The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.
Device Story
Reusable NIBP cuff accessory; used with noninvasive blood pressure measurement systems. Device consists of cuff with bladder and single-tube air hose. Operation involves wrapping bladder around patient limb, securing with hook-and-loop closure, and connecting air hose to blood pressure monitor. Used in clinical settings by healthcare providers. Output is pneumatic pressure transmission to monitor for blood pressure calculation. Benefits include reliable, reusable interface for patient blood pressure monitoring.
Clinical Evidence
Bench testing only. Biocompatibility testing (cytotoxicity, sensitization, irritation) performed per ISO 10993-1, ISO 10993-5, and ISO 10993-10. Performance testing conducted per ISO 81060-1 (non-invasive sphygmomanometers) and leakage testing (< 4 mmHg/min).
Technological Characteristics
Reusable cuff with bladder and single-tube air hose. Materials: CK-XT-78211-XXX series (PU synthetic leather cuff, TPU bladder/hose, nylon hook/loop); CK-XT-78243-XXX series (nylon cuff, TPU bladder, silica gel hose, nylon hook/loop). Pressure range 0-300 mmHg. Non-sterile. Complies with ISO 81060-1, ISO 10993-5, and ISO 10993-10.
Indications for Use
Indicated for use as an accessory with noninvasive blood pressure measurement systems for neonate, infant, child, and adult patient populations.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 17, 2019
Shenzhen Changke Connect Electronics Co., Ltd. % Kevin Wang Consultant Chonconn Medical Device Consulting Co., Ltd. No. A415, Block A. NanShan Medical devices Industrial Park Nanshan District Shenzhen, 518067 CHINA
Re: K191253
Trade/Device Name: Reusable NIBP Cuff Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXQ Dated: August 29, 2019 Received: August 29, 2019
Dear Kevin Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K191253
Device Name Reusable NIBP Cuff
Indications for Use (Describe)
The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.
| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right:5px;"> <span style="border: 1px solid black; padding: 0px; display: inline-block; width: 13px; height: 13px; text-align: center;"> ✓ </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="padding-right:5px;"> <span style="border: 1px solid black; padding: 0px; display: inline-block; width: 13px; height: 13px; text-align: center;"> </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
# 510(K) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
### Prepared Date: 2019/04/08
#### 1. Submission sponsor
Name: Shenzhen Changke Connect Electronics Co., Ltd. Address: A2-4th floor of Xiang dali Technology Park, No.87 of HengPing Road, Henggang, Longgang District, Shenzhen, P.R. China Contact person: Yahui Zhou Title: General manager E-mail: zhouyahui(@szcklt.com
Tel: +86 136 1301 2560
#### 2. Submission correspondent
Name: Chonconn Medical Device Consulting Co., Ltd.
Address: No. A415, Block A, NanShan Medical devices Industrial Park Nanshan District, Shenzhen, Guangdong, P.R. China 518067
Contact person: Kevin Wang
E-mail: kevin(@chonconn.com
Tel: +86-755 33941160
| Trade/Device Name | Reusable NIBP Cuff |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | CK-XT-78211-001, CK-XT-78211-003, CK-XT-78211-005, CK-XT-78211-006,<br>CK-XT-78211-007, CK-XT-78211-008, CK-XT-78243-000, CK-XT-78243-001,<br>CK-XT-78243-003, CK-XT-78243-005, CK-XT-78243-007, CK-XT-78243-008 |
| Common Name | Non-invasive Blood pressure cuff |
| Regulatory Class | Class II |
| Classification | 21CFR 870.1120 / Blood pressure cuff / DXQ |
| Submission type | Traditional 510(K) |
#### Subject Device Information 3.
#### 4. Predicate Device
By submission of the Traditional 510(k), Shenzhen Changke Connect Electronics Co., Ltd. is requesting clearance for Reusable NIBP Cuff. It is comparable to the following legally marketed system:
- Shenzhen Caremed Medical Technology Co., Ltd. Reusable Blood Pressure Cuff under 1. K182433.
The subject device has same intended use, same target patient population, same performance effectiveness, performance safety as the predicate devices and no question is raised regarding to effectiveness and safety.
{4}------------------------------------------------
So, the conclusion is that the subject device is substantial equivalent to the predicate.
#### 5. Device Description
The subject device is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is available in neonatal, infant, child and adult sizes.
The subject device includes reusable blood pressure cuff. All the reusable blood pressure cuff has same structure, which contains Cuff with bladder and Air Hose. Air hose has single tube.
The reusable blood pressure cuff is reusable device, Model CK-XT-78211-XXX series are made of PU synthetic leather (Cuff) and TPU (Air Hose), and the PU synthetic leather (Cuff) is the material used to contacting with the patient's intact skin. Model CK-XT-78243-XXX series are made of Nylon (Cuff) and silica gel (Air Hose), and the Nylon (Cuff) is the material used to contacting with the patient's intact skin.
