K191479 · Adeor Medical AG · HBE · Apr 11, 2021 · Neurology
Device Facts
Record ID
K191479
Device Name
Velocity Alpha Highspeed Surgical Drill System
Applicant
Adeor Medical AG
Product Code
HBE · Neurology
Decision Date
Apr 11, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Velocity Alpha™ Drill System is indicated for trepanating, incision, cutting, removal, shaping, sawing of soft and hard tissue, bone, and bone replacement materials. Applications: Cutting, removing, shaping and sawing hard and soft tissue, bone, and bone replacement materials. Areas of use: Neurosurgery and spinal surgery.
Device Story
Velocity Alpha Highspeed Surgical Drill System is an AC-powered electric motor drill for rapid cutting, sawing, drilling, and manipulation of bone and soft tissue. System comprises an electric motor handpiece, AC-powered control unit, and foot control (wired or wireless via Bluetooth). Includes various adapters, nosepieces, and optional irrigation pump. Surgeon operates device via foot control or touch screen on control unit. Variable speed (1,000–80,000 RPM) and selectable rotation direction allow adjustment to surgical requirements. Used in neurosurgery and spinal surgery. Output is mechanical rotation of surgical attachments to perform bone/tissue manipulation. Benefits include precise, high-speed surgical access and tissue removal.
Clinical Evidence
No clinical data provided. Bench testing only: functional motor speed/torque analysis, temperature monitoring, lifetime duty cycle testing, noise level analysis, electrical safety (IEC 60601-1), and electromagnetic compatibility (IEC 60601-1-2). Biocompatibility testing performed per ISO 10993.
Technological Characteristics
Electric AC-powered motor drill system. Materials: Stainless steel (handpiece/attachments), plastic (control unit housing). Speed: 1,000–80,000 RPM. Connectivity: Wired or wireless (Bluetooth) foot control. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62304. Sterilization: Steam autoclave (gravity or pre-vacuum).
Indications for Use
Indicated for trepanating, incision, cutting, removal, shaping, and sawing of soft and hard tissue, bone, and bone replacement materials in neurosurgery and spinal surgery.
Regulatory Classification
Identification
Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
April 11, 2021
adeor medical AG Ulrike Winkler Vice President QM/RA/R&D Martinshof 5 Valley, Bavaria 83626 Germany
Re: K191479
Trade/Device Name: Velocity Alpha Highspeed Surgical Drill System Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, and Their Accessories Regulatory Class: Class II Product Code: HBE Dated: March 8, 2021 Received: March 10, 2021
Dear Ulrike Winkler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Xiaolin Zheng -S
Xiaolin Zheng, Ph.D. Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191479
Device Name
Velocity Alpha Highspeed Surgical Drill System
Indications for Use (Describe)
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Adeor Medical. The logo consists of a stylized blue brain on the left, followed by the word "adeor" in black, with the word "MEDICAL" in smaller black letters below it. The brain graphic is a simplified representation of a human brain, with geometric lines forming its structure.
# 510(k) Summary
- 1. Company: adeor medical AG Martinshof 5 83626 Valley Germany
- II. Contact: Ulrike Winkler Vice President QM / RA / R&D (QMR)
- III. Preparation Date: 04th March, 2020
- IV. Proprietary Trade Name: Velocity Alpha™ Highspeed Surgical Drill System
- V. Common Name: Powered Drill System
- VI. Classification Name: Drills, Burrs, Trephines & Accessories (21 CFR 882.4310)
- VII. Classification: Class II
- VIII. Product Code: HBE
## IX. Market device claiming equivalence to: The electric drill system is substantially equivalent to the Aesculap ELAN 4 Electro Motor System cleared via K152960.
#### X. Device Description:
The Velocity Alpha™ Highspeed Surgical Drill System is an electric AC powered surgical motor drill system for rapid cutting, sawing, drilling and manipulation of soft tissue and bone.
The Subject Device has several components, such as an electric motor hand piece connected to and driven by an AC-powered control unit, a foot control unit (wired or wireless (Bluetooth)) and several adapters and nosepiece attachments, e.g. straight & angled nosepiece attachments, craniotome attachments, speed reducer attachments and sawing attachments.
