The LCCS VC-S RF Cannula is intended for use in RF heat lesion procedures for the relief of pain.
Device Story
The LCCS VC-S RF Cannula is a sterile, single-use, stainless steel needle used in conjunction with an RF electrode and RF generator to perform radiofrequency heat lesion procedures for chronic pain relief. The device features an insulated shaft with a bevel tip and a side port at the distal end, creating Y-shaped energy channels. During operation, an RF electrode is inserted into the cannula; when the electrode extends through the side port, both the active tip and the electrode section protruding from the side port deliver RF energy to the target nerve tissue. The side port also allows for the diffusion of anesthesia near the lesion site. The device is available in various lengths and gauges to accommodate different anatomical locations and patient anatomies. It is intended for use by clinicians in a clinical setting.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included insulation inspection, biocompatibility (cytotoxicity, acute systemic toxicity, sensitization, intracutaneous reactivity, pyrogenicity), accelerated aging, shelf life, and mechanical safety testing in accordance with ISO 10993-1, ISO 7864, and AAMI/ANSI ES 60601-2-2.
Technological Characteristics
Stainless steel cannula with insulated shaft; bevel tip and side port configuration; Y-shaped energy delivery channels. Available in 18G, 20G, 22G gauges and 50mm, 100mm, 150mm lengths. Single-use, sterile, ethylene oxide sterilized. Complies with ISO 10993-1, ISO 7864, and AAMI/ANSI ES 60601-2-2.
Indications for Use
Indicated for use in RF heat lesion procedures for the relief of pain in patients requiring percutaneous nerve blocks or RF denervation of nerve tissue.
Regulatory Classification
Identification
A radiofrequency lesion probe is a device connected to a radiofrequency (RF) lesion generator to deliver the RF energy to the site within the nervous system where a lesion is desired.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 24, 2020
LCCS Products Limited Nick Xu Regulatory Affairs Office 3a-7, 12/F, Kaiser Center, No.18 Center Street, Sai Ying Pun Hong Kong, China
Re: K191293
Trade/Device Name: LCCS VC-S RF Cannula Regulation Number: 21 CFR 882.4725 Regulation Name: Radiofrequency Lesion Probe Regulatory Class: Class II Product Code: GXI Dated: December 23, 2019 Received: December 26, 2019
Dear Nick Xu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Matthew Krueger, M.S.E. Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for LCCS. The logo consists of a green diamond shape on the left, with the letters "LCCS" in green to the right of the diamond. A registered trademark symbol is located in the upper right corner of the logo.
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>06/30/2020 | | Form Approved: OMB No. 0910-0120<br>Expiration Date: |
|---------------------------------------------------------------------------------------|--------------------------|------------------------------------------------------|
| Indications for Use | See PRA Statement below. | |
510(k) Number (if known)
K191293
Device Name
LCCS VC-S RF Cannula
Indications for Use (Describe)
The LCCS VC-S RF Cannula is intended for use in RF heat lesion procedures for the relief of pain.
Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881(1/14)
Page 1 of 1
{3}------------------------------------------------
### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| 1. Date: | February 28, 2019 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Submitter: | LCCS Products Limited<br>Office 3a-7, 12/F, Kaiser Center, No.18 Center Street, Sai<br>Ying Pun, Hong Kong |
| Contact person: | Nick XU<br>Regulatory Affairs<br>LCCS Products Limited<br>Office 3a-7, 12/F, Kaiser Center, No.18 Center Street, Sai<br>Ying Pun, Hong Kong<br>Telephone: +852 82321050<br>Email: xu.nick@qq.com |
- 3. Device Name:
LCCS VC-S RF Cannula
| Product Code | Classification | Regulation Section | Panel |
|--------------|----------------|-------------------------------------------------|-----------|
| GXI | Class II | 21 CFR 882.4725<br>Probe, Radiofrequency Lesion | Neurology |
#### 4. Predicate Devices:
| Predicate Device 510 (k)<br>number | Predicate Device name |
|------------------------------------|-------------------------------------------|
| K112231 | LCCS Insulated Spinal Needle (RF Cannula) |
| K123178 | Stryker® Venom™ Electrodes and Cannulae |
## 5. Device Description:
The LCCS VC-S RF Cannula is used in conjunction with the RF electrode for the use in RF heat lesion procedures. The LCCS VC-S RF Cannulas are offered in a variety of lengths, gauges to accommodate various anatomical locations and differences in patients' anatomy. The LCCS VC-S RF Cannula is a stainless steel cannula with an insulated shaft having one exposed tip and one side port located at the end of the cannula to diffuse the anesthesia closer to the lesion site and deliver the RF energy to the tissue. When RF Electrode was inserted into LCCS VC-S RF Cannula and came out from the side port, both active tip and RF Electrode section stuck out of side port can deliver Radio Frequency energy. Please refer to Figure 5-1 RF energy outlet indicator diagram.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for LCCS, which is a green color. The logo consists of a stylized diamond shape on the left, followed by the letters "LCCS" in a sans-serif font. A registered trademark symbol is located to the right of the letters. The logo is simple and modern, and the green color suggests a connection to nature or sustainability.
