HeraBEAT™ US is intended to detect fetal heart beats, display fetal heart rate, and play the fetal heart sound. HeraBEAT™ US is indicated for use by medical professionals in clinical or home care settings for singleton pregnancies from 12 weeks gestation.
Device Story
HeraBEAT™ US is a handheld, battery-powered audio Doppler device; uses 2MHz ultrasound transducer to detect fetal heart rate (FHR) and sounds. Integrates optical sensor to distinguish FHR from maternal heart rate (MHR) to eliminate crosstalk. Device transmits data via Bluetooth Low Energy (BLE) to a smartphone application. App controls device operation, displays numerical FHR, plays audio, and logs history. Used by medical professionals in clinical or home settings. Provides real-time FHR monitoring; assists clinicians in fetal assessment; enables remote monitoring capabilities.
Clinical Evidence
No clinical trials performed. Bench testing only. Evaluated FHR measurement accuracy per FDA ultrasound guidance, biocompatibility (ISO 10993-5, -10), electrical safety/EMC (IEC 60601-1, -1-11, -2-37, -1-2), photobiological safety (IEC 62471), and RF wireless coexistence. Use-life testing confirmed 3-year operational life.
Indicated for singleton pregnancies from 12 weeks gestation. For use by medical professionals in clinical or home care settings.
Regulatory Classification
Identification
A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.
Aquasonic 100 Ultrasound Transmission Gel (K802146)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 14, 2019
HeraMED Ltd. % Sheila Hemeon-Heyer, JD, RAC President and Founder Heyer Regulatory Solutions LLC 125 Cherry Lane Amherst. MA 91002
Re: K191110
Trade/Device Name: HeraBEAT™US Regulation Number: 21 CFR 884.2660 Regulation Name: Fetal ultrasonic monitor and accessories Regulatory Class: II Product Code: KNG Dated: April 24, 2019 Received: April 26, 2019
Dear Sheila Hemeon-Heyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191110
Device Name
HeraBEAT™ US
Indications for Use (Describe)
HeraBEATIM US is intended to detect fetal heart beats, display fetal heart rate, and play the fetal heart sound. HeraBEAT™ US is indicated for use by medical professionals in clinical or home care settings for singleton pregnancies from 12 weeks gestation.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Page 1 of 4
# 510(k) SUMMARY K191110
| Device Manufacturer: | HeraMED Ltd.<br>6 Meir Ariel Street<br>Netanya, 4059300, Israel | | |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|--|
| 510(k) Contact: | Sheila Hemeon-Heyer<br>President and Founder<br>Heyer Regulatory Solutions LLC<br>413-330-8578<br>Sheila@Heyer-Regulatory.com | | |
| Date Prepared: | November 13, 2019 | | |
| Name of Device: | HeraBEAT™ US | | |
| Common or Usual Name: | Fetal Ultrasonic Monitor and Accessories | | |
| Classification: | Product Code:<br>Regulation No:<br>Regulatory Class:<br>Classification Panel: | KNG<br>21 C.F.R. §884.2660<br>II<br>Obstetrics/Gynecology | |
| Predicate Device: | MD800, cleared under K112911 | | |
This predicate device has not been subject to a design related recall.
### Intended Use / Indications for Use
HeraBEAT™ US is intended to detect fetal heart beats, display fetal heart rate, and play the fetal heart sound. HeraBEAT™ US is indicated for use by medical professionals in clinical or home care settings for singleton pregnancies from 12 weeks gestation.
### Device Description
HeraBEAT™ US is a hand-held, battery powered audio Doppler device integrated with 2MHz probe, used for detecting and displaying the fetal heart rate (FHR) and FHR sound. The device uses an optical sensor to distinguish between the FHR and the maternal heart rate (MHR) to eliminate "crosstalk" in the FHR display.
HeraBEAT™ US includes the following components:
- The handheld HeraBEAT™ US device that incorporates an ultrasound transducer, . rechargeable battery, and internal microcontroller and Bluetooth Low Energy (BLE) chip for wireless data transfer from the HeraBEAT™ US device to the user's smartphone.
- The HeraBEAT™ application, which is downloaded from an app store to the user's .
{4}------------------------------------------------
smartphone. The HeraBEAT™ application is used to communicate with the HeraBEAT™ US device using wireless BLE. It controls the operation of the device and receives the FHR values for numerical display to the user. It plays the FHR sound and stores the FHR values in a history log.
