1.5T Endorectal Coil, 3.0T Endorectal Coil, ER Coil Support
K191539 · Rapid Biomedical GmbH · MOS · Jan 31, 2020 · Radiology
Device Facts
Record ID
K191539
Device Name
1.5T Endorectal Coil, 3.0T Endorectal Coil, ER Coil Support
Applicant
Rapid Biomedical GmbH
Product Code
MOS · Radiology
Decision Date
Jan 31, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Endorectal Coil is indicated for use as diagnostic device extension for GE 1.5 T MR Systems and GE 3.0 T MR Systems to produce transverse, sagittal, coronal and oblique images, spectroscopic images and/or spectra, displaying the internal structure of the prostate. These images when interpreted by a trained physician, yield information that may assist in diagnosis.
Device Story
Endorectal coil; receive-only MR coil; used with GE 1.5T/3.0T MR systems; captures RF signals from hydrogen nuclei excited by system body coil; drop-shaped housing for patient comfort; includes ER Coil Support for stabilization with five degrees of freedom; operated by radiology staff in clinical MR environment; provides high-resolution prostate images; interpreted by physicians to assist in diagnosis; improves SNR and image quality for prostate assessment.
Clinical Evidence
In vivo clinical testing performed using standard diagnostic sequences (T2 weighted, Propeller, DWI) in axial and sagittal orientations. Evaluated image quality metrics including SNR, penetration depth, contrast, resolution, and artifact robustness. Results demonstrated adequate image quality and patient comfort. No statistical metrics (p-values/CIs) reported; evidence based on qualitative assessment of DICOM clinical images.
Technological Characteristics
Receive-only single-loop coil; integrated low-noise preamplifier; fixed tuned/matched to 1.5T (63.9 MHz) or 3.0T (127.7 MHz); includes decoupling circuits. Conforms to AAMI/ANSI ES60601-1, IEC 60601-2-33, and NEMA MS 6-2008. Drop-shaped housing; includes mechanical ER Coil Support with knurled screw locking mechanism.
Indications for Use
Indicated for patients undergoing prostate MRI examinations on GE 1.5T or 3.0T MR systems to assist in diagnosis.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
MEDRAD 1.5T Pelvic Imaging System Interface Device (K053042)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized caduceus. To the right of that is the FDA logo, which is the acronym FDA in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 31, 2020
RAPID Biomedical GmbH % Mr. Christian Zimmermann Ouality Management & Official Correspondent Kettelerstrasse 3-11 97222 Rimpar, Bavaria GERMANY
Re: K191539
Trade/Device Name: 1.5T Endorectal Coil. O-HLE-015-01899 and O-HLE-015-01946. 3.0T Endorectal Coil, O-HLE-030-01900; ER Coil Support, ZUB-01955 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: September 17, 2019 Received: December 23, 2019
Dear Mr. Zimmermann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K191539
Device Name
Indications for Use (Describe)
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------|----------------------------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo for Rapid Biomedical. The word "RAPID" is in large, bold, black letters at the top. Below that, the word "Biomedical" is in smaller, bold, black letters. At the bottom, in a smaller font, it says "MR Coils - Made to Measure".
510(k)#: K [For a new submission, leave the 510(k) number blank.]
1.5T Endorectal Coil - O-HLE-015-01899 and O-HLE-015-01946 3.0T Endorectal Coil - O-HLE-030-01900
#### 510(k) Summary K191539
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92.
