T-TAS 01 Total Thrombus-formation Analysis System Instrument, PL Chip for T-TAS 01, PL Chip Reservoir Set for T-TAS 01, BAPA tube for T-TAS 01
Applicant
Fujimori Kogyo, Co., Ltd.
Product Code
JOZ · Hematology
Decision Date
Feb 14, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5700
Device Class
Class 2
Indications for Use
The T-TAS 01 Instrument is intended for use with T-TAS reagent chips in the clinical laboratory. The T-TAS 01 PL chip is intended for use in the clinical laboratory for the analysis of the platelet thrombus formation process (primary hemostatic function) in patients age 21 and older with a history of conditions associated with impaired primary hemostatic function or use of antiplatelet therapy. The test uses BAPA-anticoagulated whole blood specimens to measure platelet adhesion to a thrombogenic collagen-coated surface and aggregation, which causes an increase in flow pressure inside the PL chip. The test measures primary hemostatic function as the area under the pressure-time curve (AUC), with AUC < 260 suggesting abnormal primary hemostatic function. Additional testing may be necessary to identify the cause(s) of abnormal primary hemostatic function. The test has been evaluated in patients taking antiplatelet therapy, in patients with von Willebrand disease, and in patients with Glanzmann’s thrombasthenia. Other primary hemostasis disorders have not been evaluated. The BAPA tube for T-TAS 01 is intended to be used for the collection, transport, and storage of blood specimens for use with the T-TAS 01 system.
Device Story
T-TAS 01 is an automated tabletop platelet function system; uses BAPA-anticoagulated whole blood samples. Blood is injected into a disposable PL chip containing 26 collagen-coated microcapillary channels at 1500 s⁻¹ shear stress. Platelets adhere to collagen and aggregate, causing channel occlusion and increased flow pressure. A pressure sensor monitors the flow path; a dedicated PC calculates the area under the pressure-time curve (AUC) over 10 minutes. Output includes pressure waveform graphs and AUC values. Used in clinical laboratories by trained personnel. AUC < 260 indicates abnormal primary hemostasis; results aid clinicians in assessing platelet function in patients with bleeding disorders or those on antiplatelet therapy. System includes automated self-checks for pressure leaks and pump function.
Clinical Evidence
Clinical study of 274 subjects (142 healthy, 132 with platelet dysfunction). Primary endpoint: sensitivity and negative agreement using AUC < 260 cutoff. Negative agreement in healthy donors: 95.8% (91.1-98.0%). Sensitivity: 68.4% (ASA), 100% (clopidogrel+ASA), 100% (prasugrel+ASA), 100% (ticagrelor+ASA), 72.0% (vWD), 100% (GT). Bench testing confirmed precision (CV ≤ 15%) and interference profiles.
Technological Characteristics
Tabletop instrument with external PC control. Disposable PL chip with 26 collagen-coated microcapillary channels. Sensing: pressure sensor. Energy: electrical (instrument). Connectivity: PC-based. Sterilization: N/A (disposable chips). Software: proprietary control/analysis software.
Indications for Use
Indicated for patients age 21 and older with a history of conditions associated with impaired primary hemostatic function or use of antiplatelet therapy. Evaluated in patients on antiplatelet therapy, von Willebrand disease, and Glanzmann’s thrombasthenia.
Regulatory Classification
Identification
An automated platelet aggregation system is a device used to determine changes in platelet shape and platelet aggregation following the addition of an aggregating reagent to a platelet-rich plasma.
Predicate Devices
Dade PFA-100 Platelet Function Analyzer and Dade PFA-100 Reagents (K060489)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K191364
B Applicant
Fujimori Kogyo Co., Ltd.
C Proprietary and Established Names
Total Thrombus-formation Analysis (T-TAS) 01 System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JOZ | Class II | 21 CFR 864.5700 - Automated Platelet Aggregation System | HE - Hematology |
## II Submission/Device Overview:
A Purpose for Submission:
Clearance of a new device
B Measurand:
Platelet function
C Type of Test:
Platelet aggregation
## III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
{1}
B Indication(s) for Use:
The T-TAS 01 Instrument is intended for use with T-TAS reagent chips in the clinical laboratory.