#### 6. Intended use & Indication for use
The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.
| | 7. Comparison to the Predicate Device | | |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Features | Subject Device<br>Reusable NIBP Cuff | Predicate Device K182433<br>Reusable Blood Pressure Cuff | Remark |
| Applicant | Shenzhen Changke Connect<br>Electronics Co., Ltd. | Shenzhen Caremed Medical<br>Technology Co., Ltd. | / |
| Classification<br>Regulation | 21CRF 870.1120 | 21CRF 870.1120 | Same |
| Classification<br>and Code | Class II,<br>DXQ | Class II,<br>DXQ | Same |
| Common<br>name | Non-invasive Blood Pressure<br>Cuff | Non-invasive Blood Pressure<br>Cuff | Same |
| Intended use | The reusable blood pressure cuff is<br>an accessory used in conjunction<br>with noninvasive blood pressure<br>measurement systems. The cuff is<br>non-sterile and may be reused. It is<br>available in neonate, infant, child<br>and adult sizes. The cuff is not<br>designed, sold, or intended for use<br>except as indicated. | The reusable blood pressure cuff is<br>an accessory used in conjunction<br>with noninvasive blood pressure<br>measurement systems. The cuff is<br>non-sterile and may be reused. It is<br>available in neonate, infant, child<br>and adult sizes. The cuff is not<br>designed, sold, or intended for use<br>except as indicated. | Same |
| Patient<br>Populations | Adults/Pediatrics | Adults/Pediatrics | Same |
| Tube Number | One | One or two | Different |
| | | | (1) |
| Principles of<br>Operation | Bladder is wrapped around the<br>patient's limb and secured by hook<br>and loop closure Air hose is<br>connected to the noninvasive blood<br>pressure measurement systems | Bladder is wrapped around the<br>patient's limb and secured by hook<br>and loop closure Air hose is<br>connected to the noninvasive blood<br>pressure measurement systems | Same |
| Limb<br>Circumference<br>(Range in cm) | Conform to AHA bladder sizes<br>recommendations<br>CK-XT-78211-XXX Series:<br>Infant (10-15cm)<br>Child (14-21.5cm)<br>Small Adult (20.5-28cm)<br>Adult (27-35cm)<br>Large Adult (34-43cm)<br>Adult Thigh (42-54cm)<br><br>CK-XT-78243-XXX Series:<br>Neonatal (6-11cm)<br>Infant (10-19cm)<br>Child (18-26cm)<br>Adult (25-35cm)<br>Large Adult (33-47cm)<br>Adult Thigh (44-66cm) | Conform to AHA bladder sizes<br>recommendations<br>Neonatal (6-11cm)<br>Infant (10-19cm)<br>Child (18-26cm)<br>Small Adult (20-28cm)<br>Adult (25-35cm)<br>Adult Long (25-35cm)<br>Large Adult (33-47cm)<br>Adult Thigh (44-66cm) | Different<br>(2) |
| Pressure Range | 0-300 mmHg | 0-300 mmHg | Same |
| Sterility | Non-sterile | Non-sterile | Same |
| Max. Leakage | < 4mm Hg/ min. | < 4mm Hg/ min. | Same |
| Material | CK-XT-78211-XXX Series:<br>Cuff (Patient contacted): PU<br>Synthetic Leather<br>Bladder: TPU<br>Tubing: TPU<br>Hook: Nylon<br>Loop: Nylon<br><br>CK-XT-78243-XXX Series:<br>Cuff (Patient contacted): Nylon<br>Bladder: TPU<br>Tubing: Silica gel<br>Hook: Nylon<br>Loop: Nylon | Cuff (Patient contacted): PU<br>Synthetic Leather<br>Bladder: Transparent Polyurethane<br>(TPU Film)<br>Tubing: PVC<br>Hook: Molded Nylon<br>Loop: Nylon | Different<br>(3) |
| y | No sensitization observed;<br>Negligible (no observed primary<br>irritation) | No sensitization observed;<br>Negligible (no observed primary<br>irritation) | |
### Comparison to the Predicate Device 7
{5}------------------------------------------------
{6}------------------------------------------------
## Justification of differences:
Justifications for differences between subject device and the predicate device are shown as below:
Different (1): The predicate device includes single tube configuration. All models of subject device use single tube. This difference does not raise any safety or effectiveness issue. Different (2): The limb circumference of subject device is included in the range of the predicate device. And it is disclosed in user manual clearly for selection. Therefore, this difference does not raise any safety or effectiveness issue.
Different (3): The material of subject device and predicate device is different, but both of them comply with ISO 10993-5 and ISO 10993-10. Therefore, this difference does not result in any safety and effectiveness issue.
### Conclusion:
The subject devices share the same indications for use, device operation, overall technical and functional capabilities, meet the same standards and requirements and therefore, are substantially equivalent to the predicate device(s).
#### 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility testing
The biocompatibility evaluation for the subject device was conducted in accordance with the FDA Biocompatibility guidance, 2016 (Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process") and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- . Cytotoxicity
- Sensitization .
- Irritation .
The subject devices are considered surface contacting for a duration of exceed 24 hours but not 30 days.
### Non-clinical data
Non-clinical tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:
{7}------------------------------------------------
- . ISO 81060-1, Non-Invasive Sphygmomanometers - Part 1: Requirements and Test Methods for Non-Automated Measurement Type, First Edition 2007.
- ISO 10993-5: 2009 Biological Evaluation of Medical Devices Part 5: Tests for In Vitro . Cytotoxicity.
- . ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization
The test was selected to show substantial equivalence between the subject device and the predicate.
#### 9. Conclusion
It has been shown in this 510(k) submission that the difference between the proposed devices and the predicate devices do not raise any questions regarding safety and effectiveness. Performance testing and compliance with voluntary standards demonstrate that the proposed are substantially equivalent to the relevant aspects of the predicate devices in terms of design, components, materials, principals of operation, biocompatibility, performance characteristics, and intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.