K191201 · Hiossen, Inc. · DZE · Nov 15, 2019 · Dental
Device Facts
Record ID
K191201
Device Name
EM SA Implant System
Applicant
Hiossen, Inc.
Product Code
DZE · Dental
Decision Date
Nov 15, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EM SA Implant System (DENTURE) is intended to be place in the bone of the upper or lower jaw arches providing support to dental prosthetic devices, specifically for denture stabilization to restore a patient's chewing function. It is intended for single use only. The EM SA Implant (NARROW RIDGE) is intended to be use in the treatment of missing mandibular central and lateral incisors to support dental prosthetic devices, such as artificial teeth, in order to restore chewing function in partially edentulous patients. It is intended for single use only. It is intended for delayed loading.
Device Story
EM SA Implant System consists of titanium alloy (Ti-6Al-4V) endosseous dental implants; surface treated via sandblasting and acid etching (SA). Implants are surgically placed into upper or lower jaw bone by dental professionals. Device acts as anchor for dental prosthetic devices; Denture version provides stabilization for dentures; Narrow Ridge version supports artificial teeth for missing mandibular incisors. System includes various thread diameters and lengths. Supplied sterile via radiation. Device is non-mechanical and non-active. Clinical benefit is restoration of chewing function in edentulous or partially edentulous patients. Healthcare providers use device to provide stable foundation for prosthetic restorations.
Clinical Evidence
No clinical data submitted. Substantial equivalence established via bench testing, material analysis, and comparison to predicate devices.
Technological Characteristics
Material: Titanium alloy Ti-6Al-4V (ASTM F 136). Surface: Sandblasted and Acid etched (SA). Design: Threaded body, one-body implant with integrated abutment head. Sterilization: Radiation. Packaging: Plastic ampule in Tyvek-lidded blister tray. Non-active, non-mechanical device.
Indications for Use
Indicated for denture stabilization in upper or lower jaw arches (Denture version) or treatment of missing mandibular central and lateral incisors in partially edentulous patients (Narrow Ridge version) to support dental prosthetics and restore chewing function. Intended for single use and delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 15, 2019
Hiossen, Inc. Peter Lee QA/RA Manager 85 Ben Fairless Drive Fairless Hills, Pennsylvania 19030
Re: K191201
Trade/Device Name: EM SA Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: November 14, 2019 Received: November 15, 2019
Dear Peter Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Hiossen Implant. The word "Hiossen" is written in a sans-serif font, with the "H" in red and the rest of the letters in gray. A red swoosh appears above the "H". Below the word "Hiossen" is the word "IMPLANT" in smaller, gray letters.
85 Ben Fairless Drive airless Hills PA 19030 888-768-0001 www.hiossen.com
# Section 5 Indication for Use Statement - 1 PAGE
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K191201 Device Name EM SA Implant System
Indications for Use (Describe)
The EM SA Implant System (DENTURE) is intended to be place in the bone of the upper or lower jaw arches providing support to dental prosthetic devices, specifically for denture stabilization to restore a patient's chewing function. It is intended for single use only.
The EM SA Implant (NARROW RIDGE) is intended to be use in the treatment of missing mandibular central and lateral incisors to support dental prosthetic devices, such as artificial teeth, in order to restore chewing function in partially edentulous patients. It is intended for single use only. It is intended for delayed loading.
Type of use (Select one or both, as applicable)
Z Prescription Use (Part 21 CFR 801 Subpart D)
[ Over-The-Counter Use (21 CFR 807 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (7/17) | Page 1 of 1 | PSC Publishing Service (301) 443-6740 EF |
|----------------------|-------------|------------------------------------------|
|----------------------|-------------|------------------------------------------|
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Hiossen Implant. The word "Hiossen" is written in a stylized font, with the "H" in red and the rest of the letters in gray. The word "Implant" is written in smaller gray letters below the word "Hiossen". The logo is simple and modern, and it is likely used to promote the company's dental implant products.
