K191220 · Motus GI Medical Technologies , Ltd. · FDF · Jun 6, 2019 · Gastroenterology, Urology
Device Facts
Record ID
K191220
Device Name
Pure Vu System
Applicant
Motus GI Medical Technologies , Ltd.
Product Code
FDF · Gastroenterology, Urology
Decision Date
Jun 6, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pure Vu System is intended to connect to standard colonoscopes to help facilitate intra-procedural cleaning of a poorly prepared colon by irrigating or cleaning the colon and evacuating the irrigation fluid (water), feces and other bodly fluids and matter, e.g. blood. It is for use only by trained medical personnel located in hospitals, clinics and doctor offices.
Device Story
Pure-Vu System attaches to standard/slim colonoscopes to facilitate intra-procedural colon cleansing. System comprises reusable Workstation (WS), disposable Oversleeve (OS), and Umbilical Section (US). WS supplies irrigation (water/saline) and gas; evacuates fecal matter/fluids via OS suction channels. Physician operates via foot pedal to control cleansing, suction, and purging. Modifications include reduced WS size/weight, simplified evacuation algorithm, and improved OS interface. OS features extended hydrophilic coating (80cm) and 80A durometer polyurethane construction. Loading fixture and sealing clamp/plug redesigned for improved usability. Device enables cleaning without removing tools from endoscope working channel. Benefits include improved visualization in poorly prepared colons, potentially reducing procedure time or need for repeat colonoscopies.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including environmental conditioning, shelf-life simulation, transport simulation (ASTM D4169-16), biocompatibility (ISO 10993-5), and electrical safety/EMC (IEC 60601). Usability validation performed with 3 physicians and 15 nurses.
Technological Characteristics
Components: Reusable WS, disposable OS/US. Materials: 80A durometer polyurethane (OS). Dimensions: OS length 167cm; WS 250mm x 476mm x 358mm. Energy: 100V-240V electrical input. Principle: Irrigation/suction via external WS connected to colonoscope-mounted OS. Connectivity: Standalone. Sterilization: Clean, non-sterile. Software: Embedded control algorithm for fluid/gas delivery and suction.
Indications for Use
Indicated for intra-procedural cleaning of a poorly prepared colon in patients undergoing colonoscopy. For use by trained medical personnel in hospitals, clinics, and doctor offices.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
June 6, 2019
Motus GI Medical Technologies Ltd. % Randy Prebula Partner Hogan Lovells US LLP 555 Thirteenth Street NW Washington, DC 20004
K191220 Re:
Trade/Device Name: Pure Vu System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FDF Dated: May 4, 2019 Received: May 7, 2019
Dear Randy Prebula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shanil P. Haugen, Ph.D. Acting Assistant Division Director DHT3A: Division of Renal. Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K191220
Device Name Pure Vu System
Indications for Use (Describe)
The Pure Vu System is intended to connect to standard colonoscopes to help facilitate intra-procedural cleaning of a poorly prepared colon by irrigating or cleaning the colon and evacuating the irrigation fluid (water), feces and other bodly fluids and matter, e.g. blood.
It is for use only by trained medical personnel located in hospitals, clinics and doctor offices.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><div><span style="font-size: 20px;">☑</span></div><div>Prescription Use (Part 21 CFR 801 Subpart D)</div></div> |
| <div style="display:flex; align-items:center;"><div><span style="font-size: 20px;">☐</span></div><div>Over-The-Counter Use (21 CFR 801 Subpart C)</div></div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for "PURE-VU by MOTUS GI". The logo features a circular design on the left, composed of multiple blue dots arranged in a spiral pattern. To the right of the circular design, the text "PURE-VU" is displayed in a stylized, sans-serif font, with the letters rendered in a gradient of blue shades. Below "PURE-VU", the text "by MOTUS GI" is written in a smaller, sans-serif font, with the "MOTUS" portion in a darker gray and the "GI" portion in a lighter gray.
