K191105 · Golden Motor Technology Co., Ltd. · ITI · Jan 21, 2022 · Physical Medicine
Device Facts
Record ID
K191105
Device Name
e-Throne Folding Wheelchair, Model: ET-12F22
Applicant
Golden Motor Technology Co., Ltd.
Product Code
ITI · Physical Medicine
Decision Date
Jan 21, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Indications for Use
The e-Throne Folding Wheelchair, Model: ET-12F22 is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Device Story
Powered folding wheelchair; provides mobility for disabled/elderly users. Input: user commands via joystick controller mounted on armrest. Operation: joystick signals processed by electric motor controller to drive two rear hub motors; electromagnetic brake system manages deceleration/stops. Output: motorized movement of wheelchair. Used in indoor/outdoor environments; operated by patient. Benefits: provides independent mobility; foldable frame allows transport in vehicle trunks. Healthcare provider uses device to facilitate patient mobility; output (movement) directly controlled by patient.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2, ISO 7176-21), biocompatibility (ISO 10993), and performance testing per ISO 7176 series (stability, braking, energy consumption, obstacle climbing, fatigue strength).
Technological Characteristics
Aluminum alloy frame; PU armrests; nylon/PU foam cushions. Powered by 24V 15Ah Li-ion battery; brushless dual-drive motors. Joystick-based control system. Foldable design. Connectivity: off-board charger (100-240V AC). Standards: ISO 7176 series (stability, braking, dimensions, strength, power/control systems), IEC 60601-1, IEC 62133-2 (battery).
Indications for Use
Indicated for disabled or elderly persons with mobility limitations requiring a seated position for indoor and outdoor transportation.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
January 21, 2022
Golden Motor Technology Co., Ltd. % Raymond Luo Regulation Manager Shanghai Sungo Management Consulting Company Limited 13F, Eastern Mansion, No.1500 Century Ave. Pudong New Area, Shanghai, 200120, China
Re: K191105
Trade/Device Name: e-Throne Folding Wheelchair, Model: ET-12F22 Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: December 17, 2021 Received: December 27, 2021
Dear Raymond Luo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191105
Device Name
e-Throne Folding Wheelchair, Model: ET-12F22
Indications for Use (Describe)
The e-Throne Folding Wheelchair, Model : ET-12F22 is a motor driven, indoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K191105
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
# 1. Date of the summarv prepared: 2022-01-21
# 2. Submitter's Information
Company Name: Golden Motor Technology Co., Ltd.
Address: Room 1401-1405, Tower 5, No.18-50 Middle Chang Wu Road, Changzhou 213164 Jiangsu, China
Phone: +86 519 81004118
Contact Person: Philip Yao
E-mail: sales@goldenmotor.com
Application Correspondent: Shanghai Sungo Management Consulting Company Limited
Address:13F, Eastern Mansion, No.1500 Century Ave., Pudong New Area, Shanghai
Contact Person: Raymond Luo Tile: Regulation Manager Tel:0086-21-68828050 Email: fda.sungo@@gmail.com
# 3. Subject Device Information
Type of 510(k) submission: Traditional Common Name: Powered wheelchair Trade Name: e-Throne Folding Wheelchair, Model: ET-12F22 Classification Name: wheelchair, powered Review Panel: Physical Medicine Product Code: ITI Regulation Number:890.3860 Regulation Class: 2
| Sponsor | SUZHOU KD Medical Appliance Co. Ltd. |
|-------------------|--------------------------------------|
| Device Name | PL001 power wheelchair |
| 510(k) Number | K113463 |
| Product Code | ITI |
| Regulation Number | 890.3860 |
| Regulation Class | 2 |
# 4. Predicate Device Information
{4}------------------------------------------------
## 5. Device Description
The e-Throne Folding Wheelchair, Model: ET-12F22 is a powered wheelchair that is designed to fold when not in use so it can be more easily transported in a car trunk or similar space. It uses rechargeable lithium batteries and is controlled by a joystick and a few other simple controls which are located on the end of the armrest. The joystick assembly may be installed on either side of the chair to accommodate both right-handed and left-handed users.
The e-Throne Folding Wheelchair consists of two foldable armrests, a back rest, a seat cushion, a foldable frame, two rear drive wheels with hub motor and electromagnetic brake assemblies, two pivoting casters, a single Li-ion battery pack, an off-board battery charger, a control panel (joystick) with connect cables and an electric motor controller. The device is powered by one Li-ion battery pack (24V, 15 Ah), which can be recharged by an off-board battery charger that can be plugged into an AC socket outlet (100-240V, 50/60Hz) when the device is not in use.
