K191373 · ETHICON, Inc. · FTL · Oct 22, 2019 · General, Plastic Surgery
Device Facts
Record ID
K191373
Device Name
VICRYL Mesh
Applicant
ETHICON, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
Oct 22, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
VICRYL Mesh may be used wherever temporary wound or organ support is required (kidney, liver, spleen). VICRYL Mesh may be cut to the shape or size desired for each specific application.
Device Story
VICRYL Mesh is a synthetic absorbable sterile copolymer mesh composed of glycolide and L-lactide; identical in composition to VICRYL synthetic absorbable sutures. Device provides temporary support for wounds or organs (kidney, liver, spleen). Available in knitted (porous, stretchable) or woven configurations. Used by surgeons in clinical settings; mesh is cut to desired size/shape for specific application. Provides mechanical support during healing; material is absorbed by body over time. No changes to manufacturing, sterilization, or packaging compared to predicates.
Clinical Evidence
No clinical data; substantial equivalence is based on technological characteristics and established history of use of the predicate devices.
Technological Characteristics
Synthetic absorbable sterile copolymer (glycolide and L-lactide). Available in knitted or woven forms. Sterile. No changes to manufacturing, packaging, or sterilization processes compared to predicate devices.
Indications for Use
Indicated for patients requiring temporary wound or organ support, specifically for kidney, liver, or spleen applications.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 22, 2019
Ethicon, Inc. Melina Escobar Regulatory Affairs Specialist II Route 22 West, P.O. Box 151 Somerville, New Jersey 08876-0151
Re: K191373
Trade/Device Name: VICRYL Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTL Dated: May 22, 2019 Received: May 23, 2019
Dear Melina Escobar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K191373
Device Name VICRYL Mesh
Indications for Use (Describe)
VICRYL Mesh may be used wherever temporary wound or organ support is required (kidney, liver, spleen). VICRYL Mesh may be cut to the shape or size desired for each specific application.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" width="12" x="0" y="0"></rect> <path d="M2,2 L10,10 M2,10 L10,2" stroke="black"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" width="12" x="0" y="0"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Ethicon, a Johnson & Johnson company. The word "ETHICON" is written in large, red, sans-serif letters. Below the company name, the text "a Johnson & Johnson company" is written in a smaller, cursive font. The logo is simple and clean, with a focus on the company name.
# 510(k) Summary
Submission Date: 10/23/2019
K191373
# SUBMITTER INFORMATION:
| Company Name: | Ethicon, Inc. a Johnson & Johnson compan |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Company Address: | P.O. Box 151<br>Route 22 West<br>Somerville, NJ 08876-0151 |
| Contact Person: | Melina Escobar<br>Regulatory Affairs Specialist II<br>Phone: 908-218-2583<br>Fax: 908-218-2595<br>Email: mescob14@its.jnj.com |
| Device Trade Name: | VICRYL™ Mesh |
| Device Common Name: | VICRYL™ Mesh |
| Class: | Class II |
| Classification: | 21 CFR 878.3300 – Surgical Mesh |
| Product Code: | FTL |
## Predicate Devices:
| Device | Company | Product<br>Code | 510(k)<br>Number | Predicate for |
|---------------------------------------|------------------|-----------------|------------------|------------------------------------------------------------------------------------------------------------------------|
| VICRYLTM<br>(Polyglactin<br>910) Mesh | Ethicon,<br>Inc. | FTL | K810428 | Fundamental Scientific Technology,<br>Design, Intended Use, Materials,<br>Construction, Performance<br>Characteristics |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Ethicon, a Johnson & Johnson company. The word "ETHICON" is written in large, red, serif font. Below the company name, the words "a Johnson & Johnson company" are written in a smaller, cursive font.
| Device | Company | Product<br>Code | 510(k)<br>Number | Predicate for |
|----------------------|---------------|-----------------|------------------|------------------------------------------------------------------------------------------------------------------------|
| VICRYLTM<br>Mesh Bag | Ethicon, Inc. | FTL | K051701 | Fundamental Scientific Technology,<br>Design, Intended Use, Materials,<br>Construction, Performance<br>Characteristics |
## Device Description:
VICRYL™ (polyglactin 910) Mesh is a synthetic absorbable sterile copolymer made from glycolide and L-lactide. The copolymer is identical in composition to that used in VICRYL™ (polyglactin 910) synthetic absorbable suture.
VICRYL™ Knitted Mesh, which is more porous than VICRYL™ Woven Mesh, may be used in instances in which compliant and stretchable support material is desired.
### Indications for Use:
VICRYL™ Mesh may be used wherever temporary wound or organ support is required (kidney, liver, spleen). VICRYL™ Mesh may be cut to the shape or size desired for each specific application.
### Summary of Technological Characteristics:
VICRYL™ Mesh is substantially equivalent to the VICRYL™ Mesh (K810428) and VICRYL™ Mesh Bag (K051701) predicate devices with respect to technological characteristics. Both the subject and predicate devices are synthetic absorbable sterile copolymer. The devices function in the same manner and are designed to be used wherever temporary wound or organ support is required (kidney, liver, spleen). The subject mesh is manufactured within the existing manufacturing processes for the predicate device. There are no changes to the manufacturing, packaging, sterilization processes, or shelf life of the currently marketed device.
The subject VICRYL™ Mesh, for which this 510(k) Premarket Notification is being submitted, differs from the currently marketed device, K810428, in the labeling (Instructions for Use). The Instructions for Use has been revised to clarify the indications. Additionally, several other sections of the Instructions for Use of the subject devices have been reworded/ reformatted for clarity to address and align with evolving regulatory expectations.
### Substantial Equivalence:
{5}------------------------------------------------
VICRYL™ Mesh is substantially equivalent to the VICRYL™ Mesh (K810428) and VICRYL™ Mesh Bag (K051701) predicate devices with respect to technological characteristics. Both the subject and predicate device are synthetic absorbable sterile copolymer. The devices function in the same manner and are designed to be used wherever temporary wound or organ support is required (kidney, liver, spleen). The subject mesh is manufactured within the existing manufacturing processes for the predicate device. There are no changes to the manufacturing, packaging, sterilization processes, or shelf life of the currently marketed device.
The subject VICRYL™ Mesh, for which this 510(k) Premarket Notification is being submitted, differs from the currently marketed device. K810428, in the labeling (Instructions for Use). The Instructions for Use has been revised to clarify the indications. Additionally, several other sections of the Instructions for Use of the subject devices have been reworded/ reformatted for clarity to address evolving regulatory expectations.
## Conclusion:
VICRYL™ Mesh is substantially equivalent to the predicate devices, K810428 and K051701 in indications for use, fundamental scientific technology, and technological characteristics. It is as safe and effective as the predicates. Ethicon considers VICRYL™ Mesh to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.