The T-TAS 01 PL chip is intended for use in the clinical laboratory for the analysis of the platelet thrombus formation process (primary hemostatic function) in patients age 21 and older with a history of conditions associated with impaired primary hemostatic function or use of antiplatelet therapy. The test uses BAPA-anticoagulated whole blood specimens to measure platelet adhesion to a thrombogenic collagen-coated surface and aggregation, which causes an increase in flow pressure inside the PL chip. The test measures primary hemostatic function as the area under the pressure-time curve (AUC), with AUC < 260 suggesting abnormal primary hemostatic function. Additional testing may be necessary to identify the cause(s) of abnormal primary hemostatic function. The test has been evaluated in patients taking antiplatelet therapy, in patients with von Willebrand disease, and in patients with Glanzmann’s thrombasthenia. Other primary hemostasis disorders have not been evaluated.
The BAPA tube for T-TAS 01 is intended to be used for the collection, transport, and storage of blood specimens for use with the T-TAS 01 system.
C Special Conditions for Use Statement(s):
- Rx - For Prescription Use Only
- Additional testing may be necessary to identify the cause(s) of a primary hemostasis defect.
- The BAPA tube is used for the collection, transport, and storage of blood specimens for use with the T-TAS 01 system.
D Special Instrument Requirements:
IV Device/System Characteristics:
A Device Description:
The T-TAS 01 system is an automated platelet function system. It consists of a tabletop instrument controlled by a dedicated PC, a disposable, single-use flow chamber microchip, and anticoagulant collection tubes with benzylsulfonyl-D-Arg-Pro-4-amidinobenzylamide (BAPA). The PL Chip for T-TAS 01 contains a collagen-coated analytical path consisting of 26 microcapillary channels.
B Principle of Operation:
The assay is performed using BAPA-anticoagulated whole blood samples. BAPA inhibits thrombin and factor Xa. During the assay, the blood sample is exposed to arterial shear stresses at 1500 s⁻¹ in the presence of a collagen-coated surface, which causes platelet attachment to collagen mediated by von Willebrand factor (vWF) and platelet activation. Platelet activation causes the release of endogenous factors that recruit and activate other platelets and cause aggregation, or platelet thrombus formation (PTF) and growth.
K191364 - Page 2 of 14
{2}
PTF causes occlusion of the microcapillary channels, which increases the flow pressure within the assay chip. The process of PTF in the flow chamber is continuously monitored by a pressure sensor that tracks pressure changes in the flow path. After the start button is pressed, the blood sample is injected into the chip at a high flow rate for the purpose of moving blood samples into the chip and past the microcapillaries without creating air bubbles. After the initial step, the assay begins, using a set flow rate. The flow pressure readings subtract the base pressure and are plotted in the pressure waveform graph. The assay is completed at 10 minutes or when the occlusion pressure reaches 60 kPa, whichever occurs first. The area under the flow pressure curve (AUC) over 10 minutes is automatically calculated. The end users will see the pressure waveform graph and the AUC results.
AUC results less than 260 are associated with abnormal primary hemostasis.
C Instrument Description Information:
| Modes of Operation | Yes | No |
| --- | --- | --- |
| Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? | ☑ | ☐ |
| Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? | ☐ | ☑ |
| Software | | |
| FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types. | ☑ | ☐ |
1. Instrument Name:
T-TAS 01 Instrument
2. Specimen Identification:
Specimen identification is performed by use of a barcode scanner or manually typing the specimen information.
3. Specimen Sampling and Handling:
- The T-TAS 01 system uses BAPA-anticoagulated whole blood freshly collected with a 21 gauge or larger-bore needle.
- The whole blood is mixed with anticoagulant by gently inverting the tube 5 times.
- The whole blood sample should be incubated upright at room temperature for 30 minutes prior to testing.
- Blood samples should be measured between 30 minutes to 6 hours after collection.
- Specimens should be transported at room temperature and extreme temperatures should be avoided
- Use of pneumatic tube transport systems may cause platelet activation. Such transport systems will need to be validated by the laboratory for suitability
- Avoid using hemolyzed specimens
K191364 - Page 3 of 14
{3}
K191364 - Page 4 of 14
4. Calibration:
All calibration is performed by the manufacturer. There are no calibration activities performed by the end-user.