## 6. Intended Use / Indications for Use
The e-Throne Folding Wheelchair, Model: ET-12F22 is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
{5}------------------------------------------------
| Attribute | Subject device | Predicate device | Discussion/<br>Conclusion |
|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Golden Motor Technology Co., Ltd. | SUZHOU KD Medical Appliance Co. Ltd. | / |
| Proprietary name, model | e-Throne Folding Wheelchair, Model:<br>ET-12F22 | PL001 power wheelchair | / |
| 510(k) number | K191105 | K113463 | / |
| Device classification name | Class II | Class II | Same |
| Classification regulations | 21 CFR 890.3860 | 21 CFR 890.3860 | Same |
| Product code | ITI | ITI | Same |
| Similarities | | | |
| Indication for use | The e-Throne Folding Wheelchair, Model :<br>ET-12F22 is a motor driven, indoor and outdoor<br>transportation vehicle with the intended use to<br>provide mobility to a disabled or elderly person<br>limited to a seated position. | The device is a motor driven, indoor and<br>outdoor transportation vehicle with the<br>intended use to provide mobility to a<br>disabled or elderly person limited to a seated<br>position. | Same |
| Intended user | disabled people with mobility difficulties and<br>elderly people | disabled or elderly person limited to a seated<br>position | Same |
| Use condition | indoor and outdoor use | indoor and outdoor use | Same |
| Number of wheels | 4, including two front wheels and two rear<br>wheels | 4, including two pivoting casters and two<br>rear drive wheels | Same |
| Function of wheels | Front wheels: driven wheels suitable for<br>rotation, acceleration, retrograde<br>Rear wheels: driving wheels to control the speed | two pivoting casters: driven wheels suitable<br>for rotation, acceleration, retrograde<br>two rear drive wheels: driving wheels to | Same |
| Attribute | Subject device | Predicate device | Discussion/<br>Conclusion |
| | and direction | control the speed and direction | |
| Frame design | the frame of the wheelchair is type of aluminum<br>frame with front and rear folding structure;<br>Up-and-down turnable handrail, front and rear<br>adjustable armrest. | the frame of the wheelchair is type of<br>aluminum frame with front and rear folding<br>structure; Up-and-down turnable handrail,<br>front and rear adjustable armrest. | Same |
| Folding mechanism | The wheelchair can be folded to close the back<br>and seat parts face to face after push the two red<br>lock buttons to the central position.<br>fold the back and the armrest is rotated to the<br>down direction, and then close the back and seat<br>face to face, close the feet pedal as well. The<br>folding is completed. | the wheelchair can be folded to close the<br>back and seat parts face to face after push<br>the two lock buttons to the central position.<br>fold the back and the armrest is rotated to<br>the down direction, and then close the back<br>and seat face to face, close the feet pedal as<br>well. The folding is completed | Same |
| Max speed forward | 6 km/h, continuously adjustable | not publicly available | / |
| Maximum safe operational<br>incline degree | 9 ° | not publicly available | / |
| Movement control method | By Joystick control | By Joystick control | Same |
| Driving system | Direct drive on the rear wheels | not publicly available | / |
| Brake system | Automatic intelligent electromagnetic brake<br>system | not publicly available | / |
| Electronic controller | Brushless dual-drive rocker controller | not publicly available | / |
| Main frame material | Aluminum alloy | not publicly available | / |
| Armrest | PU | not publicly available | / |
| Seat cushion/back cushion | Nylon (Lining: PU foam) | not publicly available | / |
| Attribute | | | |
| Subject device | Predicate device | Discussion/Conclusion | |
| Differences | | | |
| Front wheel size/type | 8" x 2"/PU Solid tire | not publicly available | / |
| Rear wheel size/type | 12.5"x 2.4"/PU Solid tire | not publicly available | / |
| Max Speed backward | 0.8m/s (2.88km/h) | not publicly available | / |
| Max loading weight | 100 kg (220 lbs) | not publicly available | / |
| Battery charger | Off-board charger<br>Input: 100-240V, 50/60Hz,<br>Output: DC 24, 2A; | Off-board, Automatic Type<br>Input: 110-220 V / 50-60 Hz,<br>Output: 24 Vdc, 2A; | More wide range of input voltage in the<br>device which will not cause new safety<br>and effectiveness concerns raised. |
| Motor | brushless motor; 24VDC; 250W; 2pcs | not publicly available | / |
| Minimum braking distance<br>from maximum speed | Forward: 0.75 m | not publicly available | / |
| Battery | Li-ion; rechargeable, 24 VDC 15Ah | Li-ion, Rechargeable; 12 VDC 10Ah | same battery type, different power will<br>not affect the effectiveness and safety<br>of the device as both batteries are in<br>conformity with IEC 62133-2. |
| Maximum distance of travel<br>on the fully charged battery | 22 km | 20 km | longer distance of travel on the subject<br>device |
| Turning Radius | 650 mm | not publicly available | / |
| Maximum obstacle climbing | 50 mm | not publicly available | / |
# 7. Comparison of technological characteristics with the predicate device
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
#### 8. Summary of substantial equivalence discussion
The e-Throne Folding Wheelchair, Model: ET-12F22 complied with the requirements of ISO 7176-1:2014, ISO 7176-2:2001, ISO 7176-3:2012, ISO 7176-4:2008, ISO 7176-5:2008, ISO 7176-6:2001, ISO 7176-7:1998, ISO 7176-8:2014, ISO 7176-9:2009, ISO 7176-10:2008, ISO 7176-11:2008, ISO 7176-13:1989, ISO 7176-14:2008, ISO 7176-15:1996, ISO 7176-16:2012, ISO 7176-21:2009, ISO 7176-22:2014, ISO 7176-25:2013, IEC 60601-1:2005+A1:2012, IEC 60601-1-2: 2014, ISO 10993-1:2018, ISO10993-5:2009, ISO 10993-10:2010.
