K191448 · Ivoclar Vivadent, AG · PEM · Oct 8, 2019 · General, Plastic Surgery
Device Facts
Record ID
K191448
Device Name
Barrier Sleeves
Applicant
Ivoclar Vivadent, AG
Product Code
PEM · General, Plastic Surgery
Decision Date
Oct 8, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4370
Device Class
Class 2
Indications for Use
Barrier Sleeves are intended to serve as a disposable barrier for dental instruments and equipment. This device is nonsterile and intended for single patient use only.
Device Story
Barrier Sleeves consist of low-density polyethylene film; designed as disposable covers for dental instruments and equipment. Device slips over ends of instruments; acts as physical barrier to augment infection control; facilitates clean-up. Used in dental clinics by dental professionals. Provides protection against contamination during dental procedures; benefits patient by reducing cross-contamination risk. No electronic or software components.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing including biocompatibility (ISO 10993-5 cytotoxicity, ISO 10993-10 sensitization/irritation) and performance testing (ASTM F1670/F1670M synthetic blood penetration, ASTM F1671/F1671M viral penetration, ASTM D882 tensile strength, ASTM F1342/F1342M puncture resistance, and ASTM D1004 tear resistance).
Technological Characteristics
Material: Low-density polyethylene film. Thickness: 0.03 mm. Form factor: Custom-shaped sleeves/covers. Energy source: None (passive). Connectivity: None. Sterilization: Non-sterile. Standards: ASTM D882 (tensile), ASTM F1342 (puncture), ASTM D1004 (tear), ASTM F1670 (synthetic blood), ASTM F1671 (viral penetration).
Indications for Use
Indicated for use as a disposable barrier for dental instruments and equipment in a dental clinical setting. Intended for single patient use.
Regulatory Classification
Identification
A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.
Special Controls
*Classification.* Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
October 8, 2019
Ivoclar Vivadent, AG % Lori Aleshin Director of Quality & Regulatory Affairs Ivoclar Vivadent, Inc. 175 Pineview Drive Amherst, New York 14228
Re: K191448
Trade/Device Name: Barrier Sleeves Regulation Number: 21 CFR 878.4370 Regulation Name: Surgical Drape and Drape Accessories Regulatory Class: Class II Product Code: PEM Dated: June 26, 2019 Received: July 10, 2019
Dear Lori Aleshin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Healthh
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K191448
Device Name Barrier Sleeve
Indications for Use (Describe)
Barrier Sleeves are intended to serve as a disposable barrier for dental instruments and equipment. This device is nonsterile and intended for single patient use only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Ivoclar Vivadent. The logo is in blue and green, with the words "Ivoclar Vivadent" in a bold, sans-serif font. Below the company name is the tagline "passion vision innovation" in a smaller, sans-serif font. The logo also includes a series of green and blue dots and squares arranged in an arc above the company name.
### Submitter Information:
Lori Aleshin, Director of Quality and Requlatory Affairs lvoclar Vivadent. Inc. 175 Pineview Drive Amherst, New York 14228 716-264-2045 lori.aleshin@ivoclarvivadent.com
Company: Ivoclar Vivadent, AG Bendererstrasse 2, Schaan, FL-9494, Liechtenstein +423-235-3535
> Date Prepared: October 7, 2019
#### Device Name:
| Proprietary Name: | Barrier Sleeves |
|----------------------|--------------------------------------|
| Classification Name: | Surgical Drape and Drape Accessories |
| CFR Number: | (878.4370) |
| Device Class: | II |
| Product Code | PEM |
Predicate Device: Disposable Barrier Sleeves and Covers (K160232) by DENTSPLY International Inc.
Device Description: Barrier Sleeves are made of polyethylene film and are used as accessories to dental instruments and equipment used during dental procedures. These disposable barrier sleeves are offered in various shapes and sizes to fit over and cover the intended dental instruments and equipment. The disposable Barrier Sleeves slip over the ends of the respective devices, allowing for the attachment of the devices used during dental procedures. Barrier Sleeves act as a physical barrier, augmenting existing infection control techniques, and facilitate clean-up.
