Neodent Implant System - Temporary Abutments

K191191 · Jjgc Industria E Comercio DE Materiais Dentarios S.A. · NHA · Dec 5, 2019 · Dental

Device Facts

Record IDK191191
Device NameNeodent Implant System - Temporary Abutments
ApplicantJjgc Industria E Comercio DE Materiais Dentarios S.A.
Product CodeNHA · Dental
Decision DateDec 5, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The Neodent Implant System - Temporary Abutments are indicated to be used on Neodent implants to provide temporary support for prosthesis structure for up to 6 months.

Device Story

Neodent Implant System - Temporary Abutments are titanium alloy (Ti6Al4v-ELI) components designed for temporary prosthetic support on Neodent GM implants. Used by dental clinicians in clinical settings; devices are screw-retained and feature coronal grooves for acrylic bonding and channels for height reduction. Available in anti-rotational (single-unit) and rotational (multi-unit) versions. Clinicians select appropriate gingival height and diameter to support provisional restorations for up to 6 months. Output is a stable base for temporary prosthetics, facilitating patient chewing function during the healing phase before final restoration. Device is single-use and provided sterile.

Clinical Evidence

Bench testing only. No clinical data. Testing included dynamic fatigue testing per ISO 14801, static torsional loading, sterilization validation (ISO 1135, SAL 1x10^-6), EtO residual testing (ISO 10993-7), shelf-life testing (ASTM F1980, 5-year claim), and biological safety assessment (ISO 10993-1, -5, -12, -18).

Technological Characteristics

Material: Titanium alloy (Ti6Al4v-ELI) per ASTM F136. Design: Straight, cylindrical, screw-retained abutment with retention rings. Interface: Neodent GM. Sterilization: Ethylene oxide (SAL 1x10^-6). Dimensions: Diameters 3.5/4.5 mm; various gingival heights. Single-use.

