K191507 · Intellijoint Surgical, Inc. · OLO · Oct 1, 2019 · Neurology
Device Facts
Record ID
K191507
Device Name
Intellijoint® Navigation System
Applicant
Intellijoint Surgical, Inc.
Product Code
OLO · Neurology
Decision Date
Oct 1, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The Intellijoint® Navigation System is a computer-controlled, optical localizer intended to provide intra-operative measurements to a surgeon to aid in selection and positioning of orthopedic implant system components, where a reference to a rigid anatomical structure can be identified. The Intellijoint® Navigation System is indicated for patients undergoing orthopedic surgery, and where the use of stereotactic surgery is considered safe and effective. Intellijoint HIP® aids the surgeon in performing intra-operative measurements including measurements of limb position, joint center-of-rotation, and implant component positioning. Intellijoint KNEE™ aids the surgeon in performing intra-operative measurements including alignment of instrumentation and bony cuts for implant component positioning. Example orthopedic surgical procedures include, but are not limited to: Total Hip Arthroplasty; Minimally Invasive Hip Arthroplasty; Revision Hip Arthroplasty; Total Knee Arthroplasty
Device Story
Imageless optical navigation system for orthopedic surgery; utilizes infrared camera, camera drape, optical tracking spheres, and computer workstation. Surgeon mounts bone trackers to femur or tibia; system digitizes anatomical landmarks to provide real-time intra-operative measurements. Outputs include varus/valgus angle, flexion/extension angle, and resection depth for TKA; limb position, joint center-of-rotation, and implant component positioning for hip procedures. Used in operating room by surgeons to guide implant selection and positioning; improves surgical accuracy via real-time feedback.
Clinical Evidence
Bench testing only. Included benchtop accuracy verification per ASTM F2554-10, software functional/unit testing, and mechanical instrumentation performance tests. Validation included simulated use on anatomical bone models and cadaver labs to assess clinical accuracy and workflow performance; all requirements met.
Technological Characteristics
Imageless optical navigation system; monocular infrared camera; optical tracking spheres; computer workstation. Connectivity: standalone. Sterilization: not specified. Software: application-specific algorithms for orthopedic measurements. Standards: ASTM F2554-10 for positional accuracy.
Indications for Use
Indicated for patients undergoing orthopedic surgery where stereotactic surgery is considered safe and effective, including Total Hip Arthroplasty, Minimally Invasive Hip Arthroplasty, Revision Hip Arthroplasty, and Total Knee Arthroplasty.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
October 1, 2019
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
Intellijoint Surgical Inc. Breanne Cuddington Regulatory Affairs Associate 60 Bathurst Drive Unit 6 Waterloo, N2V 2A9 CA
Re: K191507
Trade/Device Name: Intellijoint® Navigation System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: August 30, 2019 Received: September 3, 2019
Dear Breanne Cuddington:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For, Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191507
Device Name Intellijoint® Navigation System
#### Indications for Use (Describe)
The Intellijoint® Navigation System is a computer-controlled, optical localizer intended to provide intra-operative measurements to a surgeon to aid in selection and positioning of orthopedic implant system components, where a reference to a rigid anatomical structure can be identified. The Intellijoint Navigation System is indicated for patients undergoing orthopedic surgery, and where the use of stereotactic surgery is considered safe and effective.
Intellijoint HIP aids the surgeon in performing intra-operative measurements of limb position, joint center-of-rotation, and implant component positioning.
Intellioint KNEE aids the surgeon in performing intra-operative measurements including alignmentation and bony cuts for implant component positioning.
Example orthopedic surgical procedures include, but are not limited to:
- Total Hip Arthroplasty
- Minimally Invasive Hip Arthroplasty
- Revision Hip Arthroplasty
- Total Knee Arthroplasty
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10pt">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/3 description: The image shows the logo for "Intellijoint Surgical". The logo consists of a geometric shape on the left and the company name on the right. The geometric shape is made up of three triangles, with the top triangle being dark blue, the middle triangle being light blue, and the bottom triangle being gray. The company name is written in gray, with the word "Intellijoint" on top and the word "Surgical" on the bottom.
