K191143 · C.R. Bard, Inc. · PTD · Aug 2, 2019 · General Hospital
Device Facts
Record ID
K191143
Device Name
PowerFlow Implantable Apheresis IV Port
Applicant
C.R. Bard, Inc.
Product Code
PTD · General Hospital
Decision Date
Aug 2, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Bard PowerFlow™ Implantable Apheresis IV Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for long-term therapeutic apheresis, withdrawal of blood, and infusion of medications, IV fluids, parenteral nutrition solutions, blood and blood products. The Bard PowerFlow™M Implantable Apheress IV Port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 mL/s.
Device Story
Totally implantable, low-profile, angled-access titanium port; provides repeated vascular access without daily external catheter care. System includes port body with silicone-layered septum and radiopaque polyurethane catheter. User palpates angled entry funnel; inserts 14 or 16Ga introducer needle; advances P.I.V. catheter through septum; removes needle; connects catheter to extension set for infusion/withdrawal. Used in clinical settings by healthcare providers. Enables long-term therapy delivery and blood sampling; reduces need for repeated venipuncture. Power-injectable when used with compatible catheters.
Clinical Evidence
Bench testing only. Verification tests included port subassembly radiopacity, corrosion resistance, silicone boot retention, stem tensile strength, stem-catheter connection air leak, and valve life. Biocompatibility testing included chemical characterization, cytotoxicity, sensitization, intracutaneous irritation, acute systemic toxicity, genotoxicity (Ames and chromosomal aberrations), and 13-week intramuscular/bone implantation. Results met all pre-determined acceptance criteria.
Technological Characteristics
Implantable titanium port (Ti-6AL-4V Type 1 MIM per ASTM F2885-17) with silicone-layered septum and polyurethane catheter. Sterilized via ethylene oxide (EtO). Mechanical access via angled funnel design. No software or electronic components.
Indications for Use
Indicated for patients requiring long-term, repeated vascular access for therapeutic apheresis, blood withdrawal, infusion of medications, IV fluids, parenteral nutrition, blood products, and power injection of contrast media (max 5 mL/s).
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Predicate Devices
PowerFlow™ Implantable Apheresis IV Port (K163001)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 2, 2019
C.R. Bard, Inc. Kristen Soodak Senior Regulatory Affairs Specialist 605 North 5600 West Salt Lake City, Utah 84116
Re: K191143
Trade/Device Name: PowerFlow™ Implantable Apheresis IV Port Regulation Number: 21 CFR 880.5965 Regulation Name: Subcutaneous, implanted, intravascular infusion port and catheter Regulatory Class: Class II Product Code: PTD Dated: July 2, 2019 Received: July 3, 2019
Dear Kristen Soodak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K191143
Device Name PowerFlow™ Implantable Apheresis IV Port
#### Indications for Use (Describe)
The Bard PowerFlow™ Implantable Apheresis IV Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for long-term therapeutic apheresis, withdrawal of blood, and infusion of medications, IV fluids, parenteral nutrition solutions, blood and blood products.
The Bard PowerFlow™M Implantable Apheress IV Port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 mL/s.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# PowerFlow™ Implantable Apheresis I.V. Port 510(k) Summary
21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
### Submitter Information:
Bard Access Systems 605 North 5600 West Salt Lake City, Utah 84116
Phone: 480-350-6017 Fax: 480-449-2546
Contact Person: Kristen Soodak Senior Regulatory Affairs Specialist Date of Preparation: 29 April 2019
## Subject Device Name:
| Name of Device: | PowerFlow™ Implantable Apheresis IV Port |
|------------------------------------|----------------------------------------------------------------------|
| Common or Usual Name: | Subcutaneous implanted apheresis port |
| Classification Name: | Subcutaneous, implanted, intravascular<br>infusion port and catheter |
| Requlatory Class:<br>Product Code: | Class 2<br>PTD |
| Regulation Number: | 21 CFR §880.5965 |
| | |
### Predicate Device:
| Name of Device: | PowerFlow™ Implantable Apheresis IV Port |
|-------------------------|----------------------------------------------------------------------|
| Common or Usual Name: | Subcutaneous implanted apheresis port |
| Classification Name: | Subcutaneous, implanted, intravascular<br>infusion port and catheter |
| Premarket Notification: | K163001 |
| Regulatory Class: | Class 2 |
| Product Code: | PTD |
| Regulation Number: | 21 CFR §880.5965 |
{4}------------------------------------------------
#### Device Description:
### PowerFlow™ Implantable Apheresis IV Port
The PowerFlow™ Implantable Apheresis IV Port with 9.6 Fr. ChronoFlex™ Catheter is designed to provide repeated access to the vascular system without the need for repeated venipuncture or the daily care of an external catheter. The PowerFlow™ Implantable Apheresis IV Port is a low profile totally implantable, angled access titanium port-based design and is accessed through an angled opening which consists of a funnel shaped entrance designed to guide the peripheral intravenous (P.I.V.) access needle and catheter into the subject device. The PowerFlow™ Implantable Apheresis IV Port comes with a number of kit components to aid in the implantation procedure and/or access of the device once implanted. The PowerFlow™ Implantable Apheresis IV Port and necessary kit components are provided sterile (EtO).
