K191385 · Coltene/Whaledent AG · EBF · Oct 29, 2019 · Dental
Device Facts
Record ID
K191385
Device Name
BRILLIANT COMPONEER
Applicant
Coltene/Whaledent AG
Product Code
EBF · Dental
Decision Date
Oct 29, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BRILLIANT COMPONEER is indicated for: Clinical indications - Restoration of caries defects - Reconstruction of lost tooth substance due to attrition, abrasion, erosion - Restoration of tooth fractures - Correction of anatomical malformation Cosmetic indications - Optimisation of old composite restorations - Extending incisors - Correction of malpositioned teeth - Cosmetic corrections for tooth discolouration or incorrect shading - Closure of diastema
Device Story
BRILLIANT COMPONEER is a prefabricated, polymerized, highly filled, submicron hybrid composite enamel shell used for direct veneering of front teeth and premolars. It functions as a ready-made enamel layer, eliminating the need for clinicians to model the basic tooth shape during restoration. The device is applied by a dentist in a clinical setting; the clinician selects the appropriate shade and shape shell, which is then bonded to the tooth surface. By providing a pre-formed anatomical structure, the device simplifies the restoration process, improves aesthetic outcomes, and reduces chair time. It is intended for single-patient, single-use applications.
Clinical Evidence
No clinical or animal data provided. Substantial equivalence is supported by bench testing and biocompatibility assessments. Bench testing included water sorption, solubility, flexural strength, compressive strength, elastic modulus, filler particle size distribution, surface hardness, and radiopacity, all conforming to ISO 4049. Biocompatibility testing included cytotoxicity, bacterial reverse mutation, systemic toxicity, irritation, and sensitization, conforming to ISO 10993-1 and ISO 7405.
Technological Characteristics
Prefabricated hybrid composite enamel shells composed of methacrylate resin matrix with barium glass and silica fillers. Dimensions vary by size (S, M, L, XL) and tooth type (maxillary/mandibular). Non-sterile, single-use. Performance conforms to ISO 4049. Biocompatibility conforms to ISO 10993-1. No electronic components or software.
Indications for Use
Indicated for patients requiring dental restoration of caries, tooth fractures, anatomical malformations, or reconstruction of tooth substance lost to attrition, abrasion, or erosion. Also indicated for cosmetic corrections including diastema closure, extending incisors, correcting malpositioned teeth, addressing discoloration/shading, and optimizing existing composite restorations.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 29, 2019
Coltene/Whaledent AG Wolfgang Dorner Regulatory Affairs Specialist Feldwiesenstrasse 20 Altstatten, 9450 SWITZERLAND
Re: K191385
Trade/Device Name: BRILLIANT COMPONEER Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: July 29, 2019 Received: July 31, 2019
Dear Wolfgang Dorner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K191385
Device Name BRILLIANT COMPONEER
Indications for Use (Describe) Indication
Clinical indications
- Restoration of caries defects
- Reconstruction of lost tooth substance due to attrition, abrasion, erosion
- Restoration of tooth fractures
- Correction of anatomical malformation
## Cosmetic indications
- Optimisation of old composite restorations
- Extending incisors
- Correction of malpositioned teeth
- Cosmetic corrections for tooth discolouration or incorrect shading
- Closure of diastema
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# % COLTENE
## K191385 Section 5 - 510(k) Summary
#### 5.1 Submitter:
Coltène/Whaledent AG Feldwiesenstrasse 20 9450 Altstätten Sankt Gallen Switzerland
#### Submitter Contact: 5.2
| Contact: | Wolfgang Dörner |
|----------|-------------------------------|
| Title: | Regulatory Affairs Specialist |
| Phone: | +41 71 757 5422 |
| Email: | wolfgang.doerner@coltene.com |
#### 5.3 Date Prepared
25th October 2019
#### 5.4 Device Identification
| Type of 510(k) Submission: | Traditional 510(k) |
|----------------------------|----------------------------|
| 510(k) Number: | K191385 |
| Trade Name: | BRILLIANT COMPONEER |
| Common Name: | Tooth shade resin material |
| Regulation Number: | 872.3690 |
| Product Code: | EBF |
| Class: | II |
| Classification Panel: | Dental |
#### 5.5 Primary Predicate Device
The primary predicate device is COMPONEER and is manufactured by Coltène/Whaledent AG in Switzerland. The primary predicate device was cleared in 2011 under the 510(k) number K112168.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for COLTENE. The logo consists of a blue graphic on the left and the word "COLTENE" in black, bold letters on the right. The graphic is made up of six blue squares arranged in two rows of three, with the squares in the bottom row being a lighter shade of blue than the squares in the top row.
