K191458 · Implant Direct Sybron Manufacturing, LLC · NHA · Oct 10, 2019 · Dental
Device Facts
Record ID
K191458
Device Name
Legacy SMARTBase Abutments
Applicant
Implant Direct Sybron Manufacturing, LLC
Product Code
NHA · Dental
Decision Date
Oct 10, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Legacy™ SMARTBase Abutment system is designed to be used in support of a dental implant(s) to provide support for prosthetic restorations in a partially edentulous patient. Legacy SMARTBase engaging abutments are intended for use in the mandible or maxilla in support of single unit restorations. The Legacy SMARTBase Abutment system integrates multiple components for use in both a traditional and digital dentistry workflow: scan files from Intra-oral Scanners and lab scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. The Legacy SMARTBase system consist of two major parts: the titanium base and zirconia top components make up a two-piece abutment. - Legacy SMARTBase abutment for narrow (3.2mmD) implants: Indicated for single-tooth replacement of mandibular central and lateral incisors and maxillary lateral incisors. - Legacy SMARTBase abutment for short (8mm) 3.7mmD implants: Indicated for tooth replacement of mandibular and maxillary central and lateral incisors.
Device Story
Legacy SMARTBase Abutment is a two-piece engaging dental implant abutment; titanium base and zirconia top. Used in conventional and digital dentistry workflows to support single-unit prosthetic restorations. Inputs include scan files from intra-oral or lab scanners; CAD/CAM software used to design patient-specific restorations. Device is used in dental clinics/laboratories by clinicians or lab staff. Output is a physical abutment secured to a dental implant via fixation screw. Healthcare providers use the output to anchor crowns, restoring chewing function in partially edentulous patients. Benefits include patient-specific anatomical fit and integration with digital design workflows.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing: dynamic fatigue testing per ISO 14801:2016, biocompatibility evaluation per ISO 10993 series, and steam sterilization validation per ISO 17665-1.
Technological Characteristics
Two-piece abutment: titanium base and zirconia top. Platform diameters: 3.0mm, 3.5mm, 4.5mm, 5.7mm. Collar heights: 0.25mm, 1.0mm, 2.0mm. Compatible with Legacy implant lines. Non-sterile, intended for end-user steam sterilization. Digital workflow integration via 3Shape CAD software.
Indications for Use
Indicated for partially edentulous patients requiring single-unit prosthetic restorations in the mandible or maxilla. Specific indications: narrow (3.2mmD) implants for single-tooth replacement of mandibular central/lateral incisors and maxillary lateral incisors; short (8mm) 3.7mmD implants for tooth replacement of mandibular and maxillary central and lateral incisors.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.
October 10, 2019
Implant Direct Sybron Manufacturing LLC % Yarmela Pavlovic Partner Hogan Lovells U.S. LLP 3 Embarcadero Center #1500 San Francisco, California 94111
Re: K191458
Trade/Device Name: Legacy™ SMARTBase Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA, PNP Dated: September 10, 2019 Received: September 10, 2019
Dear Yarmela Pavlovic:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below
510(k) Number (if known) K191458 Device Name
Legacy™ SMARTBase Abutments
Indications for Use (Describe)
The Legacy™ SMARTBase Abutment system is designed to be used in support of a dental implant(s) to provide support for prosthetic restorations in a partially edentulous patient. Legacy SMARTBase engaging abutments are intended for use in the mandible or maxilla in support of single unit restorations.
The Legacy SMARTBase Abutment system integrates multiple components for use in both a traditional and digital dentistry workflow: scan files from Intra-oral Scanners and lab scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. The Legacy SMARTBase system consist of two major parts: the titanium base and zirconia top components make up a two-piece abutment.
- . Legacy SMARTBase abutment for narrow (3.2mmD) implants: Indicated for single-tooth replacement of mandibular central and lateral incisors and maxillary lateral incisors.
- Legacy SMARTBase abutment for short (8mm) 3.7mmD implants: Indicated for tooth . replacement of mandibular and maxillary central and lateral incisors.
