K191523 · Kettenbach GmbH & Co. KG · KLE · Oct 9, 2019 · Dental
Device Facts
Record ID
K191523
Device Name
Visalys Restorative Primer
Applicant
Kettenbach GmbH & Co. KG
Product Code
KLE · Dental
Decision Date
Oct 9, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3200
Device Class
Class 2
Indications for Use
Restoration primer for treatment of: metal/silicate ceramics, oxide ceramic and composite surfaces - to cement prosthetic restorations - to repair fractured restorations with composite materials
Device Story
Visalys Restorative Primer is a single-component dental primer used to enhance adhesion between restorative materials (metal, composite, silicate/oxide ceramics) and luting materials. It contains adhesive monomers (10-MDP, silane methacrylate) in an ethanol base. The clinician applies the primer to the restoration surface; it does not require light-curing. The device is used in a dental clinic setting by a dentist or dental professional. By creating an adhesive surface, it enables the cementation of prosthetic restorations (crowns, bridges, inlays, onlays) and the repair of fractured restorations. The output is a chemically prepared surface that improves bond strength, ensuring the stability and longevity of the dental restoration.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including shear bond strength testing on lithium disilicate, zirconium oxide, and non-precious metal, and biocompatibility testing per ISO 10993 and ISO 7405 (cytotoxicity, mutagenicity, sensitization, chemical analysis).
Technological Characteristics
Single-component liquid primer. Composition: ethanol, 10-MDP (acidic adhesive monomer), and silane methacrylate. No BPA or BPA precursors. Storage: 2–8 °C. Application: bottle system. Non-light-cured. Biocompatibility: ISO 10993, ISO 7405.
Indications for Use
Indicated for surface treatment of metal, silicate ceramic, oxide ceramic, and composite surfaces to facilitate cementation of prosthetic restorations or repair of fractured restorations.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 9, 2019
Kettenbach GmbH & Co. KG Katja Simon Regulatory Affairs Manager Im Heerfeld 7 35713 Eschenburg, Germany
Re: K191523
Trade/Device Name: Visalys Restorative Primer Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: September 4, 2019 Received: September 9, 2019
Dear Katja Simon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K191523
Device Name Visalys® Restorative Primer
Indications for Use (Describe)
Restoration primer for treatment of:
metal/silicate ceramics, oxide ceramic and composite surfaces
- · to cement prosthetic restorations
- · to repair fractured restorations with composite materials
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for KETTENBACH. The logo consists of a blue square with six white circles arranged in a triangular pattern at the top. Below the circles, the word "KETTENBACH" is written in a bold, sans-serif font. Underneath the company name is the slogan "Simply intelligent" in a smaller, bold font. To the right of the blue square, there is a circled "R" symbol, indicating a registered trademark.
# 510(k) Summary
In accordance with the requirements of the Safe Medical Device Act, Kettenbach GmbH & Co KG herewith submits a Summary.
### A. Name and address of manufacturer:
Kettenbach GmbH & Co KG Im Heerfeld 7 35713 Eschenburg Germany Establishment Registration No.: 9681356 Owner/Operator Number: 9022134
#### Name, title and phone number of official correspondent:
Simon, Katja Regulatory Affairs Manager Im Heerfeld 7 35713 Eschenburg Germany Phone: + 49 277 4705 0 E-mail: katja.simon@kettenbach.com
#### Name, title and phone number of U.S. Agent (Contact):
Roggenbau, Wilfried InterGest North America LLC. 400 Oser Ave., Suite 1650 Hauppauge, NY 11788 Phone: 631 5010500 ext Fax: 631 5011060 Email: roggenbauw@intergestna.com
Date of preparation:
September 04th, 2019
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for KETTENBACH. The logo features a blue square with six white circles arranged in a triangular pattern at the top. Below the circles, the word "KETTENBACH" is written in a bold, sans-serif font. Underneath "KETTENBACH", the phrase "Simply intelligent" is written in a smaller, sans-serif font.
B. Device Identification: Device Trade Name: Common Name:
# Visalys® Restorative Primer
Visalys® Restorative Primer Universal adhesion primer for restoration materials
#### Classification of the device:
Device Classification Name: Product Code: Device Classification No.: Panel: Requlatory Status:
Resin tooth bonding agent KLE Part 872.3200 Dental Class II
#### C. Predicate device:
Device Trade Name: Applicant: 510(k) No.:
CLEARFIL™ CERAMIC PRIMER PLUS Kuraray Noritake Dental Inc. K150703
### D. Device Description:
Visalys® Restorative Primer with adhesive monomers (10-MDP, silane methacrylate) is for building up adhesive surfaces on restoration materials (metal, composite, as well as silicate and oxide ceramics). Visalys® Restorative Primer is a single component primer. Do not light cure Visalys® Restorative Primer. BPA or BPA precursors are not used in the manufacturing process of this device.
