The Caterpillar and Caterpillar Micro Arterial Embolization Devices
K191532 · C.R. Bard, Inc. · KRD · Feb 4, 2020 · Cardiovascular
Device Facts
Record ID
K191532
Device Name
The Caterpillar and Caterpillar Micro Arterial Embolization Devices
Applicant
C.R. Bard, Inc.
Product Code
KRD · Cardiovascular
Decision Date
Feb 4, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Caterpillar and Caterpillar Micro Arterial Embolization Devices are indicated for arterial embolization in the peripheral vasculature. The Caterpillar Micro Arterial Embolization Devices are contraindicated for use in vessels subject to cyclic bending, such as highly locomotive joints or muscle beds.
Device Story
Self-expanding arterial occlusion plugs for permanent peripheral vasculature embolization. Device components: cobalt-chrome stem, nickel-titanium fibers, platinum-iridium radiopaque marker bands, polyurethane/polyethylene occlusion membrane. System includes implant, loader, dispenser hoop, detachable delivery wire, and torque tool. Delivered via catheter to target artery; self-expands to occlude vessel. Used in clinical settings by physicians. Radiopaque markers facilitate fluoroscopic visualization during deployment. Provides permanent arterial occlusion to treat vascular conditions. System provided sterile for single use.
Clinical Evidence
No human clinical data. Evidence based on non-clinical bench testing and a chronic ovine animal study. Bench testing included dimension, radial force, luer connection, radiopacity, simulated use (delivery/deployment/detachment), fatigue resistance, material safety (corrosion, nickel leaching, particulate), MRI compatibility, and ISO 10993 biocompatibility. Ovine study evaluated migration resistance, occlusion efficiency, recanalization, and thrombogenicity, demonstrating performance equivalent or superior to control devices.
Technological Characteristics
Self-expanding occlusion plug. Materials: cobalt-chrome stem, nickel-titanium fibers, platinum-iridium radiopaque markers, polyurethane/polyethylene membrane. Delivery system includes hydrophilic (Micro) or PTFE hydrophobic (standard) coated wires. Dimensions vary by target artery diameter (1.5-7mm). Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for arterial embolization in the peripheral vasculature. Contraindicated for use in vessels subject to cyclic bending, such as highly locomotive joints or muscle beds.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font below.
February 4, 2020
ClearStream Technologies Ltd. Melanie Hadlock Regulatory Affairs Specialist II Moyne Upper Enniscorthy, Co. Wexford, Ireland
# Re: K191532
Trade/Device Name: Caterpillar™ and Caterpillar™ Micro Arterial Embolization Device
Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: December 30, 2019 Received: December 31, 2019
# Dear Melanie Hadlock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Misti Malone Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
Page 685
вн
510(k) Number (if known) K191532
## Device Name
Caterpillar™ and Caterpillar™ Micro Arterial Embolization Devices
Indications for Use (Describe)
The Caterpillar and Caterpillar Micro Arterial Embolization Devices are indicated for arterial embolization in the peripheral vasculature. The Caterpillar Micro Arterial Embolization Devices are contraindicated for use in vessels subject to cyclic bending, such as highly locomotive joints or muscle beds.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------|-----------------------------------------------------------------|
| <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Caterpillar™ and Caterpillar TM Micro Arterial Embolization Device 510(k) Summary
#### 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (1)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
### Submitter Information:
ClearStream Technologies Ltd Moyne Upper, Enniscorthy, Co. Wexford, Ireland Tel: 800.321.4254 480.894.9515 Fax: 480.966-7062 Contact Person: Melanie Hadlock, Senior Regulatory Specialist
# Date of Submission: June 7, 2019
### Subject Device:
| Device Trade Name: | Caterpillar™ and Caterpillar™ Micro Arterial Embolization Device |
|-----------------------|------------------------------------------------------------------|
| Common or Usual Name: | Device, Vascular, for Promoting Embolization |
| Classification: | Class II |
| Classification Name: | Vascular Embolization Device, KRD |
| Review Panel: | Cardiovascular |
| Regulation Number: | 21 CFR 870.3300 |
#### Predicate Device:
| Device Trade Name: | Amplatzer™ Vascular Plug II (AVP II) |
|-----------------------|----------------------------------------------|
| Common or Usual Name: | Device, Vascular, for Promoting Embolization |
| Classification: | Class II |
| Classification Name: | Vascular Embolization Device, KRD |
| Review Panel: | Cardiovascular |
| Regulation Number: | 21 CFR 870.3300 |
| 510(k) Numbers: | K071699, cleared August 15, 2007 |
{4}------------------------------------------------
### Reference Device:
| Device Trade Name: | Amplatzer™ Vascular Plug 4 (AVP4) |
|-----------------------|----------------------------------------------|
| Common or Usual Name: | Device, Vascular, for Promoting Embolization |
| Classification: | Class II |
| Classification Name: | Vascular Embolization Device, KRD |
| Review Panel: | Cardiovascular |
| Regulation Number: | 21 CFR 870.3300 |
| 510(k) Number: | K113658, cleared June 12, 2012 |
#### Device Description:
The Caterpillar™ and Caterpillar™ Micro Arterial Embolization Devices are self-expanding arterial occlusion plugs. The devices consist of the following components and are intended to be a permanent implant: cobalt-chrome stem, nickel-titanium fibers, platinum-iridium radiopaque marker bands, and a polyurethane and polyethylene occlusion membrane.