#### 5.6 Indication for Use
BRILLIANT COMPONEER is indicated for: Clinical indications
- Restoration of caries defects -
- -Reconstruction of lost tooth substance due to attrition, abrasion, erosion
- -Restoration of tooth fractures
- -Correction of anatomical malformation
## Cosmetic indications
- Optimization of old composite restorations -
- Extending incisors -
- -Correction of malpositioned teeth
- Cosmetic corrections for tooth discoloration or incorrect shading -
- Closure of diastema -
#### 5.7 Device Overview
The subject device BRILLIANT COMPONEER is a polymerized, highly filled, submicron hybrid composite enamel shell(s) for direct veneering. It simplifies the direct restoration of front teeth and premolars without having to model the basic shape of the top enamel layer. The thin, anatomical composite shells are available in different shades and shapes allowing a natural, high aesthetic restoration (Figure 1).
Image /page/4/Picture/16 description: The image shows a dental veneer kit and some individual veneers. The kit includes a variety of tools and materials for applying the veneers, such as adhesives, applicators, and different sizes and shapes of veneers. The individual veneers are thin, tooth-colored shells that are designed to be bonded to the front surface of teeth to improve their appearance. The veneers are made of a ceramic material and have a natural-looking translucency.
Figure 1: BRILLIANT COMPONEER package overview and veneers (on the right).
#### 5.8 Substantial Equivalence
A comparison of the subject device BRILLIANT COMPONEER to the primary predicate device COMPONEER (K112168) is provided in Table 1.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "COLTENE" in a bold, sans-serif font. To the left of the word is a graphic consisting of nine squares arranged in a 3x3 grid. The squares are colored in shades of blue, with the top row being the darkest and the bottom row being the lightest.
| Attributes | Subject Device | Primary Predicate Device |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | BRILLIANT COMPONEER | COMPONEER |
| Device Name | BRILLIANT COMPONEER | COMPONEER |
| Manufacturer | Coltène/Whaledent AG,<br>Switzerland | Coltène/Whaledent AG,<br>Switzerland |
| 510(k) # | K191385 | K112168 |
| Product Code | EBF | EBF |
| Classification<br>Name: | Tooth shade resin material | Tooth shade resin material |
| Regulation | 21 CFR 872.3690 | 21 CFR 872.3690 |
| Class | II | II |
| Review Panel | Dental | Dental |
| Device Image | Image: BRILLIANT COMPONEER | Image: COMPONEER |
| Physical State: | Polymerized prefabricated<br>hybrid composite enamel<br>shells. | Polymerized prefabricated<br>hybrid composite enamel<br>shells. |
| Structure: | Polymer resin composite | Polymer resin composite |
| Resin Matrix<br>Monomers: | Methacrylates | Methacrylates |
| Inorganic Filler: | Barium glass and silica | Barium glass and silica |
| Sizes: | Maxillary veneers available in<br>size S, M, L and XL.<br><br>Mandibular veneers available<br>in in sizes S or M | Maxillary veneers available in<br>size S, M, L and XL.<br><br>Mandibular veneers available<br>in in sizes S or M |
| Indications for<br>Use | Clinical indications:<br>- Restoration of caries defects<br>- Reconstruction of lost tooth<br>substance due to attrition,<br>abrasion, erosion<br>- Restoration of tooth fractures<br>- Correction of anatomical<br>malformation<br><br>Cosmetic indications<br>- Optimisation of old composite | - Restoration therapy for caries<br>- Optimization of old<br>restorations<br>- Extending incisors<br>- Malpositioned teeth<br>- Tooth fractures<br>- Tooth discoloration, incorrect<br>shading<br>- Anatomical malformation<br>- Diastema<br>- Attrition, abrasion, erosion |
| Attributes | Subject Device | Primary Predicate Device |
| | BRILLIANT COMPONEER | COMPONEER |
| | restorations<br>- Extending incisors<br>- Correction of malpositioned<br>teeth<br>- Cosmetic corrections for tooth<br>discolouration or incorrect<br>shading<br>- Closure of diastema | - Cosmetic correction |
