as the primary predicate devices.

#### 5. Performance Data:

#### a. Bench Testing

All testing performed on the Apyx Plasma/RF System was derived from the risk assessment in accordance with ISO 14971 which evaluated the safety and effectiveness

{6}------------------------------------------------

Image /page/6/Picture/2 description: The image shows the logo for Apyx Medical. The logo is primarily orange, with the word "Apyx" in a stylized font. A small blue triangle is positioned to the left of the word. Below "Apyx", the word "MEDICAL" is written in blue, using a simpler, sans-serif font.

K191542

of the design modification in accordance with Apyx Medical's Design and Development procedures. The test methodology and acceptance criteria were developed from the same standards and internal Design and Development procedures used for clearance of the primary predicate devices owned by Apyx Medical. Performance testing to assure that the subject Apyx Plasma/RF System met performance requirements was conducted in accordance with protocols to verify design specifications. The testing performed are summarized in the following table:

| Test                                          | Objective                                                                                                                                                                                                                                                                                                                                                                                                                                            | Result                                                                                                                                                                                                                                                                   |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mechanical<br>Verification &<br>Functionality | Verify Mechanical functionality of the<br>Apyx Plasma/RF Handpiece                                                                                                                                                                                                                                                                                                                                                                                   | Mechanical functionality<br>requirements met                                                                                                                                                                                                                             |
| Energy & Power<br>Output                      | Verify via Calorimetric testing that the<br>tissue effects are equivalent between<br>the devices sharing the Renuvion/J-<br>Plasma technology that are owned<br>by Apyx Medical –  Subject device (Apyx Plasma/RF<br>Handpiece); Primary Predicate device cleared<br>under K183610 & K170188<br>(Renuvion/J-Plasma® Precise<br>Open Handpiece); and Predicate device cleared under<br>K183610 & K151325<br>(Renuvion/J-Plasma® Precise<br>Handpiece) | There are no significant<br>differences in energy and<br>power outputs between the<br>subject device, Primary<br>Predicate devices owned<br>by Apyx Medical; all 3<br>devices will deliver the<br>same amount of energy to<br>tissue at equivalent<br>generator settings |

#### b. Electrical Safety and Electromagnetic Compatibility (EMC)

Electrical safety and EMC testing were conducted on the Apyx Plasma/RF System. The handpiece complies with the ANSI/AAMI/IEC 60601-1:2015/(R)2012 and A1:2012, and AAMI/ANSI/IEC 60601-2-2:2017 standards for safety and the AAMI/ANSI/IEC 60601-1-2:2014 (4th Edition) standard for EMC. The Apyx Plasma/RF System was determined to be in conformance with these standards.

#### c. Biocompatibility Testing

The biocompatibility evaluation for the Apyx Plasma/RF Handpiece was conducted in accordance with the June 16, 2016 FDA Guidance, "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and the AAMI/ANSI/ISO 10993-1:2009/(R2013) standard,

{7}------------------------------------------------

Image /page/7/Picture/2 description: The image shows the logo for Aryx Medical. The logo is orange and blue. The word "Aryx" is in orange, with a blue triangle in place of the letter "A". Below the word "Aryx" is the word "MEDICAL" in blue.

"Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process". The battery of testing was either conducted or was previously submitted to the FDA in 510(k)s cleared under the predicate devices (K183610, K151325 & K170188). The battery of evaluations included the following testing:

- Cvtotoxicity
Sensitization

- Systemic toxicity ■
- ■
- 트 Irritation

■

- Hemolysis
- Material Mediated Pyrogen/Pyrogenicity

The Apyx Plasma/RF Handpiece (subject device) is an external communicating device with an indirect blood path contact for a duration of less than 24 hours. Biocompatibility has been established per ISO 10993 guidelines for this category. Biocompatibility verification testing was satisfactorily conducted for the subject device.