The overall implanted system consists of three primary components: the port body with a silicone layered septum, an attachable radiopaque polyurethane catheter lock which secures the catheter to the port body stem. The method of implantation and access of the subject PowerFlow™ Implantable Apheresis IV Port is the exact same as the predicate PowerFlow™ Implantable Apheresis IV Port device. After the implanted device has been identified and access is prepped per institutional policy, the user palpates the uniquely shaped angled entry funnel. Once the funnel is palpated, providing the location of the introducer needle access path, the 14 or 16Ga introducer needle is inserted into the funnel. After the Introducer Needle Stop is reached, the Introducer Needle is pulled back slightly and the P.I.V. Catheter is advanced forward. The P.I.V. Catheter is then advanced through the silicone layered septum and the Introducer Needle is removed. After needle removal, the P.I.V. Catheter is attached to the appropriate extension set and secured for the necessary infusion or withdrawal procedure.
The PowerFlow™ Implantable Apheresis IV Port can be used for routine vascular access infusion or withdrawal using a BD Insyte™ Autoguard™ Shielded IV Catheter. For power injection infusion procedures, the subject device can be accessed with a power injection rated IV catheter to create a power-injectable system.
### Intended Use:
The PowerFlow™ Implantable Apheresis IV Port is intended to be an implanted vascular access device designed to provide long-term, repeated access to the vascular system.
{5}------------------------------------------------
#### Indications for Use of Device:
The Bard PowerFlow™ Implantable Apheresis IV Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for long-term therapeutic apheresis, withdrawal of blood, and infusion of medications, IV fluids, parenteral nutrition solutions, blood and blood products.
The Bard PowerFlow™ Implantable Apheresis IV Port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 mL/s.
## Technological Comparison to Predicate Devices:
The PowerFlow™ Implantable Apheresis IV Port is identical to the predicate device, the PowerFlow™ Implantable Apheresis IV Port (clearance to market via K163001 on April 17, 2017) in the following ways:
- Same intended use .
- Same indications for use .
- Same fundamental scientific technology .
- . Same operating principle and method of action
- Same packaqing configuration .
- . Same sterility assurance level and method of sterilization
- . Same material type (Ti-6AL-4V)
The following change has been made between the subject device and the predicate device:
- . The port base incorporates an alternative titanium formulation (Ti-6AL-4V Type 1 MIM per ASTM F2885-17) and manufacturing method (Metal Injection Molding). The predicate device used Ti-6AL-4V-ELI Grade 23 per ASTM F136 and machining as the manufacturing method.
### Performance Tests:
To demonstrate substantial equivalence of the subject device to the predicate device, its technological characteristics and performance criteria were evaluated. Using FDA Guidance Documents on non-clinical testing of medical devices and internal risk assessment procedures, the following verification tests were performed on the subject device:
- Port Subassembly Radiopacity .
- Corrosion Resistance .
{6}------------------------------------------------
- . Silicone Boot Retention
- . Stem Tensile Strength
- Stem-Catheter Connection Air Leak .
- . Valve Life
The results demonstrate that the subject device met all pre-determined acceptance criteria and that it performs equivalently to the predicate device.
Cytotoxicity was conducted on the subject device, and chemical characterization testing was conducted on the predicate and subject device to confirm toxicological equivalence. Additionally, testing conducted on a component made of the same MIM titanium material was leveraged. This testing included Chemical Characterization, Cytotoxicity, Sensitization, Intracutaneous irritation, Acute systemic toxicity, Genotoxicity (Ames), Genotoxicity (Chromosomal Aberrations), Implantation - 13 week intramuscular, and Implantation - 13 week bone. This information demonstrates that the subject device is biocompatible and does not elicit any substances at levels of concern as a result of this change.
The subject device is sterilized using ethylene oxide and was adopted into the same validated sterilization cycle as the predicate device. Endotoxin (LAL) and EO residual testing will be routinely performed.
## Summary of Substantial Equivalence:
The subject device met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. Therefore, the subject device is considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.