5. Quality Control:
The following three types of System Checks (SCs) are automatically performed but focus primarily on the functionality of the instrument:
| SC Type | Frequency | Function |
| --- | --- | --- |
| Simple SC | Automatically performed during each test | Check for pressure leaks inside the pumps. Check the solenoid valve, pump and tubing. |
| Automatic SC | Automatically performed every time the system is powered on or the PL chip is selected from the measurement screen. | Check for pressure leaks inside the pumps. Check the solenoid valve, pump and tubing. |
| Manual SC | Monthly (recommended) | Check for pressure leaks within the overall system, including the tubing from the pumps to the nozzles. |
Testing of blood samples from a normal healthy donor is needed to confirm reagent integrity and performance of the PL chip. It is recommended to test a control donor blood sample in duplicate with each new shipment of PL chips received or to verify the performance of the system.
V Substantial Equivalence Information:
A Predicate Device Name(s):
Dade PFA-100 Platelet Function Analyzer and Dade PFA-100 Reagents
B Predicate 510(k) Number(s):
K060489
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K191364 | K060489 |
| --- | --- | --- |
| Device Trade Name | Total Thrombus-formation Analysis (T-TAS) 01 System | Dade® PFA-100® Platelet Function Analyzer
Dade® PFA-100® Reagents |
{4}
| Device & Predicate Device(s): | K191364 | K060489 |
| --- | --- | --- |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The T-TAS 01 Instrument is intended for use with T-TAS reagent chips in the clinical laboratory.
The T-TAS 01 PL chip is intended for use in the clinical laboratory for the analysis of the platelet thrombus formation process (primary hemostatic function) in patients age 21 and older with a history of conditions associated with impaired primary hemostatic function or use of antiplatelet therapy. The test uses BAPA-anticoagulated whole blood specimens to measure platelet adhesion to a thrombogenic collagen-coated surface and aggregation, which causes an increase in flow pressure inside the PL chip. The test measures primary hemostatic function as the area under the pressure-time curve (AUC), with AUC < 260 suggesting abnormal primary hemostatic function. Additional testing may be necessary to identify the cause(s) of abnormal primary hemostatic function. | The Dade® PFA-100® Platelet Function Analyzer and Dade® PFA-100® Reagents are in vitro diagnostic devices intended to aid in the detection of platelet dysfunction in citrated human whole blood. |
K191364 - Page 5 of 14
{5}
K191364 - Page 6 of 14
| Device & Predicate Device(s): | K191364 | K060489 |
| --- | --- | --- |
| | The test has been evaluated in patients taking antiplatelet therapy, in patients with von Willebrand disease, and in patients with Glanzmann’s thrombasthenia. Other primary hemostasis disorders have not been evaluated.
The BAPA tube for T-TAS 01 is intended to be used for the collection, transport, and storage of blood specimens for use with the T-TAS 01 system. | |
| Platelet Adhesion Matrix | Collagen | Same |
| Calibration | Factory calibrated | Same |
| Internal Quality Control | Self-test performed at least once per shift at the start of each shift | Same |
| External Quality Control | External quality control uses known control donor blood samples | Same |
| Assay Temperature | 36°C | 37.9 ± 1°C |
| Chip/Cartridge Warm-up Time in Instrument Prior to Assay | Same | < 2.5 minutes |
| Recommended Operating Temperature | 68°F to 86°F
(20°C to 30°C) | 64°F to 90°F
(18°C to 29°C) |
| General Device Characteristic Differences | | |
| Specimen Type | BAPA-anticoagulated whole blood | Buffered sodium citrate-anticoagulated whole blood |
| Detection Principle | Measures integrated area under the curve of pressure over time as platelets adhere to a thrombogenic surface under high shear flow | Measures time required for platelets to adhere and aggregate, resulting in occlusion of an aperture under high shear flow conditions. |
{6}
| Device & Predicate Device(s): | K191364 | K060489 |
| --- | --- | --- |
| | conditions. | |
| Mechanism of Platelet Activation | vWF-mediated binding to collagen at high shear rate | Soluble exogenous agonist (epinephrine) + vWF-mediated binding to collagen at high shear rate |
| Shear Rate During Assay | 1500 s^{-1} | 5000 s^{-1} |
| Other Reagents Necessary for Assay | Mineral oil | Trigger solution |
| Sample Volume | 300–330 μL (320 μL recommended) | 800 μL |
| Sample Incubation Time Prior to Assay | 30 minutes at room temperature | 15 minutes at room temperature |
| Maximum Allowable Sample Stability Time Prior to Assay | 6 hours at room temperature | 4 hours at room temperature |
| Assay Units | AUC | Closure time (seconds) |
| Recommended Cutoff | AUC < 260 | > 170 seconds |
| Interpretation of Result | AUC ≥ 260 indicates the overall primary hemostatic function is normal.
