The Lyra with DPOAE is intended for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions. The target population for Lyra with DPOAE includes all ages. The Lyra with TEOAE is intended for use in the audiologic evaluation of ear disorders using Transient Evoked Otoacoustic Emissions. The target population for Lyra with TEOAE includes all ages. The Lyra System is to be used by trained personnel only, such as audiologists, ENT surgeons, doctors, hearing healthcare professionals or personnel with a similar level of education. The device should not be necessary knowledge and training to understand its use and how results should be interpreted.
Device Story
Lyra is an audiometric device for screening/diagnosing inner ear abnormalities; utilizes DPOAE (pure tone pairs) or TEOAE (short duration stimuli) to assess cochlear function. Hardware unit connects to PC running IA OAE suite software; powered via USB. Sensitive microphone in OAE probe captures ear canal responses to stimuli. Used by audiologists, ENT surgeons, or trained technicians in clinical settings. Software processes signals to provide PASS-REFER results or frequency band assessments. Healthcare providers use output to document ear disorders and guide clinical decision-making regarding hearing loss. Benefits include objective, non-invasive screening for patients of all ages.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, safety, EMC, and software validation.
Technological Characteristics
Hardware unit with OAE probe; USB-powered via PC. Software: IA OAE suite. Standards: IEC 60601-1 (safety), IEC 60645-6 (OAE performance). Stimuli: DPOAE (pure tones, 500Hz-10kHz) or TEOAE (short duration, 500Hz-5500Hz). A/D resolution: 24-bit. Artifact rejection: 0 to +60 dB SPL.
Indications for Use
Indicated for audiologic evaluation and documentation of ear disorders in all ages using Distortion Product Otoacoustic Emissions (DPOAE) or Transient Evoked Otoacoustic Emissions (TEOAE).
Regulatory Classification
Identification
An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol on the left is a stylized representation of a human figure. The text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The words "U.S. FOOD & DRUG" are on the top line, and the word "ADMINISTRATION" is on the bottom line.
July 15, 2019
Interacoustics A/S Erik Nielsen Director, Regulatory & Compliance Audiometer Alle 1 Middelfart, DK-5500 Dk
Re: K191372
Trade/Device Name: Lyra Regulation Number: 21 CFR 874.1050 Regulation Name: Audiometer Regulatory Class: Class II Product Code: EWO Dated: May 13, 2019 Received: May 22, 2019
Dear Erik Nielsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191372
Device Name Lyra
#### Indications for Use (Describe)
The Lyra with DPOAE is intended for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions. The target population for Lyra with DPOAE includes all ages.
The Lyra with TEOAE is intended for use in the audiologic evaluation of ear disorders using Transient Evoked Otoacoustic Emissions. The target population for Lyra with TEOAE includes all ages.
The Lyra System is to be used by trained personnel only, such as audiologists, ENT surgeons, doctors, hearing healthcare professionals or personnel with a similar level of education. The device should not be necessary knowledge and training to understand its use and how results should be interpreted.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Interacoustics. The logo consists of a stylized globe-like symbol above the word "Interacoustics" in a bold, sans-serif font. The globe symbol is made up of curved lines that suggest the shape of a globe. The logo is in black and white.
FDA 510(k) K191372 Lyra Summary as required by 21 CFR 807.92.
#### Administrative Information
Submitter: Interacoustics A/S Audiometer Allé 1
5500 Middelfart Denmark Tel: +45 6371 3555 Fax: +45 6371 3522
Contact Person: Erik Nielsen Director of Regulatory Affairs erni@demant.com
Date Summary Prepared: May 13, 2019
Device Identification
Trade Name: Lyra™ Common Name: audiometry, otoacoustic emission device Device Classification Name: Audiometer Device classification: Class II Panel: Ear Nose & Throat Classification Regulation: 874.1050 Product Code: EWO
Primary Predicate Device: Titan™(TEOAE), cleared on 06/20/2013 via K130795
Secondary Predicate Device: Titan™(DPOAE), cleared on 05/05/2011 via K103760
#### Device Description
The device is audiometric equipment used for assisting of inner ear abnormalities. Lyra features a hardware unit connecting to a PC installed with IA OAE suite software designated for use with Lyra. The PC software provides a user interface designed to integrate in the standard Microsoft Windows environment. Lyra can be purchased with various licenses allowing you to perform different hearing screening tests.
Distortion product otoacoustic emissions (DPOAE) technology uses pairs of pure tones presented in sequence to screen patients for cochlear hearing loss. Responses to the stimulus are predictable and therefore can be measured via a sensitive microphone placed in the patient's ear canal.
Transient otoacoustic emissions (TEOAE) technology uses a short duration stimulus to screen patients for cochlear hearing loss. Responses to the stimulus are predictable and therefore can be measured via a sensitive microphone placed in the patient's ear canal. The response can be divided into frequency bands for assessment.
Device Intended Use / Device indications for use
The Lyra with DPOAE is intended for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions. The target population for Lyra with DPOAE
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Interacoustics. The logo consists of the word "Interacoustics" in a bold, sans-serif font. Above the word is a symbol that resembles a globe with intersecting lines. The logo is black and white.
includes all ages.
The Lyra with TEOAE is intended for use in the audiologic evaluation and documentation of ear disorders using Transient Evoked Otoacoustic Emissions. The target population for Lyra with TEOAE includes all ages.
The Lyra System is to be used by trained personnel only, such as audiologists, ENT surgeons, doctors, hearing healthcare professionals or personnel with a similar level of education. The device should not be used without the necessary knowledge and training to understand its use and how results should be interpreted.
