K191209 · Dongguan Mars Medical Products Co., Ltd. · DXQ · Jun 18, 2019 · Cardiovascular
Device Facts
Record ID
K191209
Device Name
Aneroid Type Blood Pressure Kit
Applicant
Dongguan Mars Medical Products Co., Ltd.
Product Code
DXQ · Cardiovascular
Decision Date
Jun 18, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1120
Device Class
Class 2
Indications for Use
The device is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure. The models MS-200, MS-200D and MS-200M are intended to be manually attached to a patient and manually inflated along with a manual method.
Device Story
Manual non-invasive aneroid sphygmomanometer; models MS-200 (wall), MS-200D (desk), MS-200M (mobile). Device consists of an inflation cuff, bladder, manual inflation bulb, and aneroid gauge. Operator wraps cuff around patient's upper arm; manually inflates/deflates using bulb. Gauge displays pressure readings. Used in clinical or home settings by medical professionals or patients. No automated processing; relies on manual auscultatory method (though stethoscope not included in subject models). Provides blood pressure monitoring capability; aids clinical decision-making through manual pressure assessment.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (ISO 10993-1, -5, -10) and performance testing (ISO 81060-1).
Technological Characteristics
Manual aneroid sphygmomanometer. Materials: Aluminum/ABS gauge, steel bracket, nylon/cotton cuff, PVC/Latex/Neoprene bladder/tubing, TPU coil tubing, chrome-plated aluminum connectors. Pressure range: 0-300 mmHg; Accuracy: ±3 mmHg. Standards: ISO 81060-1, ISO 10993 series. No software or energy source.
Indications for Use
Indicated for medical professionals or home users for manual measurement of systolic and diastolic blood pressure via upper arm cuff inflation/deflation.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 18, 2019
Dongguan Mars Medical Products Co., Ltd % Jinghua Zhou Regulation Control Manager Guangzhou Junyi Information Technology Co., Ltd. Room 215, Huaming Building, Chebei Road Guangzhou. 511660 CHINA
Re: K191209
Trade/Device Name: Aneroid Type Blood Pressure Kit Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXQ Dated: May 6, 2019 Received: May 6. 2019
Dear Jinghua Zhou:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K191209
Device Name Aneroid Type Blood Pressure Kit
Indications for Use (Describe)
The device is intended to be used by medical professionals or in the measurement of systolic and diastolic pressure. The models MS-200, MS-200M are intended to be manually attached to a patient and manually inflated along with a manual method.
| Y Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5 - 510(k) Summary
Date of Summary Preparation: April 30, 2019
### 1. Submitter's Identifications
Submitter's Name: Dongguan Mars Medical Products Co., Ltd Address: 381, Manfeng West Road, Manfeng Group, Dani Village, Shatian Town, Dongguan City, Guandgong 523993, China Contact Person: Lily Liao Contact Title: Sales representative Contact E-mail Address: lily@mars.com.tw Telephone: +86-769-85834233 Fax: +86-769-85834232
#### 2. Correspondent's Identifications
Correspondent's Name: Guangzhou Junyi Information Technology Co., Ltd. Address: Room 215, Huaming Building, Chebei Road, Guangzhou, P.R. China ZIP Code: 511660 Contact Person: Jinghua Zhou Contact Title: Regulation Control Manager Contact E-mail Address: admanzhou@126.com Telephone: +86-20-82329549 Fax: +86-20-82329549
#### 3. Name of the Device
Device Classification Name: Blood Pressure Cuff Product Name: Blood Pressure Cuff Trade Name: Aneroid Type Blood Pressure Kit Model: MS-200, MS-200D, MS-200M Classification Panel: Cardiovascular Product Code: DXQ Device Classification: Class II
### 4. The Predicate Devices
K780715 Aneroid Type Blood Pressure Kit
#### 5. Device Description
Aneroid Type Blood Pressure Kit, models MS-200D and MS-200M are a manual non-invasive aneroid sphygmomanometer which respectively uses an inflation cuff wrapped around the upper arm. The cuff is inflated and deflated by a manual inflation bulb.