# 510(k) Summary
K191201
| (1) Submitter Information:<br>Submitted by: | Hiossen, Inc.<br>85 Ben Fairless Drive<br>Fairless Hills, PA 19030 |
|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Contact Person:<br>Telephone Number:<br>Fax Number: | Peter Lee<br>267-759-7031<br>267-759-7031 |
| Date Prepared: | November 15, 2019 |
| (2) Device Name:<br>• Proprietary Name:<br>• Classification Name:<br>• CFR Number:<br>• Device Class:<br>• Product Code: | EM SA Implant System<br>Endosseous dental implant<br>872.3640<br>Class II<br>DZE |
# (3) Predicate Devices:
| | Device | 510(k) | Manufacturer |
|-------------------|------------------------|---------------------|--------------|
| Primary Predicate | MS SA Implant System | K122171 | OSSTEM |
| Reference Device | MS System | K083067/<br>K072959 | OSSTEM |
| Reference Device | Hiossen Implant System | K140934 | HIOSSEN |
## (4) Description of Device:
The EM SA Implant System is a dental implant made of titanium alloy (Ti-6AI-4V) and is supplied sterile. The surface treatment of EM SA Implant System is SA (Sandblasted and Acid etched) and is intended to be surgically placed in the bone of the upper or lower jaw arches.
The EM SA Implant System is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. It is substantially equivalent in design, function and intended use to the OSSTEM Implant Co., Ltd.'s MS SA Implant System (K122171).
| Device | Thread Diameter(mm) | Thread Length(mm) |
|-------------------------------------|---------------------|-----------------------------|
| EM SA Implant System (Denture) | Ø2.0, Ø2.5, Ø3.0 | 8.5, 10.0, 11.5, 13.0, 15.0 |
| EM SA Implant System (Narrow Ridge) | Ø2.5, Ø3.0 | 8.5, 10.0, 11.5, 13.0, 15.0 |
# (5) Indication For Use:
The EM SA Implant System (DENTURE) is intended to be place in the bone of the upper or lower jaw arches providing support to dental prosthetic devices, specifically for denture stabilization to restore a patient's chewing function. It is intended for single use only.
The EM SA Implant (NARROW RIDGE) is intended to be use in the treatment of missing mandibular central and lateral incisors to support dental prosthetic devices, such as artificial teeth, in order to restore chewing function in partially edentulous patients. It is intended for single use only. It is intended for delayed loading. EM SA IMPLANT SYSTEM TRADITIONAL 510(k) Confidential
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Hiossen Implant. The word "Hiossen" is written in a stylized font, with the "H" in red and the rest of the letters in gray. The word "Implant" is written in smaller gray letters below "Hiossen". There is a registered trademark symbol next to the word "Hiossen".
85 Ben Fairless Drive Fairless Hills, PA 19030 888-768-0001 www.hiossen.com
# (6) Substantial Equivalence:
HIOSSEN believes the EM SA Implant System is substantially equivalent to the primary predicate device, OSSTEM MS SA Implant System in November 2012 (510(k) number K122171).
