← Product Code [JWH](/productcode/JWH) · K191192

# SurgTech GENOLL Total Knee System (K191192)

_Surgtech, Inc. · JWH · Aug 1, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K191192

## Device Facts

- **Applicant:** Surgtech, Inc.
- **Product Code:** [JWH](/productcode/JWH.md)
- **Decision Date:** Aug 1, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Surg Tech GENOLL™ system is intended to reduce or relieve pain and restore function and motion to the knee joint. Total knee replacement is indicated for patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery. This device is intended for cemented use only.

## Device Story

SurgTech GENOLL™ Total Knee System is a posterior stabilized (PS) knee prosthesis for total knee arthroplasty. Components include CoCr alloy femoral component and tibial baseplate, plus conventional polyethylene tibial bearing insert and patella. Used by orthopedic surgeons in clinical settings to replace damaged knee joints; intended to reduce pain and restore function. Device is implanted via surgical procedure; requires cemented fixation. Benefits include pain relief and improved joint mobility for patients with severe degenerative or traumatic knee conditions. Instrumentation for implantation is included.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including baseplate fatigue (ASTM F1800), constraint (ASTM F1223), contact area (ASTM F2083, ASTM F1672), PS post fatigue (ASTM F1814), tibial insert/baseplate disassembly (ASTM F2083, ASTM F1814), and bacterial endotoxin testing (USP <85>).

## Technological Characteristics

Materials: CoCr alloy (femoral component, tibial baseplate), conventional polyethylene (tibial insert, patella). Design: Posterior stabilized (PS) semi-constrained cemented prosthesis. Standards: ASTM F1800, F1223, F2083, F1672, F1814; USP <85>. Connectivity: None (mechanical implant).

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Zimmer-NexGen LPS Flex ([K991581](/device/K991581.md))
- Otis Biotech-Prolixus ([K170534](/device/K170534.md))

## Reference Devices

- SurgTech-MALUC ([K162125](/device/K162125.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 1, 2019

SurgTech, Inc. % Kellen Hills Quality and Regulatory Consultant Orchid Design 80 Shelton Technology Center Shelton, Connecticut 06484

Re: K191192

Trade/Device Name: SurgTech GENOLL™ Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: April 29, 2019 Received: May 3, 2019

Dear Kellen Hills:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Raquel Peat, PhD, MPH, USPHS Director OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K191192

Device Name SurgTech GENOLL™ Total Knee System

### Indications for Use (Describe)

The Surg Tech GENOLL™ system is intended to reduce or relieve pain and restore function and motion to the knee joint. Total knee replacement is indicated for patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery. This device is intended for cemented use only.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

GENOLL Total Knee System

{3}------------------------------------------------

| (a)(1) | Date Prepared:                     | April 29, 2019                                                                                                       |
|--------|------------------------------------|----------------------------------------------------------------------------------------------------------------------|
|        | Submitted By:                      | SurgTech, Inc.<br>24600 Center Ridge Road, Suite 195<br>Westlake, OH 44145                                           |
|        | Phone:                             | 440-899-2922                                                                                                         |
|        | Contact:                           | Robert Gutierrez                                                                                                     |
|        | Prepared By:                       | Orchid Design<br>80 Shelton Technology Center,<br>Shelton, CT 06484                                                  |
|        | Phone:                             | (203) 922 0105                                                                                                       |
| (a)(2) | Proprietary Name:                  | SurgTech GENOLLTM Total Knee System                                                                                  |
|        | Common Name:                       | Total Knee Prosthesis                                                                                                |
|        | Classification Name and Reference: | 21CFR 888.3560 - Knee joint<br>patellofemorotibial<br>polymer/metal/polymer semi-<br>constrained cemented prosthesis |
|        | Product Code:                      | JWH                                                                                                                  |
| (a)(3) | Predicate Devices:                 |                                                                                                                      |
|        | Primary:                           | Zimmer-NexGen LPS Flex (K991581)                                                                                     |
|        | Additional:                        | Otis Biotech-Prolixus (K170534);                                                                                     |
|        | Reference:                         | SurgTech-MALUC (K162125);                                                                                            |

## (a)(4) Device Description:

The SurgTech GENOLL"Total Knee System is used for total knee arthroplasty in skeletally mature individuals. The system consists of a CoCr alloy Posterior Stabilized (PS) femoral component (10 sizes, Left and Right), a CoCr alloy tibial baseplate (10 sizes), a conventional polyethylene PS tibial bearing insert (5 sizes with 5 thicknesses each) and a conventional polyethylene patella (3 sizes). Instrumentation necessary for proper implantation is also included.

The purpose of this submission is to gain initial marketing authorization in the United States.

(a)(5) Indications for Use:

> The SurgTech GENOLL™ system is intended to reduce or relieve pain and restore function and motion to the knee joint. Total knee replacement is indicated for patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary

{4}------------------------------------------------

traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery. This device is intended for cemented use only.

- (a)(6) Comparison of Technological Characteristics:
The SurgTech GENOLL Total Knee System is substantially equivalent to the previously cleared predicate devices based on similarities in intended use, design, materials, packaging, sterilization and mechanical performance. The technological characteristics do not raise any new questions of safety and efficacy.

- (b)(1) Non-clinical testing:
The subject devices were evaluated for the following performance characteristics:

- Baseplate Fatigue (ASTM F1800)
- Constraint (ASTM F1223) ●
- Contact Area (ASTM F2083, ASTM F1672)
- PS Post Fatigue (ASTM F1814) ●
- 0 Tibial Insert/Baseplate Disassembly (ASTM F2083, ASTM F1814)
- Bacterial endotoxins (USP <85>)

## (b)(2) Clinical testing:

Clinical testing was not required to demonstrate substantial equivalence in this premarket notification.

## (b)(3) Conclusions:

Based on the information provided in this premarket notification, we believe that the subject SurgTech GENOLL Total Knee System demonstrates substantial equivalence to the identified predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K191192](https://fda-staging.innolitics.com/device/K191192)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).