| Primary | Blisters with single veneers | Blisters with single veneers |
| Packaging | | |
| Usage | Single patient, single use | Single patient, single use |
| Sterility | Non-sterile | Non-sterile |
| Biocompatibility | Conforms with ISO 10993-1 | Conforms with ISO 10993-1 |
| Performance | Conforms with ISO 4049 | Conforms with ISO 4049 |
| Compressive | 285 MPa | 392 MPa |
| Strength: | | |
| Flexural | 125.4 MPa | 127 MPa |
| Strength: | | |
| Elastic | 9.8 GPa | 9.0 GPa |
| Modulus: | | |
| Radiopacity: | 3.24 mm of Al | 2 mm of Al |
| Water sorption: | 18.2 µg / mm³ | 16 µg / mm³ |
| Solubility: | 0.3 µg / mm³ | 0.9 µg / mm³ |
## Table 1: Comparison of BRILLIANT COMPONEER and COMPONEER
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "COLTENE" in a bold, sans-serif font. To the left of the word is a graphic consisting of six blue rectangles arranged in two rows of three. The rectangles in the top row are a darker shade of blue than the rectangles in the bottom row, creating a gradient effect. The word "COLTENE" is in black.
#### 5.9 Non-Clinical Performance Testing
As part of demonstrating substantial equivalence of the subject device BRILLIANT COMPONEER to the primary predicate device COMPONEER, Coltène/Whaledent AG performed extensive testing of the finished device in accordance with the applicable parts of the following standards, as well as according to the company's additional internal test protocols.
- 1. EN ISO 10993-1:2018 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
- 2. EN ISO 14971:2012 Medical devices Application of risk management to medical devices.
- 3. FDA General Guidance: 2016 - Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- 4. ISO 7405:2018 Dentistry Evaluation of biocompatibility of medical devices used in dentistry
- 5. ISO 4049:2009 Dentistry-Polymer-based filling, restorative and luting materials
The test results showed that BRILLIANT COMPONEER is biocompatible and performs as intended based on the biocompatibility and bench tests according to the applicable standards. In Table 2 a list of the conducted tests is provided.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "COLTENE" in bold, black letters. To the left of the word is a graphic of six blue squares arranged in two rows. The top row has three squares, and the bottom row has three squares. The squares are different shades of blue, with the top left square being the darkest and the bottom right square being the lightest.
## Table 2: List of tests completed for BRILLIANT COMPONEER
| Biocompatibility |
|------------------------------------------------------------|
| Chemical Characterization of Extractable Organic Compounds |
| In vitro Cytotoxicity Assay |
| Bacterial Reverse Mutation Assay |
| Acute Systemic Toxicity |
| Subchronic Systemic Toxicity |
| Biological Evaluation Report |
| Irritation Assessment |
| Acute Systemic Toxicity Assessment |
| Sensitization Assessment |
| Bench Testing |
| Water sorption |
| Solubility |
| Flexural strength |
| Compressive Strength |
| Elastic Modulus |
| Filler Particle Size Distribution |
| Surface Hardness / Vickers hardness |
| Radiopacity |
| Water sorption |
| Solubility |
Animal and clinical performance testing was not included.
#### 5.10 Statement of Substantial Equivalence
BRILLIANT COMPONEER has a similar intended use, indications for use, physical and chemical attributes, as the predicate device COMPONEER. Any minor differences in the materials used to make the subject device, when compared to the predicate device, have been successfully evaluated by Coltène/Whaledent AG through extensive performance and biocompatibility testing on the device. In conclusion, the subject device BRILLIANT COMPONEER has been determined to be substantially equivalent to the predicate device COMPONEER.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.