← Product Code [MBI](/productcode/MBI) · K191204

# Infinity Fixation System (K191204)

_Conmed Corporation · MBI · Jul 3, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K191204

## Device Facts

- **Applicant:** Conmed Corporation
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Jul 3, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The InfinityTM Fixation System is intended to provide suspension fixation for soft tissue to bone in the repair of the natural ligament or tendon disruption or reconstruction of a ligament using soft tissue grafts. Examples of such procedures include anterior cruciate ligament, posterior cruciate ligament, medial collateral ligament, and lateral collateral ligament.

## Device Story

Infinity Fixation System comprises metal buttons and nonabsorbable Hi-Fi suture for suspensory fixation of tendon to bone during ligament reconstruction. Device used in orthopedic surgical procedures (arthroscopic or open) by surgeons. System provides mechanical fixation of soft tissue grafts to bone; surgeon manipulates button and suture via disposable driver to secure graft. Output is physical stabilization of soft tissue to bone, facilitating healing. Benefits include secure graft fixation during ligament repair. Provided sterile, single-use.

## Clinical Evidence

Bench testing only. Testing included reliability, ultimate fixation strength, cyclic testing, biocompatibility, shelf-life, transportation, pyrogen/bacterial endotoxin testing, and sterilization validation. No clinical data presented.

## Technological Characteristics

System consists of metal buttons and nonabsorbable Hi-Fi suture. Features include femoral button with pull tab, tibial buttons, free loop suture, and cradle. Employs adjustable loop and free loop techniques for suspensory fixation. Sterilized via ethylene oxide (EO).

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- ConMed Linvatec XO Button ([K070780](/device/K070780.md))
- GraftMax™ Button, ALB (Adjustable Loop Button) and GraftMax™ Button, BTB (Bone-Tendon-Bone) ([K151037](/device/K151037.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

ConMed Corporation Orjada Dervishleri Regulatory Affairs Specialist 525 French Road Utica, New York 13502

Re: K191204

Trade/Device Name: InfinityTM Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: June 21, 2019 Received: June 21, 2019

Dear Orjada Dervishleri:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

July 3, 2019

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Laurence Coyne, Ph.D. Assistant Director DHT6C: Division of Stereotaxic. Trauma and Restorative Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

Device Name InfinityTM Fixation System

### Indications for Use (Describe)

The InfinityTM Fixation System is intended to provide suspension fixation for soft tissue to bone in the repair of the natural ligament or tendon disruption of a ligament using soft tissue grafts. Examples of such procedures include anterior cruciate ligament, posterior cruciate ligament, medial collateral ligament, and lateral ligament.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span>      X     </span> </div> |
|----------------------------------------------|----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <div> <span> </span> </div>            |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Conmed Corporation. The logo consists of a blue square with a white curved shape inside, followed by the word "CONMED" in black, with a horizontal line underneath. Below the line, the word "CORPORATION" is written in smaller, gray letters.

# 510(k) SUMMARY

In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, CONMED Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) Number K191204

#### I. SUBMITTER

CONMED Corporation 11311 Concept Blvd Largo, Florida 33773

Phone: 727-399-5491 Fax: 727-399-5264

Contact Person: Orjada Dervishleri Date Prepared: May 3rd, 2019

#### II. DEVICE NAME

| Device Name:         | Infinity™ Fixation System                      |
|----------------------|------------------------------------------------|
| Classification Name: | Fastener, fixation, nondegradable, soft tissue |
| Regulatory Class:    | Class II, per 21 CFR Part 888. 3040            |
| Product Codes:       | MBI                                            |

#### III. PREDICATE/ LEGALLY MARKET DEVICE

| Device Name:  | ConMed Linvatec XO Button |
|---------------|---------------------------|
| Company Name: | ConMed Linvatec           |
| 510(k) #:     | K070780                   |

#### IV. PREDICATE DEVICE

| Device Name:  | GraftMax™ Button, ALB (Adjustable Loop Button) and<br>GraftMax™ Button, BTB (Bone-Tendon-Bone) |
|---------------|------------------------------------------------------------------------------------------------|
| Company Name: | ConMed Corporation                                                                             |
| 510(k) #:     | K151037                                                                                        |

#### V. DEVICE DESCRIPTION

The CONMED Infinity™ Fixation System is a family of metal buttons and nonabsorbable Hi-Fi suture for suspensory fixation of tendon to bone used in ligament reconstruction procedures. The Infinity™ Fixation System is provided individually packaged, single-use, sterile. The anchor, suture, and disposable driver are EO Sterilized.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Conmed Corporation. The logo features a blue abstract shape on the left, resembling a stylized "C". To the right of the shape is the company name "CONMED" in bold, black letters. Below the company name is the word "CORPORATION" in smaller, lighter font, underlined with a thin blue line.