| | Proposed Device | Primary Predicate | Reference Device | Reference Device |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | EM SA Implant | MS SA System | MS System | HIOSSEN IMPLANT |
| | System (Denture) | (Denture) | | SYSTEM |
| Manufacturer | Hiossen, Inc. | Osstem Co., Ltd. | Osstem Co., Ltd. | Hiossen, Inc. |
| 510(K) No. | New device | K122171 | K083067/K072959 | K140934 |
| Design | | | | |
| Intended use | The EM SA Implant<br>System (DENTURE)<br>is intended to be<br>place in the bone of<br>the upper or lower<br>jaw arches providing<br>support to dental<br>prosthetic devices,<br>more specifically<br>denture stabilization<br>to restore a patient's<br>chewing function. It is<br>intended for single<br>use only. | The MS System<br>(Denture) is intended<br>to be place in the<br>bone of the upper or<br>lower jaw arches to<br>provide support the<br>prosthetic devices to<br>restore the patient's<br>chewing function,<br>including the denture<br>stabilization. The MS<br>System (Denture) is<br>intended for single<br>use only. | The MS System<br>(Denture) is intended<br>to be place in the<br>bone of the upper or<br>lower jaw arches to<br>provide support the<br>prosthetic devices to<br>restore the patient's<br>chewing function,<br>including the denture<br>stabilization. The MS<br>System (Denture) is<br>intended for single<br>use only. | The HIOSSEN<br>Implant System is<br>indicated for use in<br>partially or fully<br>edentulous mandibles<br>and maxillae, in<br>support of single or<br>multiple-unit<br>restorations including;<br>cemented retained,<br>screw retained, or<br>overdenture<br>restorations, and final<br>or temporary<br>abutment support for<br>fixed bridgework. It is<br>intended for delayed<br>loading. Ultra wide<br>Fixture System is<br>intended to be used<br>in the molar region. |
| Structure | • Threaded Body<br>Design<br>• One Body Implant<br>• 2/4mm gingival<br>height neck<br>• Ball-shaped head<br>for O-Ring<br>attachment. | • Threaded Body<br>Design<br>• One Body Implant<br>• 2/4mm gingival<br>height neck<br>• Ball-shaped head<br>for O-Ring<br>attachment. | • Threaded Body<br>Design<br>• One Body Implant<br>• 2/4mm gingival<br>height neck<br>• Ball-shaped head<br>for O-Ring<br>attachment. | • Internal Hex-<br>connected<br>• Submerged Fixture<br>• Tapered body<br>shape & Straight<br>body shape |
| Thread | 2.0, 2.5, 3.0 | 2.5, 3.0 | 2.0 ~ 3.0 | 3.5~6.8 |
| Diameters | | | | |
| Thread<br>Lengths | 8.5, 10.0, 11.5, 13.0,<br>15.0 | 10.0, 13.0, 15.0 | 8.5, 10.0, 13.0, 15.0 | 6.2~18.2 |
| Material | Titanium alloy Ti-6Al-<br>4V (ASTM F 136) | Titanium alloy Ti-6Al-<br>4V (ASTM F 136) | Titanium alloy Ti-6Al-<br>4V (ASTM F 136) | Pure Titanium Grade<br>4 (ASTM F67) |
| Surface | SA (Sandblasted and<br>Acid etched). | - SA(Sandblasted<br>and Acid etched) | RBM (Resorbable<br>Blast Media) | - SA(Sandblasted<br>and Acid etched) |
| Sterilization | Radiation Sterile | Radiation Sterile | Radiation Sterile | Radiation Sterile |
| Packaging | Secured in plastic<br>ampule, housed in<br>Tyvek-lidded blister<br>tray, placed in a<br>tamper-evident outer<br>package. | Secured in plastic<br>ampule, housed in<br>Tyvek-lidded blister<br>tray, placed in a outer<br>box. | Secured in plastic<br>ampule, housed in<br>Tyvek-lidded blister<br>tray, placed in a outer<br>box. | Secured in plastic<br>ampule, housed in<br>Tyvek-lidded blister<br>tray, placed in a<br>tamper-evident outer<br>package. |
| S.E. | The proposed device is substantially equivalent with respect to the intended use, structure,<br>material, surface, sterilization and design with the legally marketed primary predicate device as<br>well as the surface, sterilization and packaging of the legally marketed reference device. The<br>slight difference in thread diameter and length is a simple size addition. The proposed device<br>does not pose any new or increased risks as compared to the legally marketed primary<br>predicate and reference devices. | | | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Hiossen Implant. The word "Hiossen" is written in a stylized font, with the "H" in red and the rest of the letters in gray. The word "Implant" is written in gray below "Hiossen". The logo is simple and modern, and it is likely used to promote the company's dental implant products.