AUC < 260 is considered abnormal and indicates impaired primary hemostatic function (reduced platelet thrombus formation). | Closure time > 170 seconds is considered abnormal and indicates platelet dysfunction |
VI Standards/Guidance Documents Referenced:
CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition
CLSI EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline – Second Edition
CLSI EP28-A3c, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline – Third Edition
CLSI EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline
K191364 - Page 7 of 14
{7}
CLSI GP34-A, Validation and Verification of Tubes for Venous and Capillary Blood Specimen Collection; Approved Guidance
ISO 14971, Medical devices – Applications of Risk Management to Medical Devices
CLSI GP39-A6, Tubes and Additives for Venous and Capillary Blood Specimen Collection; Approved Standard – Sixth Edition
EN 61326-1, Electrical equipment for measurement, control, and laboratory use – EMC requirements – Part 1: General requirements
EN 61326-2-6, Electrical equipment for measurement, control, and laboratory use – EMC requirements – Part 2-6: Particular requirements for in vitro diagnostic (IVD) medical equipment
## VII Performance Characteristics (if/when applicable):
## A Analytical Performance:
### 1. Precision:
#### a. Repeatability Study
The repeatability study was performed at a single laboratory. Three unique venous blood samples representing different AUC levels (low, middle, high) were tested on either two of the three T-TAS 01 instruments. A normal donor provided the sample with normal primary hemostatic ability (high). Two donors taking aspirin provided specimens with primary hemostatic ability around the assay cutoff (middle) and abnormal primary hemostatic ability (low). Each sample was analyzed in duplicate on three PL chip lots and two T-TAS 01 instruments by three different trained operators. The results met the predefined acceptance criteria.
| Operator | A | | | | | | B | | | | | | C | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| T-TAS 01 instrument | 4 | | | 5 | | | 5 | | | 6 | | | 4 | | | 6 | | |
| PL chip lot | 1 | 2 | 3 | 1 | 2 | 3 | 1 | 2 | 3 | 1 | 2 | 3 | 1 | 2 | 3 | 1 | 2 | 3 |
| Replicate | AB | AB | AB | AB | AB | AB | AB | AB | AB | AB | AB | AB | AB | AB | AB | AB | AB | AB |
| Sample | N | Mean | Within-Run (SD, %CV) | Between-Operator (SD, %CV) | Between-Lot (SD, %CV) | Between-Instrument (SD, %CV) | Total (SD, %CV) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| High | 36 | 428.1 | 10.7, 2.5 | 2.0, 0.5 | 4.7, 1.1 | 1.6, 0.4 | 11.9, 2.8 |
| Middle | 36 | 237.3 | 31.7, 13.4 | 6.4, 2.7 | 10.5, 4.4 | 0.0, 0.0 | 34.0, 14.3 |
| Low | 36 | 130.7 | 18.4, 14.1 | 11.8, 9.0 | 13.5, 10.3 | 0.0, 0.0 | 25.7, 19.6 |
#### b. Reproducibility
The reproducibility study was conducted at three sites over five days. At each site, four unique whole blood samples were tested in five duplicate measurements with one PL
K191364 - Page 8 of 14
{8}
chip lot and one T-TAS 01 instrument. The four donor types used in the study are as follows, considering inter-individual variability in response to aspirin:
| Donor | AUC Result Category (AUC < 260 = Abnormal) | Aspirin Use | Aspirin Response |
| --- | --- | --- | --- |
| 1 | Normal | No | N/A |
| 2 | Normal | Yes | Poor Response |
| 3 | High Abnormal | Yes | Medium Response |
| 4 | Low Abnormal | Yes | High Response |
The study evaluated the within-run and between-day precision. Since different donors were recruited at each site for the precision study, results were not pooled across days or donors due to natural inter- and intra-individual variability in response to aspirin. The pre-defined acceptance criteria were. All results met the pre-defined acceptance criteria.