### 6. Pre-Clinical & Ex-Vivo Studies:

- a. Tissue Temperature Profile Equivalency Pre-clinical testing was conducted to establish and compare the tissue temperature profiles of the devices owned by Apyx Medical and that share the Renuvion/J-Plasma technology: the subject device (Apyx Plasma/RF Handpiece), the predicate devices (Renuvion/J-Plasma® Precise Open Handpiece and Renuvion/J-Plasma® Precise Handpiece). Testing involved device activation on porcine skin while capturing the skin temperatures using a thermal video camera. Testing demonstrated that the three devices heat tissue in an equivalent manner.
- b. Maximum Temperature Study for Renuvion/J-Plasma Technology Family A preclinical study was conducted in a live porcine model to quantify the internal and external tissue temperatures (via a thermal video camera) resulting from the use of the Renuvion/J-Plasma® technology for soft tissue coagulation and to determine the effect of multiple variables on the tissue temperatures. These temperatures were recorded after performing six consecutive passes of the Renuvion/J-Plasma® technology under the same area of tissue to represent a "worst case" scenario. However, even in this "worst case" scenario, the epidermal surface temperature remained below 41°C and within safe limits (below 47°C) *. The results of this testing demonstrate that safe skin temperatures are maintained with the Renuvion/J-Plasma® Technology and skin temperature monitoring is not indicated due to this unique mechanism of action. This is different from devices such as the Reference Device (Thermi Temperature Controlled Radiofrequency (RF) System, K173582 & Thermi Injectable RF Electrodes, K170116) that work on the principle of bulk tissue heating in order to achieve coagulation of soft tissue and where monopolar energy is primarily directed into the dermis thus necessitating skin temperature monitoring.

{8}------------------------------------------------

Image /page/8/Picture/2 description: The image shows the logo for Apyx Medical. The logo is orange and features the company name in a stylized font. The word "MEDICAL" is written in blue below the company name. 

## K191542

- c. Tissue Thermal Effect Equivalency Test for Apyx Plasma RF Device An ex-vivo study was conducted on several tissue types (i.e., muscle, liver, kidney) to compare the tissue thermal effect of the Apyx Plasma/RF Handpiece and the predicate device. The quantitative results of this testing demonstrated that the tissue effects of all the subject device models (APYX-27-TP, APYX-15-TP and APYX-15-SP) are equivalent to the predicate device's effect over a range of power and flow settings that are clinically relevant.

| Test                                                      | Objective                                                                                                                                                                                                                                                                                                                             | Result                                                                                                                                                                                                                                                                                                                                                                                         |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tissue<br>Temperature<br>Profile<br>Equivalency           | Compare the tissue temperature<br>profiles of the devices owned by<br>Apyx Medical and that share the<br>Renuvion/J-Plasma® technology<br>(subject device & primary predicate<br>devices, ) via device activation on<br>porcine skin while capturing the<br>skin temperatures using a thermal<br>video camera.                        | The subject device (Apyx Plasma/RF<br>Handpiece) and  predicate devices<br>(Renuvion/J-Plasma® Precise Open<br>Handpiece & Renuvion/J-Plasma®<br>Precise Handpiece) all heat tissue in<br>an equivalent manner and are<br>substantially equivalent in in terms of<br>tissue effect for the coagulation of<br>soft tissue.                                                                      |
| Maximum<br>Internal &<br>External<br>Temperature<br>Study | Quantify the internal and external<br>tissue temperatures (via a thermal<br>video camera) resulting from the<br>use of the Renuvion/J-Plasma®<br>technology for soft tissue<br>coagulation in a worst-case<br>scenario and to determine the<br>effect of multiple variables on the<br>tissue temperatures in a live<br>porcine model. | The porcine skin epidermal surface<br>temperature remained below 41°C<br>and within safe limits (below 47°C) *<br>even with 6 passes in the same area.<br>The results of this testing<br>demonstrate that safe skin<br>temperatures are maintained with the<br>Renuvion/J-Plasma® Technology and<br>skin temperature monitoring is not<br>indicated due to this unique<br>mechanism of action. |
| Thermal<br>Tissue Effect<br>Comparison                    | Verify that the Apyx Plasma/RF<br>Handpiece produces equivalent ex-<br>vivo tissue effect when compared<br>to the Predicate Device.                                                                                                                                                                                                   | The depth and lateral spread (i.e.,<br>average of length and width) of<br>thermal effect measured for the Apyx<br>Plasma/RF handpiece were<br>equivalent to the predicate device's<br>thermal effect as measured over a<br>range of power and flow settings<br>which include worst-case clinical<br>scenarios.                                                                                 |

The pre-clinical and ex-vivo testing discussed above is summarized in the table below:

* Chen SS, Wright NT, Humphrey JD. Heat-induced changes in the mechanics of a collagenous tissue: isothermal free shrinkage. Journal of Biomechanical Engineering 1997:109:372-378.