Indications for Use: Barrier Sleeves are intended to serve as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only
| Intended Use | Predicate device | Barrier Sleeves |
|-----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications | Disposable Barrier Sleeves and Covers are intended<br>to serve as a disposable barrier for dental instruments<br>and equipment. This device is non-sterile and in-<br>tended for single patient use only. | Barrier Sleeves are intended to serve<br>as a disposable barrier for dental<br>instruments and equipment. This de-<br>vice is non-sterile and intended for<br>single patient use only. |
| Precaution<br>Measures/<br>Contraindications | None | None |
| Summary of Indications | The indications and contraindications for the predicate and the new device are<br>the same. | |
| Technology | Predicate device | Barrier Sleeves |
| Working principle | Disposable Barrier Sleeves and<br>Covers, are made of<br>polyethylene film and are used<br>as accessories to dental<br>instruments and equipment<br>used during dental procedures.<br>These disposable barrier<br>sleeves and covers are offered<br>in various shapes and sizes to<br>fit over and cover the intended<br>dental instruments and<br>equipment. The Disposable<br>Barrier Sleeves slip over the<br>ends of the respective devices,<br>allowing for the attachment<br>and/or protrusion (where<br>applicable) of those parts of the<br>devices used during dental<br>procedures. The Disposable<br>Barrier Cover is placed over<br>equipment, such as dental chair<br>head rests or trays, to cover<br>those parts.<br>Disposable Barrier Sleeves<br>and Covers act as a physical<br>barrier, augmenting existing<br>infection control techniques and<br>facilitate clean-up. | Disposable Barrier Sleeves are made of<br>polyethylene film and are used as<br>accessories to dental instruments and<br>equipment used during dental procedures.<br>These disposable barrier sleeves are<br>offered in various shapes and sizes to fit<br>over and cover the intended dental<br>instruments and equipment. The<br>Disposable Barrier Sleeves slip over the ends<br>of the respective de- vices, allowing for the<br>attachment of those parts of the devices used<br>during dental procedures.<br>Disposable Barrier Sleeves act as a physical<br>barrier, augmenting existing infection control<br>techniques, and facilitate clean-up. |
| Summary Principles of<br>operation | The working principle for both groups of devices is the same: the disposable<br>barrier sleeves are placed over the devices. The form fits over the intended<br>devices. The working principle is equivalent. | |
| Delivery forms/dosage | There are several sleeves and<br>covers for dental instruments<br>and equipment. | There are several sleeves for dental<br>instruments and equipment. |
| Summary of Delivery form<br>content | In the Clearance letter K160232 several sleeves and covers are listed. For the<br>new devices (Bluephase Sleeves and Vivapen Sleeves) there are supposed to<br>be 2 articles, which differ in the size of the sleeve. The sleeves perfectly fit over<br>Ivoclar Vivadent`s curing lights (Bluephase) and Vivapen (applicator for<br>liquids). | |
| Shelf life | Supplier report is 5 years | |
| Summary of shelf life | Shelf life for the predicate is not known. The claimed shelf life for the new<br>devices (Bluephase and Vivapen Sleeves) is 5 years as recommended by the<br>manufacturer of the sleeves. Storage Stability report was provided. | |
| Principles of operation | The disposable barrier sleeves and covers fit over and cover the intended dental<br>instruments and equipment. | |
| Summary of Operation | The principle of operation is the same. | |
| Material Composition | Predicate device | Barrier Sleeves |
| Material | Polyethylene film | Polyethylene film |
| Material composition | Low density polyethylene film | Low density polyethylene film |
| Summary of Material<br>Composition | No difference. The predicate and the new device are manufactured by the<br>same supplier (Minitube AB, Sweden) and made of the same material. | |
| Biocompatibility | Predicate device | Barrier Sleeves |
| Cytotoxicity - ISO 10993-5 | Non-cytototoxic | Non-cytototoxic |
| Sensitization/Irritation – ISO<br>10993-10 | Non-sensitizing, non-irritating | Non-sensitizing, non-irritating |
| Summary of Biocompatibility | No difference. The predicate and the new device are manufactured by the<br>same supplier (Minitube AB, Sweden) and made of the same material. The<br>results of Biocompatibility testing are included in this submission. | |
| Performance Properties | Predicate device | Barrier Sleeves |
| Synthetic Blood Penetration<br>ASTM F1670/F1670M | Pass | Pass |
| Synthetic Blood Penetration at<br>seams and non-continuous<br>components ASTM<br>F1670/F1670M | Pass | Pass |
| Viral penetration ASTM<br>F1671/F1671M | Pass | Pass |
| Viral penetration at seams and<br>non- continuous components<br>ASTM F1671/F1671M | Pass | Pass |
| Summary | No difference. The predicate and the new device are manufactured by the<br>same supplier (Minitube AB, Sweden) and made of the same material. The<br>results of Performance testing are included in this submission. | |
| Mechanical Properties | Predicate device | Barrier Sleeves |
| Tensile strength ASTM D882 | 0.03 mm - acceptable | 0.03 mm - acceptable |
| Puncture Resistance ASTM<br>F1342/F1342M | 0.03 mm - acceptable | 0.03 mm - acceptable |
| Tear Resistance ASTM D1004 | 0.03 mm – acceptable | 0.03 mm - acceptable |
| Summary | No difference. The predicate and the new device are manufactured by the same supplier (Minitube AB, Sweden) and made of the same material. The results of Mechanical Properties testing are included in this submission. | |
| Device Characteristics | Predicate device | Barrier Sleeves |
| Thickness | 0.03 mm (Minitube AB, Sweden) | 0.03 mm (Minitube AB, Sweden) |
| Shape | Custom design to fit the intended dental instruments and equipment they cover | Custom design to fit the intended dental instruments and equipment they cover |
| Dimensions | Determined by the size and shape of the dental instruments and equipment they cover | Determined by the size and shape of the dental instruments and equipment they cover |
| Color | clear | clear |
| Sterile | Not sterile | Not sterile |
| Single Use | Single use device | Single use device |
| Summary | The new and the predicate devices are made of the same material by the same supplier (Minitube AB, Sweden). Both devices are not sterile, for single use and differ only in the shape and the dimensions as the need to fit over different types of devices | |
### Comparison to Predicate:
{4}------------------------------------------------
passion vision innovation
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Ivoclar Vivadent, a dental company. The logo features the company name in a stylized font, with the words "passion vision innovation" written below. Above the company name is a series of green and blue dots arranged in an arc. The number 3 is in the upper right corner of the image.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the number 4 in the upper right corner. The image also contains the logo for Ivoclar Vivadent. The logo is blue and green and contains the words "Ivoclar Vivadent: passion vision innovation".