Site 1:
| Donor | N | Mean | Within-Run (SD, %CV) | Between-Day (SD, %CV) | Total (SD, %CV) |
| --- | --- | --- | --- | --- | --- |
| 1 | 25 | 413.4 | 12.8, 3.1 | 16.7, 4.0 | 21.1, 5.1 |
| 2 | 25 | 366.5 | 25.8, 7.7 | 0.0, 0.0 | 25.8, 7.7 |
| 3 | 25 | 140.8 | 23.9, 17.0 | 27.9, 19.8 | 36.7, 26.1 |
| 4 | 25 | 50.0 | 15.2, 30.3 | 12.9, 25.9 | 19.9, 39.9 |
Site 2:
| Donor | N | Mean | Within-Run (SD, %CV) | Between-Day (SD, %CV) | Total (SD, %CV) |
| --- | --- | --- | --- | --- | --- |
| 1 | 25 | 417.7 | 20.0, 4.8 | 13.1, 3.1 | 23.9, 5.7 |
| 2 | 25 | 347.5 | 27.5, 7.9 | 10.7, 3.1 | 29.5, 8.5 |
| 3 | 25 | 176.0 | 26.0, 14.8 | 21.9, 12.5 | 34.0, 19.3 |
| 4 | 25 | 78.4 | 22.7, 29.0 | 20.8, 26.5 | 30.8, 39.3 |
Site 3:
| Donor | N | Mean | Within-Run (SD, %CV) | Between-Day (SD, %CV) | Total (SD, %CV) |
| --- | --- | --- | --- | --- | --- |
| 1 | 25 | 411.1 | 11.9, 2.9 | 10.0, 2.4 | 15.5, 3.8 |
| 2 | 25 | 336.4 | 20.2, 6.0 | 23.2, 6.9 | 30.7, 9.1 |
| 3 | 25 | 162.6 | 28.9, 17.7 | 3.6, 2.2 | 29.1, 17.9 |
| 4 | 25 | 54.6 | 10.2, 18.7 | 13.6, 24.8 | 17.0, 31.1 |
2. Linearity:
Not applicable
3. Analytical Specificity/Interference:
The interference studies were conducted to determine whether various substances commonly encountered in the intended use population affect the T-TAS 01 PL assay results. The analytical specificity for the T-TAS 01 PL assay was determined by spiking venous whole
K191364 - Page 9 of 14
{9}
blood samples taken from a normal donor (AUC ≥ 260) and a donor taking aspirin (AUC < 260). There were at least four replicate measurements for each control and test blood sample. The following endogenous and exogenous interfering substances were evaluated and showed no significant interference up to the specified concentration in the venous whole blood samples collected in the BAPA tubes.
| Compound | Concentration | Compound | Concentration |
| --- | --- | --- | --- |
| Acetaminophen | 7.8 mg/dL | Heparin | 525 U/mL |
| Bilirubin | 40 mg/dL | L-Thyroxine | 0.0858 mg/dL |
| Caffeine | 21.6 mg/dL | Metformin | 2.4 mg/dL |
| Captopril | 0.528 mg/dL | Omeprazole | 1.68 mg/dL |
| Catechin | 5 mg/dL | Pravastatin | 0.414 mg/dL |
| Cilostazol | 1.25 mg/dL | Propranolol | 0.202 mg/dL |
| Dabigatran | 0.047 mg/dL | Rivaroxaban | 0.044 mg/dL |
| Dextran 40 | 2400 mg/dL | Streptokinase | 50,000 U/dL |
| Diltiazem | 0.18 mg/dL | Theophylline | 6 mg/dL |
| Dipyridamole | 0.25 mg/dL | Triglycerides | 750 mg/dL |
| Fish Oil | 25.6 mg/dL | Warfarin | 7.5 mg/dL |
| Ibuprofen | 0.438 mg/dL | | |
Cilostazol, dipyridamole, ibuprofen and tirofiban are known to reduce platelet activity in a dose-dependent manner. The maximum tirofiban concentration without interference was not determined.