## General Information
Date of summary preparation: 2019-06-03
#### 1. Manufacturer
RAPID Biomedical GmbH Kettelerstrasse 3-11 97222 Rimpar, Bavaria, Germany FEI: 3005049692
#### 2. Distributed by
GE Healthcare 3200 Grandview Blvd Waukesha, WI 53188 FEI: 2124823
#### 3. Contact Person
Mr. Christian Zimmermann RAPID Biomedical GmbH Kettelerstr. 3-11 97222 Rimpar, Bavaria, Germany Phone: +49-9365-8826-0 Fax: +49-9365-8826-99 email: christian.zimmermann@rapidbiomed.de
- 4. Type of Submission Traditional 510(k) Premarket Notification
- Classification and Device Name 5. Classification Panel: Radiology Classification Name: Magnetic Resonance Diagnostic Device Accessory Device Class: Class II [21 CFR § 892.1000] Product Code: MOS Product Nomenclature: Coil, Magnetic Resonance, Specialty Common Name: Special Purpose Coil Trade Names: 1.5T Endorectal Coil and 3.0T Endorectal Coil
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for Rapid Biomedical. The word "RAPID" is in large, bold, black letters, with the word "Biomedical" underneath in a smaller font. Below that, the text "MR Coils - Made to Measure" is visible in a smaller font size.
510(k)#: K [For a new submission, leave the 510(k) number blank.]
MR Coils - Made to Measure
1.5T Endorectal Coil - O-HLE-015-01899 and O-HLE-015-01946 3.0T Endorectal Coil - O-HLE-030-01900
### Safety and Effectiveness Information Supporting Substantial Equivalence
#### Indications for Use
The Endorectal Coil is indicated for use as diagnostic device extension for GE 1.5 T MR Systems and GE 3.0 T MR Systems to produce transverse, sagittal, coronal and oblique images, spectroscopic images and/or spectra, displaying the internal structure of the prostate.
These images when interpreted by a trained physician, yield information that may assist in diagnosis.
#### Device Description
The Endorectal Coil (1.5T Endorectal Coil O-HLE-015-01899, 1.5T Endorectal Coil O-HLE-015-01946, 3.0T Endorectal Coil O-HLE-030-01900 and ER Coil Support ZUB-01955) is designed for use with a magnetic resonance (MR) system. The coil is designed to work in unison with the Body Coil (BC) of the MR system, which will excite the hydrogen (1H) nuclei with radio frequency (RF) magnetic fields, so that the coil may receive the resultant RF signal from the excited nuclei. The coil is designed as a reusable receive-only coil for high resolution MR examination of the prostate.
The coil housing is minimum sized and drop-shaped for better patient comfort. It features a flat top to minimize the distance of the inside receive coil electronics to the prostate. The coil is receive-only (Rx) and consists of a single loop coil element with an integrated low noise preamplifier and a connector to the GE 1.5 T MR-System or GE 3.0 T MR-System. The coil is fixed tuned and matched to the typical loading condition of a prostate examination at the Larmor frequency of 1H at 1.5 T (63.9 MHz) or 3.0 T (127.7 MHz), respectively. Decoupling circuits are integrated in the single loop element providing a decoupling from the Body Coil of the MR System during transmission of the RF excitation pulse.
It is recommended to employ an Endorectal Coil Model in combination with the additional available ER Coil Support. The ER Coil Support is designed for use with any Endorectal Coil Model (1.5T Endorectal Coil O-HLE-015-01899, 1.5T Endorectal Coil O-HLE-015-01946 and 3.0T Endorectal Coil O-HLE-030-01900). It supports stabilizing the Endorectal Coil in any position required by each individual endorectal MR examination. The ER Coil Support features a collet for acceptance of the Endorectal Coil. The Endorectal Coil is fixated inside the collet by tightening of a knurled screw. It offers five degrees of freedom to align the position of the collet with the required spatial position of the Endorectal Coil housing. Two additional knurled screws allow the lock the ER Coil Support in the desired alignment.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for "Rapid Biomedical". The word "RAPID" is in large, bold, black letters. Below that, the word "Biomedical" is in smaller, bold, black letters with a line underneath. Underneath that, the text "MR Coils - Made to Measure" is in a smaller, black font.
510(k)#: K [For a new submission, leave the 510(k) number blank.]