## 510(k) Summary
This special 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92.
| Applicant Information: | Motus GI Medical Technologies Ltd.22 Keren<br>Ha'yesod Str.<br>Tirat Carmel, 3902638Israel<br>Tel.: +972-4-6214446<br>Fax: +972-4-6214442 |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Ravit Peled |
| Phone Number: | +972-4-6214447 |
| Fax Number: | +972-4-6214442 |
| Establishment<br>Registration #: | 3011816755 |
| Date Prepared: | May 4th, 2019 |
| Trade Name(s): | Pure Vu System |
| CommonName: | Pure Vu System |
| Classification Name: | Endoscope and accessories |
| Classification: | Regulation No: 876.1500 Class: II<br>Panel: Gastroenterology and Urology |
| Predicate Device(s): | Pure Vu System (K181437) |
| Intended Use: | The Pure-Vu System is intended to connect to standard colonoscopes to<br>help facilitate intra-proceduralcleaning of a poorly prepared colon by<br>irrigating or cleaning the colon and evacuating the irrigation fluid (water),<br>feces and other bodily fluids and matter, e.g. blood.<br>It is for use only by trained medical personnel located in hospitals,<br>clinics, and doctors' offices. |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for "PURE-VU by MOTUS GI". The logo features a circular design made up of small blue dots on the left side. To the right of the dots, the words "PURE-VU" are written in a stylized blue font. Below "PURE-VU", the words "by MOTUS GI" are written in a smaller, gray font.
| Reason for Submission: | The purpose of this special 510(k) is to improve the overall ease of<br>use of the Pure-Vu system by reduce the size and weight of the Work<br>Station (WS) for better access in the procedure room and easier<br>mobility. |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The fundamental scientific technology of the device remains<br>unchanged. Modifications have been made as listed below: |
| | • Pure-Vu Workstation modifications to improve the usability of<br>the system. The key improvements are reducing the size and<br>weight of the WS, simplifying the evacuation algorithm during<br>cleansing and improve the interface with the disposable<br>Oversleeve. The embedded software was updated to comply<br>with the WS design modifications. |
| | • Disposable Oversleeve modifications include extending the<br>length of the hydrophilic lubricious coating from 60 cm to 80<br>cm on the distal end of the Pure-Vu Oversleeve to enhance<br>advancement, Outer and inner sleeves are made of 80A<br>durometer polyurethane as opposed to combination of 80A<br>and 70A durometer polyurethane and added an In-Line<br>connector to ease management and assembly of the<br>Umbilical tubing. All manufacturing locations and key<br>processes remain the same. |
| | • The loading fixture was modified to reduce its size and weight.<br>The sealing clamp and plug used on the distal end of the<br>device was modified to be a one-piece assembly that is now<br>disposable to improve ease of loading. No new materials were<br>introduced and there were no modifications to the compressor<br>box (the only electrical assembly part of the fixture) that<br>generates the inflation pressure used to open the sleeve for<br>loading. In addition, the unloading process is performed with<br>the disposable sealing clamp as opposed to an off the shelf<br>Kelly clamp used for unloading in the predicate device. |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for "PURE-VU by MOTUS GI". The logo features a circular design on the left, composed of multiple blue dots arranged in a spiral pattern. To the right of the circular design is the text "PURE-VU" in a bold, sans-serif font, with the letters in varying shades of blue. Below "PURE-VU" is the text "by MOTUS GI" in a smaller, lighter font.