{9}------------------------------------------------

Image /page/9/Picture/2 description: The image shows the logo for Apyx Medical. The logo features the word "Apyx" in a combination of blue and orange colors, with a small blue triangle replacing the dot over the "i". Below the word "Apyx", the word "MEDICAL" is written in blue, in a smaller, bolder font.

# K191542

### 7. Substantial Equivalence:

| Feature/<br>Characteristic                            | Subject Device<br>Apyx Plasma/RF HP                                                                                                                                                                                                                                                                                                                                               | Primary Predicate Device<br>Renuvion/J-Plasma Precise Open HP (K183610 & K170188)                                                                                                                                                                                                                                                                                                                   |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use                                | The Apyx Plasma/RF Handpiece is intended to be used with compatible electrosurgical generators for the delivery of radiofrequency energy and/or helium plasma for cutting, coagulation and ablation of soft tissue during open surgical procedures.<br>The Apyx RF/Plasma Handpiece is compatible with the Electrosurgical Generators BVX-200H and BVX-200P owned by Apyx Medical | The Renuvion/J-Plasma® Precise Open Handpiece is intended to be used with compatible J-Plasma electrosurgical generators for the delivery of radiofrequency energy and/or helium plasma for cutting, coagulation and ablation of soft tissue during open surgical procedures.<br>The Renuvion/J-Plasma® Precise Open Handpiece is compatible with electrosurgical Generators BVX-200H and BVX-200P. |
| Energy<br>Source                                      | RF Generator; Only with Generators owned by Apyx Medical                                                                                                                                                                                                                                                                                                                          | RF Generator; Only Generators owned by Apyx Medical                                                                                                                                                                                                                                                                                                                                                 |
| Energy Type                                           | Helium gas plasma;                                                                                                                                                                                                                                                                                                                                                                | Helium gas plasma; Monopolar coagulation                                                                                                                                                                                                                                                                                                                                                            |
| Electrical<br>Currents<br>Transmitted                 | 150mA – 250mA                                                                                                                                                                                                                                                                                                                                                                     | 150mA – 250mA                                                                                                                                                                                                                                                                                                                                                                                       |
| Output                                                | Monopolar                                                                                                                                                                                                                                                                                                                                                                         | Monopolar                                                                                                                                                                                                                                                                                                                                                                                           |
| Device<br>Activation                                  | Hand activation button or optional footswitch                                                                                                                                                                                                                                                                                                                                     | Hand activation button or optional footswitch                                                                                                                                                                                                                                                                                                                                                       |
| System<br>Components                                  | The Apyx Plasma/RF System consists of:<br>RF GeneratorDisposable HPFoot pedalPower cordGas regulatorGas cylinder                                                                                                                                                                                                                                                                  | The Renuvion/J-Plasma Precise Open System consists of:<br>RF GeneratorDisposable HPFoot pedalPower cordGas regulatorGas cylinder                                                                                                                                                                                                                                                                    |
| User Interface                                        | Straight                                                                                                                                                                                                                                                                                                                                                                          | Straight                                                                                                                                                                                                                                                                                                                                                                                            |
| Shaft Design<br>& Energy<br>Delivery<br>Configuration | 15cm and 27cm long with a side port configuration (single or twin) and indicator lines on the shaft                                                                                                                                                                                                                                                                               | 15cm and 4.4cm long with a distal port delivery configuration                                                                                                                                                                                                                                                                                                                                       |
| Feature/<br>Characteristic                            | Subject Device<br>Apyx Plasma/RF HP                                                                                                                                                                                                                                                                                                                                               | Primary Predicate Device<br>Renuvion/J-Plasma Precise Open HP (K183610 & K170188)                                                                                                                                                                                                                                                                                                                   |
| Shaft Outer<br>Diameter                               | 3mm                                                                                                                                                                                                                                                                                                                                                                               | 5mm                                                                                                                                                                                                                                                                                                                                                                                                 |
| Electrode<br>Configuration                            | Non-extendable                                                                                                                                                                                                                                                                                                                                                                    | Extendable                                                                                                                                                                                                                                                                                                                                                                                          |
| Plasma<br>Settings                                    | Maximum 40 watts, 1-5 L/min gas flow                                                                                                                                                                                                                                                                                                                                              | Maximum 40 watts, 1-5 L/min gas flow                                                                                                                                                                                                                                                                                                                                                                |
| Compatibility                                         | Only with Electrosurgical Generators BVX-200H and BVX-200P owned by Apyx Medical                                                                                                                                                                                                                                                                                                  | Only with Electrosurgical Generators BVX-200H and BVX-200P owned by Apyx Medical                                                                                                                                                                                                                                                                                                                    |
| Connector                                             | Company Proprietary – per design change reflected in K183610 that establishes both pneumatic seal and electrical connection                                                                                                                                                                                                                                                       | Company Proprietary - per design change reflected in K183610 that establishes both pneumatic seal and electrical connection                                                                                                                                                                                                                                                                         |

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for Apyx Medical. The logo is primarily orange, with the word "Apyx" in a stylized font. A blue triangle is incorporated into the "A" of "Apyx." Below "Apyx," the word "MEDICAL" is written in blue, indicating the company's focus on the medical field.

