K191376 · Motion Concepts · IPL · Aug 13, 2019 · Physical Medicine
Device Facts
Record ID
K191376
Device Name
Modular Power Standing System
Applicant
Motion Concepts
Product Code
IPL · Physical Medicine
Decision Date
Aug 13, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3900
Device Class
Class 2
Indications for Use
The Modular Power Standing System is appropriate for use by individuals who drive a power wheelchair and cannot stand up on their own. The Modular Power Standing System allows such users to change position including from seating to standing, standing to seating, or any position in between. The device is appropriate for indoor and outdoor use. Motion Concepts makes no claims as to the therapeutic effectiveness of the products. Our only claims relate to the ability of the products to provide safe and reliable powered repositioning on the equipment onto which they are installed.
Device Story
Modular Power Standing System (MPSS) is a seating system for attachment to power wheelchair bases; provides power stand-up, tilt (0°-45°), recline (90°-178°), and seat elevation (6.5 inches). Operated via switches or powerbase joystick. Uses electro-mechanical linear actuators to drive linkages; includes knee and chest supports for standing stability. Used in private residences, chronic care facilities, indoors, and outdoors. Safety features include drive lock-out, reduced drive mode when positioned beyond limits, and front castor locks for standing stability. Electrical components are 24VDC with current limiting. Benefits include improved reach, accessibility, and repositioning for users unable to stand independently.
Clinical Evidence
No clinical data. Bench testing only. Includes biocompatibility (ISO 10993-5, ISO 10993-10), mechanical performance (ISO 7176 series), electrical safety, and EMC (ISO 7176-21).
Technological Characteristics
Materials: laser-cut steel, steel tube, machined aluminum, powder-coated. Actuation: 24VDC electro-mechanical linear actuators. Connectivity: interfaces with powerbase joystick or switches. Standards: ISO 7176 series (stability, brakes, energy, dimensions, speed, fatigue, climatic, obstacle, control systems, EMC). Biocompatibility: ISO 10993-5/10. Software: Moderate level of concern.
Indications for Use
Indicated for individuals who drive a power wheelchair and cannot stand up on their own, including those with spinal cord injury, spina bifida, cerebral palsy, multiple sclerosis, muscular dystrophy, polio, rheumatism, or other conditions requiring power wheelchair use and inability to shift weight.
Regulatory Classification
Identification
A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat.
Motion Concepts Modular Power Positioning System (K150574)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The FDA part is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out in a clear, sans-serif font.
August 13, 2019
Motion Concepts Dona Bhamra Senior Quality and Regulatory Affairs Manager 84 Citation Drive, Unit #1 Concord, L4K 3C1 Canada
Re: K191376
Trade/Device Name: Modular Power Standing System Regulation Number: 21 CFR 890.3900 Regulation Name: Standup Wheelchair Regulatory Class: Class II Product Code: IPL Dated: May 17, 2019 Received: May 23, 2019
Dear Dona Bhamra:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Vivek J. Pinto. PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K191376
### Device Name Modular Power Standing System (MPSS)
## Indications for Use (Describe)
The Modular Power Standing System is appropriate for use by individuals who drive a power wheelchair and cannot stand up on their own. The Modular Power Standing System allows such users to change position including from seating to standing, standing to seating, or any position in between. The device is appropriate for indoor and outdoor use. Motion Concepts makes no claims as to the therapeutic effectiveness of the products. Our only claims relate to the ability of the products to provide safe and reliable powered repositioning on the equipment onto which they are installed.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92
#### l. SUBMITTER
II.
III.
