K191331 · Braveheart Wireless, Inc. · DRG · Jun 12, 2019 · Cardiovascular
Device Facts
Record ID
K191331
Device Name
Life Sensor Cardiac Monitor
Applicant
Braveheart Wireless, Inc.
Product Code
DRG · Cardiovascular
Decision Date
Jun 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2910
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Life Sensor Cardiac Monitor (CM) is a wireless monitoring system intended for use by healthcare professionals for monitoring of physiological data within healthcare settings. This includes heart rate and electrocardiography (ECG). Data is transmitted wirelessly from Life Sensor Electrode to an application on iOS device where it is displayed for review by healthcare professionals. The device is intended for use on general care patients 18 years or older and by prescription only.
Device Story
Wireless monitoring system for ECG and heart rate; consists of single-use adhesive Life Sensor Patch, battery-powered Life Sensor Module, and Life Sensor App on iOS device. Module captures physiological signals; transmits encrypted data via Bluetooth Low Energy (BLE) to iOS app. App displays data and provides visual alarms for heart rate limits, lead-off, asystole, tachycardia, bradycardia, low battery, or communication loss. Used in healthcare settings by healthcare professionals; provides real-time surveillance to assist clinical decision-making. Benefits include continuous patient monitoring without tethering to bedside equipment.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including biocompatibility (ISO 10993-1), electrical safety/EMC (IEC 60601-1, IEC 60601-2-27, IEC 60601-1-2), software verification/validation, shelf life, packaging integrity, and usability testing (IEC 62366).
Technological Characteristics
Wireless patch design; Bluetooth Low Energy (BLE) 2.4 GHz connectivity. Powered by non-rechargeable lithium-ion battery. Single-channel ECG sensing. Complies with IEC 60601-1, IEC 60601-2-27, IEC 60601-1-2, and IEC 62366. Software level of concern: moderate.
Indications for Use
Indicated for general care patients 18 years or older for monitoring heart rate and ECG in healthcare settings. Contraindicated for critical care patients, patients with active implantable medical devices (pacemakers, ICDs, LVADs), use in MR environments, or during surgical procedures with electro-surgical equipment.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym and name on the right. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is in blue text.
June 12, 2019
BraveHeart Wireless, Inc. % Thomas Schorre Official Correspondent Accelerated Device Approval Services, LLC 6800 S.W. 40th Street. Ste. 444 Ludlum, Florida 33155-3708
Re: K191331
Trade/Device Name: Life Sensor Cardiac Monitor Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency physiological signal transmitter and receiver Regulatory Class: Class II Product Code: DRG, DRT, DRX Dated: May 14, 2019 Received: May 16, 2019
Dear Thomas Schorre:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Nicole Goodsell Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K191331
Device Name Life Sensor Cardiac Monitor
## Indications for Use (Describe)
The Life Sensor Cardiac Monitor (CM) is a wireless monitoring system intended for use by healthcare professionals for monitoring of physiological data within healthcare settings. This includes heart rate and electrocardiography (ECG). Data is transmitted wirelessly from Life Sensor Electrode to an application on iOS device where it is displayed for review by healthcare professionals. The device is intended for use on general care patients 18 years or older and by prescription only.
The device is contraindicated for use on critical care patients with active implantable medical devices such as pacemakers, implanted cardioverter defibrillator (ICD), and left ventricular assist devices (LVAD); for use in magnetic resonance (MR) environments; for use during surgical procedures when electro-surgical equipment is optional. The Life Sensor Cardiac Monitor does not detect or diagnose medical conditions.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5
# 510(k) Summary for Life Sensor Platform
# (As required by 21 CFR 807.92)
### Submitter Information l.
Date Prepared:
| Submitter | BraveHeart Wireless, Inc.<br>11 Perimeter Road<br>Nashua, NH 03063 |
|-----------------|------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Balaji Sudabattula<br>VP, Quality and Regulatory Affairs<br>Phone: 385-988-0625<br>Email: balaji.sudabattula@braveheart.life |
April 10, 2019
### II. Device Identification and Classification
| Device Trade Name:<br>Model Name:<br>Common Name: | Life Sensor Platform<br>Life Sensor Cardiac Monitor<br>Wireless Remote Monitoring System |
|---------------------------------------------------|------------------------------------------------------------------------------------------|
| Classification Name: | Transmitters and Receivers, Physiological Signal,<br>Radiofrequency |
| Classification Panel: | Cardiovascular |
| CRF Section: | 870.2910, 870.2300, 870.2360 |
| Device Class: | |
| Product Code: | DRG, DRT, DRX |
### III. Predicate Device
| Manufacturer: | Life Signals, Inc |
|---------------|------------------------------------|
| Model Name: | LifeSignals WiPatch & WiApp System |
| 510(k) #: | K172011 |
This predicate has not been subject to a design-related recall.
{4}------------------------------------------------
### IV. Device Description
Life Sensor Electrode
Bluetooth
iPad installed with Life Sensor App
The Life Sensor Cardiac Monitor (CM) Platform is a wireless data collection system that monitors physiological data (electrocardiograph and heart rate) and consists of the following sub-systems:
Life Sensor Electrode (includes Life Sensor Patch and Life Sensor Module) Life Sensor Firmware Life Sensor Application
# Life Sensor Electrode
- Life Sensor Patch: The Life Sensor Patch is the single-use, self-adhesive interface to a patient that is attached to the patient's upper torso region just above the heart.
- . Life Sensor Module: The Life Sensor Module attached to the Life Sensor Patch, contains the battery powered electronics and sensing apparatus necessary to operate the system. The Life Sensor Module automatically performs all the processing functions related to capturing the required physiological data from the body and performs encrypted, bi-directional communication to the Life Sensor Application, using Bluetooth Low Energy (BLE), when in range of the Life Sensor Application installed on a paired iOS device.