## Compatibility
Legacy SMARTBase engaging abutments are compatible at the implant level with Legacy (3.0mm, 3.5mm, 4.5mm and 5.7mm platform diameter) implants, excluding 6mm length implants.
| Implant Line | Body Diameter | Platform Diameter | Implant Length |
|------------------------------------------------|-------------------------------------------------|----------------------------|-------------------------------|
| Legacy1 | 3.7mm, 4.2mm, 4.7mm, 5.7mm | 3.5mm, 4.5mm, 5.7mm | 8mm, 10mm, 11.5mm, 13mm, 16mm |
| Legacy2, 3, 4, simplyLegacy2,<br>simplyLegacy3 | 3.2mm, 3.7mm, 4.2mm, 4.7mm, 5.2mm, 5.7mm, 7.0mm | 3.0mm, 3.5mm, 4.5mm, 5.7mm | 16mm |
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
#### K191458
### 510(k) SUMMARY
### Implant Direct Sybron Manufacturing, LLC's Legacy SMARTBase Abutments
#### Submitter
Implant Direct Sybron Manufacturing LLC 3050 East Hillcrest Drive Thousand Oaks, CA 91362
Phone: 818-444-3306
Contact Person: Reina Choi, Regulatory Affairs Manager
Date Prepared: October 9, 2019
Name of Device: Legacy™ SMARTBase Abutments
Classification Name: Endosseous Dental Implant Abutment (21 C.F.R. § 872.3630)
Regulatory Class: Class II
Primary Product Code: NHA
Secondary Product Code: PNP
Primary Predicate: Implant Direct Sybron Manufacturing, LLC's InterActive SMARTBase Abutments (K181359)
Reference Devices: Implant Direct Sybron Manufacturing, LLC's Spectra-System Dental Implants 2008 (K090234); 3Shape A/S's 3Shape Abutment Designer Software (K151455)
#### Device Description
The Legacy SMARTBase Abutment is a two-piece engaging dental implant abutment comprised of a titanium base and a zirconia top (which can be supplied with the base or acquired separately by the customer). The abutments are offered in three widths (narrow, regular, and wide), platform diameters of 3.0mm, 3.5mm, 4.5mm and 5.7mm, and collar (titanium base) heights of 0.25, 1.0, and 2.0 mm in order to accommodate different patient anatomies. The device is supplied with fixation screws that function as an extension of the implant to which the SMARTBase is secured, and is used with several accessories in conventional and digital workflows to fabricate the patient-specific restorations, including scan adapters, implant analogs, and off-axis tools.
The Legacy SMARTBase Abutments allow for patient-specific designs through conventional and digital restoration materials and methods. The final restorations are designed and produced under the direction of a clinical professional and are based on requirements provided to Implant Direct or the preferred laboratory in digital or stone model form. The restoration (crown) is designed to fit on top of the SMARTBase abutment using off-the-shelf 3Shape software (K151455). The reference device, 3Shape
{4}------------------------------------------------
Abutment Designer Software (product code PNP), provides the digital design as an accessory to the physical dental abutment. The CAD design requires loading of the Implant Direct abutment design library via the 3Shape server to the 3Shape Software in order to design the zirconia top component within the established design limitations and specifications. The 3Shape software provides a digital design output file that is used for fabricating the finished device.
The digital workflow includes the following products (not subject devices to this submission):
- Ceramic material: Zenostar MT ●
- Cement: EMBRACE Wetbond Resin Cement (K071278) ●
- Intra-oral scanner: 3M Tru-Definition (K122467), ITero Scanner (K131101) .
- Lab scanner: 3Shape D700 & 3Shape Scan-it Restoration Dental System (510(k) exempt, . product code NOF)
- . Abutment design software: 3Shape Abutment Designer™ Software (K151455)
- Milling machine: Wieland-Zenotec Select & Zenotec CAM .
The device is single-use and supplied non-sterilization by the end user. It is an externallycommunicating device which comes in permanent contact (>30 days) with a patient's tissue/bone.
# Intended Use / Indications for Use
The Legacy SMARTBase Abutment system is designed to be used in support of a dental implant(s) to provide support for prosthetic restorations in a partially edentulous patient. Legacy SMARTBase engaging abutments are intended for use in the mandible or maxilla in support of single unit restorations.