The device has an optional irrigation pump that can supply irrigation through sterile tubing. Speed is variable from 1.000 RPM up to 80.000 RPM in order to adjust the speed to the surgeon's requirements. The direction of rotation can be pre-selected to the left and to the right.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Adeor Medical. On the left side of the logo is a blue image of a brain. To the right of the brain is the word "adeor" in black, with the word "MEDICAL" in smaller black letters underneath.
#### XI. Intended Use / Indications for Use:
The Velocity Alpha™ Drill System is indicated for trepanating, incision, cutting, removal, shaping, sawing of soft and hard tissue, bone, and bone replacement materials.
| Applications: | Cutting, removing, shaping and sawing hard and<br>soft tissue, bone, and bone replacement materials. |
|---------------|------------------------------------------------------------------------------------------------------|
| Areas of use: | Neurosurgery and spinal surgery. |
- XII. Comparison of the Technological Characteristics and Substantial Equivalence: The technological characteristics of subject device are based on the same highspeed motor drill technology as the predicate devices.
| Item | New device | Predicate device | Equiva-<br>lence |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Trade Name | Velocity Alpha™ Highspeed Surgical Drill<br>System | Aesculap ELAN 4 Electro | - |
| Picture | Image: Velocity Alpha Highspeed Surgical Drill System | Image: Aesculap ELAN 4 Electro | - |
| Intended Use | The Velocity Alpha™ Drill System is indicated<br>for trepanating, incision, cutting, removal,<br>shaping, sawing of soft and hard tissue,<br>bone, and bone replacement materials.<br>Applications: Cutting, removing, shaping and<br>sawing hard and soft tissue, bone, and bone<br>replacement materials.<br>Areas of use: Neurosurgery and spinal<br>surgery. | The Elan 4 Electro motor<br>system is intended for high<br>speed cutting, sawing and<br>drilling of bone in the fields of<br>Spine, ENT, Neuro and<br>Maxillofacial surgery. | Similar |
| Control Unit | | | |
| Maximum speed | 80,000 rpm | 80,000 rpm | Identical |
| Voltage | 100-240 V | 110-240 V | Identical |
| Frequency | 50-60 Hz | 50-60 Hz | Identical |
| Irrigation Pump | Yes | Yes | Identical |
| Motor | | | |
| High speed<br>motor (min /<br>max) | 1.000 / 80.000 rpm | 10.000 / 80.000 rpm | Identical |
| Motor rotation | Right and left speed rotation | Right and left speed rotation | Identical |
| Materials | Stainless Steel | Stainless Steel | Identical |
| Foot Control | | | |
| Control buttons | 4 buttons for<br>• pump on/off<br>• Forward/reverse | 3 buttons:<br>• Pump on/off<br>• Forward/reverse | Substantially<br>equivalent -<br>no question to |
| | • Change programs<br>• Motor control (on/off and variable) | • Foot pad | safety and<br>effectiveness |
| Power supply: | • Wireless via 3xAA batteries<br>• cable | Cable | Substantially<br>equivalent -<br>no question to<br>safety and<br>effectiveness |
| Features | | | |
| Control<br>mechanism | Foot or touch screen | Foot or touch screen | Identical |
| Ability to connect<br>different motor<br>handpieces | no | yes | Identical |
| Motor<br>connections | 1 | 2 | Identical |
| Footswitch<br>connections | 1 | 1 | Identical |
| Control unit | Touch screen | Touch screen | Identical |
| Applied<br>standards | IEC60601-1<br>IEC60601-1-2<br>IEC62304 | IEC60601-1<br>IEC60601-1-2<br>IEC62304 | Identical |
| Materials | | | |
| Control unit with<br>housing | Plastic material | Plastic material | Identical |
| Motor with cable | Stainless steel | Stainless steel | Identical |
| attachments /<br>nosepieces | Stainless steel | Stainless steel | Identical |
| Hygiene / Maintenance | | | |
| Cleaning | The foot control is water-tight according to<br>IPX8, 1 m depth of immersion, 1 hour (water-<br>tight in accordance with IEC 60529).<br>Control Unit and foot control (cabled or<br>Bluetooth): cleaning wiped with moist tissue.<br>Wiping disinfection.<br><br>Attachments: validated manual and<br>mechanical cleaning and disinfection. For<br>more details see IFU page 30 to 35).<br><br>Cutters/burrs: only sterile packed single-use<br>cutters/burrs are being used (not part of this<br>submission, see HiCUT Instruments<br>K130755). | Similar cleaning procedure.<br>(see annex IFU ELAN4<br>Electric System) | Substantially<br>equivalent -<br>no question<br>in safety and<br>effectiveness |