- Life Sensor Firmware:
The Life Sensor Firmware is the software installed on the Life Sensor Module. The Life Sensor Firmware manages bi-directional communication between the Life Sensor Module and the Life Sensor Application. Encrypted data is transmitted to and from the Life Sensor Module and the Life Sensor Application.
- . Life Sensor Application: The Life Sensor Application, installed on a paired iOS device, interacts with the Life Sensor Firmware and manages the upload, processing, and display of the physiological data transmitted by the Life Sensor Module.
When installed on a compatible hardware, the Life Sensor Application establishes connection to any of the Life Sensor Electrode through pairing process.
{5}------------------------------------------------
Life Sensor Application has visual alarms to alert clinical personnel when heart rate of the person being monitored is outside the set limits, lead off, asystole, tachycardia and bradycardia, when the battery is low, or when there is loss of communication between the Life Sensor Application and the Life Sensor Electrode.
#### V. Indications for Use
The Life Sensor Cardiac Monitor (CM) is a wireless monitoring system intended for use by healthcare professionals for monitoring of physiological data within healthcare settings. This includes heart rate and electrocardiography (ECG). Data is transmitted wirelessly from Life Sensor Electrode to an application on iOS device where it is displayed for review by healthcare professionals. The device is intended for use on general care patients 18 years or older and by prescription only.
The device is contraindicated for use on critical care patients with active implantable medical devices such as pacemakers, implanted cardioverter defibrillator (ICD), and left ventricular assist devices (LVAD); for use in magnetic resonance (MR) environments; for use during surgical procedures when electro-surgical equipment is optional. The Life Sensor Cardiac Monitor does not detect or diagnose medical conditions.
| Comparison | Life Sensor Platform<br>(subject) | WiPatch & WiApp System |
|--------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor | BraveHeart Wireless, Inc.<br>Nashua, NH | LifeSignals<br>Fremont, CA |
| Intended use | The Life Sensor Platform is a<br>wireless monitoring system<br>intended for use by healthcare<br>professionals for surveillance of<br>physiological data within<br>healthcare settings. This includes<br>heart rate and<br>electrocardiography (ECG). Data<br>is transmitted wirelessly from Life<br>Sensor Patch to an application on<br>iOS device where it is displayed<br>for review by healthcare<br>professionals. The device is<br>intended for use on general care | The LifeSignals WiPatch &<br>WiApp system is intended for<br>spot check and short-term<br>monitoring of ECG & heart<br>rate of patients at rest or<br>patients who can be<br>transported within the range<br>of the antenna network. The<br>WiPatch & WiApp system has<br>visual and audio alarms to<br>alert clinical personnel when<br>heart rate falls outside the set<br>limits. The device is intended<br>for use on general care<br>patients of 18 years or older.<br>The device is not intended for |
### VI. Summary Table of Substantial Equivalence
{6}------------------------------------------------
| Comparison | Life Sensor Platform<br>(subject) | WiPatch & WiApp System |
|-----------------------------|---------------------------------------------------------|----------------------------------|
| | patients 18 years or older and by<br>prescription only. | use on critical care patients. |
| Intended<br>Population | Adults ≥ 18 | Adults ≥ 18 |
| Intended Use<br>Environment | Healthcare facility | Healthcare facility |
| Heart Rate | Yes | Yes |
| ECG Display | Single channel | ECG-A and ECG-B |
| Sensor | Wireless patch design - | Wireless patch design RF/Wi-Fi |
| | RF/Bluetooth 2.4 GHz | 2.4 GHz 802.11b |
| Wear duration | 72 hours maximum | 72 hours maximum |
| Hardware | Standard commercial iOS tablet | Standard commercial iOS tablet |
| Platform | | |
| Technical Specifications | | |
| Display Range | 30 to 200 BPM | 30 to 250 BPM |
| Accuracy | ±10 or ±5/min, whichever is | ±3/min |
| | greater. | |
| Battery | Lithium ion (non-chargeable) | Two zinc air coin cell batteries |
Life Sensor CM Platform and the predicate device (LifeSignals WiPatch & WiApp System, 510(k) #: K172011) monitor and display ECG and heart rate. Healthcare professional can set limits for heart rate to alarm the healthcare provider if a patient's heart rate falls outside the set limits. Both transmit data to an iPad for review by healthcare professional.
Minor difference is that the predicate device uses Wi-Fi while the subject device uses Bluetooth Low Energy (BLE) to communicate with iPad.
### VII. Summary of Performance Testing
The following non-clinical testing has been completed to demonstrate safety and effectiveness of the Life Sensor Cardiac Monitor Platform.
- . Biocompatibility testing, including in-vitro cytotoxicity, irritation and sensitization, performed according to the recommendations of ISO 10993-1:2018, Biological evaluation of medical devices – Part: Evaluation and testing.
- . Electrical safety and EMC testing. The device complies with IEC 60601-1, IEC 60601-2-27, IEC 60601-1-2.
{7}------------------------------------------------
- . Software verification and validation testing was conducted, and documentation is provided as recommended by FDA's Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The software for this device is determined as "moderate" level of concern because a failure or latent flaw could lead to a minor injury to the patient through incorrect information or through the action of the care provider.
- . Bench testing, including shelf life, packaging integrity, operation and storage conditions.
- Usability testing. The device complies with IEC 62366.
# VIII. Conclusion
The Life Sensor Cardiac Monitor Platform is substantially equivalent with respect to safety and effectiveness to the legally marketed predicate device for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.