HeraBEAT™US is supplied with a battery charger (power supply), two tubes of commercially available Aquasonic 100 Ultrasound Transmission Gel (cleared under K802146), carry case and User Manual.
| Device & Predicate Device(s): | K191110 | K112911 |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| General Device Characteristics | | |
| Device Name | HeraBEAT™ US | MD800 |
| Indications for Use | HeraBEAT™ US is intended<br>to detect fetal heart beats,<br>display fetal heart rate, and<br>play the fetal heart sound.<br>HeraBEAT™US is indicated<br>for use by medical<br>professionals in clinical or<br>home care settings for<br>singleton pregnancies from<br>12 weeks gestation. | The Fetal Doppler is a<br>hand-held, battery<br>powered audio Doppler<br>device used for<br>detecting fetal heart<br>beats. |
| Intended patient population | Pregnant women | Pregnant women |
| Intended device operator | Healthcare professionals | Healthcare<br>professionals |
| Gestational age | 12 weeks | 12 weeks |
| Environment of use | Professional and home<br>healthcare environment | Professional and home<br>healthcare environment |
| FHR measurement method | Doppler ultrasound | Doppler ultrasound |
| Device Features | | |
| Battery/Power Supply | Lithium ion rechargeable<br>battery supplied with charger | 2 AA alkaline batteries,<br>assumed replaceable |
| Display | Smart phone application | LCD |
| Bluetooth | Yes | - |
| Technical Specifications | | |
| Working mode | Continuous wave | Continuous wave |
| Ultrasound nominal frequency | 2 MHz | 2 MHz |
| FHR measuring range | 50-240 bpm | 50-210 bpm |
| Resolution | 1bpm | 1 bpm |
| Accuracy | +/- 2bpm | +/- 2bpm |
### Substantial Equivalence Discussion
The HeraBEAT™US device has the same intended use and similar technological characteristics and principles of operation as the MD800 fetal ultrasonic monitor. Different technological features include the use of Bluetooth technology, different battery/power supply,
{5}------------------------------------------------
different ingress protection, and different display modalities. The technological differences between the HeraBEAT™US device and the predicate device do not raise different questions of safety and effectiveness.
## Non-clinical Performance Data
### Bench Performance Testing
Clinical measurement accuracy was conducted per the acoustic output measurement methodology as recommended in the FDA guidance document "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers," dated September 9, 2008.
### Biocompatibility
HeraBEAT™ US is a surface contact device with a short-term (<24 hours) contact duration with intact skin. Per the recommendations of the 2016 FDA quidance document "Biological Evaluation of Medical Devices - Part 1: Evaluation and testing Within A Risk Management Process," the following tests were performed:
- Cytotoxicity Test per ISO 10995-5:2009 .
- Intracutaneous Irritation Test per ISO 10993-10:2010 ●
- Guinea Pig Maximization Sensitization Test per ISO 10993-10:2010 ●
# Software Validation
The HeraBEAT™ US software meets the requirements of IEC 62304:2015-06 (ed. 1.1) Medical device software - Software life cycle processes. Software documentation, including software verification and validation testing is in accordance with FDA's 2005 guidance document. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.".
### Electrical Safety and EMC Testing
The HeraBEAT™ US device was tested to and found to be in compliance with the following safety tests:
- . IEC 60601-1:2005/A1:2012 (ed. 3.1) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-11:2015 (ed. 2.0) Medical electrical equipment Part 1-11: General . requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-2-37:2015 (ed. 2.1) Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical
{6}------------------------------------------------
diagnostic and monitoring equipment
- . IEC 60601-1-2:2014 (ed. 4.0) Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 62471:2006-07 (ed. 1.0) Photobiological safety of lamps and lamp systems. The ● device was confirmed to meet the EXEMPT category of light emissions.
### Use-life testing:
Devices underwent simulated repeated use conditions and were evaluated for key performance and FHR measurement accuracy. The results demonstrated that the subject device can be used for three years under normal use conditions.
### RF Wireless Transmission
The HeraBEAT™ US Bluetooth communication between the device and mobile app was subjected to wireless co-existence and quality of service (QoS) testing in accordance with FDA's guidance document, "Radio Frequency Wireless Technology in Medical Devices," August 13, 2013, and was found to be compliant for the device's intended use and indications for use.
# Conclusion
The results of the performance testing described above demonstrate that the HeraBEAT™ US is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.