The Legacy SMARTBase Abutment system integrates multiple components for use in both a traditional and digital dentistry workflow: scan files from Intra-oral Scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories. The Legacy SMARTbase system consist of two major parts: the titanium base and zirconia top components make up a two-piece abutment.
- Legacy SMARTBase abutment for narrow (3.2mmD) implants: Indicated for single-tooth . replacement of mandibular central and lateral incisors and maxillary lateral incisors.
- . Legacy SMARTBase abutment for short (8mm) 3.7mmD implants: Indicated for tooth replacement of mandibular and maxillary central and lateral incisors.
# Compatibility
Legacy "" SMARTBase engaging abutments are compatible at the implant level with Legacy (3.0mm, 3.5mm, 4.5mm and 5.7mm platform diameter) implants, excluding 6mm length implants.
| Implant Line | Body Diameter | Platform Diameter | Implant Length |
|------------------------------------------------|----------------------------------------------------|-------------------------------|-------------------------------------|
| Legacy1 | 3.7mm, 4.2mm, 4.7mm, 5.7mm | 3.5mm, 4.5mm,<br>5.7mm | 8mm, 10mm,<br>11.5mm, 13mm,<br>16mm |
| Legacy2, 3, 4, simplyLegacy2,<br>simplyLegacy3 | 3.2mm, 3.7mm, 4.2mm, 4.7mm,<br>5.2mm, 5.7mm, 7.0mm | 3.0mm, 3.5mm,<br>4.5mm, 5.7mm | |
The Legacy SMARTBase Abutments have the same intended use as the predicate InterActive SMARTBase Abutments, namely to aid in prosthetic rehabilitation by anchoring/supporting a restoration to a dental implant.
{5}------------------------------------------------
The Legacy SMARTBase Abutments also have similar indications for use as the predicate device and the differences do not alter the fundamental therapeutic/surgical use of the device because they either narrow its use as compared to the predicate or serve to clarify appropriate conditions of use. While the predicate device is more broadly indicated for use in both partially and fully edentulous patients, and accordingly can be used to support multi-unit restorations in addition to single-unit restorations, the target population/conditions to be treated with the Legacy SMARTBase Abutments are fully encompassed by those of the predicate device.
Additionally, the indications for use include reference to the lab scanners used with the SMARTBase Abutments. This component is also used with the previously cleared CAD/CAM system and was assessed in the previously performed CAD/CAM software verification and validation, which support both the primary predicate and the subject device. Therefore, the inclusion of the lab scanners is merely an editorial change for clarity and does not raise different questions of safety or effectiveness.
The other main difference between the two indications statements is that the predicate is compatible with the company's InterActive dental implant line as opposed to the Legacy implant line. This difference also does not raise different questions of safety or effectiveness when the device is used as labeled, because the same abutment compatibility with the company's Legacy implant line has previously been cleared in the reference device.
## Summary of Technological Characteristics
Both the subject and predicate devices are based on the funciple of providing support for dental implants in order to enable a dental restoration to be inserted into the patient's mouth and rehabilitate a patient's chewing function. At a high level, the subject and predicate devices are based on the following same technological elements:
- . Both are two-piece abutments comprised of a Titanium base and a zirconia top which can be modified to patient-specific requirements.
- . Both are used in the conventional and digital workflows, with similar auxiliary devices, to facilitate fabrication of a patient-specific dental restoration.
- . Both feature the same fundamental design (e.g., size and diameter ranges, post height).
The main technological difference between the subject and predicate devices is the abutment- implant interface: The subject device mates with the Legacy implant line whereas the predicate device mates with the InterActive implant line.
A table comparing the key features of the subject and predicate devices is provided below.
### Performance Data
Non-clinical testing was performed on the proposed device, including mechanical strength (fatigue), biocompatibility, cleaning and steam sterilization validation, and software verification. Successful test results indicated that the Legacy SMARTBase Abutments will perform as intended, and support the device's substantial equivalence.