Hemolysis significantly affects assay AUC results due to shear-induced platelet activation and aggregation. Therefore, a statement is included in the package insert to alert the user not to use hemolyzed specimens when performing the T-TAS 01 PL assay.
There was no significant hemodilution effect on AUC results up to 20%.
There was no significant effect of under-filling the BAPA blood collection tube on AUC results up to 50%.
4. Assay Reportable Range:
The reportable range and the analytical measurement range for the T-TAS 01 assay were determined upon the range of clinical samples as 0.3–467.7. However, the numeric output of the PL assay has not been evaluated for correlation to disease severity.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
a. Shelf Life/Stability
i. Reagent stability for T-TAS 01 PL chip
Open-pouch stability studies were isochronously performed to characterize the stability of PL chips following removal from the sealed pouch and storage at ambient temperature. Venous whole blood samples were collected from a single healthy donor
K191364 - Page 10 of 14
{10}
with normal primary hemostatic ability into BAPA-anticoagulated tubes. Three PL chip lots were tested and two PL chips from each lot were removed from their pouches and stored at ambient temperature for the following durations prior to testing: <5 minutes, 4 hour, 6 hours, 8 hours and 10 hours. Each chip was tested with normal donor blood in quadruplicate. The “<5 minute” condition serves as the control condition for comparison. The data support a stability claim of up to 8 hours at ambient temperature after PL chip removal from the pouch.
Closed-pouch stability studies were isochronously performed to characterize the stability of PL chips when stored in sealed pouches under refrigeration (2–8°C). Venous whole blood samples were collected into BAPA-anticoagulated tubes from two healthy donors with normal primary hemostatic ability and a native donor with abnormal AUC result. Three PL chip lots were tested. Two PL chips from each lot were tested following the tested storage condition at the following durations after their manufacture: 0–13 months. Each chip was tested with each blood sample in quadruplicate. The control condition was a fresh PL chip lot that was newly manufactured prior to the testing timepoint. The data support a stability claim of 12 months when stored at 2–8°C.
ii. Blood sample stability
The stability of BAPA-anticoagulated blood specimens stored at room temperature was tested by collecting venous whole blood samples from two healthy normal donors and two subjects taking aspirin with abnormal primary hemostatic function. Blood samples were stored upright at room temperature for 30 minutes, 1 hour, 2 hours, 4 hours, 6 hours, 8 hours, and 24 hours prior to testing. Three T-TAS 01 instruments and one lot of PL chips were used in the test. Six replicate measurements were performed for each tested time point. The control condition was established as 30 minutes after sample collection. Results were averaged prior to analysis. Each condition was compared to the control condition by calculating the % of control for the average of the replicates. The results demonstrate that BAPA-anticoagulated venous blood samples are stable for up to 6 hours after collection.
6. Detection Limit:
Not applicable
7. Assay Cut-Off:
The T-TAS PL 01 assay cut off (AUC < 260) was established in a group of 122 apparently healthy adults (34 females, 88 males, ages 20–45). The recruited participants did not take antiplatelet or nonsteroidal anti-inflammatory drugs (NSAID). They did not have any history of significant bleeding or diagnosis of a primary hemostatic abnormality.
8. Accuracy (Instrument):
Not applicable
9. Carry-Over:
K191364 - Page 11 of 14
{11}
Not applicable
## B Comparison Studies:
### 1. Method Comparison with Predicate Device:
The method comparison study was performed to compare the performance of T-TAS 01 PL assay to the predicate device PFA-100 assay. Both assays evaluated 25 patients with vWD and three patients with GT. The predicate device involves the measurement of the collagen/epinephrine (Col/Epi) cartridge and collagen/ADP (Col/ADP) cartridge. The result interpretation was according to the PFA-100 test cartridge package insert. The following table demonstrates the percent positive, percent negative and overall agreement between the T-TAS 01 PL assay and the PFA-100 assay for the vWD and GT patients.