### E. Indications for use:
Restoration primer for treatment of:
metal/silicate ceramics, oxide ceramic and composite surfaces
- · to cement prosthetic restorations
- · to repair fractured restorations with composite materials
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for KETTENBACH. The logo consists of a blue square with six white circles arranged in a triangular shape. Below the square, the word "KETTENBACH" is written in white, followed by the tagline "Simply intelligent devices" in black. There is also a registered trademark symbol in the upper right corner of the blue square.
# F. Comparison of technological characteristic with the predicate devices
| | Predicate Devices | Substantial Equivalent Device | Conclusion |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product | CLEARFIL™ CERAMIC PRIMER PLUS | Visalys® Restorative Primer | |
| Manufacturer | Kuraray Noritake Dental Inc. | Kettenbach GmbH & Co. KG | |
| 510(k)<br>numbers | K150703 | K191523 | |
| Product<br>Description | CLEARFIL™ CERAMIC PRIMER PLUS is a<br>dental universal prosthetic primer that<br>provides an enhanced adhesive surface to<br>ceramic (e.g. conventional porcelain, lithium<br>disilicate, zirconia), hybrid ceramics (e.g.<br>ESTENIA C&B), composite resin and metal. | Visalys® Restorative Primer with adhesive<br>monomers (10-MDP, silane methacrylate) is<br>for building up adhesive surfaces on<br>restoration materials (metal, composite, as well<br>as silicate and oxide ceramics). Visalys®<br>Restorative Primer is a single component<br>primer that does not have to be light-cured. | Similar |
| Indication for<br>use | Indications:<br>CLEARFIL™ CERAMIC PRIMER PLUS is<br>indicated for the following uses:<br>[1] Surface treatment of prosthetic<br>restorations made of ceramic, hybrid<br>ceramics, composite resin or metal<br>[2] Intraoral repairs of fractured restorations<br>made of ceramics, hybrid ceramics,<br>composite resin or metal | Indications:<br>Restoration primer for treatment of<br>metal/silicate ceramics, oxide ceramic and<br>composite surfaces<br>• to cement prosthetic restorations<br>• to repair fractured restorations with composite<br>materials | Similar |
| Storage<br>temperature | 2 – 8 °C/36 – 46 °F | 2 – 8 °C/35 – 46 °F | Similar |
| Application<br>system | Bottle | Bottle | Similar |
| Composition | • 3-Methacryloxypropyl trimethoxysilane<br>• 10-Methacryloyloxydecyl dihydrogen<br>phosphate (MDP)<br>• Ethanol | The main components of the Visalys®<br>Restorative Primer are<br>ethanol, acidic adhesive monomer (10-MDP)<br>and silane methacrylate. | Similar |
| | Predicate Devices | Simply intelligent<br>Substantial Equivalent Device | Conclusion |
| Product | CLEARFIL™ CERAMIC PRIMER PLUS | Visalys® Restorative Primer | |
| Manufacturer | Kuraray Noritake Dental Inc. | Kettenbach GmbH & Co. KG | |
| 510(k) numbers | K150703 | K191523 | |
| Shear bond strengths on different restorative materials<br>(ISO 16506 / Chapter A 2.2.5): | | | |
| Lithium<br>disilicate | (8.9 MPa) | (10.2 MPa) | Similar |
| Zirconium<br>oxide | (8.4 MPa) | (7.5 MPa) | Similar |
| Non-precious<br>metal | (8.3 MPa) | (12.5 MPa) | Similar |
| Products used<br>in the testing | CLEARFIL™ Ceramic Primer Plus<br>(K150703) was tested with the resin<br>product Panavia V5 dental cement<br>(K150704), manufactured by Kuraray | Visalys® Restorative Primer (K191523)<br>was tested with the resin product Visalys<br>CemCore (K191527), manufactured by<br>Kettenbach GmbH & Co. KG. | |
| Product description and composition for the predicate device has been extracted from the instruction for use. Values given in brackets<br>have been determined by Kettenbach | | | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for KETTENBACH. The logo consists of a blue square with six white circles arranged in a triangular pattern at the top. Below the circles is the word "KETTENBACH" in white, and below that is the phrase "Simply intelligent" in black. There is also a registered trademark symbol in the upper right corner.
Table 1: Visalys® Restorative Primer product description and characteristic to predicate device
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Kettenbach GmbH & Co. KG. The logo features a blue square with a white graphic of six circles arranged in a triangular shape at the top. Below the graphic, the word "KETTENBACH" is written in white, followed by the tagline "Simply intelligent" in black.