- . Dynamic Fatique testing on the worst-case device configurations per ISO 14801;2016 (consistent with, FDA's Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments). Results confirmed that the Legacy SMARTBase abutments perform equivalently to the predicate device.
{6}------------------------------------------------
- Biocompatibility evaluation was conducted according to ISO 10993-1 and FDA's corresponding ● June 2016 quidance, for an externally-communicating device with permanent (>30 day) contact with patient bone/tissue/blood. Tested was performed in accordance with the following standards: ISO 10993-2, ISO 10993-3, ISO 10993-5, ISO 10993-10, ISO 10993-11, ISO 10993-12 and ISO 10993-33 to demonstrate that the devices met biocompatibility requirements for its intended use.
- The proposed devices are provided non-sterile and intended to be steam sterilized by the end user. Steam sterilization validation according to ISO 14947 and ISO 17665-1 to demonstrate attainment of a sterility assurance level of 10° using the same steam sterilization processes as the predicate device.
- . Software validation ensured the ability of the system to scan the articulator and design the model to configure to the user's need to successfully create abutments employing 3Shape software within established design limitations and specification and validation testing performed in support of the previously cleared InterActive SMARTBase Abutments (K181359) was also relied upon as the subject device employs the identical off-the-shelf abutment designer software and milling unit.
Clinical data is not needed to characterize performance and establish substantial equivalence. The nonclinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Thus, clinical testing has not been conducted on this product.
### Conclusion
The Legacy SMARTBase Abutments have the same intended use and principles of operation, and similar indications for use and technological characteristics, as its predicate device. The minor differences in indications do not alter the intended therapeutic/surgical use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the Legacy SMARTBase Abutment system and its predicate device raise no new issues of safety or effectiveness, and have been further addressed by performance data which demonstrates that the device performs in a substantially equivalent manner as the predicate device. Thus, the Legacy SMARTBase Are substantially equivalent to the predicate InterActive SMARTBase Abutment System (K181359).
{7}------------------------------------------------
| | Legacy SMARTBase<br>Abutment<br>(Subject Device) | InterActive SMARTBase<br>Abutment (K181359)<br>(Primary Predicate) | Spectra-System Dental<br>Implants 2008 (K090234)<br>(Reference Device) | 3Shape Software<br>(K151455)<br>(Reference Device) |
|--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for<br>Use | The Legacy SMARTBase Abutment<br>system is designed to be used in<br>support of a dental implant(s) to provide<br>support for prosthetic restorations in a<br>partially edentulous patient.<br>Legacy SMARTBase engaging<br>abutments are intended for use in the<br>mandible or maxilla in support of single<br>unit restorations.<br>The Legacy SMARTBase Abutment<br>system integrates multiple components<br>for use in both a traditional and digital<br>dentistry workflow: scan files from Intra-<br>oral Scanners and lab scanners, CAD<br>software, CAM software, ceramic<br>material, milling machine and<br>associated tooling and accessories. The<br>Legacy SMARTBase system consist of<br>two major parts: the titanium base and<br>zirconia top components make up a two-<br>piece abutment.<br>• Legacy SMARTBase abutment<br>for narrow (3.2mmD) implants:<br>Indicated for single-tooth<br>replacement of mandibular<br>central and lateral incisors and<br>maxillary lateral incisors.