#### VI. INTENDED USE/ INDICATIONS FOR USE

- VII. The Infinity™ Fixation System is intended to provide suspension fixation for soft tissue to bone in the repair of the natural ligament or tendon disruption or reconstruction of a ligament using soft tissue grafts. Examples of such procedures include anterior cruciate ligament, posterior cruciate ligament, medial collateral ligament, and lateral collateral ligament.
- COMPARISON OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE VIII. PREDICATE DEVICE

The following table represents a summary of the technological characteristics between the proposed and the predicate device.

|                                      | InfinityTM Fixation System<br>Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | ConMed Linvatec XO Button<br>Predicate Device | GraftMax Button, ALB with<br>Cradle, GraftMax Button,<br>BTB with Cradle<br>Predicate Device                                                                                                                                                                                                                                                                                                                                                |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description                   | The InfinityTM Fixation System is a family of<br>metal buttons and non-absorbable Hi-Fi<br>suture for suspensory fixation of soft tissue<br>grafts to bone used in ligament<br>reconstruction procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                          | Same                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Intended Use /<br>Indication for Use | The InfinityTM Fixation System is intended to<br>provide suspension fixation for soft tissue to<br>bone in the repair of the natural ligament or<br>tendon disruption or reconstruction of a<br>ligament using soft tissue grafts. Examples of<br>such procedures include anterior cruciate<br>ligament, posterior cruciate ligament, medial<br>collateral ligament, and lateral collateral<br>ligament.                                                                                                                                                                                                                                                                                               | Same                                          | The device may be used in<br>either arthroscopic or open<br>surgical procedures. After the<br>suture is anchored to the bone,<br>it may be used to reattach soft<br>tissue, such as ligaments,<br>tendons, or joint capsules to<br>the bone. The suture anchor<br>system thereby stabilizes the<br>damaged soft tissue, in<br>conjunction with appropriate<br>postoperative immobilization,<br>throughout the healing period.               |
| Contraindications                    | 1. Do not use the Infinity™ Fixation System<br>for fixation of bone-patellar tendon-bone<br>(BTB) grafts.<br>2. The Infinity™ Femoral Adjustable Loop<br>Button is contraindicated for fixation of<br>soft tissue in the tibial tunnel of ACL and<br>PCL repairs.<br>3. Insufficient quantity or quality of cortical<br>bone for fixation.<br>4. Blood supply limitations and/or previous<br>infections, which may tend to retard<br>healing.<br>5. Patients with active sepsis.<br>6. Conditions which tend to limit the<br>patient's ability or willingness to follow<br>directions during the healing period.<br>7. Foreign body sensitivity, known or<br>suspected allergies to implant materials. | Same                                          | 1.Insufficient quantity or<br>quality of cortical bone for<br>fixation.<br>2.Blood supply limitations<br>and/or previous infections,<br>which may tend to retard<br>healing.<br>3.Patients with active sepsis.<br>4.Conditions which tend to<br>limit the patient's ability or<br>willingness to follow<br>directions during the healing<br>period.<br>5.Foreign body sensitivity,<br>known or suspected allergies<br>to implant materials. |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Conmed Corporation. The logo features a blue square with a white abstract shape on the left. To the right of the square is the word "Conmed" in black, with the word "Corporation" in a smaller font size underneath. A thin blue line separates the two words.

| Components                       | Femoral button with pull tab<br>Tibial buttons (2)<br>Free loop Suture<br>Cradle                                                                                                                                     |                                                                                       | Same                                                                                            |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics | Providing Suspension Fixation<br>Pre-attached with sutures<br>Lead strand suture to manipulate and position<br>the button but not implanted<br>Adjustable Loop technique<br>Free Loop<br>Adaptor for the button body | Providing Suspension<br>Fixation<br>Pre-attached with sutures<br>Fixed Loop technique | Providing Suspension<br>Fixation<br>Pre-attached with sutures<br>Adaptor for the button<br>body |

#### IX. PERFORMANCE DATA

Testing has been completed to demonstrate that the Infinity™ Fixation System performs as intended and is substantially equivalent to the predicate device. Bacterial endotoxin testing was conducted and met the endotoxin limits.

Completed testing includes the following:

Verification Testing

- Reliability ●
- Ultimate Fixation Strength ●
- Transportation ●
- Cyclic
- Pyrogen ●
- Biocompatibility ●
- Shelf-life ●

Validation Testing

- User Validation ●
- Packaging ●
- Labeling ●
- Sterilization ●

- X. CONCLUSION
The Infinity™ Fixation System is either substantially equivalent or identical in design, materials, intended use, principles of operation, and technical characteristics to the predicate ConMed Linvatec XO Button. Based upon the findings of our performance testing, the differences present no new issues of safety and efficacy, and the Infinity™ Fixation System is substantially equivalent to the ConMed Linvatec XO Button (K070780).

---

**Source:** [https://fda-staging.innolitics.com/device/K191204](https://fda-staging.innolitics.com/device/K191204)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
