← Product Code [EWO](/productcode/EWO) · K191372

# Lyra (K191372)

_Interacoustics A/S · EWO · Jul 15, 2019 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K191372

## Device Facts

- **Applicant:** Interacoustics A/S
- **Product Code:** [EWO](/productcode/EWO.md)
- **Decision Date:** Jul 15, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.1050
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat
- **Attributes:** Pediatric

## Indications for Use

The Lyra with DPOAE is intended for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions. The target population for Lyra with DPOAE includes all ages. The Lyra with TEOAE is intended for use in the audiologic evaluation of ear disorders using Transient Evoked Otoacoustic Emissions. The target population for Lyra with TEOAE includes all ages. The Lyra System is to be used by trained personnel only, such as audiologists, ENT surgeons, doctors, hearing healthcare professionals or personnel with a similar level of education. The device should not be necessary knowledge and training to understand its use and how results should be interpreted.

## Device Story

Lyra is an audiometric device for screening/diagnosing inner ear abnormalities; utilizes DPOAE (pure tone pairs) or TEOAE (short duration stimuli) to assess cochlear function. Hardware unit connects to PC running IA OAE suite software; powered via USB. Sensitive microphone in OAE probe captures ear canal responses to stimuli. Used by audiologists, ENT surgeons, or trained technicians in clinical settings. Software processes signals to provide PASS-REFER results or frequency band assessments. Healthcare providers use output to document ear disorders and guide clinical decision-making regarding hearing loss. Benefits include objective, non-invasive screening for patients of all ages.

## Clinical Evidence

No clinical data. Substantial equivalence established via bench testing, safety, EMC, and software validation.

## Technological Characteristics

Hardware unit with OAE probe; USB-powered via PC. Software: IA OAE suite. Standards: IEC 60601-1 (safety), IEC 60645-6 (OAE performance). Stimuli: DPOAE (pure tones, 500Hz-10kHz) or TEOAE (short duration, 500Hz-5500Hz). A/D resolution: 24-bit. Artifact rejection: 0 to +60 dB SPL.

## Regulatory Identification

An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.

## Predicate Devices

- Titan (TEOAE) ([K130795](/device/K130795.md))
- Titan (DPOAE) ([K103760](/device/K103760.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol on the left is a stylized representation of a human figure. The text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The words "U.S. FOOD & DRUG" are on the top line, and the word "ADMINISTRATION" is on the bottom line.

July 15, 2019

Interacoustics A/S Erik Nielsen Director, Regulatory & Compliance Audiometer Alle 1 Middelfart, DK-5500 Dk

Re: K191372

Trade/Device Name: Lyra Regulation Number: 21 CFR 874.1050 Regulation Name: Audiometer Regulatory Class: Class II Product Code: EWO Dated: May 13, 2019 Received: May 22, 2019

Dear Erik Nielsen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Srinivas Nandkumar, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K191372

Device Name Lyra

#### Indications for Use (Describe)

The Lyra with DPOAE is intended for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions. The target population for Lyra with DPOAE includes all ages.

The Lyra with TEOAE is intended for use in the audiologic evaluation of ear disorders using Transient Evoked Otoacoustic Emissions. The target population for Lyra with TEOAE includes all ages.

The Lyra System is to be used by trained personnel only, such as audiologists, ENT surgeons, doctors, hearing healthcare professionals or personnel with a similar level of education. The device should not be necessary knowledge and training to understand its use and how results should be interpreted.

| Type of Use (Select one or both, as applicable)                               |                                                                              |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span>    Prescription Use (Part 21 CFR 801 Subpart D)   </div> | <div> <span> </span>    Over-The-Counter Use (21 CFR 801 Subpart C)   </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Interacoustics. The logo consists of a stylized globe-like symbol above the word "Interacoustics" in a bold, sans-serif font. The globe symbol is made up of curved lines that suggest the shape of a globe. The logo is in black and white.

FDA 510(k) K191372 Lyra Summary as required by 21 CFR 807.92.

#### Administrative Information

Submitter: Interacoustics A/S Audiometer Allé 1

5500 Middelfart Denmark Tel: +45 6371 3555 Fax: +45 6371 3522

Contact Person: Erik Nielsen Director of Regulatory Affairs erni@demant.com

Date Summary Prepared: May 13, 2019

Device Identification

Trade Name: Lyra™ Common Name: audiometry, otoacoustic emission device Device Classification Name: Audiometer Device classification: Class II Panel: Ear Nose & Throat Classification Regulation: 874.1050 Product Code: EWO

Primary Predicate Device: Titan™(TEOAE), cleared on 06/20/2013 via K130795

Secondary Predicate Device: Titan™(DPOAE), cleared on 05/05/2011 via K103760

#### Device Description

The device is audiometric equipment used for assisting of inner ear abnormalities. Lyra features a hardware unit connecting to a PC installed with IA OAE suite software designated for use with Lyra. The PC software provides a user interface designed to integrate in the standard Microsoft Windows environment. Lyra can be purchased with various licenses allowing you to perform different hearing screening tests.