The Caterpillar™ and Caterpillar™ Micro Arterial Embolization Device Systems are packaged as a single unit with the implant, loader, dispenser hoop, detachable delivery wire, and torque tool. The Caterpillar™M Micro delivery wire is coated with a hydrophilic coating. While the Caterpillar™ delivery wire has a PTFE hydrophobic coating. The system is provided sterile and non-pyrogenic and is intended for single use only.
The Caterpillar™ and Caterpillar™ Micro Arterial Embolization Device is designed for a specific arterial diameter range. The artery diameter range and required delivery catheter (ID)) for deployment are provided in the table below.
| Product Name | Product<br>Reference | Target Artery<br>Diameter (mm) | Delivery Catheter<br>Compatibility: Inner<br>Diameter (in/mm) | Marker to<br>Marker<br>Length<br>(mm)1 | Maximum<br>Deployed<br>Length<br>(mm)2 | Delivery<br>Wire<br>Length<br>(cm) |
|-----------------------|----------------------|--------------------------------|---------------------------------------------------------------|----------------------------------------|----------------------------------------|------------------------------------|
| Caterpillar™<br>Micro | 027 | 1.5 - 4 | 0.027 / 0.686 | 7 | 16 | 170 |
| Caterpillar™ | 038 | 3 - 6 | 0.038 / 0.965 | 17 | 26 | 155 |
| Caterpillar™ | 056 | 5 - 7 | 0.056/ 1.422 | 18 | 37 | 155 |
The Marker to Marker Length is the distal radiopaque marker band to the most proximal radiopaque marker 1. band.
2. The Maximum Deployed Length is the length from the distal fibers in the minimum target artery diameter.
# Indications for Use:
The Caterpillar™ and Caterpillar™ Micro Arterial Embolization Devices are indicated for arterial embolization in the peripheral vasculature. The devices are contraindicated for use in vessels subject to cyclic bending, such as locomotive joints or muscle beds.
{5}------------------------------------------------
K191532 Page 3 of 4
### Comparison to Predicate device:
The Caterpillar™ and Caterpillar™ Micro Arterial Embolization Device have the following similarities to the predicate device, the Amplazter™ Vascular Plug II, AVP II (K071699, cleared August 15, 2007):
- Same intended use
- . Similar target population
- Similar technological characteristics
- Similar user interface
- Same mechanism of action
- Same sterility assurance level and method of sterilization ●
- Similar materials ●
The reference device Amplazter™ Vascular Plug 4. AVP4 (K113658, cleared June 12, 2012) has the above similarities and was used to support the methods used for characterization of delivery system performance based on similarities of the delivery system.
The subject devices, the Caterpillar™ and Caterpillar™ Micro Arterial Embolization Device, have the following changes:
- Contraindication ●
- A new implant design
- A modified delivery system ●
# Performance Data:
To demonstrate substantial equivalence of the subject device to the predicate devices, its technological characteristics and performance criteria were evaluated. Using FDA Guidance Documents on non-clinical testing of medical devices, consensus standards, and internal risk assessment procedures, the following tests were performed on the subject device:
- . Dimension
- Implant Length o
- O Catheter Compatibility
- O Delivery Wire Length and Outer Diameter
- Radial Force ●
- Luer Connection Testing ●
- Radiopacity ●
- Simulated Use
- Visual Inspection o
- Delivery, Load, Track and Deployment Forces O
- Accuracy of Deployment O
- Detachment Time and Detachment Mechanism Reliability o
- Tensile and Torsional Strength o
- Recapture and Resheathing for Removal o
- Delivery System Removal (withdrawal) ●
- Fatigue Resistance
- Pulsatile and Pinching Compression O
- . Material Safety Testing
- Corrosion Resistance O
{6}------------------------------------------------
- Nickel Leaching o
- Particulate O
- MRI Compatibility and Safety ●
- Packaging Testing .
- Biocompatibility per ISO 10993 ●
- Cytotoxicity, sensitization (guinea pig maximization test), intracutaneous reactivity, acute o systemic toxicity, material-mediated pyrogenicity, genotoxicity (in vitro bacterial reverse mutation assay and in vitro mouse lymphoma assay), hemolysis (direct and indirect), and complement activation.
- Biocompatibility endpoints for subchronic toxicity, implantation, chronic toxicity, and in vivo O thrombogenicity were addressed within the chronic ovine animal study.
- Biocompatibility endpoints for genotoxicity, chronic toxicity, and carcinogenicity were evaluated O with chemical characterization testing and toxicological assessment
### Pre-Clinical Animal Study:
A chronic ovine study was performed to evaluate the chronic safety and performance of the Caterpillar and Caterpillar Micro Arterial Embolization Devices. The animal study included evaluations for migration resistance, ease of delivery, occlusion efficiency, recanalization, deliverability, hemostasis after procedure, thrombogenicity, device safety, and freedom from complications for the study. The results demonstrated that the study device performance was equivalent or superior to control devices across each of the evaluated endpoints.
### Conclusions:
The subject devices, the Caterpillar™ and Caterpillar™ Micro Arterial Embolization Device, met all predetermined acceptance criteria of design verification as specified by applicable standards, guidance, test protocols and/or customer inputs. The Caterpillar™ and Caterpillar™ Micro Arterial Embolization Device are substantially equivalent to the legally marketed predicate device, the Amplatzer™ Vascular Plug II, AVP II (K071699, cleared August 15, 2007) and the reference device Amplatzer™ Vascular Plug 4, AVP 4 (K113658, cleared June 12, 2012).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.