| FDA Establishment Registration Number: | 9615350 |
|----------------------------------------|---------------------------------------------------------------------------|
| Address: | Motion Concepts<br>84 Citation Drive, Unit 1<br>Concord, Ontario, L4K 3C1 |
| Phone: | 905-695-0134 |
| Fax: | 905-695-0138 |
| Contact person: | Dona Bhamra |
| Date Summary Prepared: | 1-Aug-2019 |
| <b>DEVICE</b> | |
|----------------------------|--------------------------------------|
| Device Proprietary Name: | Modular Power Standing System (MPSS) |
| Common Name: | Standup Wheelchair |
| Classification Regulation: | 21 CFR, 890.3900 |
| Product Code: | IPL |
| Classification of Device: | Class II |
| <b>PREDICATE DEVICE</b> | |
|-------------------------|--------------------------------------------------|
| Predicate Device: | Levo C3 Power Wheelchair |
| | K083017 |
| | Product Code IPL |
| | 21 CFR 890.3900 |
| Reference Device: | Motion Concepts Modular Power Positioning System |
| | K150574 |
| | Product Code ITI |
| | 21 CFR 890.3860 |
#### IV. DEVICE DESCRIPTION
The Modular Power Standing System (MPSS) is a seating system which is added to a wheelchair powerbase to provide four basic functions: power stand-up, power tilt, power recline including shear reduction, and power elevate. It can be operated in private residences, chronic care facilities, indoor and outdoors. Note the MPSS in itself does not include any wheelchair base components such as wheelchair frame, drive controls, wheels, brakes, batteries, suspension etc.
The Stand function is used to orient the user in an upright position. The mechanism consists of linkages driven by linear actuators. The system includes both a knee support to help stabilize the user in the standing position.
The Tilt function is used to change the seating inclination angle of user. The mechanisms consist of linkages driven by linear actuators. The range of tilt is 0° to 45°.
{4}------------------------------------------------
510(k) Summary Modular Power Standing System
The Recline function causes the position of the occupant's back to change by changing the position of the backrest with respect to the seat pan. The Shear Reduction works in conjunction with Recline to reduce the shear movement between the user and the backrest. The mechanisms consist of linkages driven by linear actuators. The range of Recline is 90° to 178°.
The Elevating seat module allows the user to elevate the entire seat. The mechanisms consist of linkages driven by linear actuators. The range of elevation is 6.5 inches.
The maximum occupant weight for the system is 250 lb. The Modular Power Standing System is assembled using primarily laser-cut steel parts, steel tube, machined aluminum, and mounting hardware.
The various power positioning modules may be activated via two options: using switches or through the powerbase manufacturer supplied joystick.
Safety features include a drive lock-out and reduced drive mode which are activated when any of the power functions are activated beyond pre-set limits. The system also includes front castor locks to provide additional front stability when in stand mode. Electrical components are maximum 24 volts and include current limiting within the seat control box. Stability of the Modular Power Standing System was tested on the powerbase selected for this application. These tests were conducted to ensure the safety of the power wheelchair was not compromised by the addition of the Modular Power Standing System.
#### V. INDICATIONS FOR USE
The Modular Power Standing System is appropriate for use by individuals who drive a power wheelchair and cannot stand up on their own. The Modular Power Standing System allows such users to change position including from seating to standing to seating, or any position in between. The device is appropriate for indoor and outdoor use. Motion Concepts makes no claims as to the therapeutic effectiveness of the products. Our only claims relate to the ability of the provide safe and reliable powered repositioning on the equipment onto which they are installed.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The new Modular Power Standing System was designed to add the standing function option to the existing Motion Concepts Modular Power Positioning System functions of tilt, recline, and elevate. This new standing feature functions similarly to the Levo C3 K083017, which will therefore serve as the Predicate device. In order to further demonstrate safety and effectiveness the Modular Power Position System K150574 will also be used as a reference device.
## Predicate device: Levo C3 (K083017)
Both the subject device Modular Power Standing System (MPSS) and the predicate device Levo C3 Power Wheelchair are systems which provide a stand-up feature, allowing users to reposition themselves from seating to standing and vice versa. Both systems include both a knee support to help stabilize the user in the standing position.
The two systems are equivalent with respect to the following:
- ー indications for use
- ー target population
{5}------------------------------------------------
510(k) Summary Modular Power Standing System
- safety features l
- general design
- use of materials l
- standards met
The seat size (width/depth/back height) differs in that the MPSS does not offer the smaller seat sizes offered on the predicate device. This difference does not affect the effectiveness of the MPSS over the range offered.
## Reference Device Motion Concepts Modular Power Position System (K150574)
Both the subject device Modular Power Standing System (MPSS) and reference device Modular Positioning System are seating systems that provide three basic functions: power recline including shear reduction and power elevate.
The two systems are equivalent with respect to the following:
- safety features l
- electrical safety
- general design
- use of materials
- | energy used
- motor type used
- standards met ।
The seat size (width/depth/back height) differs in that the MPSS does not offer the larger seat sizes offered on the referenced device. This difference does not affect the effectiveness of the MPSS over the range offered.