| Technological<br>Characteristics: | The Pure-Vu System enables colon cleansing during colonoscopy using a standard or slim colonoscope with a length of 1630mm – 1710mm and a slim colonoscope outer diameter range of 11.7mm – 13.3mm and a standard colonoscope outer diameter range of 12.8mm – 13.7mm. The Oversleeve, which fits over the colonoscope and is connected to an external Workstation, generates fluid to break up feces. The fecal matter & fluids are removed through the suction channels of the Oversleeve into an external waste receptacle.<br>The Pure-Vu System consists of the following main components:<br>Oversleeve (OS) and Umbilical Section (US) - The disposable Oversleeve is mounted on Standard or Slim commercially available colonoscopes to allow a physician to cleanse the colon and is connected to the external Workstation via a disposable US. Workstation (WS) – The Workstation [WS] is reusable and supplies an irrigation mixture of water/saline and gas, and evacuates fecal material & fluids. The Workstation [WS] includes the following components: A monitoring & Control Unit that controls the delivery of irrigation fluids and gas into the colon, and suction of fluids and feces from the colon. Inlet Module that includes pumps and regulators enabling fluid & gas flow into the cleansing device. Outlet Module that includes pumps to evacuate fecal matter and fluids from the colon. A foot pedal activates the cleansing, suction and purging function, and switch between cleansing modes used by the physician. Loading fixture- The loading fixture is reusable and aids in assembling the Oversleeve onto the colonoscope. Unloading aids - A disposable 60cc lure lock syringe and 3-way stopcock used in unloading the Oversleeve from the colonoscope are provided in the disposable packaging. |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for "PURE-VU by MOTUS GI". The logo is primarily blue, with a circular pattern of blue dots on the left side. The text "PURE-VU" is in a sans-serif font, with a gradient effect from dark to light blue. "by MOTUS GI" is in a smaller font size and is located below the "PURE-VU" text.
| Performance Data: | Verification and Validation Testing tests were conducted for all modifications<br>to the Pure Vu System component as follows: |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 1) Pure- Vu Disposable parts (Oversleeve and Workstation connector) |
| | All disposable units undergone preconditioning simulations tests performed by<br>Carmel Environmental Laboratory (Israel) including but not limited to<br>Environmental conditioning and Shelf life simulation to demonstrate that the<br>Pure Vu Disposables functionality meets the requirements following a<br>simulated aging of one year. |
| | Simulation (bench test) was conducted and to support the sub system and<br>system's use as intended and its substantial equivalence as follows:<br>Dimensions Test Steering (Angulation) System Test Pressure Test Bond Strength Test |
| | Disposable package undergone environmental conditioning and transport<br>simulation (by DDL labs) in order to demonstrate efficiency and integrity of the<br>packaging and their accompanying labels following the simulation.<br>In addition, biocompatibility testing was done to determine if there was any<br>toxicological risk to the patient as per ISO 10993-5. |
| | 2) Pure-Vu Workstation (WS)<br>The Modified WS undergone the following tests:<br>Environmental conditions and distribution cycle simulation performed<br>by Hermon Laboratory (Israel) as per ASTM D4169-16 along with<br>shipping integrity. After the preconditioning simulations, the Pure Vu<br>WS undergone verification testing including visual inspection and<br>functionality tests. The embedded updated software to comply with the WS design<br>modifications was tested via software validation. Safety and EMC tests per IEC 60601 Usability and validation testing for the WS and disposable device has<br>been performed in two separate studies with 3 physicians and 15<br>nurses. |
| | 3) Pure-Vu Loading and Unloading<br>Validation testing for the new sealing plug used during loading and<br>unloading was performed. Environmental conditions and distribution cycle simulation performed<br>by DDL (USA) as per ASTM D4169-16 along with shipping integrity.<br>After the preconditioning simulations, the Pure Vu WS undergone<br>verification testing including visual inspection and functionality tests. |
| | Design verification and validation testing concluded that the design changes<br>have no impact on the Pure-Vu System performance. |
| | |
L
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for "PURE-VU by MOTUS GI". The logo is in blue and gray. To the left of the text is a circular design made up of many small blue circles.