#### Technological Characteristics
Lyra features a hardware unit connecting to a PC installed with IA OAE suite software designated for use with Lyra. Power to the Lyra is provided from the USB connection to the PC.
A comparison between the new and predicate devices shows that the technological characteristics and indications for use are equivalent. The device employs similar technology to accomplish the same tasks as the predicates. A detailed table is provided below.
| Description | Titan with TEOAE440 (k130795) | Lyra |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Type | Audiometer – Audiometric equipment | Same |
| Regulation Number | 21 CFR 874.1050<br>(otoacoustic emission device) | Same |
| Classification Product Code | EWO | Same |
| Indications for Use | The Titan with TEOAE440 is intended<br>for use in the audiologic evaluation<br>and documentation of ear disorders<br>using Transient Evoked Otoacoustic<br>Emissions. | Same |
| Target Population | The devices are suitable for all<br>populations including new-born<br>infants | Same |
| Intended User | The Titan System is to be used by<br>trained personnel only such as<br>audiologists, ENT surgeons, doctors,<br>hearing healthcare professionals or<br>personnel with a similar level of<br>education. | Same (or used by a trained<br>technician under the<br>supervision of a professional) |
| Anatomical Sites | Examination of Ear | Same |
| Safety Standards | IEC 60601-1 | Same |
| Performance Standards | IEC 60645-6 | Same |
| Device Type | Screening and diagnostic | Same |
| System Configuration | Dedicated hardware unit with no<br>display or controls. OAE probe<br>permanently connected to hardware<br>unit. Hardware unit operated through<br>a connected PC. | Same for clinical use.<br>(Titan also has possibility for<br>handheld use as it also has<br>display and controls on<br>device) |
| TEOAE Stimulus | | |
| Frequency Range | 500 to 5500Hz | Same |
#### Equivalence Predicate Chart 1 (primary):
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Interacoustics. The logo consists of a stylized globe-like symbol above the word "Interacoustics" in a bold, sans-serif font. A horizontal line underlines the text, emphasizing the company name. The logo is simple, clean, and professional.
| Stimuli Type | Non-Linear and Linear<br>Short duration signal<br>According to IEC 60645-3 | same |
|----------------------------------------------|----------------------------------------------------------------------------|------|
| Level | 30 to 90 dB peSPL | same |
| Level Step | 1 dB SPL | same |
| Transducer | Dedicated OAE Probe | same |
| Probe Detection | Auto detection | same |
| Recording | | |
| A/D Resolution | 24 bit | same |
| Artifact Reject System | 0 -> +60 dB SPL or off | same |
| Automatic test with display<br>of PASS-REFER | Yes | same |
# Equivalence Predicate Chart 2 (secondary):
| Description | Titan With DPOAE440<br>(k103760) | Lyra |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Type | Audiometer – Audiometric equipment | Same |
| Regulation Number | 21 CFR 874.1050<br>(otoacoustic emission device) | Same |
| Classification<br>Product Code | EWO | Same |
| Indications for Use | The Titan with DPOAE440 is intended for use<br>in the audiologic evaluation and<br>documentation of ear disorders using<br>Distortion Product Otoacoustic Emissions. | Same |
| Target Population | The devices are suitable for all populations<br>including new-born infants | Same |
| Intended User | The Titan System is to be used by trained<br>personnel only such as audiologists, ENT<br>surgeons, doctors, hearing healthcare<br>professionals or personnel with a similar level<br>of education. | Same (or used by a trained<br>technician under the supervision<br>of a professional) |
| Anatomical Sites | Examination of Ear | Same |
| Safety Standards | IEC 60601-1 | Same |
| Performance<br>Standards | IEC 60645-6 | Same |
| Device Type | Screening and diagnostic | Same |
| System<br>Configuration | Dedicated hardware unit with no display or<br>controls. OAE probe permanently connected to<br>hardware unit. Hardware unit operated<br>through a connected PC. | Same for clinical use.<br>(Titan also has possibility for<br>handheld use as it also has<br>display and controls on device) |
| DPOAE Stimulus | | |
| Frequency range<br>(f2) | 500Hz — 10kHz | Same |
| Stimuli Type | 2 pure tones | same |
| Level | 30 dB SPL to 80 dB SPL | Same |
| Level Step | 1 dB SPL | same |
| Transducer | Dedicated OAE Probe | same |
| Probe Detection | Auto detection | same |
| Recording | | |
| A/D Resolution | 24 bit | same |
| Artifact Reject System | 0 -> +60 dB SPL or off | same |
| Automatic test with display of<br>PASS-REFER | Yes | same |
{6}------------------------------------------------
## Summary of Non-Clinical Testing
Design verification and validation were performed according to current standards for OAE to assure the device meets its performance specifications. EMC and Safety was performed in compliance with recognized standards IEC 60601-1 series, Medical Electrical Equipment - General requirements for basic safety and essential performance. The product meets the requirements from the international standard for OAE measurements IEC 60645 series. Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in medical Devices." The software for this device was considered as a "minor" level of concern since a malfunction of, or a latent design flaw in, the Software Device could not lead to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury. Internal validation and comparison tests were performed and demonstrate that Lyra fulfil the requirements and is valid for its intended medical purpose.
## Summary of Clinical Testing
Not applicable. Not required to establish substantial equivalence.
#### Conclusion
We have compared the intended use and performance characteristics with the predicate device. The Lyra was tested according to current standards and there were found no significant differences between the devices.
The Lyra conforms to the current standards. After analyzing bench testing, safety, EMC, and software validation (with risk analysis) testing we conclude that the Lyra is found to be substantially equivalent to the predicate devices in technological characteristics and indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.