<br>• Legacy SMARTBase abutment<br>for short (8mm) 3.7mmD<br>implants: Indicated for tooth<br>replacement of mandibular and<br>maxillary central and lateral<br>incisors. | InterActive/SwishActive Implant<br>System consists of two-piece<br>implants for one-stage or two- stage<br>surgical procedures. These implants<br>are intended for use in partially and<br>fully edentulous upper and lower<br>jaws in support of single or multiple-<br>unit restorations and terminal or<br>intermediate SMARTBase support<br>for fixed bridgework. The system<br>integrates multiple components of<br>the digital dentistry workflow: scan<br>files from Intra-Oral Scanners, CAD<br>software, CAM software, ceramic<br>material, milling machine and<br>associated tooling and accessories.<br>The SMARTBase Abutments consist<br>of two major parts. Specifically, the<br>titanium base and zirconia top<br>components make up a two- piece<br>abutment.<br>Implants can be indicated for<br>immediate loading when good<br>primary stability has been<br>achieved and with appropriate<br>occlusal loading.<br>• Narrow Diameter (3.2. 3.3mm)<br>Implants: Indicated for single- tooth<br>replacement of mandibular central<br>and lateral incisors and maxillary<br>lateral incisors. Also indicated for<br>multiple tooth replacements or<br>denture stabilization. | Spectra-System Dental<br>Implants 2008 are comprised of<br>dental implant fixtures and<br>prosthetic devices that<br>compose a two-piece implant<br>system. The implants are<br>intended for use in the<br>mandible and maxilla, in<br>support of single unit or multiple<br>unit cement or screw-receiving<br>restorations and for the<br>retention and support of<br>overdentures. The implants are<br>intended for immediate<br>placement and function for the<br>support of single tooth or<br>multiple- tooth restorations,<br>recognizing bone stability and<br>appropriate occlusal load<br>requirements. | The 3Shape Abutment<br>Designer Software is intended<br>as an aid to the restoration of<br>chewing function in partially or<br>fully edentulous mandibles and<br>maxillae. The 3Shape<br>Abutment Designer Software is<br>intended for use by a dental<br>practitioner or dental laboratory<br>staff for designing the patient<br>specific component of a two-<br>piece, one-piece, or hybrid<br>dental implant abutment. The<br>single or multi-unit abutment<br>design is intended to be used<br>by the manufacturer of an<br>endosseous dental implant<br>abutment to create the final<br>device. |
| Compatibility | • Implant Direct Legacy1: body<br>diameter 3.7 - 5.7mm, platform<br>diameter 3.5, 4.5, 5.7mm 3.5,<br>4.5, 5.7mm; | • Implant Direct interactive (3.2,<br>3.7, .3, 5.0mm body diameter,<br>3.0 and 3.4mm)<br>• Implant Direct SwishActive | N/A | N/A |
| | Legacy SMARTBase<br>Abutment<br>(Subject Device) | InterActive SMARTBase<br>Abutment (K181359)<br>(Primary Predicate) | Spectra-System Dental<br>Implants 2008 (K090234)<br>(Reference Device) | 3Shape Software<br>(K151455)<br>(Reference Device) |
| | • Implant Direct Legacy2, 3, 4,<br>simplyLegacy2 & 3: body<br>diameter 3.2 - 7.0mm; platform<br>diameter 3.0, 3.5, 4.5, 5.7mm,<br>excluding 6mm length | (3.3, 4.1, 4.8mm body<br>diameter, 3.0 and 3.4mm<br>platform) | | |
| Prosthesis<br>Attachment | Screw- or cement-retained | Screw- or cement-retained | Screw- or cement-retained | N/A |
| Restoration | Single-unit | Single-unit and multi-unit | Single-unit and multi-unit | Single-unit and multi-unit |
| Abutment<br>Angulation | 0° to 30° | 0° to 30° | 0° to 30° | N/A |
| Abutment<br>Collar Heights | 0.25mm – 2mm | 1.0mm – 2.0mm | 1.0mm – 3.0mm | N/A |
| Post Height | | 4mm minimum | N/A | N/A |
| Abutment<br>material<br>and surface<br>treatment | | Titanium & Zirconia; abutments are titanium and anodized gold and pink (with<br>grooves machined for cement adhesion) | Titanium alloy | N/A |
| Cement<br>Adhesive | | EMBRACE (K071278) | N/A | N/A |
| Screw Material | | Titanium | Titanium | N/A |
| Implant<br>Interface<br>Platform | 3.0mm, 3.5mm, 4.5mm, and 5.7mm<br>platform diameter | 3.0mm and 3.4mm platform diameter | 3.0, 3.5, 4.5, and 5.7mm<br>diameter at the interface | N/A |
| Sterility | | Supplied non-sterile; steam sterilized by end user prior to use | Supplied non-sterile; steam<br>sterilized by end user prior to use | N/A |
| Number of<br>uses | | Single use | Single use | N/A |
## Substantial Equivalence Table
{8}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.