Distortion product otoacoustic emissions (DPOAE) technology uses pairs of pure tones presented in sequence to screen patients for cochlear hearing loss. Responses to the stimulus are predictable and therefore can be measured via a sensitive microphone placed in the patient's ear canal.

Transient otoacoustic emissions (TEOAE) technology uses a short duration stimulus to screen patients for cochlear hearing loss. Responses to the stimulus are predictable and therefore can be measured via a sensitive microphone placed in the patient's ear canal. The response can be divided into frequency bands for assessment.

Device Intended Use / Device indications for use

The Lyra with DPOAE is intended for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions. The target population for Lyra with DPOAE

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Interacoustics. The logo consists of the word "Interacoustics" in a bold, sans-serif font. Above the word is a symbol that resembles a globe with intersecting lines. The logo is black and white.

includes all ages.

The Lyra with TEOAE is intended for use in the audiologic evaluation and documentation of ear disorders using Transient Evoked Otoacoustic Emissions. The target population for Lyra with TEOAE includes all ages.

The Lyra System is to be used by trained personnel only, such as audiologists, ENT surgeons, doctors, hearing healthcare professionals or personnel with a similar level of education. The device should not be used without the necessary knowledge and training to understand its use and how results should be interpreted.

#### Technological Characteristics

Lyra features a hardware unit connecting to a PC installed with IA OAE suite software designated for use with Lyra. Power to the Lyra is provided from the USB connection to the PC.

A comparison between the new and predicate devices shows that the technological characteristics and indications for use are equivalent. The device employs similar technology to accomplish the same tasks as the predicates. A detailed table is provided below.

| Description                 | Titan with TEOAE440 (k130795)                                                                                                                                                                         | Lyra                                                                                                                           |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Type                        | Audiometer – Audiometric equipment                                                                                                                                                                    | Same                                                                                                                           |
| Regulation Number           | 21 CFR 874.1050<br>(otoacoustic emission device)                                                                                                                                                      | Same                                                                                                                           |
| Classification Product Code | EWO                                                                                                                                                                                                   | Same                                                                                                                           |
| Indications for Use         | The Titan with TEOAE440 is intended<br>for use in the audiologic evaluation<br>and documentation of ear disorders<br>using Transient Evoked Otoacoustic<br>Emissions.                                 | Same                                                                                                                           |
| Target Population           | The devices are suitable for all<br>populations including new-born<br>infants                                                                                                                         | Same                                                                                                                           |
| Intended User               | The Titan System is to be used by<br>trained personnel only such as<br>audiologists, ENT surgeons, doctors,<br>hearing healthcare professionals or<br>personnel with a similar level of<br>education. | Same (or used by a trained<br>technician under the<br>supervision of a professional)                                           |
| Anatomical Sites            | Examination of Ear                                                                                                                                                                                    | Same                                                                                                                           |
| Safety Standards            | IEC 60601-1                                                                                                                                                                                           | Same                                                                                                                           |
| Performance Standards       | IEC 60645-6                                                                                                                                                                                           | Same                                                                                                                           |
| Device Type                 | Screening and diagnostic                                                                                                                                                                              | Same                                                                                                                           |
| System Configuration        | Dedicated hardware unit with no<br>display or controls. OAE probe<br>permanently connected to hardware<br>unit. Hardware unit operated through<br>a connected PC.                                     | Same for clinical use.<br>(Titan also has possibility for<br>handheld use as it also has<br>display and controls on<br>device) |
| TEOAE Stimulus              |                                                                                                                                                                                                       |                                                                                                                                |
| Frequency Range             | 500 to 5500Hz                                                                                                                                                                                         | Same                                                                                                                           |

#### Equivalence Predicate Chart 1 (primary):

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Interacoustics. The logo consists of a stylized globe-like symbol above the word "Interacoustics" in a bold, sans-serif font. A horizontal line underlines the text, emphasizing the company name. The logo is simple, clean, and professional.

| Stimuli Type                                 | Non-Linear and Linear<br>Short duration signal<br>According to IEC 60645-3 | same |
|----------------------------------------------|----------------------------------------------------------------------------|------|
| Level                                        | 30 to 90 dB peSPL                                                          | same |
| Level Step                                   | 1 dB SPL                                                                   | same |
| Transducer                                   | Dedicated OAE Probe                                                        | same |
| Probe Detection                              | Auto detection                                                             | same |
| Recording                                    |                                                                            |      |
| A/D Resolution                               | 24 bit                                                                     | same |
| Artifact Reject System                       | 0 -> +60 dB SPL or off                                                     | same |
| Automatic test with display<br>of PASS-REFER | Yes                                                                        | same |

