BONEBRIDGE System, Bone Conduction Implant Kit (BCI 602 Implant Kit), BCI 602 Sizer Kit, BCI 602 Lifts (1mm)

K191457 · Med-El Elektromedizinische Geraete GmbH · PFO · Sep 18, 2019 · Ear, Nose, Throat

Device Facts

Record IDK191457
Device NameBONEBRIDGE System, Bone Conduction Implant Kit (BCI 602 Implant Kit), BCI 602 Sizer Kit, BCI 602 Lifts (1mm)
ApplicantMed-El Elektromedizinische Geraete GmbH
Product CodePFO · Ear, Nose, Throat
Decision DateSep 18, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.3340
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Bonebridge is intended to improve hearing for paticats with conductive or mixed hearing losses, bilatcral fitting, and single-sided deafness.

Device Story

Bonebridge is an active implantable bone conduction hearing system. It consists of an implantable component and an external audio processor. The system captures sound via the audio processor, which transmits signals to the implanted transducer. The transducer converts these signals into mechanical vibrations, which are conducted through the skull bone to the inner ear, bypassing the outer and middle ear. It is used by patients with conductive/mixed hearing loss or single-sided deafness. The device is surgically implanted by a physician. Patients use the external audio processor to receive sound amplification. The system aims to improve hearing and speech recognition for patients who may not benefit from or cannot use traditional air-conduction hearing aids.

Technological Characteristics

Active implantable bone conduction hearing system. Consists of an implantable transducer (BCI 602), sizer kit, and lifts. Operates via bone conduction to stimulate the inner ear. Class II device (21 CFR 874.3340).

Indications for Use

Indicated for patients 12+ years old with conductive or mixed hearing loss (PTA BC threshold ≤ 45 dB HL) or single-sided deafness (profound sensorineural loss in one ear, normal hearing in opposite ear with PTA AC threshold ≤ 20 dB HL). Bilateral fitting indicated for symmetric conductive/mixed loss. Also indicated for patients who are candidates for AC CROS hearing aids but cannot/will not use them. Contraindications: chronic vestibular/balance disorders, progressive hearing loss, poor speech recognition potential, skin/scalp conditions precluding processor attachment, or skull abnormalities preventing implant placement.

Regulatory Classification

Identification

An active implantable bone conduction hearing system is a prescription device consisting of an implanted transducer, implanted electronics components, and an audio processor. The active implantable bone conduction hearing system is intended to compensate for conductive or mixed hearing losses by conveying amplified acoustic signals to the cochlea via mechanical vibrations on the skull bone.

Special Controls

In combination with the general controls of the FD&C Act, the active implantable bone conduction hearing system is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must characterize any adverse events observed during implantation and clinical use, and must also demonstrate that the device performs as intended under anticipated conditions of use. (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including the following: (i) Performance data must validate force output in a clinically relevant model. (ii) Impact testing in a clinically relevant anatomic model must be performed. (iii) Mechanical integrity testing must be performed. (iv) Reliability testing consistent with expected device life must be performed. (3) The patient-contacting components of the device must be demonstrated to be biocompatible. (4) Performance data must demonstrate the sterility of the patient-contacting components of the device. (5) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life. (6) Performance data must demonstrate the wireless compatibility, electromagnetic compatibility, and electrical safety of the device. (7) Software verification, validation, and hazard analysis must be performed. (8) Labeling must include: (i) A summary of clinical testing conducted with the device that includes a summary of device-related complications and adverse events; (ii) Instructions for use; (iii) A surgical guide for implantation, which includes instructions for imaging to assess bone dimensions; (iv) A shelf life, for device components provided sterile; (v) A patient identification card; and (vi) A patient user manual.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text. September 18, 2019 MED-EL Elektromedizinische Geraete GmbH Ines Santiago Regulatory Affairs Specialist Fuestenweg 77a Innsbruck, 6020 AT Re: K191457 Trade/Device Name: BONEBRIDGE System. Bone Conduction Implant Kit (BCI 602 Implant Kit), BCI 602 Sizer Kit, BCI 602 Lifts (1mm) Regulation Number: 21 CFR 874.3340 Regulation Name: Active implantable bone conduction hearing system Regulatory Class: Class II Product Code: PFO Dated: August 20, 2019 Received: August 23, 2019 Dear Ines Santiago: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Srinivas Nandkumar, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) Device Name BONEBRIDGE or Bonebridge ## Indications for Use (Describe) INTENDED USE: The Bonebridge is intended to improve hearing for paticats with conductive or mixed hearing losses, bilatcral fitting, and single-sided deafness. ## INDICATIONS: The Bonebridge bone conduction hearing implant system is intended for the following patients and indications: - Patients 12 years of age or older. - Patients who have a conductive or mixed hearing loss and still can benefit from sound amplification. The pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5, 1, 2 and 3 kHz) should be better than or equal to 45 dB HL. - Bilateral fitting of the Bonebridge is intended for patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2 and 3 kHz, or less than 15 dB at individual frequencies. - Patients who have profound sensorineural hearing loss in one ear and normal hearing in the opposite ear (i.e., single-sided deafness or "SSD"). The pure tone average air conduction hearing thresholds of the hearing ear should be better than or equal to 20 dB HL (measured at 0.5, 1, 2 and 3 kHz). -The Bonebridge for SSD is also indicated for any patient who is indicated for an air-conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS. - Prior to receiving the device, it is recommended that an individual has experience with appropriately fit air conduction or bone conduction hearing aids. CONTRAINDICATIONS: - Chronic or non-revisable vestibular or balance disorders. - Abnormally progressive hearing loss. - Evidence of conditions that would prevent good speech recognition potential as determined by good clinical judgment. - Skin or scalp conditions that may preclude attachment of the audio processor or that may interfere with the use of the audio processor. -Skull size or abnormality that would preclude appropriate placement of the Bonebridge implant as determined by CT scan. Documented in IU18-019 (see attached). Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ ## This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...