The intended uses for both devices are the same. Mainframes of two devices are folded by the same way as well. The design principles of the controller and driving system are the same, and both meet the requirements of the ISO 7176-14:2008. Software validation is carried out on both control systems. Brake system and speed control are designed in the same way as well, and both meet the requirements of the ISO 7176-3:2012. Maximum obstacle climbing are different from each other while such differences will not impact the safety and effectiveness of the subject device or raise new safety and effectiveness concerns as well as both meet the requirements of the ISO 7176-2:2001. The biocompatibility of the predicate device and subject device meet the requirements of the ISO 10993-5:2009 & ISO 10993-10:2010.
The flame retardant test of the seat cushion/back cushion and armrest of both subject device and predicate device is carried out according to the ISO 7176-16 test. Therefore, both devices are assured to be under the same safety level.
In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
#### 9. Summary of non-clinical testing (Performance testing-bench)
The device has been evaluated the safety and performance by lab bench testing or other non-clinical way as follows:
#### Risk Management
Risk analysis and management was implemented in accordance with ISO 14971:2007.
#### Biocompatibility
The biocompatibility evaluation for the e-Throne Folding Wheelchair, Model: ET-12F22 was carried out in accordance with 1S0 10993-1: 2018, including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2010) and irritation (ISO 10993-10:2010).
page 6 of 8
{9}------------------------------------------------
### Electrical safety and electromagnetic compatibility (EMC)
Electrical Safety Testing was conducted on the e-Throne Folding Wheelchair, Model: ET-12F22 in accordance with IEC 60601-1:2005 (3rd Edition).
Electromagnetic Compatibility Testing was conducted on the e-Throne Folding Wheelchair, Model: ET-12F22 in accordance with IEC 60601-1-2:2014 and ISO 7176-21:2009.
Li-ion battery pack used on the e-Throne Folding Wheelchair, Model: ET-12F22 was conducted in accordance with IEC 62133-2:2017.
# Performance Testing
Performance Testing was conducted on the e-Throne Folding Wheelchair, Model: ET-12F22 in accordance with the following standards:
- ISO 7176-1: 2014 Wheelchairs Part 1: Determination of static stability .
- . ISO 7176-2: 2017 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs
- ISO 7176-3: 2012 Wheelchairs Part 3: Determination of effectiveness of brakes ●
- ISO 7176-4: 2008 Wheelchairs Part 4: Energy consumption of electric wheelchairs ● and scooters for determination of theoretical distance range
- ISO 7176-5: 2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space
- ISO 7176-6: 2018 Wheelchairs Part 6: Determination of maximum speed, . acceleration and deceleration of electric wheelchairs
- . ISO 7176-7: 1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
- . ISO 7176-8: 2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strength
- ISO 7176-9: 2009 Wheelchairs Part 9: Climatic tests for electric wheelchairs .
- . ISO 7176-10: 2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs
- ISO 7176-11: 2012 Wheelchairs -- Part 11: Test dummies ●
- ISO 7176-13: 1989 Wheelchairs Part 13: Determination of coefficient of friction of ● test surfaces.
- ISO 7176-14: 2008 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods
- . ISO 7176-15: 1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling.
- . ISO 7176-16: 2012 Wheelchairs -- Part 16: Resistance to ignition of postural support devices
- ISO 7176-25: 2013 Wheelchairs Batteries and chargers for powered wheelchairs ● page 7 of 8
{10}------------------------------------------------
## Software Verification and Validation Testing
Software Verification Testing was performed to evaluate the functionality of the design, materials, and operational principles of the subject device. This includes the following: Software verification testing was conducted on the subject device as recommended by the FDA's guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
Level of Concern: The software for this device was considered as a moderate hazard.
## Animal Study
Animal testing was not required for this submission.
# Clinical Testing
Clinical testing was not required to demonstrate substantial equivalence to the predicate device.
## 10. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, mode of operation, and intended use of e-Throne Folding Wheelchair, Model: ET-12F22 is substantially equivalent to the predicate devices quoted above. The differences between the subject device and its predicate devices do not introduce a new intended use and do not raise new issues of safety and effectiveness. Verification and Validation testing demonstrated that no adverse effects have been introduced by these differences and that the device performs as intended. From the results of nonclinical testing described, it can be concluded that the subject device e-Throne Folding Wheelchair, Model: ET-12F22 is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.