This RF cannula is provided as a sterile, single use, disposable device. This RF Cannula is offered in a variety of length and Needle outside Dimension to accommodate various anatomical locations and differences in patient anatomy.
RF energy outlet
RF energy outlet
Figure 5-1 RF energy outlet indicator diagram
The LCCS VC-S RF Cannula is consisted of:
- Protection Sleeve a)
- Needle Tube b)
- Insulating layer c)
- Needle hub d)
- Stylet e)
- Stylet Hub f)
- 6. Intended Use/Indications for Use:
The LCCS VC-S RF Cannula is intended for use in RF heat lesion procedures for the relief of pain.
- 7. Technology:
The LCCS VC-S RF Cannula employs the same fundamental scientific technology as its predicate devices.
LCCS VC-S RF Cannula is the same usage as its predicated devices, which is used in conjunction with RF denervation probes and RF Generator to create Radiofrequency (RF) Lesions of nerve tissue or for use in percutaneous nerve blocks.
Radiofrequency heat lesion procedure is a common technique used in the treatment of chronic pain. The LCCS VC-S RF Cannula and its predicate devices are offered in a variety of lengths, gauges to accommodate various anatomical locations and differences in patients' anatomy. The LCCS VC-S RF Cannula is a stainless steel cannula with an insulated shaft having one exposed tip and one side port located at the end of the cannula to diffuse the anesthesia closer to the lesion site and deliver the RF energy to the tissue. The tip and the side port constitute Y-shaped energy channels
{5}------------------------------------------------
and RF energy can be delivered both from these two channels.
- 8. Determination of Substantial Equivalence:
Comparison to Predicate Devices:
Below table is the summary comparison of features of the LCCS VC-S RF Cannula and the predicate devices.
| Feature | Proposed Device:<br>LCCS VC-S RF<br>Cannula | Predicate Device:<br>LCCS Insulated Spinal<br>Needle (RF Cannula)<br>(K112231) | Predicate Device:<br>Stryker® Venom™<br>Electrodes and Cannulae<br>(K123178) | Discussion of<br>Differences |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended/<br>Indication<br>s for use | The LCCS VC-S RF<br>Cannula is intended for<br>use in RF heat lesion<br>procedures for the<br>relief of pain. | LCCS Insulated Spinal<br>Needle (RF Cannula) is<br>indicated for use in RF<br>heat lesion procedures for<br>the relief of pain. | The Stryker RF electrodes<br>and cannulae, in combination<br>with the Stryker RF<br>Generator/Multigen, are<br>intended for coagulation of<br>soft tissues in orthopedic,<br>spinal, and neurosurgical<br>applications. These products<br>are also used for selective<br>denervation and tissue<br>destruction procedures which<br>may be performed on the<br>lumbar, thoracic, and cervical<br>regions of the peripheral<br>nerves, and nerve roots for the<br>relief of pain. Examples<br>include, but are not limited to,<br>Facette Denervation,<br>Trigeminus Neuralgia,<br>Peripheral Neuralgia and<br>Rhizotomy. | Identical |
| Material<br>type | Stainless Steel | Stainless Steel | Stainless Steel | Identical |
| Needle<br>Gauge | 18G, 20G, 22G | 18G, 20G, 22G | 18G, 20G | Identical. The<br>Needle Gauge of<br>proposed device is<br>identical to the<br>predicate device:<br>LCCS Insulated<br>Spinal Needle (RF<br>Cannula)<br>(K112231) |
| Needle<br>Tube<br>Length | 50mm,<br>100mm,<br>150mm | 50mm,<br>100mm,<br>150mm | 100mm,<br>150mm | Identical. The<br>Needle Tube Length<br>of proposed device<br>is identical to the<br>predicate device:<br>LCCS Insulated<br>Spinal Needle (RF<br>Cannula)<br>(K112231) |
| Feature | Proposed Device:<br>LCCS VC-S RF<br>Cannula | Predicate Device:<br>LCCS Insulated Spinal<br>Needle (RF Cannula)<br>(K112231) | Predicate Device:<br>Stryker® Venom™<br>Electrodes and Cannulae<br>(K123178) | Discussion of<br>Differences |
| Structure | A bevel tip and one<br>side port locate at the<br>distal end of the<br>cannula. There are<br>two channels to<br>deliver the RF energy. | A bevel tip locates at the<br>distal end of the cannula.<br>There is no side port and<br>only one channel to<br>deliver the RF energy. | A bevel tip and a side port are<br>created at the distal end of the<br>cannula. | Identical. The<br>proposed device is<br>identical to the<br>predicate device:<br>Stryker® Venom™<br>Electrodes and<br>Cannulae (K123178) |
| Method of<br>Sterilization | Single use, Ethylene<br>Oxide sterilization | Single use, Ethylene<br>Oxide sterilization | Single use, Ethylene Oxide<br>sterilization | Identical |
| Biocompati<br>bility | Non-Cytotoxicity<br>Non-Acute Systemic<br>Toxicity<br>Non-Intracutaneous<br>Reactivity<br>Non-Material Pyrogen<br>Non- Skin<br>Sensitization | Non-Cytotoxicity<br>Non-Intracutaneous<br>Reactivity<br>Non- Skin Sensitization | Unknown | Equivalent. The<br>subject device is also<br>comply with ISO<br>10993-11:2012 with<br>non-Material Pyrogen<br>and non- Non-Acute<br>Systemic Toxicity. |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for LCCS. The logo consists of a green diamond shape on the left, with the letters "LCCS" in green to the right of the diamond. There is a small "R" in a circle in the upper right corner of the logo, indicating that the logo is a registered trademark.