The elevate range on the referenced device is less than that offered on the MPSS, however the effectiveness is maintained as the standing function improves the reach and accessibility available to the user.
#### VII. PERFORMANCE DATA
।
The following performance data has been provided in support of the substantial equivalence determination.
## Biocompatibility Testing
- Bio-Compatibility testing requirements were evaluated for all potential body (skin) contacting O materials. Cytotoxicity testing was performed per 'ISO 10993 Part 5: Testing for in vitro cytotoxicity', and Dermal Sensitization testing and/or Primary Skin Irritation testing was performed per 'ISO 10993 Par10: Tests for irritation and skin sensitization'. Testing was conducted on all body contact materials on the MPSS, including: Meshtex fabric, Startex fabric, Spacetex 4000U™ fabric, O-Vinyl fabric, Style 6499 Polyester fabric, Self-Skinning Skin PU Foam (Bolasto F380 (RN7548)).
#### l Software Verification and Validation Testing
- o Software verification and validation testing have been conducted and documentation is provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device is considered to have a "Moderate level of Concern" because "prior to mitigation of hazards, a failure of the Software Device could result in Minor Injury, either to a patient or to a user of the device."
- -Electrical Safety and electromagnetic compatibility (EMC)
{6}------------------------------------------------
510(k) Summary
Modular Power Standing System
- o Electromagnetic Compatibility testing has been conducted on the Modular Power Standing System. The system complies with ISO 7176-21, Requirements and Test Methods for Electromagnetic Compatibility of Electrically Powered Wheelchairs and Scooters.
### -Mechanical and acoustic testing
- Mechanical testing of the Modular Power Standing System was carried out to cover functional o verification and device performance. Testing established correct functionality according to the relevant ISO 7176 standards. No acoustic testing was required to demonstrate device safety and effectiveness of the subject device.
। Animal Study
- o Animal performance testing was not required to demonstrate the safety and effectiveness of the subject device.
- -Clinical Studies
- Clinical testing was not required to demonstrate the safety and effectiveness of the subject device. o
The following table provides a comparison of technological characteristics with the predicate device (Levo X3) and the reference device (Modular Power Positioning System) to demonstrate substantial equivalence.
| | Subject Device | Predicate (A) | Reference device (B) | |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Motion Concepts Modular<br>Power Standing System | Levo C3 | Motion Concepts Modular<br>Power Positioning System | Comparison |
| 510k number | N/A | K083017 | K150574 | N/A |
| Product Code | IPL 890.3900 | IPL 890.3900 | ITI 890.3860 | The subject device<br>is identical to (A) |
| Indications for<br>Use | The Modular Power Standing<br>System is appropriate for use by<br>individuals who drive a power<br>wheelchair and cannot stand up<br>on their own. The Modular<br>Power Standing System allows<br>such users to change position<br>including from seating to<br>standing, standing to seating, or<br>any position in between. The<br>device is appropriate for indoor<br>and outdoor use. Motion<br>Concepts makes no claims as to<br>the therapeutic effectiveness of<br>the products. Our only claims<br>relate to the ability of the<br>products to provide safe and<br>reliable powered repositioning<br>on the equipment onto which<br>they are installed. | The LEVO C3 power<br>wheelchair with optional<br>seating and standing position<br>function may be of interest<br>for any individuals who<br>needs a power wheelchair<br>and cannot stand up on their<br>own such as people with<br>spinal cord injury, spina<br>bifida, cerebral palsy,<br>multiple sclerosis, muscular<br>dystrophy, polio,<br>rheumatism, etc | N/A | The subject device<br>is equivalent to (A). |
| | Subject Device<br>Motion Concepts Modular<br>Power Standing System | Predicate (A)<br>Levo C3 | Reference device (B)<br>Motion Concepts Modular<br>Power Positioning System | |
| Picture (in<br>standing mode) | | | N/A | The subject device<br>is equivalent to (A)<br>for standing<br>function. |