| Device | Proposed Device<br>EM SA Implant<br>System<br>(Narrow Ridge) | Primary Predicate<br>MS SA Implant<br>System<br>(Narrow Ridge) | Reference Device<br>MS System | Reference Device<br>HIOSSEN IMPLANT<br>SYSTEM |
|--------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Hiossen, Inc. | Osstem Co., Ltd. | Osstem Co., Ltd. | Hiossen, Inc. |
| 510(K) No. | New device | K122171 | K083067 | K140934 |
| Design | Image: Proposed Device EM SA Implant System (Narrow Ridge) | Image: Primary Predicate MS SA Implant System (Narrow Ridge) | Image: Reference Device MS System | Image: Reference Device HIOSSEN IMPLANT SYSTEM |
| Intended use | The EM SA Implant<br>System (NARROW<br>RIDGE) is intended<br>to be use in the<br>treatment of missing<br>mandibular central<br>and lateral incisors to<br>support dental<br>prosthetic devices,<br>such as artificial<br>teeth, in order to<br>restore chewing<br>function in partially<br>edentulous patients.<br>It is intended for<br>single use only. It is<br>intended for delayed<br>loading. | The MS SA Implant<br>System (Narrow<br>Ridge) is intended to<br>use in the treatment<br>of missing<br>mandibular central<br>and lateral incisors to<br>support prosthetic<br>device, such as<br>artificial teeth, in<br>order to restore<br>chewing function in<br>partially edentulous<br>patients. The MS SA<br>Implant (Narrow<br>Ridge) is intended for<br>single use only. It is<br>intended for delayed<br>loading. | The MS Implant<br>System (Narrow<br>Ridge) is intended to<br>use in the treatment<br>of missing<br>mandibular central<br>and lateral incisors to<br>support prosthetic<br>device, such as<br>artificial teeth, in<br>order to restore<br>chewing function in<br>partially edentulous<br>patients. MS Implant<br>(Narrow Ridge) is<br>intended for single<br>use only. It is<br>intended for delayed<br>loading | The HIOSSEN<br>Implant System is<br>indicated for use in<br>partially or fully<br>edentulous<br>mandibles and<br>maxillae, in support<br>of single or multiple-<br>unit restorations<br>including; cemented<br>retained, screw<br>retained, or<br>overdenture<br>restorations, and final<br>or temporary<br>abutment support for<br>fixed bridgework. It is<br>intended for delayed<br>loading. Ultra wide<br>Fixture System is<br>intended to be used |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Hiossen Implant logo. The word "Hiossen" is written in a stylized font, with the "H" in red and the rest of the letters in gray. The word "Implant" is written in gray below "Hiossen". There is a registered trademark symbol next to the word "Hiossen".
| | | | | in the molar region. |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Structure | • Threaded Body<br>Design<br>• One Body Implant<br>• 2.5/4mm gingival<br>height neck<br>• Integrated abutment<br>head | • Threaded Body<br>Design<br>• One Body Implant<br>• 2.5/4mm gingival<br>height neck<br>• Integrated abutment<br>head | • Threaded Body<br>Design<br>• One Body Implant<br>• 2/4mm gingival<br>height neck<br>• Integrated abutment<br>head | • Internal Hex-<br>connected<br>• Submerged Fixture<br>• Tapered body<br>shape & Straight<br>body shape |
| Thread<br>Diameters | 2.5, 3.0 | 2.5, 2.9 | 2.5, 3.0 | 3.5~6.8 |
| Thread<br>Lengths | 8.5, 10.0, 11.5, 13.0,<br>15.0 | 8.5,10.0, 13.0, 15.0 | 8.5, 10.0, 13.0, 15.0 | 6.2~18.2 |
| Material | Titanium alloy Ti-6Al-<br>4V (ASTM F 136) | Titanium alloy Ti-6Al-<br>4V (ASTM F 136) | Titanium alloy Ti-6Al-<br>4V (ASTM F 136) | Pure Titanium Grade<br>4 (ASTM F67) |
| Surface | - SA(Sandblasted<br>and Acid etched) | - SA(Sandblasted<br>and Acid etched) | RBM (Resorbable<br>Blast Media) | - SA(Sandblasted<br>and Acid etched) |
| Sterilization | Radiation Sterile | Radiation Sterile | Radiation Sterile | Radiation Sterile |
| Packaging | Secured in plastic<br>ampule, housed in<br>Tyvek-lidded blister<br>tray, placed in a<br>tamper-evident outer<br>package. | Secured in plastic<br>ampule, housed in<br>Tyvek-lidded blister<br>tray, placed in a<br>outer box. | Secured in plastic<br>ampule, housed in<br>Tyvek-lidded blister<br>tray, placed in a<br>outer box. | Secured in plastic<br>ampule, housed in<br>Tyvek-lidded blister<br>tray, placed in a<br>tamper-evident outer<br>package. |