The proposed device has three models, MS-200/MS-200D/MS-200M. The three models have same intended use, same design and measuring principle, same measuring range and same main components. The only differences among the three
{4}------------------------------------------------
models are assembly accessories for fixation and installation of the aneroid sphygmomanometer.
| Table 1 Model Differences Comparison | | | | |
|--------------------------------------|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| Model | Type | Main<br>components | Assembly accessories | Picture |
| MS-200 | Wall<br>mounted<br>type | | Same parts:<br>● Four (4) 1" Screws<br>● Four (4) drywall<br>anchors<br>● Two (2) ¼" lock<br>screws | Image: MS-200 |
| MS-200D | Desk<br>type | ● Aneroid<br>gauge<br>● Wall clock<br>mount<br>bracket | Different parts only<br>used for Desk type:<br>one lower bracket<br>M6 screw & washer<br>one base and base pad | Image: MS-200D |
| MS-200M | Mobile<br>type | ● Adult size<br>cuff with<br>bladder and<br>inflation bulb<br>● Metal leur<br>connector set<br>● Black<br>plastic<br>double-ended<br>male<br>connector<br>● Coil tubing<br>● Mini<br>screwdriver | ● Floor Stand<br>● Five(5)Caster<br>Wheels<br>● Metal Basket<br>● 3-piece Pole<br>Assembly(black pole,<br>silver pole, spring)<br>● Height Adjustment<br>Knob<br><b>Floor Stand<br/>Hardware</b><br>● One (1) 4.8/M8<br>Screw Bolt with 2<br>Washers<br>● Two (2) 3/8"<br>Screws with<br>Integrated Washers +<br>Two (2)<br>● Washers<br>● Two (2) ½" Screws<br><b>Wall Mount<br/>Hardware</b><br>Same as the “Same<br>parts” | Image: MS-200M |
# Table 1 Model Differences Comparison
{5}------------------------------------------------
#### Main components:
Aneroid gauge: indicate the measurement result
Wall clock mount bracket: fix the aneroid gauge
Adult size cuff with bladder and inflation bulb: the cuff is inflated and deflated by the manual inflation bulb
Metal leur connector set: connect the black plastic double-ended male connector and the rubber tube
Black plastic double-ended male connector: connect the metal leur connector set and the coil tubing
Coil tubing: connect the aneroid gauge and the black plastic double-ended male connector
Mini screwdriver: adjust and set to zero when the needle of the aneroid gauge is not within the zero square position
Assembly accessories: fix and install the aneroid sphygmomanometer as wall type, or desk type, or mobile type
#### Cuff types and dimensions:
The device includes an Adult sized nylon cuff. Additional cuff sizes are available through your dealer. The types and dimensions of the optional cuff are as below:
| Size | Dimensions<br>(inches) | Arm Girth<br>(inches) | Dimensions<br>(Centimeters) | Arm Girth<br>(Centimeters) |
|-------------|------------------------|-----------------------|-----------------------------|----------------------------|
| Infant | 3.54"W x 11.8"L | 5.12"-7.48" | 9.00cm W x 30.0cm L | 13.0cm-19.0cm |
| Child | 4.33"W x 14.2"L | 7.64"-10.5" | 11.00cm W x 36.0cm L | 19.4cm-26.7cm |
| Adult | 5.71"W x 21.3"L | 10.0"-16.0" | 14.50cm W x 54.0cm L | 25.4cm-40.6cm |
| Large Adult | 6.89"W x 24.6"L | 13.5"-20.0" | 17.50cm W x 62.5cm L | 34.4cm-50.8cm |
| Thigh | 8.46"W x 30.5"L | 16.0"-26.0" | 21.50cm W x 77.5cm L | 40.6cm-66.0cm |
#### Materials:
Aneroid gauge: the enclosure of gauge is made of Aluminum (bottom) and ABS (panel)
Wall clock mount bracket: steel
Adult size cuff with bladder and inflation bulb: the cuff outside layer is made of nylon cloth or cotton cloth (optional), and inside bladder is made of PVC, or Latex (Rubber), or Latex Free (Neoprene) (optional), and the inflation bulb is made of PVC or Latex (Rubber), the tubes connecting to the inside bladder is made of PVC, or Latex (Rubber), or Latex Free (Neoprene)
Metal leur connector set: chrome plated aluminum
Black plastic double-ended male connector: PVC
Coil tubing: TPU
Mini screwdriver: metal
#### 6. Intended Use of Device
The device is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure. The models MS-200, MS-200D and MS-200M are intended to be manually attached to a patient and manually inflated along with a manual method.