# 510(k) Summary
# 1. Submitter Information
| Submitter: | Intellijoint Surgical Inc. |
|----------------|-----------------------------------------------------------------|
| Address: | 809 Wellington St. N, Unit 2<br>Kitchener, ON<br>Canada N2H 5L6 |
| Telephone: | (519) 342-3178 |
| Fax: | (226) 317-0471 |
| Contact: | Breanne Cuddington |
| Date Prepared: | 30 August 2019 |
# 2. Device Information
| Trade Name: | Intellijoint® Navigation System |
|----------------------|-----------------------------------|
| Common Name: | Orthopedic Stereotaxic Instrument |
| Classification: | Class II per 21 CFR 882.4560 |
| Classification Name: | Orthopedic Stereotaxic Instrument |
| Product Code: | OLO |
# 3. Purpose of Submission
The purpose of this submission is to gain clearance for updates to the indications for use for the Intellijoint HIP Generation 2B System, which has been previously cleared in K171525. The updated product will be referred to as the Intellijoint® Navigation System throughout the submission.
# 4. Predicate Device Information
The Intellijoint® Navigation System described in this substantially equivalent to the following predicate devices:
## Primary Predicate:
| Device Name | Manufacturer | 510(k) No. |
|--------------------|-----------------|------------|
| KneeAlign 2 System | OrthAlign, Inc. | K103829 |
## Additional Predicate:
| Device Name | Manufacturer | 510(k) No. |
|---------------------------------------|----------------------------|------------|
| Intellijoint HIP Generation 2B System | Intellijoint Surgical Inc. | K171525 |
{4}------------------------------------------------
tellijoint
# 5. Device Description
The Intellijoint® Navigation System (IINS) is an imageless optical navigation system intended for use in orthopedic surgery. The IINS consists of an infrared Camera, Camera Drape, optical tracking Spheres, a computer Workstation and associated hardware.
This submission provides the addition of Total Knee Arthroplasty (TKA) to the indications for use (cleared in K171525), allowing real-time measurements of varus/valgus angle, flexion/extension angle, and resection depth for distal femur and proximal tibia cuts during TKA procedures. The modifications include the addition of TKA-specific mechanical instruments and software. The updated product will be referred to as the Intellijoint® Navigation System throughout this submission.
# 6. Indications for Use
The Intellijoint® Navigation System is a computer-controlled, optical localizer intended to provide intra-operative measurements to a surgeon to aid in selection and positioning of orthopedic implant system components, where a reference to a rigid anatomical structure can be identified. The Intellijoint® Navigation System is indicated for patients undergoing orthopedic surgery, and where the use of stereotactic surgery is considered safe and effective.
Intellijoint HIP® aids the surgeon in performing intra-operative measurements including measurements of limb position, joint center-of-rotation, and implant component positioning.
Intellijoint KNEE™ aids the surgeon in performing intra-operative measurements including alignment of instrumentation and bony cuts for implant component positioning.
Example orthopedic surgical procedures include, but are not limited to:
- Total Hip Arthroplasty ।
- Minimally Invasive Hip Arthroplasty ।
- Revision Hip Arthroplasty -
- -Total Knee Arthroplasty
## 7. Comparison of Technological Characteristics
The substantial equivalence of the Intellijoint® Navigation System compared to the predicate devices is based on the following technological characteristics:
| Property | Intellijoint® Navigation System | Primary Predicate<br>KneeAlign® 2 System | Additional Predicate<br>Intellijoint HIP® Generation 2B System |
|------------------------|---------------------------------|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| 510(k) No. | K191507 | K103829 | K171525 |
| Tracking method | Optical | Inertial | Optical |
| Imaging Modality | Imageless | Imageless | Imageless |
| Camera System | Monocular | N/A | Monocular |
| Registration<br>Method | HIP<br>Same as K171525 | HIP<br>N/A | HIP<br>i. Camera rigidly fixed to pelvis.<br>ii. Digitization of Anterior Pelvic Plane and/or frontal / sagittal planes. |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for "IntelliJoint Surgical". The logo consists of a geometric shape on the left, resembling a stylized joint, with three overlapping triangles in different shades of blue and gray. To the right of the geometric shape is the word "IntelliJoint" in a sans-serif font, with the word "Surgical" in a smaller font size underneath.