| Picture (in<br>tilt/recline/eleva<br>te mode) | | | | The subject device<br>is equivalent to (B)<br>for tilt, recline, and<br>elevate functions. |
| Target<br>Population | Paraplegics or quadriplegics<br>People with spinal cord<br>injury, spina bifida, cerebral<br>palsy, multiple sclerosis,<br>muscular dystrophy, polio,<br>rheumatism, and other<br>diseases and conditions<br>which cause the individual<br>to require a power<br>wheelchair and not be able<br>to shift his/her weight | The LEVO C3 power<br>wheelchair with optional<br>seating and standing position<br>function may be of interest<br>for any individuals who<br>needs a power wheelchair<br>and cannot stand up on their<br>own such as people with<br>spinal cord injury, spina<br>bifida, cerebral palsy,<br>multiple sclerosis, muscular<br>dystrophy, polio,<br>rheumatism, etc.. | Quadriplegics<br>persons with ALS, MS,<br>spinal muscular<br>atrophy, and any other<br>disability which causes<br>the individual to<br>require a power<br>wheelchair and not be<br>able shift his/her<br>weight | The subject device<br>is equivalent to (A). |
| Design | Link based mechanisms<br>powered by electro-mechanical<br>linear actuators | Link based mechanisms<br>powered by electro-<br>mechanical linear actuators | Link based mechanisms<br>powered by electro-<br>mechanical linear actuators | The subject device<br>is equivalent to (A)<br>and (B). |
| Materials | steel tube and plate, aluminum,<br>powder-coated | steel tube and plate,<br>aluminum, powder-coated | steel tube and plate,<br>aluminum, powder-coated | The subject device<br>is equivalent to (A)<br>and (B). |
| Weight Capacity | 250 lbs | 310 lbs | 300 lbs (250lbs with<br>elevate) | The subject device<br>is equivalent to (B)<br>with elevate. |
| Performance -<br>Tilt | 45° | 35° | 50° | The subject is<br>functionally<br>equivalent to (A)<br>and (B) |
| Performance -<br>Recline | 178° | Info Not available | 168° | The subject device<br>is functionally<br>equivalent to (A)<br>and (B). It is noted<br>that the slight |
| | Subject Device<br>Motion Concepts Modular<br>Power Standing System | Predicate (A)<br>Levo C3 | Reference device (B)<br>Motion Concepts Modular<br>Power Positioning System | Comparison |
| Performance –<br>Elevation | 6.5" | N/A | 12" | increase in<br>maximum amount<br>of recline compared<br>to (B) is required to<br>provide an effective<br>stand option<br>The subject device<br>has less elevation<br>than (B), This is<br>offset by the stand<br>feature which offers<br>alternatives for<br>comfort, positioning<br>and versatility. |
| Seat width range | 16 to 21 inches | 12.5 to 20.5 inches | 15 to 24 inches | The subject device<br>has smaller width<br>range than (A) or (B) |
| Seat depth range | 16 to 20 inches | 13.75 to 25 inches | 15 to 22 inches | The subject device<br>has smaller depth<br>range than (A) or (B) |
| Back height<br>range | 20 to 28 inches | 12 to 21.5" | 18 to 30 inches | The subject device<br>has smaller back<br>height range than<br>(B) |
| Mechanical<br>Safety | - wheelchair remains stable<br>when fully tilted, reclined<br>seat, elevated, standing<br>speed is reduced when<br>patient is positioned<br>beyond pre-set limits.<br>- drive lock-out prevents user<br>from driving power chair<br>while tilted beyond a pre-<br>set limits<br>- tilt limit is available<br>- Front caster lock to reduce<br>risk of tipping forward while<br>in stand-up position | - The speed is reduced to<br>half speed as soon as<br>the patient is not in the<br>sitting position | - wheelchair remains<br>stable with fully<br>elevated, tilted, and<br>reclined seat - center-<br>of-gravity shift further<br>enhances stability<br>- drive lock-out prevents<br>user from driving<br>power chair while tilted<br>beyond a pre-set limit<br>speed is reduced when<br>patient is positioned<br>beyond pre-set limits.<br>- tilt limit is available | The subject device<br>includes safety<br>features equivalent<br>to those listed for<br>both (A) and (B) |
| Where Used | - private residences<br>- chronic-care facilities<br>- indoors, outdoors | The product provides high<br>indoor and outdoor mobility<br>on surfaces like tar, grass<br>and gravel | - private residences<br>chronic-care facilities<br>indoors, outdoors<br>- | The subject device<br>is equivalent to (A)<br>and (B). |