{6}------------------------------------------------
# 7. Summary of Substantial Equivalence
# Table 2 Comparison to Predicate Device
| | Proposed Device | Predicate device | Comparison |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k Number | ------- | K780715 | ------- |
| Product Code | DXQ | DXQ | Same |
| Proprietary Name | Aneroid Type Blood Pressure Kit | Aneroid Type Blood Pressure Kit | Same |
| Model | MS-200, MS-200D, MS-200M | MS-115, MS-116, MS-117, MS-200, MS-200D, MS-200M | ------- |
| Manufacturer | Dongguan Mars Medical Products Co., Ltd | Dongguan Mars Medical Products Co., Ltd | ------- |
| Indications for Use | The device is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure. The models MS-200, MS-200D and MS-200M are intended to be manually attached to a patient and manually inflated along with a manual method. | The device is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure. The models MS-115, MS-116, MS-117 are intended to be manually attached to a patient and manually inflated along with a manual method and Auscultatory Korotkoff sounds method. The models MS-200, MS-200D and MS-200M are intended to be manually attached to a patient and manually inflated along with a manual method. | Same<br>The proposed device models are contained in those of the predicate device. And the measurement method of models MS-200/200D/200M is same. |
| Type of Use | Prescription Use | Over-The-Counter Use | Different<br>The difference does not raise safety and effectiveness of the |
5-4
{7}------------------------------------------------
| K191209 |
|-----------------------------------------|
| Dongguan Mars Medical Products Co., Ltd |
| | | | proposed device |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Operating Principle | Aneroid Type Blood Pressure Kit, models<br>MS-200, MS-200D and MS-200M are a<br>manual non-invasive aneroid<br>sphygmomanometer which respectively<br>uses an inflation cuff wrapped around the<br>upper arm. The cuff is inflated and<br>deflated by a manual inflation bulb. | Aneroid Type Blood Pressure Kit, models<br>MS-115, MS-116, MS-117, MS-200,<br>MS-200D and MS-200M are a manual<br>non-invasive aneroid sphygmomanometer<br>which respectively uses an inflation cuff<br>wrapped around the upper arm. The cuff is<br>inflated and deflated by a manual inflation<br>bulb. | Same |
| Basic Design | The Aneroid Type Blood Pressure Kit is a<br>non-invasive blood pressure<br>measurement system for monitoring<br>blood pressure levels. This<br>non-automated sphygmomanometer uses<br>an occluding cuff, an aneroid<br>sphygmomanometer to measure pressure. | The Aneroid Type Blood Pressure Kit is a<br>non-invasive blood pressure measurement<br>system for monitoring blood pressure levels.<br>This non-automated sphygmomanometer<br>uses an occluding cuff, an aneroid<br>sphygmomanometer to measure pressure<br>(applied for models<br>MS-115/116/117/200/200D/200M) and a<br>stethoscope for detecting Korotkoff sounds<br>(applied for models MS-115/116/117). | Same<br>The basic design of<br>the proposed device is<br>same as the predicate<br>device models<br>MS-200/200D/200M. |
| Measuring Method | Aneroid | MS-115/116/117: Aneroid / Auscultatory<br>MS-200/200D/200M: Aneroid | Same<br>The measuring<br>method of the<br>proposed device is<br>same as the predicate<br>device models<br>MS-200/200D/200M. |
