← Product Code [HSX](/productcode/HSX) · K191325

# EMPOWR Partial Knee (K191325)

_Encore Medical L.P. · HSX · Jul 16, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K191325

## Device Facts

- **Applicant:** Encore Medical L.P.
- **Product Code:** [HSX](/productcode/HSX.md)
- **Decision Date:** Jul 16, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3520
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The EMPOWR Partial Knee System is indicated for patients with: - Painful and/or disabling knee joints due to osteoarthritis or traumatic arthritis. - Previous tibial condyle or plateau fractures with loss of anatomy or function. - Varus or valgus deformities. - Revision procedures where other treatments or devices have failed. - These devices are indicated for cemented use only.

## Device Story

EMPOWR Partial Knee is a unicompartmental knee prosthesis; replaces damaged joint surfaces in patients with osteoarthritis, traumatic arthritis, or fractures. System comprises cobalt chrome femoral component, titanium tibial baseplate, and Vitamin E polyethylene insert. Device is unconstrained in anteroposterior and mediolateral directions; allows internal/external rotation. Used by orthopedic surgeons in clinical/surgical settings to restore knee function and reduce pain. Output is a physical implant system; clinical decision-making based on patient anatomy and pathology. Benefits include restoration of joint function and pain relief.

## Clinical Evidence

No clinical data. Substantial equivalence established via non-clinical bench testing, including FEA for femoral strength, tibial baseplate strength testing, range of motion assessment, tibio-femoral contact area testing, and static locking mechanism disassembly.

## Technological Characteristics

Unicompartmental knee system. Materials: Cobalt chrome (femoral), titanium (tibial baseplate), Vitamin E polyethylene (insert). Design: 2-peg femoral, keel/2-peg tibial baseplate. Unconstrained articulation. Cemented use only. Sterilization and specific ASTM standards not provided.

## Regulatory Identification

A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Zimmer® Unicompartmental Knee ([K033363](/device/K033363.md))

## Reference Devices

- EPIK Knee System ([K020741](/device/K020741.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 16, 2019

Encore Medical, L.P. Teffany Hutto Manager, Regulatory Affairs 9800 Metric Blvd. Austin, Texas 78758

Re: K191325

Trade/Device Name: EMPOWR Partial Knee Regulation Number: 21 CFR 888.3520 Regulation Name: Knee Joint Femorotibial Metal/Polymer Non-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: HSX Dated: May 15, 2019 Received: May 16, 2019

Dear Teffany Hutto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Raquel Peat, PhD, MPH, USPHS Director OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K191325

Device Name EMPOWR Partial Knee

#### Indications for Use (Describe)

The EMPOWR Partial Knee System is indicated for patients with:

- · Painful and/or disabling knee joints due to osteoarthritis or traumatic arthritis.
- · Previous tibial condyle or plateau fractures with loss of anatomy or function.
- Varus or valgus deformities.
- · Revision procedures where other treatments or devices have failed.
- · These devices are indicated for cemented use only.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

Date: July 16, 2019

Manufacturer: DJO Surgical (Legal Name: Encore Medical, L.P.) 9800 Metric Blvd Austin, TX 78758

Contact Person: Teffany Hutto Manager, Regulatory Affairs Phone: (512) 834-6255 Fax: (760) 597-3466 Email: teffany.hutto(@djoglobal.com

| Product             | Common Name          | Classification | Product Code |
|---------------------|----------------------|----------------|--------------|
| EMPOWR Partial Knee | Partial Knee Implant | Class II       | HSX          |

| Product Code | Regulation and Classification Name                                                              |
|--------------|-------------------------------------------------------------------------------------------------|
| HSX          | Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer per 21 CFR<br>888.3520 |

### Description:

The EMPOWR Partial Knee is a unicompartmental knee system that is unconstrained in the anteroposterior and mediolateral directions, allowing unconstrained internal/external rotation between the femoral and tibial components. This system includes a cobalt chrome femoral component, titanium tibial baseplate, and Vitamin E Polyethylene insert.

### Indications for Use:

The EMPOWR Partial Knee System is indicated for patients with:

- Painful and/or disabling knee joints due to osteoarthritis or traumatic arthritis. o
- Previous tibial condyle or plateau fractures with loss of anatomy or function. ●
- Varus or valgus deformities. ●
- o Revision procedures where other treatments or devices have failed.

These devices are indicated for cemented use only.

#### Predicate Devices:

| Device                        | Manufacturer         | 510(k) Number | Predicate Type   |
|-------------------------------|----------------------|---------------|------------------|
| Zimmer® Unicompartmental Knee | Zimmer Biomet        | K033363       | Primary          |
| EPIK Knee System              | Encore Medical, L.P. | K020741       | Reference Device |

{4}------------------------------------------------

### Comparable Features to Predicate Device(s):

- Indications for Use ●
- Material of Femoral and Tibial Baseplate components. (Femoral component Cobalt o Chrome, Tibial Baseplate - Titanium)
- Tibial baseplate component sizing and dimensions
- Insert thicknesses ●
- Femoral Component Size offerings
- 2-peg femoral design
- Keel and 2-peg tibial baseplate design ●
- Unconstrained tibial insert articulation

## Key Differences in Subject Device to Predicate:

- Femoral component profile and articular surface geometry ●
- o Tibial baseplate peg and keel geometry

Non-Clinical Testing: Mechanical testing has demonstrated the device's ability to perform under expected conditions. This testing includes:

- FEA for Femur Component Strength
- Tibial Baseplate Component Strength Testing ●
- Range of Motion Assessment ●
- Tibial-femoral Contact Area Testing ●
- Static Locking Mechanism Disassembly

All testing has determined that the device is substantially equivalent to the predicate devices.

Endotoxin Assessment: DJO Surgical conducts device testing to assure that pyrogen limit specifications are met via the Kinetic Chromogenic method for bacterial endotoxin testing.

Clinical Testing: Clinical testing was not required

Conclusions: All testing and evaluations demonstrate that the device is substantially equivalent to the predicates identified.

---

**Source:** [https://fda-staging.innolitics.com/device/K191325](https://fda-staging.innolitics.com/device/K191325)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