Indications for Use

Indicated for patients requiring dental implants in the upper or lower jaw for prosthetic support (artificial teeth) to restore chewing function. Used in single or two-stage procedures for single or multiple unit restorations. Temporary abutments provide support for prosthesis structures for up to 6 months.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 5, 2019 JJGC Industria e Comercio de Materiais Dentarios S.A. % Jennifer Jackson Director of Regulatory Affairs Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810 Re: K191191 Trade/Device Name: Neodent Implant System - Temporary Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: November 1, 2019 Received: November 4, 2019 Dear Jennifer Jackson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K191191 Device Name Neodent Implant System - Temporary Abutments #### Indications for Use (Describe) The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The Neodent Implant System - Temporary Abutments are indicated to be used on Neodent implants to provide temporary support for prosthesis structure for up to 6 months. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------|---------------------------------------------| | <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## K191191 – 510(k) Summary ### ADMINISTRATIVE INFORMATION | Sponsor | JJGC Indústria e Comércio de Materiais Dentários SA<br>(dba Neodent)<br>Av. Juscelino Kubitschek de Olivera, 3291<br>Curitiba, Parana, Brazil 81270-200<br>Registration No.: 3008261720<br>Owner/Operator No.: 10031702 | | | |--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--| | Contact Person | Jennifer M. Jackson, MS<br>Director of Regulatory Affairs,<br>Straumann USA<br>E-mail: jennifer.jackson@straumann.com<br>Telephone (978) 747-2509 | | | | Date Prepared | 04/Dec/2019 | | | | Preparer / Alternate Contact | Mariana Soares Hartmann<br>Regulatory Affairs Analyst<br>E-mail: mariana.hartmann@neodent.com | | | | DEVICE NAME AND CLASSIFICATION | | | | | Trade/ Proprietary Name | Neodent Implant System - Temporary Abutments | | | | Common Name | Endosseous dental implant abutment | | | | Classification Name | Endosseous dental implant abutment | | | | Classification Regulations | 21 CFR 872.3630, Class II | | | | Product Code | NHA | | | | Classification Panel | Dental Products Panel | | | | Reviewing Branch | Dental Devices Branch | | | | PREDICATE DEVICE INFORMATION | | | | | Primary Predicate Device | K163194, Neodent Implant System - GM Line, JJGC Indústria e<br>Comércio de Materiais Dentários S.A. | | | | Reference Predicate Devices | K093027 - STRAUMANN RC TEMPORARY ABUTMENTS<br>Straumann Manufacturing, INC. | | | {4}------------------------------------------------ K092814 – STRAUMANN DENTAL ABUTMENTS, Straumann Manufacturing, INC. K103084 - Neodent Graft Screw, JJGC Indústria e Comércio de Materiais Dentários S.A. #### INDICATIONS FOR USE The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The Neodent Implant System - Temporary Abutments are indicated to be used on Neodent implants to provide temporary support for prosthesis structure for up to 6 months. #### SUBJECT DEVICE DESCRIPTIONS - Temporary Abutment for provisional restoration employed for up to 6 months; - Intended for single use; - Provided sterile via ethylene oxide gas; - Constructed of titanium alloy (Ti6Al4v-ELI) per ASTM F136; - Coronal geometry has grooves to facilitate bonding of acrylic material; - Abutment is screw-retained; - . Coronal geometry contains channels to facilitate reduction of abutment height; - Provided in anti-rotational and rotational versions supporting single and multi-unit restorations, respectively; - . Provided in multiple gingival heights; - Implant-to-abutment interface compatible with GM implants of the Neodent Implant System. #### TECHNOLOGICAL CHARACTERISTIC COMPARISON TABLE | | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | | |------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Neodent Implant System | K163194<br>Neodent Implant System - GM Line | EQUIVALENCE DISCUSSION | | Indications<br>for Use | The Neodent Implant System is intended<br>to be surgically placed in the bone of the<br>upper or lower jaw to provide support for<br>prosthetic devices such as artificial teeth,<br>to restore chewing function. It may be<br>used with single-stage or two-stage<br>procedures, for single or multiple unit<br>restorations, and may be loaded<br>immediately when good primary stability<br>is achieved and with appropriate occlusal<br>loading. The Neodent Implant System -<br>Temporary Abutments are indicated to<br>be used on Neodent implants to provide<br>temporary support for prosthesis<br>structure for up to 6 months. | Indications for Use for GM implants and<br>conventional abutments:<br>The Neodent Implant System is intended<br>to be surgically placed in the bone of the<br>upper or lower jaw to provide support<br>for prosthetic devices such as artificial<br>teeth, to restore chewing function. It<br>may be used with single-stage or two-<br>stage procedures, for single or multiple<br>unit restorations, and may be loaded<br>immediately when good primary stability<br>is achieved and with appropriate occlusal<br>loading.<br>Indications for Use for GM Titanium Base<br>abutments:<br>Titanium Base Abutment is a titanium<br>base placed onto Neodent dental<br>implants to provide support for | Equivalent<br>The subject device is a temporary<br>abutment for use with GM implants,<br>therefore only the first paragraph of the<br>Indications of the primary predicate are<br>relevant. The subject submission does<br>not include Titanium Base Abutments or<br>ProPEEK Abutments. | | | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | EQUIVALENCE DISCUSSION | | | Neodent Implant System | K163194<br>Neodent Implant System - GM Line | | | | | customized prosthetic restorations. It is<br>used with a coping and crown, or crown<br>alone, and is indicated for cement-<br>retained single or multi-unit<br>restorations, or screw-retained single<br>restorations. | | | | | All digitally designed copings and/or<br>crowns for use with the Neodent<br>Titanium Base Abutment System are<br>intended to be sent to Straumann for<br>manufacture at a validated milling<br>center. | | | | | Indications for Use for GM Pro Peek<br>Abutments: | | | | | The Pro PEEK Abutments are indicated to<br>be used on Neodent implants to provide<br>temporary support for prosthesis<br>structure for up to 6 months. They can<br>be used in one or two stage procedures<br>and also immediate load when there is<br>good primary stability. | | | Interface | GM | GM | Same | | Design | Straight, cylindrical with retention rings | Straight and cylindrical | Equivalent | | Reusable | No | No | Same | | Gingival Height | 0.8; 1.5; 2.5 and 3.5 mm | 0.8; 1.5; 2.5; 3.5; 4.5 and 5.5 mm | Equivalent<br>Subject devices within range established<br>by predicate devices | | Diameter ( $Ø$ ) | 3.5 and 4.5 mm | 3.3 and 4.5 mm | Equivalent | | Angulation | Straight | Straight, 17° and 30° | Same<br>Subject devices are straight which is the<br>same as the straight predicate devices. | | Material | TAV (ASTM F136) | Implants:<br>Titanium grade 4<br>Conventional abutments:<br>Titanium alloy Ti-6Al-4V ELI<br>CAD/CAM abutments:<br>Titanium alloy Ti-6Al-4V ELI<br>Temporary abutments:<br>Titanium alloy Ti-6Al-4V PEEK | Same<br>Subject devices are the same material as<br>the TAV predicate devices. | | Duration of use<br>(maximum) | 180 days | Lifetime | Equivalent<br>Subject devices within range established<br>by primary predicate. Subject devices<br>same as for reference predicates per<br>K093027 and K092814 | | Sterilization<br>Method | Ethylene Oxide to an SAL of 1x10-6 | Ethylene Oxide to an SAL of 1x10-6 | Same | {5}------------------------------------------------ {6}------------------------------------------------ The subject devices and the primary predicate device K163194 have equivalent Indications for Use, similar design, range of gingival height, same raw material and same sterilization method. #### PERFORMANCE DATA Dynamic fatigue test per ISO 14801 was performed to determine the fatigue strength for the worst-case implant construct assembled with GM Temporary Abutments for multi-unit restorations, according to FDA Guidance. Torsion testing was performed to evaluate the GM Temporary Abutments Screws under static torsional loading. Sterilization of the subject abutments via ethylene oxide gas using the overkill method has been validated according to the requirements of ISO 1135. A minimum Sterility Assurance Level (SAL) of 1 x 106 has been validated. Ethylene oxide residuals have been assessed per ISO 10993-7. Residuals are within accepted limits. Accelerated and Real Time Aging tests were performed according to ASTM F1980. The Shelf Life of the subject devices is 5-years. Biological Safety Assessment guided by ISO 10993-1. Cytotoxicity testing was performed per ISO 10993-5. Chemical characterization was performed per ISO 10993-18. Biocompatibility sample preparation was performed per ISO 10993-12. #### CONCLUSION The subject devices and the primary predicate devices K163194 have the same intended use, similar designs and technological characteristics, and are made of the same materials. The data included in this submission demonstrate that the subject device is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...