K191507
| Property | Intellijoint®<br>Navigation System | Primary Predicate | Additional Predicate |
|----------|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| | | KneeAlign® 2 System | Intellijoint HIP®<br>Generation 2B System |
| | KNEE | KNEE | KNEE |
| | i. Bone Tracker mounted<br>rigidly to Femur or<br>Tibia<br>ii. Digitization of Femoral<br>or Tibial Landmarks | i. Navigation unit mounted<br>rigidly to Femur or Tibia<br>ii. Digitization of Femoral<br>or Tibial landmarks | N/A |
# 8. Performance Data
This submission provides the addition of TKA to the indications for use cleared in 510(k) K171525 and adds the Intellijoint KNEE™ Software Application to the Intellijoint® Navigation System. Other modifications include the addition of TKA specific mechanical instruments to the Intellijoint® Navigation System.
The following tests were performed to support the addition of Intellijoint KNEE™ and to demonstrate the substantial equivalence of the Intellijoint® Navigation System to its predicate devices:
| Test | Summary | Result |
|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Verification | | |
| Benchtop Accuracy | Verified clinical accuracy requirements using<br>calibrated benchtop test fixtures. | All accuracy<br>requirements were<br>met. |
| Tracking System Accuracy and Robustness | Accuracy of Intellijoint KNEE was verified<br>according to the methodology in ASTM F2554-<br>10 - Standard Practice for Measurement of<br>Positional Accuracy of Computer Assisted<br>Surgical Systems. Testing simulated normal<br>conditions, and a variety of worst-case use<br>scenarios and realistic tracking disturbances. | All accuracy<br>specifications and<br>robustness<br>requirements were<br>met. |
| Software Functional and Unit Tests | Verified that the Intellijoint KNEE Software<br>Application satisfies functional requirements<br>and performs as intended. Algorithms and<br>measurement calculations were also verified in<br>these tests. | Software satisfied<br>all requirements<br>and specifications. |
| Mechanical Instrumentation Functional and Performance Tests | Provides confirmation that mechanical<br>instrumentation satisfies functional and<br>performance requirements. | All functional and<br>performance<br>requirements met. |
{6}------------------------------------------------
Image /page/6/Picture/3 description: The image shows the logo for "Intellijoint Surgical". The logo consists of a blue and gray triangle shape on the left, followed by the word "intellijoint" in gray. Below "intellijoint" is the word "surgical" in a smaller, lighter gray font.
| Validation | | |
|--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Anatomical Phantom<br>Simulated Use and<br>Clinical Accuracy | Simulated use testing was performed on bone<br>models by orthopedic surgeons in a simulated<br>TKA procedure following a typical workflow.<br>This test validated that Intellijoint KNEE<br>satisfies user needs, intended use and clinical<br>accuracy requirements. Accuracy was assessed<br>by comparing simulated use measurements with<br>ground truth values. | All user needs and<br>clinical accuracy<br>requirements were<br>met. |
| Cadaver Simulated Use | Simulated use testing was also performed in<br>cadaver labs. These tests validated that<br>Intellijoint KNEE satisfies clinical use<br>requirements and performs as intended when:<br>- Operated by a surgeon<br>- Used on human specimens<br>- Used in a realistic operating room<br>environment | All clinical use<br>requirements were<br>met. |
The testing demonstrated that the Intellijoint® Navigation System is substantially equivalent to the legally marketed predicate devices for its intended use.
# 9. Conclusion
Based on the indications for use, technological characteristics, performance testing, and comparison to the predicate devices, the Intellijoint® Navigation System has been shown to be substantially equivalent to the legally marketed predicate devices identified in this submission and does not present any changes to safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.