# Ah!Yes Ob Coco Personal Lubricant (K191411)

_Bio-Tech Lubricants, Ltd. · NUC · Feb 6, 2020 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K191411

## Device Facts

- **Applicant:** Bio-Tech Lubricants, Ltd.
- **Product Code:** [NUC](/productcode/NUC.md)
- **Decision Date:** Feb 6, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

AH!YES® OB COCO: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. AH! YES® OB COCO is not compatible with natural rubber latex, polyurethane or polyisoprene condoms.

## Device Story

AH! YES OB COCO is a non-sterile, plant-oil based personal lubricant; consists primarily of sunflower oil and fractionated coconut oil. Packaged in polyethylene single-use applicator. Used by consumers for intimate sexual activity; applied vaginally or penically to supplement natural lubrication. Opaque paste at room temperature; melts at body temperature. Not a contraceptive or spermicide. Not compatible with natural rubber latex, polyurethane, or polyisoprene condoms. Benefits include enhanced ease and comfort during sexual activity.

## Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993-1 (Cytotoxicity ISO 10993-5, Sensitization ISO 10993-10, Vaginal Irritation ISO 10993-10, Acute Systemic Toxicity ISO 10993-11) confirmed non-cytotoxic, non-irritating, non-sensitizing, and non-systemically toxic. Shelf-life validated to 24 months via accelerated aging (ASTM F1980-16) and real-time storage. Condom compatibility tested per ASTM D7661-10; confirmed incompatible with latex, polyisoprene, and polyurethane.

## Technological Characteristics

Plant-oil based lubricant (sunflower, coconut, jojoba, grapeseed oils, Vitamin E, Dermofeel Viscolid). Opaque paste form factor. Packaged in polyethylene single-use applicator. Non-sterile. Shelf-life validated per ASTM F1980-16. Microbial limits: TAMC <100 cfu/g, TYMC <10 cfu/g, pathogens absent per USP <61>/<62>. Specific gravity 0.85-1.0.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- YES Oil-based Personal Lubricant ([K162569](/device/K162569.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 6, 2020

Bio-Tech Lubricants Ltd Karen Cornelius Regulatory Affairs Manager C/O The Timber Barn, Selborne Road Greatham, Liss, Hampshire GU33 6HG UNITED KINGDOM

K191411 Re: Trade/Device Name: AH! YES OB COCO Personal Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: December 23, 019 Received: December 23, 2019

Dear Karen Cornelius:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/trainingand-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory- assistance/contact-usdivision-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K191411

Device Name AH!YES OB COCO Personal Lubricant

#### Indications for Use (Describe)

AH! YES® OB COCO: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. AH! YES® OB COCO is not compatible with natural rubber latex, polyurethane or polyisoprene condoms.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510 (k) Summary – K191411

#### 1. Submitter Information

| Applicant: | Bio-Tech Lubricants Limited                                                                           |
|------------|-------------------------------------------------------------------------------------------------------|
| Address:   | Epsilon House, Enterprise Road<br>Southampton Science Park,<br>Southampton, SO6 7NS<br>United Kingdom |
| Phone:     | +44 (0) 3450 94 11 41                                                                                 |
| Email:     | sarah@ahyes.org                                                                                       |

#### 2. Correspondent Information

| Contact: | Karen Cornelius<br>Bio-Tech Lubricants Limited                                               |
|----------|----------------------------------------------------------------------------------------------|
| Address: | C/O The Timber Barn,<br>Selborne Road, Greatham,<br>Hampshire.<br>GU33 6HG<br>United Kingdom |
| Phone:   | +44 (0) 3450 94 11 41                                                                        |
| Email:   | Karen@yesyesyes.org                                                                          |

#### 3. Date prepared: February 04, 2020

#### 4. Device Information

| Device Name:       | AH! YES OB COCO Personal Lubricant |
|--------------------|------------------------------------|
| Common Name:       | Personal Lubricant                 |
| Regulation Number: | 21 CFR 884.5300                    |
| Regulation Name:   | Condom                             |
| Regulatory Class:  | Class II                           |
| Product Code:      | NUC (lubricant, personal)          |

### 5. Predicate Device Information

| Device Name:      | YES Oil-based Personal Lubricant |
|-------------------|----------------------------------|
| 510(k) Number:    | K162569                          |
| Manufacturer:     | Bio-Tech Lubricants Limited      |
| Regulatory Class: | Class II                         |
| Product Code:     | NUC (lubricant, personal)        |

The predicate device has not been subject to a design-related recall.

#### 6. Device Description

AH! YES OB COCO is a non-sterile, plant-oil based personal lubricant that consists primarily of sunflower oil and fractionated coconut oil and is not compatible with condoms. It is neither a contraceptive nor a spermicide. The formulation is an opaque paste that melts at body temperature. The product is packaged in a polyethylene single use applicator.

{4}------------------------------------------------

The device specifications for AH! YES OB COCO are listed in the table below:

| Property                                             | Specification                                                           |
|------------------------------------------------------|-------------------------------------------------------------------------|
| Appearance                                           | Opaque paste below melt point, oily pale yellow liquid above melt point |
| Color                                                | Pale yellow                                                             |
| Odor                                                 | Slight odor, strengthening with age                                     |
| Specific Gravity                                     | 0.85 - 1.0                                                              |
| Total aerobic microbial count (TAMC) per USP <61>    | Less than 100 cfu/g                                                     |
| Total yeast and mold count (TYMC) per USP <61>       | Less than 10 cfu/g                                                      |
| Presence of Pathogens per USP <62>                   | Specification                                                           |
| Staphylococcus aureus                                | Absent                                                                  |
| Candida albicans                                     | Absent                                                                  |
| Gram negative bacteria (e.g. Pseudomonas aeruginosa) | Absent                                                                  |

# Table 1: Device Specifications for "AH! YES OB COCO personal lubricant"

# 7. Indications for Use

AH!YES® OB COCO: A personal lubricant, for vaginal and penile application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. AH! YES® OB COCO is not compatible with natural rubber latex, polyurethane or polyisoprene condoms.