| Subject Type | N | Percent Positive Agreement (95% CI) | Percent Negative Agreement (95% CI) | Overall Agreement (95% CI) |
| --- | --- | --- | --- | --- |
| vWD | 25 | 72% (51–88%) | 100.0% (40–100%) | 88% (69–97%) |
| GT | 3 | 100% (43.9–100.0%) | N/A | 100% (43.9–100.0%) |
### 2. Matrix Comparison:
Not applicable
## C Clinical Studies:
### 1. Clinical Sensitivity:
Six investigational sites recruited 142 healthy subjects, 57 patients on aspirin (81-mg ASA) monotherapy, 47 subjects taking dual antiplatelet therapy (18 subjects on clopidogrel and ASA, 15 subjects on prasugrel and ASA, and 14 subjects on ticagrelor and ASA), 25 vWD patients and three GT patients. Among vWD patients, there were 12 patients with Type 1, 10 patients with Type 2 and three patients with Type 3. The study was to evaluate the sensitivity and specificity of the T-TAS 01 PL assay to detect conditions associated with the abnormal primary hemostatic function. At each investigational site, blood samples were tested in duplicate using BAPA-anticoagulated whole blood samples. Measurements were performed within four hours of sample collection. The first measurement was used for the data analysis. The following table provides a summary for each subject group.
| Parameter | N | Value | 95% CI |
| --- | --- | --- | --- |
| Sensitivity (ASA) | 57 | 68.4% | 55.5-79.0% |
| Sensitivity (clopidogrel + ASA DAPT) | 18 | 100.0% | 81.5-100.0% |
| Sensitivity (prasugrel + ASA DAPT) | 15 | 100.0% | 78.2-100.0% |
| Sensitivity (ticagrelor + ASA DAPT) | 14 | 100.0% | 76.8-100.0% |
| Sensitivity (vWD) | 25 | 72.0% | 51–88% |
K191364 - Page 12 of 14
{12}
| Parameter | N | Value | 95% CI |
| --- | --- | --- | --- |
| Sensitivity (GT) | 3 | 100.0% | 43.9-100% |
| Negative agreement (healthy) | 142 | 95.8% | 91.1–98.0 % |
2. Clinical Specificity:
See Clinical Sensitivity above.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
D Clinical Cut-Off:
Not applicable
E Expected Values/Reference Range:
Three clinical sites recruited 142 healthy adult subjects to establish the reference range for the T-TAS 01 PL assay. At each testing site, blood samples were tested in duplicate using BAPA-anticoagulated whole blood samples. Measurements were performed within four hours of sample collection. The first measurement was used for the data analysis. The central 90% reference interval is 270.0–447.7. Measurements from a separate group of 20 healthy subjects validated the established reference interval (302.5–429.1).
F Other Supportive Instrument Performance Characteristics Data:
1. BAPA blood collection tube validation
a. BAPA tube vacuum performance
The BAPA tube vacuum performance was conducted to demonstrate the stability of the tube vacuum at ambient temperature. BAPA tubes from three production lots were stored at room temperature. The tube vacuum was measured by allowing the tube to fill with water, measuring the weight of the water inside the tube and converting the weight (gram) to volume (mL). Five replicate measurements were performed for each condition. The study is on-going. Current results support the vacuum present in the BAPA tube for T-TAS 01 is stable for 10 months.
b. BAPA tube anticoagulant fill
The BAPA tube anticoagulant fill study was conducted to determine the consistency of BAPA anticoagulant added to the BAPA tube for the T-TAS 01 system. Three lots of the BAPA tubes were used and ten replicate measurements were performed for each lot. The results demonstrated that the amount of BAPA anticoagulant added to the BAPA tube for the T-TAS 01 system is consistent.
c. Sterilization
K191364 - Page 13 of 14
{13}
The BAPA tube was sterilized by gamma irradiation. Sterilization validation was conducted and concluded that there is a 1 in 1,000,000 possibility that the product is not sterile.
d. Transportation stress test
Transport stress test was conducted to demonstrate the stability of BAPA tubes after transportation. The randomly selected samples were put on the oscillation device for three days (frequency at 150 times/min) to simulate the bumping situation in transportation. The device was also exposed to high temperature (40°C) and high humidity (80% relative humidity) for 30 days. The test results support the stability of BAPA tubes after transportation.
2. Restriction of Hazardous Substances Directive (RoHS)
The following substances in the T-TAS 01 do not exceed any threshold as set forth by the RoHS directive: cadmium, hexavalent chromium, lead, mercury, polybrominated biphenyls, polybrominated diphenyl ethers
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K191364 - Page 14 of 14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.