| Electrical Safety | - electrical components are<br>24 volts maximum<br>current limit built into relay<br>box | Info Not available | - electrical components<br>are 24 volts maximum<br>current limit built into<br>relay box | Identical to (B) |
| Energy Used | 24 VDC Wheelchair Batteries | 24 VDC Wheelchair Batteries | 24 VDC Wheelchair<br>Batteries | Identical to (A) and<br>(B). |
| | Subject Device | Predicate (A) | Reference device (B) | |
| | Motion Concepts Modular | Levo C3 | Motion Concepts Modular | Comparison |
| | Power Standing System | | Power Positioning System | |
| Power stand-up<br>feature | Chest support and knee support<br>standard | Chest support and knee<br>support standard | Not Applicable | The subject device<br>is equivalent to (A) |
| Motor Type | Linear actuator,<br>Gear reduction screw type | Info Not available | Linear actuator,<br>Gear reduction screw type | Identical to (B) |
| Motor Voltage | 24VDC | Info Not available | 24VDC | Identical to (B) |
| Motor Maximum<br>Current | 4 Amps | Info Not available | 4 Amps | Identical to (B) |
| Power | 96W | Info Not available | 96W | Identical to (B) |
| Standards Met | ISO 7176 see details below | ISO 7176<br>ANSI/RESNA | RESNA WC-1:2009<br>RESNA WC-2:2009 | The subject device<br>is equivalent to (A)<br>and (B). |
{7}------------------------------------------------
Page 5 of 8
510(k) Summary
Modular Power Standing System Motion opts
{8}------------------------------------------------
Motion opts
# 510(k) Summary
Modular Power Standing System
{9}------------------------------------------------
Motion
# 510(k) Summary
Modular Power Standing System
## Standards Met
| Standard | Description | FDA Recognition<br>Number |
|------------------|------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| ISO 7176-1:2014 | Wheelchairs - Part 1: Determination of static stability | 16-195 |
| ISO 7176-2:2017 | Wheelchairs - Part 2: Determination of dynamic stability of electrically powered wheelchairs | 16-202 |
| ISO 7176-3:2012 | Wheelchairs - Part 3: Determination of effectiveness of brakes | 16-192 |
| ISO 7176-4:2008 | Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range | 16-162 |
| ISO 7176-5:2008 | Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space | 16-163 |
| ISO 7176-6:2018 | Wheelchairs - Part 6: Determination of maximum speed of electrically powered wheelchairs | 16-204 |
| ISO 7176-7:1998 | Wheelchairs - Part 7: Measurements of seating and wheel dimensions | 16-196 |
| ISO 7176-8:2014 | Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths | 16-197 |
| ISO 7176-9:2009 | Wheelchairs - Part 9: Climatic tests for electric wheelchairs | 16-167 |
| ISO 7176-10:2008 | Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs | 16-164 |
| ISO 7176-11:2012 | Wheelchairs - Part 11: Test dummies | 16-190 |
| ISO 7176-13:1989 | Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces | 16-25 |
| ISO 7176-14:2008 | Wheelchairs - Part 14: Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods | 16-165 |
{10}------------------------------------------------
Page 8 of 8
MotionConcepts
## 510(k) Summary Modular Power Standing System
| Standard | Description | FDA Recognition<br>Number |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| ISO 7176-15:1996 | Wheelchairs – Part 15: Requirements for information disclosure, documentation and<br>labelling | 16-27 |
| ISO 7176-16:2012 | Wheelchairs – Part 16: Resistance to ignition of postural support devices | 16-191 |
| ISO 7176-21:2009 | Wheelchairs – Part 21: Requirements and test methods for electromagnetic compatibility<br>of electrically powered wheelchairs and scooters, and battery chargers | 16-166 |
| ISO 7176-22:2014 | Wheelchairs – Part 22: Set-up procedures | 16-198 |
| ISO 7176-30:2018 | Wheelchairs – Part 30: Wheelchairs for changing occupant posture – Test methods and<br>requirements | N/A |
| ISO 10993-5:2009 | Part 5: Biological Evaluation of Medical Devices – Tests For In Vitro Cytotoxicity | 2-245 |
| ISO 10993-10:2010 | Part 10: Biological Evaluation of Medical Devices – Tests for irritation and skin sensitization | 2-174 |
| EN 1021-1/-2:2014 | Testing of Ignitability For Upholstered Furniture…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.