| Inflation system | Manual inflation with air pump bulb | Manual inflation with air pump bulb | Same |
| Deflation system | Manual deflation | Manual deflation | Same |
| Main Components | Aneroid gauge<br>Wall clock mount bracket<br>Adult size cuff with bladder and inflation bulb<br>Metal leur connector set<br>Black plastic double-ended male<br>connector<br>Coil tubing<br>Mini screwdriver | MS-115/116/117:<br>Aneroid gauge<br>Inflation bulb<br>Binaural<br>Y-Tubing<br>Adult size cuff<br>Stethoscope<br>MS-200/200D/200M:<br>Aneroid gauge<br>Wall clock mount bracket<br>Adult size cuff with bladder and inflation bulb<br>Metal leur connector set<br>Black plastic double-ended male connector<br>Coil tubing<br>Mini screwdriver | Same<br>The main components of proposed device models are same as those models<br>MS-200/200D/200M of predicate device. |
| Materials | Aneroid gauge: the enclosure of gauge is made of Aluminum (bottom) and ABS (panel)<br>Wall clock mount bracket: steel<br>Adult size cuff with bladder and inflation bulb: the cuff outside layer is made of nylon cloth or cotton cloth (optional), and inside bladder is made of PVC, or Latex (Rubber), or Latex Free (Neoprene) (optional), and the inflation bulb is made of PVC or Latex (Rubber), the tubes connecting to the inside bladder is made of PVC, or Latex (Rubber), or Latex Free | MS-115/116/117: Aneroid gauge: the enclosure of gauge is made of steel<br>Binaural: stainless steel and nature latex<br>Y-Tubing: PVC<br>Adult size cuff: the cuff outside layer is made of nylon cloth or cotton cloth (optional), and inside bladder is made of PVC, or Latex (Rubber), or Latex Free (Neoprene) (optional), and the inflation bulb is made of PVC or Latex (Rubber)<br>Stethoscope: stainless steel<br>MS-200/200D/200M: Aneroid gauge: the | Same<br>The materials of the proposed device are same as those of the predicate device models<br>MS-200/200D/200M. |
{8}------------------------------------------------
K191209 Dongguan Mars Medical Products Co., Ltd
{9}------------------------------------------------
| K191209 |
|-----------------------------------------|
| Dongguan Mars Medical Products Co., Ltd |
| | (Neoprene)<br>aluminum<br>Black<br>connector: PVC<br>Coil tubing: TPU<br>Mini screwdriver: metal | Metal leur connector set: chrome plated<br>plastic<br>double-ended<br>male | enclosure of gauge is made of Aluminum<br>(bottom) and ABS (panel)<br>Wall clock mount bracket: steel<br>Adult size cuff with bladder and inflation<br>bulb: the cuff outside layer is made of nylon<br>cloth or cotton cloth (optional), and inside<br>bladder is made of PVC, or Latex (Rubber),<br>or Latex Free (Neoprene) (optional), and the<br>inflation bulb is made of PVC or Latex<br>(Rubber), the tubes connecting to the inside<br>bladder is made of PVC, or Latex (Rubber),<br>or Latex Free (Neoprene)<br>Metal leur connector set: chrome plated<br>aluminum<br>Black plastic double-ended male connector:<br>PVC | | | |
|---------------------|--------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|-----------------------------------------|--------------------------------------------------------|
| | | | Coil tubing: TPU<br>Mini screwdriver: metal | | | |
| Measuring<br>range | Pressure: 0 - 300 mmHg | | Pressure: 0 - 300 mmHg | | | Same |
| Accuracy | Pressure: ± 3 mmHg | | Pressure: ±3 mmHg | | | Same |
| Storage condition | -10°C ~ +60°C, 10% ~ 95% RH | | -10°C ~ +60°C, 10% ~ 95% RH | | | Same |