K191542

## K191542

Since the predicate devices were cleared with the results of non-clinical testing, non-clinical testing was also used to support substantial equivalence for the subject Apyx Plasma/RF System. The subject device has the same indications for use, operational principle and very similar technological characteristics as the predicate devices that were cleared in K183610 & K170188 (Renuvion®J-Plasma® Precise & Precise Open Handpieces). The conclusions drawn from the bench and pre-clinical testing demonstrate that the Apyx Plasma/RF is at least as safe and effective as the legally marketed predicate device for the same indications. Functional and mechanical verification testing demonstrate that the Apyx Plasma/RF System should perform as intended in the specified use conditions. The results of the bench, pre-clinical and ex-vivo tissue studies demonstrate that the Apyx Plasma/RF Handpiece performs comparably to the predicate device.

#### 8. Conclusion:

The Apyx Plasma/RF Handpiece is a modification to the primary predicate device that was cleared in K183610 & K170188 (Renuvion/J-Plasma® Precise Open Handpiece). There are no changes to indications for use, principle of operation, mechanism of action or sterilization methods. The subject device (Apyx Plasma/RF Handpiece) like the primary predicate device (Renuvion J-Plasma® Precise Open Handpiece, K183610 & K170188) is intended for cutting, coagulation, and ablation of soft tissue and both devices utilize the same energy source (Electrosurgical Generators BVX-200H and BVX-200P) with no changes to the principle of operation.

Technological differences between the Apyx Plasma/RF Handpiece (subject device), the Renuvion/J-Plasma® Precise Open Handpiece (predicate device cleared in K183610 & K170188), and the Renuvion/J-Plasma® Precise Handpiece (predicate device cleared in

{11}------------------------------------------------

Image /page/11/Picture/2 description: The image shows the logo for Apyx Medical. The logo is orange and blue. The word "Apyx" is in orange, with a blue triangle in place of the "A". Below the word "Apyx" is the word "MEDICAL" in blue.

K183610) that all share the Renuvion/J-Plasma® technology, do not raise any new concerns of safety or efficacy as:

- 1. Bench testing demonstrated that all three devices owned by Apyx Medical will deliver the same amount of enerqy to tissue at equivalent generator settings.
- 2. Testing utilizing porcine skin demonstrated that there is no difference in the external skin temperature profiles of these 3 devices owned by Apyx Medical.
- Pre-clinical testing demonstrated safe internal and external tissue temperatures 3. for the Renuvion/J-Plasma® Technology.
- 4. Since the energy delivered to the tissue has been demonstrated to be equivalent, it can be concluded that the histologic effect of the energy on the tissue will be equivalent between all 3 devices owned by Apyx Medical.
- 5. Ex-vivo tissue testing for thermal equivalency demonstrated that depth and lateral spread (i.e., average of length and width) of thermal effect measured for the Apyx Plasma/RF handpiece are equivalent to the predicate device over a range of power and flow settings which include worst-case clinical scenarios.

The subject device's safety and performance have been confirmed by results of the performance bench, ex-vivo tissue, and animal testing and the design change has gone through the design controls process to demonstrate that the device performs as intended. Therefore, there are no differences that would raise new or different questions regarding safety or effectiveness as the tissue effects of all the subject device models (APYX-27-TP, APYX-15-TP and APYX-15-SP) are equivalent to the predicate device's effect over a range of power and flow settings that are clinically relevant.

Results of bench and animal testing demonstrate that the Apyx Plasma/RF Handpiece is as and effective as the predicate devices (Renuvion/J-Plasma® Precise Open safe Handpiece (K183610 & K170188) & Renuvion/J-Plasma® Precise Handpiece (K183610).

---

**Source:** [https://fda-staging.innolitics.com/device/K191542](https://fda-staging.innolitics.com/device/K191542)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