| Sterilization | Gravity method: 15min, 132°C, drying time:<br>20min<br><br>Pre-vacuum method: 4min, 132°C, drying time:<br>16min<br><br>For more details see IFU page 35. | • Steam Autoclave/Gravity<br>Air Displacement 132°C/<br>15 minutes wrapped or<br>unwrapped 30 minutes dry<br>time<br>• Steam autoclave / Pre-<br>vacuum: 132°C/ 3<br>minutes wrapped or<br>unwrapped, 30 minutes<br>dry time. | Substantially<br>equivalent -<br>no question<br>in safety and<br>effectiveness |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Adeor Medical. On the left is a blue graphic of a brain. To the right of the brain is the word "adeor" in black, with the word "MEDICAL" in smaller black letters underneath.
#### XIII. Substantial Equivalence Discussion:
The technologic characteristics of the subject device are based on the same electric motor drill technology as the predicate devices. The adeor® Velocity Alpha™ Highspeed Surgical
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Adeor Medical. The logo consists of a stylized blue brain on the left and the word "adeor" in black on the right. Below the word "adeor" is the word "MEDICAL" in smaller black letters.
Drill System is substantially equivalent to the predicate devices based upon the comparison of the technological characteristics.
Any difference that exists between the Velocity Alpha™ Highspeed Surgical Drill System and the predicate devices has no negative effects on safety or effectiveness of the subject device when used as labelled.
#### XIV. Performance data testing:
| Test | Description | Acceptance criteria | Results |
|-----------------------------------------------|--------------------------------------------------------------------------------|------------------------------------------------------------------------|---------|
| Functional (motor speed /<br>torque analysis) | Analyzed motor speed / torque<br>curve | Torque was reached as<br>specified | Pass |
| Temperature | Analyzed motor temperature<br>during activity | Temperature was in<br>compliance to specifications. | Pass |
| Lifetime | Analyzed System lifetime during<br>regular duty cycle | The System functioned<br>according to specifications. | Pass |
| Noise level | Analyzed Noise level during<br>activity | The Noise level did not<br>exceed the limit. | Pass |
| Electrical Safety | Electric powered instruments<br>evaluated for electrical safety | Be aligned with IEC60601-<br>1:2005 Electrical Safety | Pass |
| Electromagnetic Compatibility | Electric powered instruments<br>evaluated for electromagnetic<br>compatibility | Be aligned with IEC60601-1-<br>2:2014 Electromagnetic<br>compatibility | Pass |
#### XV. Wireless Foot control:
For the wireless foot switch (VAFCBT) software verification/validation of the functions of the foot switch was conducted according to IEC 62304:2006. Additionally, EMC testing was performed to evaluate the risk of communication loss according to IEC 60601-1-2:2007 and Electrical Safety Tests have been done according to IEC 60601-1-1:2005.
#### XVI. Software:
Software validation meets the requirements according to IEC 62304:2006: Medical device software - Part 1: Guidance on the application of ISO 14972 to medical device software.
#### XVII. Biocompatibility:
Biocompatibility was evaluated on attachments/nosepieces which have patient contact and meets the requirements of ISO10993. In addition, Biocompatibility testing per EN ISO 10993-1 was performed.
### XVIII. Animal and Clinical Performance Data:
No clinical data is required for this device classification submission.
#### XIX. Conclusion:
Substantial equivalence of the adeor® Velocity Alpha™ Highspeed Surgical Drill System can be assumed based on the similarity of the devices' technological characteristics, functionality and indications of use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.