For permanent cementation of tooth structure to: · crowns, bridges, inlays, veneers and adhesive bridges in combination with Visalys® CemCore · root posts in combination with Visalys® CemCore and Visalys® Core · indirect core build-ups in combination with Visalys® CemCore For build-up adhesive bond to structure with Visalys® CemCore and Visalys® Core to fabricate: · adhesive core build-ups
Device Story
Visalys Tooth Primer is a self-etching, single-component dental primer containing acidic polymers (10-MDP). It is applied by a dental professional to tooth structure to facilitate adhesive bonding of Visalys CemCore and Visalys Core materials. The primer functions without light curing and can be used with or without selective enamel etching. It serves as a chemical interface to improve bond strength for restorative procedures including crowns, bridges, veneers, and core build-ups. The device is supplied in a bottle and stored at 2-8°C.
Clinical Evidence
No clinical studies were performed. Evidence is based on in vitro bench testing, including shear bond strength (ISO 29022) and biocompatibility assessments (cytotoxicity ISO 10993-5, mutagenicity OECD 487, sensitization ISO 10993-10, and chemical analysis ISO 10993-18).
Technological Characteristics
Self-etching, single-component primer containing acidic polymers (10-MDP). Applied via bottle system. Does not require light curing. Biocompatibility verified per ISO 10993 and ISO 7405. Storage temperature 2-8°C.
Indications for Use
Indicated for permanent cementation of crowns, bridges, inlays, veneers, adhesive bridges, root posts, and indirect core build-ups, and for fabrication of adhesive core build-ups in combination with specified Visalys products.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 29, 2019
Kettenbach GmbH & Co. KG Katja Simon Regulatory Affairs Manager Im Heerfeld 7 Eschenburg, 35713 Germany
Re: K191524
Trade/Device Name: Visalys Tooth Primer Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: August 27, 2019 Received: August 30, 2019
Dear Katja Simon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Srinivas Nandkumar Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K191524
Device Name Visalys® Tooth Primer
Indications for Use (Describe)
For permanent cementation of tooth structure to: · crowns, bridges, inlays, veneers and adhesive bridges in combination with Visalys® CemCore
· root posts in combination with Visalys® CemCore and Visalys® Core
· indirect core build-ups in combination with Visalys® CemCore
For build-up adhesive bond to structure with Visalys® CemCore and Visalys® Core to fabricate:
· adhesive core build-ups
Type of Use (Select one or both, as applicable)
|X | Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image features the logo for KETTENBACH, a company whose slogan is "Simply intelligent". The logo consists of a blue square with the company name in white, stacked on top of the slogan, which is in black. Above the company name is a graphic of six white circles arranged in a triangular shape. To the right of the blue square is the registered trademark symbol.
# 510(k) Summary
In accordance with the requirements of the Safe Medical Device Act, Kettenbach GmbH & Co KG herewith submits a Summary.
## A. Name and address of manufacturer:
Kettenbach GmbH & Co KG Im Heerfeld 7 35713 Eschenburg Germany Establishment Registration No.: 9681356 Owner/Operator Number: 9022134
#### Name, title and phone number of official correspondent:
Simon, Katja Regulatory Affairs Manager Im Heerfeld 7 35713 Eschenburg Germany Phone: + 49 277 4705 0 E-mail: katja.simon@kettenbach.com
#### Name, title and phone number of U.S. Agent (Contact):
Roggenbau, Wilfried InterGest North America LLC. 400 Oser Ave., Suite 1650 Hauppauge, NY 11788 Phone: 631 5010500 ext Fax: 631 5011060 Email: roggenbauw@intergestna.com
Date of preparation:
August 26th, 2019
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for KETTENBACH. The logo features a blue square with six white circles arranged in a triangular pattern at the top. Below the circles, the word "KETTENBACH" is written in a bold, sans-serif font, also in white. Underneath "KETTENBACH," the tagline "Simply intelligent" is displayed in a smaller, black, sans-serif font. To the right of the blue square, there is a circled "R" symbol, indicating a registered trademark.
Visalvs® Tooth Primer B. Device Identification:
Device Trade Name: Common Name:
Visalys® Tooth Primer self-etching single component primer for the tooth hard substance
## Classification of the device:
Device Classification Name: Product Code: Device Classification No.: Panel: Requlatory Status:
Resin tooth bonding agent KLE Part 872.3200 Dental Class II
## C. Predicate device:
Device Trade Name: Applicant: 510(k) No.:
Panavia™ V5 Tooth Primer Kuraray Noritake Dental Niigata Plant K150704
#### D. Device Description:
Visalys® Tooth Primer is a self-etching single-component primer with acidic polymers (10-MDP) for adhesive bonding of Visalys® CemCore and Visalys® Core to the dental hard substance. The primer can be applied without additional etching or after selective etching the enamel. Do not light cure Visalys® Tooth Primer. BPA or BPA precursors are not used in the manufacturing process of this device.