| Substantial<br>Equivalence<br>Discussion: | The Pure Vu System has the same intended use, indications, principles of operation<br>and similar technological characteristics as its predicate device. The differences<br>between the device and its predicate device due to the modifications as detailed in<br>this submission do not raise any new issues of safety or effectiveness.<br>Performance data demonstrate that the Pure-Vu System is substantially equivalent. |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | The Pure-Vu System is substantially equivalent to the predicate device.<br>The intended of use of this product meets the requirements of 21 CFR 801.4 |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for PURE-VU by MOTUS GI. The logo features a circular design on the left, composed of many small blue circles. To the right of the circular design is the text "PURE-VU" in a blue gradient color. Below the text "PURE-VU" is the text "by MOTUS GI" in a smaller font size.
| Characteristics Comparison: | | |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Modified Device | Predicate Device |
| Manufacturer | Motus GI Medical Technologies Ltd. | |
| Description | Pure-Vu System | |
| Intended Use | Same | The Pure-Vu System is intended to<br>connect to standard colonoscopes to<br>help facilitate intra-procedural cleaning of<br>a poorly prepared colon by irrigating or<br>cleaning the colon and evacuating the<br>irrigation fluid (water), feces and other<br>bodily fluids and matter, e.g. blood. |
| | | It is for use only by trained medical<br>personnel located in hospitals, clinics |
| Environment of Use | Same | Hospitals, clinics and doctors' offices |
| Prescriptive | Same | Yes, only trained medical personnel |
| Disposable | Same | Single patient, single use |
| Distal tip design | Same | Multi irrigation hole Two distal suction holes |
| Principle of<br>operation | Same | Distal attachment to an endoscope,<br>sleeve ensuring attachment along entire<br>length, suction and irrigation tubes<br>running along the endoscope, suction<br>and irrigation head at the distal tip.<br>Enables irrigation and suction at any time<br>during the procedure without removing<br>any tools, which may be inserted in the<br>endoscope's working channel. |
| Operational<br>Procedures | 1) Same<br>2) Same<br>3) Same | 1) Attachment to a standard and<br>Slim colonoscope<br>2) Intra-procedure colon cleansing<br>during standard colonoscopy<br>3) Evacuation of water and feces |
| System<br>Components | The Pure-Vu System consists of the<br>same main components as the<br>predicate device but each one has had<br>minor modifications. The detailed<br>information on the modifications doneplease refer to: Section 9: "Design<br>Control Activities" | 1) Pure-Vu Workstation<br>2) Pure-Vu Standard and Slim<br>colonoscope Oversleeves<br>3) Pure-Vu WS Connector<br>4) Pure-Vu Loading Fixture |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for "PURE-VU by MOTUS GI". The logo features a circular design on the left, composed of multiple blue dots arranged in a circular pattern. To the right of the circular design, the word "PURE-VU" is written in large, blue, sans-serif font. Below the "PURE-VU" text, the words "by MOTUS GI" are written in a smaller, gray, sans-serif font.
| Oversleeve outer<br>diameter | Same | 21 mm |
|---------------------------------------------------|----------------------------------------------------|-----------------------------------------------------------------|
| Irrigation &<br>suction system | Same | Irrigation: 4 nozzle x 0.7 mm²<br>Suction: 2 nozzles x 12.5 mm² |
| Disposable length | Same | 167 cm attached to colonoscope |
| Air / Water<br>pressure<br>specification<br>(bar) | Same<br>Same | Up to 23 psi<br>Suction specifications: - 0.5 Bar |
| Flow rate<br>(cc / min) | Same | Water - Up to 645 cc/min<br>Air - up to 1350 cc/min |
| Dimensions<br>Workstation | W250mm \ D476mm (516 with pump<br>head) \ H-358 mm | W460 \ D480 (520 with pump head) \ H340<br>mm |
| Weight Workstation | 18Kg | 25Kg |
| Electrical Input | Same | 100V-240V<br>50/60 Hz |
| Material | Same | Complies with ISO 10993 |
| Sterilization | Same | Clean, Non-sterile |
| Safety Standards | Same | Complies with:<br>• IEC 60601-1<br>• IEC 60601-1-2 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.