From the comparison summary table, the LCCS VC-S RF Cannula is substantially equivalent to the predicated devices with regard to intended use, technological characteristics, safety and effectiveness.
- . The devices are all intended for used in conjunction with the RF denervation Probes for the use in RF heat lesion procedures.
- The devices are all offered in a variety of lengths, gauges to accommodate ● various anatomical locations and differences in patient anatomy.
- The proposed device and predicated devices are all have the same material of the needle, the needle of the devices are all made of stainless steel.
- The proposed device has the same structure of a bevel tip and a side port at the distal end of the cannula with its predicated device Stryker® Venom™ Electrodes and Cannulae (K123178). The proposed device and this predicated device both can provide two channels to deliver the RF energy.
- The devices are all offered with single use, and the methods of sterilization ● are all Ethylene Oxide sterilization.
- 9. Summary of Non-Clinical Tests:
LCCS VC-S RF Cannula has been evaluated for insulation inspection, biocompatibility, accelerated aging test as well as shelf life, sterilization, and mechanical safety, and has been found to conform to applicable medical device safety standards. The LCCS VC-S RF Cannula complies with below voluntary standards:
{7}------------------------------------------------
- 1. AAMI/ANSI ES 60601-2-2, Medical electrical equipment Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
- 2. ISO 7864. Sterile hypodermic needles for single use Requirements and test methods
- 3. ISO 10993-1, Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
- 4. ANSI AAMI ISO 11607-1, Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
- 5. ANSI AAMI ISO 11135, Sterilization of health care products Ethylene oxide -Requirements for development, validation and routine control of a sterilization process for medical devices.
- 6. ISO 14971, Medical devices Applications of risk management to medical devices
- 7. ASTM F1980-16, Sterile hypodermic needles for single use Requirements and test methods
The following quality assurance measures are applied to the development of the system:
- . Risk Analysis
- Performance testing (Verification)
- Safety testing (Verification) ●
The LCCS VC-S RF Cannula is provided with sterilization and is biocompatible.
| Components | Test | Results | Conclusions |
|---------------------------------------|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| Needle tube,<br>Stylet,<br>Needle Hub | Cytotoxicity | The extract did not show potential toxicity<br>to L-929 cells. | Non-Cytotoxic |
| | Acute Systemic<br>Toxicity | The sample extracts showed no significant<br>evidence of causing acute systemic toxicity<br>in the mouse. | Non-Acute<br>Systemic<br>Toxicity |
| | Sensitization | The sample extract showed no significant<br>evidence of causing skin sensitization in the<br>guinea pig. | Non-<br>Sensitization |
| | Intracutaneous<br>Reactivity | The test result showed that the extract of the<br>test article did not induce Intracutaneous<br>reactivity in rabbit. | Non-<br>Intracutaneous<br>Reactivity |
| | Pyrogen | The animals' body temperature changed<br>less than 0.5 degree C. The test article met<br>the requirement of the absence of pyrogens. | Non- Pyrogen |
The Biocompatibility test details information as below:
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for LCCS, which is a green diamond shape with the letters LCCS in green next to it. The letters are in a sans-serif font. There is a registered trademark symbol next to the letters. Below the logo is the text "10. Conclusion".
LCCS Products Limited considers the LCCS VC-S RF Cannula to be as safe, as effective, and performance is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.