# 8. Comparison of Intended Use and Technological Characteristics with the Predicate Device

The table below lists the a comparison of the indications for use and technological characteristics of the subject and predicate device.

# Table 2: Comparator Table for Subject Device – The AH!YES OB COCO Personal Lubricant and Predicate Device YES Oil-based Personal Lubricant

|                               | AH!YES OB COCO<br>Personal Lubricant<br>(K191411)                                                                                                                                                                                                                                                                                                        | YES Oil-based Personal<br>Lubricant (K162569)                                                                                                                                                                                                                                                                                                                         |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Classification<br>Name | Lubricant, Personal                                                                                                                                                                                                                                                                                                                                      | Lubricant, Personal                                                                                                                                                                                                                                                                                                                                                   |
| Product Code                  | NUC                                                                                                                                                                                                                                                                                                                                                      | NUC                                                                                                                                                                                                                                                                                                                                                                   |
| Indications for Use           | AH!YES® OB COCO: A personal<br>lubricant, for vaginal and penile<br>application, intended to moisturize<br>and lubricate, to enhance the ease<br>and comfort of intimate sexual<br>activity and supplement the body's<br>natural lubrication. AH! YES OB<br>COCO is not compatible with<br>natural rubber latex, polyurethane<br>or polyisoprene condoms | YES OB personal lubricant is a<br>personal lubricant, for vaginal<br>and penile application, intended<br>to moisturize and lubricate, to<br>enhance the ease and comfort of<br>intimate sexual activity and<br>supplement the body's natural<br>lubrication. This product is not<br>compatible with natural rubber<br>latex, polyurethane or<br>polyisoprene condoms. |

{5}------------------------------------------------

| Water soluble                      | No                                                                                                   | No                                                                                              |
|------------------------------------|------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Contains water                     | No                                                                                                   | No                                                                                              |
| Oil based lubricant                | Yes                                                                                                  | Yes                                                                                             |
| Primary ingredients                | Sunflower seed oil, Coconut oil,<br>Jojoba seed oil, Dermofeel<br>Viscolid, Grapeseed Oil, Vitamin E | Sunflower seed oil, Shea butter,<br>Sweet almond oil, Bees wax,<br>Cocoa seed butter, Vitamin E |
| Sterile                            | No                                                                                                   | No                                                                                              |
| Condom Compatibility               | Not compatible with NRL,<br>polyisoprene or polyurethane<br>condoms                                  | Not compatible with NRL,<br>polyisoprene or polyurethane<br>condoms                             |
| Biocompatibility Tested            | Yes                                                                                                  | Yes                                                                                             |
| Total Microbial Count              | Less than 100 cfu/g                                                                                  | Less than 10 cfu/g                                                                              |
| Fungal/Yeast/Mold<br>Limits        | Less than 10 cfu/g                                                                                   | Less than 10 cfu/g                                                                              |
| Absence of Pathogenic<br>Organisms | Absent                                                                                               | Absent                                                                                          |
| Packaging materials                | Plastic / Polyethylene                                                                               | Plastic / Polyethylene                                                                          |

The subject and predicate device have similar indications for use and have the same intended use. The subject and predicate device have different technological characteristics, including different formulations and specifications. The different technological characteristics of the subject device do not raise different types of safety and effectiveness questions.

# 9. Summary of Non-Clinical Performance Testing

# Biocompatibility

Biocompatibility studies, including Acute Systemic Toxicity, Vaginal Irritation Testing, Cytotoxicity and Sensitization testing were performed in accordance with the 2016 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" and ISO 10993- 1:2009 as follows:

- Cytotoxicity (ISO 10993-5:2009) ●
- Sensitization (ISO 10993-10:2010) ●
- Vaginal Irritation (ISO 10993-10:2010) ●
- Acute Systemic Toxicity (ISO 10993-11:2006) .

The results of this testing demonstrated that the AH!YES OB COCO personal lubricant is non-cytotoxic, non-irritating, non-sensitizing, and non-systemically toxic.

# Shelf-Life

The subject devices are non-sterile personal lubricants with a 24-month shelf-life in accordance with the results of accelerated aging studies per ASTM F1980-16 and real time storage. All device specifications listed in Table 1 were evaluated for shelf-life. The subject device met the device specifications at all time points.

# Condom Compatibility

The compatibility of the subject device with natural rubber latex, polyisoprene and polyurethane condoms was evaluated in accordance with ASTM D7661-10 Standard Test Method for Determining 

{6}------------------------------------------------

Compatibility of Personal Lubricants with Natural Rubber Latex Condoms. The results of this test demonstrate that the device is not compatible with condoms.

# 10. Conclusion

The results of the performance testing described above demonstrate that the AH!YES OB COCO Personal Lubricant is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda-staging.innolitics.com/device/K191411](https://fda-staging.innolitics.com/device/K191411)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).