#### Infection Control:
The device is sold in a non-sterile condition.
#### Non-Clinical Performance Data:
The sponsor provided biocompatibility testing for the subject device. Direct contact with tissue is not intended. Therefore ISO 10993-1 is not applicable. Biocompatibility testing included Cytotoxicity study using the ISO Elution Method ISO 10993-5, the ISO 10993-10 Guinea Pig Maximization Sensitization Test and the ISO 10993-10 Intracutaneous Study in Rabbits-Irritation test. A Toxicological statement was also included. The performance of the proposed device, Barrier Sleeves, met the requirements of the biocompatibility testing conducted to support substantial equivalence with the primary predicate, Disposable Barrier Sleeves and Covers (K160232).
The performance testing (Bench) of the subject device, Barrier Sleeves, met the requirements of the non-clinical bench testing conducted to support substantial equivalence with the primary predicate, Disposable Barrier Sleeves and Covers (K160232). Below is a summary of the testing performed:
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Ivoclar Vivadent, a dental company. The logo features the company name in a stylized font, with the words "passion vision innovation" written below. Above the company name is a series of green and blue dots and squares arranged in an arc shape.
- ASTM F1670/F1670M Standard Test Method for Resistance of Materials Used in ● Protective Clothing to Penetration by Synthetic Blood
- . ASTM F1671/ F1671M Standard Test Method for Resistance of Materials Used in Protective Clothing to Penetration by Blood-Borne Pathogens Using Phi-X174 Bacteriophage Penetration as a Test System
| Mechanical<br>Property | Standard | Proposed Device<br>Barrier Sleeves | Predicate Device<br>Disposable Barrier<br>Sleeves and Covers<br>(K160232) |
|------------------------|--------------------------|-------------------------------------|---------------------------------------------------------------------------|
| Tensile Strength | ASTM D882 | 0.03mm - Comparable<br>to Predicate | Acceptable for<br>0.03mm |
| Puncture<br>Resistance | ASTM<br>F1342/<br>F1342M | 0.03mm - Comparable<br>to Predicate | Acceptable for<br>0.03mm |
| Tear Resistance | ASTM<br>D1004 | 0.03mm Comparable<br>to Predicate | Acceptable for<br>0.03mm |
The range of mechanical properties for the samples is reported for the proposed devices. Barrier Sleeves. Below is a summary of the testing performed:
- . ASTM D882 Standard Test Methods for Tensile Properties of Thin Plastic Sheeting
- ASTM F1342/ F1342M Standard Test Method for Protective Clothing Material ● Resistance to Puncture
- ASTM D1004 Standard Test Method for Tear Resistance (Graves Tear) of Plastic Film and Sheeting
The range of mechanical properties for the samples of the proposed devices, Barrier Sleeves, is comparable to those reported for the predicate device. Disposable Barrier Sleeves and Covers (K160232). The results indicate the subject device passed performance testing under the conditions of the tests, and support the substantial equivalence of the subject device, Barrier Sleeves to its primary predicate, Disposable Barrier Sleeves and Covers (K160232).
#### Clinical Performance Data:
No data from human clinical studies have been included to support the substantial equivalence of the proposed devices, Barrier Sleeves as clinical studies are not required for this medical device.
#### Conclusions Regarding Substantial Equivalence;
The proposed device, Barrier Sleeves, consists of dental barrier sleeves and covers which are intended to serve as a disposable barrier for dental instruments and equipment. These barrier sleeves are non-sterile and are intended for single patient use only. The proposed device, Barrier Sleeves, have the same intended use and indications for use and incorporate the same fundamental technology and working principle, as the primary predicate device, Disposable Barrier Sleeves and Covers, cleared under premarket notification (K160232). Test data to verify the
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the Ivoclar Vivadent logo. The logo is blue and green. The word "Ivoclar" is on top of the word "Vivadent".
ssion vision innovation
performance of the proposed device, Barrier Sleeves, have been provided, including cytotoxicity, skin sensitization, oral mucosa irritation, synthetic blood penetration, synthetic blood penetration at seams and non-continuous components, viral penetration and viral penetration at seams and non-continuous components. The range of mechanical properties has been provided for puncture resistance, tear resistance, and tensile strength. The test results, combined with the design, intended use and mechanical property comparisons with the primary predicate device, support the substantial equivalence of the proposed Barrier Sleeves with its primary predicate device Disposable Barrier Sleeves and Covers (K160232).
{9}------------------------------------------------
Image /page/9/Picture/1 descriptio…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.