1.5T Endorectal Coil - O-HLE-015-01899 and O-HLE-015-01946 3.0T Endorectal Coil - O-HLE-030-01900
### Equivalency Information
RAPID Biomedical believes that the SUBJECT DEVICE is substantially equivalent to the cleared PREDICATE DEVICE which is described in the following submission.
| PREDICATE DEVICE | Clearance Number | Date |
|-------------------------------------------------------|------------------|------------|
| MEDRAD 3.0T eCoil Imaging System | K060401 | 03/15/2006 |
| MEDRAD 1.5T Pelvic Imaging System<br>Interface Device | K053042 | 11/17/2005 |
#### Summary of Technological Characteristics of the SUBJECT DEVICE as compared with the PREDICATE DEVICE
The proposed labelling is adjusted compared to the PREDICATE DEVICE with respect to differing MR Systems, changes in the currently applicable standards and for better usability.
Overall, the indications for use, the intended use and the coil technology and safety are substantially equivalent.
#### General Safety and Effectiveness Concerns
The safety and performance parameters for SUBJECT DEVICE according to the FDA Guidance Document "Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices, issued on Nov. 18, 2016" are equivalent or better than for the PREDICATE DEVICE.
### Nonclinical Tests
Nonclinical tests were performed on a phantom to evaluate the SUBJECT DEVICE concerning applicable imaging parameters Signal to Noise Ratio (SNR) and Image Uniformity. Tests were conducted following NEMA Standards. Test results show an enhanced SNR performance of the SUBJECT DEVICE, while determined Image Uniformity compares well to the results from the PREDICATE DEVICE.
The SUBJECT DEVICE is conforming to:
- AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, c1:2009/(r)2012 and . a2:2010/(r)2012 (consolidated text) medical electrical equipment - part 1: general requirements for safety and essential performance (iec 60601-1;2005, mod), (General II (ES/EMC)). FDA Recognition number: #19-4.
- IEC 60601-2-33 Ed. 3.2 b:2015: Medical electrical equipment Part 2-33: Particular . requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis. FDA Recognition number: #12-295.
- NEMA Standards Publication MS 6-2008 (R2014) Determination of Signal-to-Noise . Ratio and Image Uniformity for Single-Channel Non-Volume Coils in Diagnostic MR Imaging. FDA Recognition number: #12-195.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Rapid Biomedical. The logo has the word "RAPID" in large, bold, black letters on the top line. The word "Biomedical" is in a smaller font size and is located below the word "RAPID". The phrase "MR Coils - Made to Measure" is located at the bottom of the logo.
510(k)#: K [For a new submission, leave the 510(k) number blank.]
1.5T Endorectal Coil - O-HLE-015-01899 and O-HLE-015-01946 3.0T Endorectal Coil - O-HLE-030-01900
All safety tests were performed on GE MR Systems. The test results are approved by GE for proving the full system compatibility.
#### Clinical Tests
Clinical tests were performed in order to check image quality in vivo. Typical sequences for diagnosis were applied (T2 weighted, Propeller, DWI) with standard scan parameters and image orientations (axial and sagittal). Clinical tests show that the SUBJECT DEVICE provides adequate image quality (SNR, penetration depth, contrast, resolution and robustness against artifacts) as well as scan time and patient comfort are good.
Sample clinical images in DICOM format are provided as part of the discussion of the clinical tests submitted.
#### Conclusion as to Substantial Equivalence
Testing was performed to support this claim of substantial equivalence and to show that the modifications do not raise any new questions pertaining to safety and effectiveness.
The modifications did not affect the Indications for Use, the Intended Use and did not alter the Fundamental Scientific Technology.
RAPID Biomedical therefore believes the SUBJECT DEVICEs and the PREDICATE DEVICEs to be substantially equivalent (SE).
Rimpar, 2019-06-04
Signature:
Name:
Christian Zimmermann (Quality Management & Official Correspondent)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.