# LENS 4K Camera Control Unit - Wifi, LENS 4K Camera Control Unit - Non-Wifi, LENS 4K Camera Head (K191177)

_Smith and Nephew, Inc. · GCJ · Aug 23, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K191177

## Device Facts

- **Applicant:** Smith and Nephew, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Aug 23, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The LENS 4K Camera Control Unit and LENS 4K Camera Head are intended to provide illumination, visualization and capture of still and motion pictures of surgical sites.

## Device Story

Integrated camera, LED light source, and image management system for arthroscopic and endoscopic surgery. Device consists of a 4K Camera Control Unit (CCU) and 4K Camera Head. CCU connects to camera head, light guides, couplers, and endoscopes to visualize and capture surgical site images/video. Used in OR/clinical settings by surgeons. CCU available in Wi-Fi and non-Wi-Fi versions; non-Wi-Fi version stores data to USB or external image management system. Wi-Fi version supports optional tablet application for remote capture, annotation, and sharing. Output displayed for surgeon visualization and archived for clinical documentation. Benefits include high-resolution 4K visualization of surgical sites.

## Clinical Evidence

Bench testing only. Performance data included video output verification, camera head surface temperature monitoring, software verification and validation, electromagnetic emissions and immunity testing, and electrical safety testing.

## Technological Characteristics

Integrated 4K CCU and camera head. Camera head features 3-chip CMOS sensor and 304L stainless steel housing. CCU includes LED light source and image management. Connectivity via USB, Wi-Fi (optional), and tablet application. Compatible with endoscopes, light guides, and couplers. Software-based image processing.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Smith & Nephew LENS Integrated System and LENS Camera Head ([K153606](/device/K153606.md))

## Reference Devices

- Smith & Nephew 660HD-E Image Management System
- Tablet Application ([K190367](/device/K190367.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Smith and Nephew, Inc. Kathleen Solomon Principal Regulatory Affairs Specialist 150 Minuteman Road Andover, Massachusetts 01810

August 23, 2019

# Re: K191177

Trade/Device Name: LENS 4K Camera Control Unit - Wifi. LENS 4K Camera Control Unit - Non-Wifi, LENS 4K Camera Head Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ, HRX Dated: July 25, 2019 Received: July 26, 2019

## Dear Kathleen Solomon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Neil R.P. Ogden, MS Assistant Director, THT4A4 DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K191177

Device Name Smith & Nephew LENS 4K System

#### Indications for Use (Describe)

The LENS 4K Camera Control Unit and LENS 4K Camera Head are used in diagnostic and operative procedures for arthroscopic and endoscopic procedures to provide illumination, visualization and capture of surgical sites within articular cavities, body cavites, hollow organs and canals. Additionally, the LENS 4K Camera Control Unit and LENS 4K Camera Head are indicated for use in endoscopic surgical procedures in the thoracic cavity when used with an appropriately indicated thoracoscope.

Type of Use (Select one or both, as applicable)

| <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)     </span> </div> |
|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)     </span> </div>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Endoscopy
Smith & Nephew, Inc.
150 Minuteman Road
Andover, MA 01810
USA

mith-nephew.com

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION

as required by the Safe Medical Devices Act of 1990 and codified in 21 CFR 807.92 upon which the substantial equivalence is based.

# Smith & Nephew LENS 4K System (Camera Control Unit and Camera Head)

Date Prepared: August 19, 2019

## A. Submitter's Name:

Smith & Nephew, Inc., Endoscopy Division 150 Minuteman Road, Andover MA. 01810

## B. Company Contact:

Kathleen Solomon Principal Regulatory Affairs Specialist T 978-749-1605 F 978-749-1443

#### C. Device Name:

| Trade Name:          | LENS 4K Camera Control Unit-Wifi, LENS 4K Camera Control Unit<br>- Non-Wifi LENS 4K Camera Head |
|----------------------|-------------------------------------------------------------------------------------------------|
| Common Name:         | LENS 4K System (4K camera control unit and 4K camera head)                                      |
| Classification Name: | Endoscopes and Accessories per 876.1500                                                         |
| Regulatory class:    | II                                                                                              |
| Product Code:        | GCJ, HRX                                                                                        |

#### D. Predicate Devices

The Smith & Nephew LENS 4K is substantially equivalent in intended use and fundamental scientific technology to the following legally marketed device in commercial distribution: Smith & Nephew LENS Integrated System and LENS Camera Head, K153606.