### G. Technological Characteristics Summary:
Visalys® Restorative Primer is considered substantially equivalent to Clearfil Ceramic Primer Plus. There is no significant difference in intended use or technology. Therefore Visalys® Restorative Primer is substantially equivalent to the above listed predicate device. Differences in respect to chemical composition and resulting mechanical and physical properties have been evaluated.
#### H. Summary of Non-Clinical Performance Testing:
The following in vitro bench tests were performed on the Visalys® Restorative Primer to verify physical properties and performance in support of substantial equivalence:
- . Shear Bond strength on restorative materials acc. to ISO 16506 "Dentistry -Polymer-based luting materials containing adhesive component shear bond strength test" (Chapter A 2.2.5).
The performance of the Visalys® Restorative Primer satisfactorily met the requirements of the non-clinical bench testing conducted to support substantial equivalence.
Biocompatibility tests have been performed to assure biological safety in accordance with the ISO 10993 family and ISO 7405. Tests in respect of cytotoxicity (ISO 10993-5), mutagenicity (OECD 487), sensitization (ISO 10993-10) and a chemical analysis (10993-18) showed, that Visalys® Restorative Primer, universal adhesion primer for restoration materials biocompatibility data is comparable or superior to the predicate device. Therefore no toxicological risks and resulting hazards for patients, users and third parties can be concluded.
### Rationale for using ISO 16506 (Chapter A 2.2.5) as test standard
The Medical Device Product Classification for the stated product is KLE (Agent, Tooth Bonding, Resin). However, the stated product Visalys® Restorative Primer acts as a coupling agent between restoration material (e.g. ceramics, metals) and luting material (e.g. Visalys® CemCore) and not between luting material and tooth structure. For the later indication, Visalys® Tooth Primer can be used. The recognized test standards ISO 29022 and ISO/TS 11405 as well as ISO/TS 16506 describe different bonding strength measurements. Unfortunately, all three standards only show methods to determine bond strength to tooth structure. As no product-specific standard exists for the subject device, testing was conducted under ISO/TS 16506 due to the following reason:
Shear bond strength is described in ISO 29022 and in ISO/TS 16506. The main difference is the loading of the test specimen after application of the luting material in ISO/TS 16506 (force of 150 N ± 2N, Ch 7.5.1.2), whereas no force is applied in the test method described in ISO 29022. Visalys® Restorative Primer is used on restorations such as crowns, bridges, inlays and onlays. After applying a luting material on the restoration, it is placed in the patient's mouth onto a tooth,
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image is a logo for KETTENBACH. The logo is set against a blue square background. The logo consists of seven white circles arranged in a triangular shape, with six circles forming the base and one circle at the apex. Below the circles, the word "KETTENBACH" is written in bold, white letters. Below the word, the phrase "Simply intelligent" is written in black letters.
core or into a cavity by applying a force onto the restoration to maintain a small laver thickness of the luting material. Therefore, the shear bond test method described in ISO/TS 16506 is a more realistic measurement for this kind of product (Visalys® Restorative Primer) due to the loading of the test specimen, which simulates the above-mentioned clinical procedure. ISO/TS 11405 gives a general overview of techniques for examination of adhesion. Once aqain, only procedures for testing of the adhesion to tooth structure are included (title of the standard: "Testing of adhesion to tooth structure"). Annex A describes methods for the measurement of bond strength. As explained above, shear bond strength of Visalys® Restorative Primer was already tested according to ISO/TS 16506. An additional bond strength measurement is e.g. micro-tensile bond strength, as described in Annex A, A.2.4. Unfortunately, such measurement cannot be performed with restorative materials because there is no practicable way to cut restorative materials (e.g. metals like titanium) into the desired dimension of 1 mm x 1 mm. Additionally, no force can be applied on the luting material to simulate the clinical procedure (see above). For a detailed description of micro-tensile bond strength and the challenging test procedure to prepare test specimen, see S. Armstrong et al. (1)
Armstrong, S. et al. Academy of Dental Materials quidance on in vitro 1. testing of dental composite bonding effectiveness to dentin / enamel using microtensile bond strength (µTBS) approach. Dent. Mater. 33, 133-143 (2016).
#### -Clinical Performance Data
No data from human clinical studies has been included to support the substantial equivalence of Visalys® Restorative Primer.
### J. Conclusion Regarding Substantial Equivalence:
Kettenbach GmbH & Co KG believes that Visalys® Restorative Primer, universal adhesion primer for restoration materials is substantially equivalent to Clearfil Ceramic Primer Plus (K150703). It does not introduce new indications for use, has similar technological characteristics and does not introduce new potential hazards or safety risks.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.