## E. Indications for use:
For permanent cementation of tooth structure to:
- · crowns, bridges, inlays, veneers and adhesive bridges in combination with Visalys® CemCore
- · root posts in combination with Visalys® CemCore and Visalys®Core
- · indirect core build-ups in combination with Visalys® CemCore
For build-up adhesive bond to structure with Visalys® CemCore and Visalys® Core to fabricate:
- · adhesive core build-up
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Kettenbach. The logo consists of a blue square with the word "KETTENBACH" in white letters at the bottom. Above the word is a symbol of seven white circles arranged in a triangular shape. Below the blue square are the words "Simply intelligent" in black letters.
### F. Comparison of technological characteristic with the predicate devices
| | Predicate Devices | Substantial Equivalent Device | Conclusion |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product | Panavia V5 TOOTH PRIMER | Visalys® Tooth Primer | |
| Manufacturer | Kuraray | Kettenbach GmbH & Co. KG | |
| 510(k) | K150704 | K191524 | |
| Product Description | The Tooth Primer is a self-etching primer to tooth<br>structure that<br>accelerates the polymerization of the Paste | Visalys® Tooth Primer is a self-etching single-<br>component primer with acidic polymers (10-MDP) for<br>adhesive bonding of Visalys®CemCore and Visalys®<br>Core to the dental hard substance. The primer can be<br>applied without additional etching or after selective<br>etching the enamel. Visalys® Tooth Primer does not<br>have to be cured with light. | Similar |
| Indications for use | PANAVIA V5 is indicated for the following uses:<br>[1] Cementation of crowns, bridges, inlays and onlays<br>[2] Cementation of veneers<br>[3] Cementation of adhesion bridges and splints<br>[4] Cementation of prosthetic restorations on implant<br>abutments and frames<br>[5] Cementation of posts and cores<br>[6] Amalgam bonding | For permanent cementation of tooth structure to:<br>• crowns, bridges, inlays, onlays, veneers and adhesive<br>bridges in combination with Visalys® CemCore<br>• root posts in combination with Visalys® CemCore and<br>Visalys® Core<br>• indirect core build-ups in combination with Visalys®<br>CemCore<br>For build-up adhesive bond to structure with Visalys®<br>CemCore and Visalys® Core to fabricate:<br>• adhesive core build-ups | Similar |
| Storage temperature | 2-8°C/36-46°F | /2-8°C/36-46°F | Similar |
| Application system | Bottle | Bottle | Similar |
| A comparison of parameters in respect to processing time/mechanical properties was performed. The results demonstrated the substantial equivalence to the<br>predicate device. | | | |
Table 1: Visalys® Tooth Primer product description and characteristic to predicate device
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Kettenbach. The logo features a blue square with six white circles arranged in a triangular pattern at the top. Below the circles, the word "KETTENBACH" is written in a bold, sans-serif font. Underneath the company name, the tagline "Simply intelligent" is displayed in a smaller, less bold font. To the right of the blue square, there is a circled "R" symbol, indicating a registered trademark.
## G. Technological Characteristics Summary:
Visalys® Tooth Primer is considered substantially equivalent to Panavia V5 Tooth Primer. There is no significant difference in intended use or technology. Therefore Visalys® Tooth Primer is substantially equivalent to the above listed predicate device. Differences in respect to chemical composition and resulting mechanical and physical properties have been evaluated.
# H. Summary of Non-Clinical Performance Testing:
The following in vitro bench tests were performed on the Visalys® Tooth Primer to verify physical properties and performance in support of substantial equivalence:
- . Shear Bond strength on enamel and dentin according to ISO 29022 "Denistry -Adhesion - Notched edge shear bond strength test"
The performance of the Visalys® Tooth Primer satisfactorily met the requirements of the non-clinical bench testing conducted to support substantial equivalence.
Biocompatibility tests have been performed to assure biological safety in accordance with the ISO 10993 family and ISO 7405. Tests in respect of cytotoxicity (ISO 10993-5), mutagenicity (OECD 487), sensitization (ISO 10993-10) and a chemical analysis (10993-18) showed, that Visalys® Tooth Primer, self-etching single component primer for the tooth hard substance's biocompatibility data is comparable or superior to the predicate device. Therefore no toxicological risks and resulting hazards for patients, users and third parties can be concluded.
## I. Clinical Performance Data
No data from human clinical studies has been included to support the substantial equivalence of Visalys® Tooth Primer.
## J. Conclusion Regarding Substantial Equivalence:
Kettenbach GmbH & Co KG believes that Visalys® Tooth Primer, self-etching single component primer for the tooth hard substance is substantially equivalent to Panavia V5 tooth primer (K150704). It does not introduce new indications for use, has similar technological characteristics and does not introduce new potential hazards or safety risks.
The fact that the indication for use for Visalys® Tooth Primer does not include bonding to amalgam does not interfere with the substantial equivalency to the predicate device, Panavia V5 tooth primer (K150704) since by international convention, usage of mercurycontaining dental amalgam is to be reduced. Likewise, it was not tested (Press Release U.S. Department of State. PRN: 2013/1353.
https://2009-2017.state.gov/r/pa/prs/ps/2013/11/217295.htm)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.