# Equivalence Predicate Chart 2 (secondary):

| Description                                  | Titan With DPOAE440<br>(k103760)                                                                                                                                                                   | Lyra                                                                                                                        |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Type                                         | Audiometer – Audiometric equipment                                                                                                                                                                 | Same                                                                                                                        |
| Regulation Number                            | 21 CFR 874.1050<br>(otoacoustic emission device)                                                                                                                                                   | Same                                                                                                                        |
| Classification<br>Product Code               | EWO                                                                                                                                                                                                | Same                                                                                                                        |
| Indications for Use                          | The Titan with DPOAE440 is intended for use<br>in the audiologic evaluation and<br>documentation of ear disorders using<br>Distortion Product Otoacoustic Emissions.                               | Same                                                                                                                        |
| Target Population                            | The devices are suitable for all populations<br>including new-born infants                                                                                                                         | Same                                                                                                                        |
| Intended User                                | The Titan System is to be used by trained<br>personnel only such as audiologists, ENT<br>surgeons, doctors, hearing healthcare<br>professionals or personnel with a similar level<br>of education. | Same (or used by a trained<br>technician under the supervision<br>of a professional)                                        |
| Anatomical Sites                             | Examination of Ear                                                                                                                                                                                 | Same                                                                                                                        |
| Safety Standards                             | IEC 60601-1                                                                                                                                                                                        | Same                                                                                                                        |
| Performance<br>Standards                     | IEC 60645-6                                                                                                                                                                                        | Same                                                                                                                        |
| Device Type                                  | Screening and diagnostic                                                                                                                                                                           | Same                                                                                                                        |
| System<br>Configuration                      | Dedicated hardware unit with no display or<br>controls. OAE probe permanently connected to<br>hardware unit. Hardware unit operated<br>through a connected PC.                                     | Same for clinical use.<br>(Titan also has possibility for<br>handheld use as it also has<br>display and controls on device) |
| DPOAE Stimulus                               |                                                                                                                                                                                                    |                                                                                                                             |
| Frequency range<br>(f2)                      | 500Hz — 10kHz                                                                                                                                                                                      | Same                                                                                                                        |
| Stimuli Type                                 | 2 pure tones                                                                                                                                                                                       | same                                                                                                                        |
| Level                                        | 30 dB SPL to 80 dB SPL                                                                                                                                                                             | Same                                                                                                                        |
| Level Step                                   | 1 dB SPL                                                                                                                                                                                           | same                                                                                                                        |
| Transducer                                   | Dedicated OAE Probe                                                                                                                                                                                | same                                                                                                                        |
| Probe Detection                              | Auto detection                                                                                                                                                                                     | same                                                                                                                        |
| Recording                                    |                                                                                                                                                                                                    |                                                                                                                             |
| A/D Resolution                               | 24 bit                                                                                                                                                                                             | same                                                                                                                        |
| Artifact Reject System                       | 0 -> +60 dB SPL or off                                                                                                                                                                             | same                                                                                                                        |
| Automatic test with display of<br>PASS-REFER | Yes                                                                                                                                                                                                | same                                                                                                                        |

{6}------------------------------------------------

## Summary of Non-Clinical Testing

Design verification and validation were performed according to current standards for OAE to assure the device meets its performance specifications. EMC and Safety was performed in compliance with recognized standards IEC 60601-1 series, Medical Electrical Equipment - General requirements for basic safety and essential performance. The product meets the requirements from the international standard for OAE measurements IEC 60645 series. Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in medical Devices." The software for this device was considered as a "minor" level of concern since a malfunction of, or a latent design flaw in, the Software Device could not lead to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury. Internal validation and comparison tests were performed and demonstrate that Lyra fulfil the requirements and is valid for its intended medical purpose.

## Summary of Clinical Testing

Not applicable. Not required to establish substantial equivalence.

#### Conclusion

We have compared the intended use and performance characteristics with the predicate device. The Lyra was tested according to current standards and there were found no significant differences between the devices.

The Lyra conforms to the current standards. After analyzing bench testing, safety, EMC, and software validation (with risk analysis) testing we conclude that the Lyra is found to be substantially equivalent to the predicate devices in technological characteristics and indications for use.

---

**Source:** [https://fda-staging.innolitics.com/device/K191372](https://fda-staging.innolitics.com/device/K191372)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).