| Operating condition | +10°C ~ +40°C, 15% ~ 85% RH | | +10°C ~ +40°C, 15% ~ 85% RH | | | Same |
| Cuff dimension | Size<br>Infant<br>Child | Dimension<br>3.54"W x 11.8"L<br>9.00cm W x 30.0cm L<br>4.33"W x 14.2"L | Size<br>Infant<br>Child | Dimension<br>3.54"W x 11.8"L<br>9.00cm W x 30.0cm L<br>4.33"W x 14.2"L | | Same |
| | 11.00cm W x 36.0cm L | | | 11.00cm W x 36.0cm L | | |
| | Adult | 5.71"W x 21.3"L<br>14.50cm W x 54.0cm L | | Adult | 5.71"W x 21.3"L<br>14.50cm W x 54.0cm L | |
| | Large<br>Adult | 6.89"W x 24.6"L<br>17.50cm W x 62.5cm L | | Large<br>Adult | 6.89"W x 24.6"L<br>17.50cm W x 62.5cm L | |
| | Thigh | 8.46"W x 30.5"L<br>21.50cm W x 77.5cm L | | Thigh | 8.46"W x 30.5"L<br>21.50cm W x 77.5cm L | |
| | Size | Arm Girth | | Size | Arm Girth | |
| Arm Girth | Infant | 5.12" – 7.48"<br>13.0cm – 19.0cm | | Infant | 5.12" – 7.48"<br>13.0cm – 19.0cm | Same |
| | Child | 7.64" – 10.5"<br>19.4cm – 26.7cm | | Child | 7.64" – 10.5"<br>19.4cm – 26.7cm | |
| | Adult | 10.0" – 16.0"<br>25.4cm – 40.6cm | | Adult | 10.0" – 16.0"<br>25.4cm – 40.6cm | |
| | Large<br>Adult | 13.5" – 20.0"<br>34.4cm – 50.8cm | | Large<br>Adult | 13.5" – 20.0"<br>34.4cm – 50.8cm | |
| | Thigh | 16.0" – 26.0"<br>40.6cm – 66.0cm | | Thigh | 16.0" – 26.0"<br>40.6cm – 66.0cm | |
| Standard | ISO 81060-1 | | | ANSI/AAMI SP9 | | Similar<br>Standards have been<br>updated and replaced |
| | ISO 10993-1 | | | ISO 10993-1 | | |
| | ISO 10993-5 | | | ISO 10993-5 | | |
| | ISO 10993-10 | | | ISO 10993-10 | | |
{10}------------------------------------------------
K191209 Dongguan Mars Medical Products Co., Ltd
{11}------------------------------------------------
# 8. Substantial Equivalence discussion:
Comparing with the predicate device (K780715), Aneroid Type Blood Pressure Kit Model MS-115, MS-116, MS-117, MS-200, MS-200D and MS-200M, modified Aneroid Type Blood Pressure Kit Model MS-200, MS-200D and MS-200M have same product name, classification and product code, product physical design and operation, same aneroid gauge, packaging materials, performance parameter ranges, physical properties, mechanical properties, biocompatibility. The differences between those two devices are as following:
- · Type of use for the proposed device is Prescription Use, and the predicate device is OTC. The difference does not raise safety and effectiveness of the proposed device.
- · Performance standard complied with the proposed device is ISO 81060-1:2007, and the predicate device is ANSI/AAMI SP9. Standards have been updated and replaced, the proposed device have performed test complying with ISO 81060-1:2007.
## 9. Non-Clinical Testing
Non-clinical testing included biocompatibility and bench testing. The tests listed in the ISO10993 and ISO81060-1, were included. Results of testing were acceptable.
## 10. Clinical Study
No clinical study is included in this submission.
## 11. Conclusion:
Aneroid Type Blood Pressure Kit, models MS-200, MS-200D and MS-200M have the same intended use and similar technological characteristics as the above predicate devices. Moreover, information contained in this submission supplied demonstrates that any differences in their characteristics do not raise any new questions of safety or effectiveness. Thus, Aneroid Type Blood Pressure Kit, models MS-200, MS-200D and MS-200M are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.