## E. Description of Device

The Smith & Nephew LENS 4K System ( 4K Camera Control Unit (CCU) and LENS 4K Camera Head), is an integrated camera/ LED light source/ image management device that provides visualization, illumination and image/video capture combined in a single console. The LENS 4K CCU works in conjunction with a camera head, appropriate light guides, couplers and endoscopes.

{4}------------------------------------------------

Endoscopy
Smith & Nephew, Inc.
150 Minuteman Road
Andover, MA 01810
USA

smith-nephew.com

The LENS 4K CCU will be offered in two versions, one with Wi-Fi and one without Wi-Fi. The non-Wi-Fi version does not have the ability to communicate to a tablet but captures the data to an attached USB storage device or to the Smith & Nephew 660HD-E Image Management System, which is a Class 1 device per LMB for display and sharing for display and sharing.

Additionally, the LENS 4K System will also work with an optional Tablet Application, which was cleared in K190367. The Tablet application, when downloaded to a mobile device, acts as a tool to remotely capture video recordings and image files wirelessly from the LENS 4K System. The application provides the ability to view and annotate by voice or graphics or to text. Sharing of patient data from the Tablet is provided via the ability to archive to a USB storage device on the CCU, email, printing, or export to the local tablet camera roll

# F. Intended Use

●

The LENS 4K Camera Control Unit and LENS 4K Camera Head are intended to provide illumination, visualization and capture of still and motion pictures of surgical sites.

# Indications for Use:

The LENS 4K Camera Control Unit and LENS 4K Camera Head are used in diagnostic and operative procedures for arthroscopic and endoscopic procedures to provide illumination, visualization and capture of still and motion pictures of surgical sites within articular cavities, body cavities, hollow organs and canals. Additionally, the LENS 4K Camera Control Unit and LENS 4K Camera Head are indicated for use in endoscopic surgical procedures in the thoracic cavity when used with an appropriately indicated thoracoscope.

# G. Comparison of Technological Characteristics

The proposed Smith & Nephew LENS 4K Camera Control Unit and LENS 4K Camera Head has the following similarities as the predicate device cleared in K153606. In that:

- The proposed and predicate devices have similar indications for use/intended use.
- . Both the predicate and proposed utilizes the same principle of operation and fundamental scientific technology
- Incorporates the same basic design
	- o Camera control unit utilizes the same chassis design (size and shape) and same material
	- Camera head utilizes the same housing and silicone button design o
	- Camera Control Unit incorporates the same material
- Video formats are similar in that both the LENS 4K and predicate use 1080p/1080i ● switchable formats.
- Both the predicate and the proposed are designed and compatible with the Tablet Application

{5}------------------------------------------------

Endoscopy
Smith & Nephew, Inc.
150 Minuteman Road
Andover, MA 01810
USA

T 978-749-1000
F 978-749-1443
www.smith-nephew.com

The differences between the proposed LENS 4K System and the predicate device LENS Integrated System are:

- The LENS 4K System has 4K video technology whereas the LENS Integrated System is HD video.
- LENS 4K camera head is a 3-chip CMOS sensor comprised of 304L stainless steel housing. The predicate LENS camera heads is a single chip sensor and PEEK housing.

The differences between the proposed and predicate device do not constitute a new intended use, there are no differences in technological characteristics and does not introduce new types of safety or effectiveness questions.

# H. Performance Data

The following testing was performed on the LENS 4K System to demonstrate that it has met the established performance specifications and therefore, is substantially equivalent to the predicate device cleared in K153606:

- Video Output ●
- Camera Head Surface Temperature ●
- Software Verification and Validation ●
- Emissions and Immunity Testing ●
- Electrical Safety Testing .

#### I. Conclusion:

The Smith & Nephew LENS 4K System met all specified criteria and did not raise new safety and/or effectiveness questions. The substantial equivalence of the proposed device is based on similar indications for use, fundamental technology including design, and operational principles. Based on the similarities to the predicate and the performance data, the Smith & Nephew LENS 4K System is substantially equivalent to its predicate (K153606).

---

**Source:** [https://fda-staging.innolitics.com/device/K191177](https://fda-staging.innolitics.com/device/K191177)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).