| S.E. | The proposed device is substantially equivalent with respect to the intended use, structure,<br>material, surface, sterilization and design with the legally marketed primary predicate device<br>as well as the surface, sterilization and packaging of the legally marketed reference device.<br>The slight difference in thread diameter and length is a simple size addition. The proposed<br>device does not pose any new or increased risks as compared to the legally marketed primary<br>predicate and reference devices | | | |
## Non-Clinical Performance Data
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence include data from the following tests:
## Biocompatibility Testing
The EM SA Implant System is manufactured using the same manufacturing process and same well known and well established material as the primary predicate device, the MS SA Implant System, K122171, therefore we reason it was not necessary to re-test biocompatibility in order to support the biological safety of the EM SA Implant System. Furthermore, as described in International Standard Organization (ISO) standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing when a new material is used that has not been identified in a primary predicate device. The proposed device is manufactured from standard raw material that is used in the primary predicate device and other currently marketed dental implant system. Therefore, no additional biocompatibility testing is required to establish substantial equivalence.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for Hiossen Implant. The word "Hiossen" is written in a sans-serif font, with the "H" in red and the rest of the letters in gray. There is a red curved line above the "H". Below "Hiossen" is the word "IMPLANT" in smaller gray letters. There is a registered trademark symbol next to the word "Hiossen".
ww.hiossen.com
## Sterilization Validation
The EM SA Implant System (manufactured using the same manufacturing process, material and utilizes the same sterilize barrier system) is gamma irradiated under the same conditions and process as the primary predicate device, the MS SA Implant System, K122171 and validated following ISO 11137 [2006] Sterilization of health care products - Requirements for Validation and Routine control - Radiation Sterilization quidelines, therefore we reason it was not necessary to re-test validation in order to support the sterilization validity of the EM SA Implant System.
#### Shelf Life
The EM SA shelf life is identical to the reference device HIOSSEN Implant System which has been validated for eight (8) years following ISO 11607-1 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems. The EM SA utilizes the exact same packaging materials & methods and is a medical grade titanium, nonmechanical, non-active device, therefore, degradation in performance characteristics is not likely over the established shelf life period. Therefore, no shelf-life validation or accelerated aging testing was performed.
### Surface Treatment Characterization Testing
The EM SM Implant System is manufactured using the same manufacturing process, material and well known and well established surface treatment, SA (Sand blasted, Acid etched) as the predicate device, the MS SA Implant system, K122171, therefore no additional character testing was necessary to support the equivalency of the EM SA Implant System.
### Fatique Testing
The EM SA Implant System is manufactured using the same manufacturing process, material and very similar design, as the MS SA which was fatique tested in accordance with ISO 14801:2003 Dentistry -Fatigue Test for Endosseous Dental Implants, we reason it was not necessary to re-test fatigue.
#### Clinical Performance Testing
No clinical performance report(s) is being submitted.
#### (7) Conclusion:
In accordance with the Federal Food Drug and Cosmetic Act, 21 CFR Part 807, and based on the data and information provided in this premarket notification, the proposed device, EM SA Implant System has established substantially equivalency to the primary predicate and reference device. An extensively comparison of the same design, intended use, structure, general shape and size, material and application method confirms the EM SA share many similarities with the primary predicate MS SA Implant System (K122171) and utilize the same sterilization method and packaging materials to the reference device Hiossen Implant System (K140934). The additional thread diameter and length is a simple size addition, a risk and mitigation measure for root-form endosseous dental implant in a FMEA risk analysis did not disclose any safety or health risk.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.