The Edwards Pressure Cable is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. The Edwards Pressure Cable is intended for use with compatible monitors, and Edwards pressure monitoring sensors and compatible catheters.
Device Story
Reusable cable connecting Edwards pressure monitoring sensors to compatible monitors; functions as sub-system module. Contains integrated electronics, software, and algorithms to process sensor signals; calculates hemodynamic parameters including CO/CI, CPO/CPI, CVP, blood pressures (DIA, MAP, MPAP, SYS), SV/SVI, SVR/SVRI, and SVV. Used in hospital critical care settings by clinicians. Output displayed on compatible monitors to assist in hemodynamic assessment and goal-directed therapy decision-making; benefits patient through continuous monitoring of cardiovascular status.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 60601-1-6, IEC 62366), alarm systems (IEC 60601-1-8), and software verification (IEC 62304).
Technological Characteristics
Reusable cable module; integrates electronics, software, and algorithms for hemodynamic parameter calculation. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 62304, and IEC 62366. Software classified as Moderate Level of Concern.
Indications for Use
Indicated for critical care patients requiring continuous assessment of cardiac function, fluid status, vascular resistance, and pressure. May be used for hemodynamic monitoring in conjunction with perioperative goal-directed therapy protocols in hospital environments.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Edwards LifeSciences, LLC Anne Lo Specialist, Regulatory Affairs One Edwards Way Irvine, California 92614
June 13, 2019
Re: K191089
Trade/Device Name: Edwards Pressure Cable Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DOK Dated: April 22, 2019 Received: April 24, 2019
Dear Anne Lo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K191089
Device Name Edwards Pressure Cable
#### Indications for Use (Describe)
The Edwards Pressure Cable when used with an Edwards' pressure monitoring sensor and connected to a compatible monitor, offers continuous assessment of hemodynamic parameters. The Edwards Pressure Cable is indicated for use in critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SECTION 5 – 510(k) SUMMARY
| Edwards Pressure Cable | |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter | Edwards Lifesciences, LLC |
| Contact Person | Anne Lo |
| Date Prepared | June 6, 2019 |
| Trade Name | Edwards Pressure Cable |
| Common Name | Pressure Cable |
| Classification<br>Name | Programmable diagnostic computer |
| Regulation Class /<br>Product Code | 21 CFR 870.1425/Class II/DQK |
| Predicate Device | HemoSphere Pressure Cable cleared in the HemoSphere Advanced Platform<br>510(k) (K180881, cleared November 16, 2018) |
| Device Description | The Edwards Pressure Cable is a reusable device that connects with a compatible<br>monitor on one end, and an Edwards' pressure monitoring sensor on the other end<br>to continuously measure hemodynamic parameters.<br>The Edwards Pressure Cable acts as a sub-system module to a compatible monitor<br>and has all the electronics, software and algorithms built into the cable to allow for<br>monitoring and calculating patient hemodynamic parameters.<br>Parameters available on the Edwards Pressure Cable include:<br>Continuous cardiac output/continuous cardiac index (CO/CI)<br>Cardiac power output/cardiac power index (CPO/CPI)<br>Central venous pressure (CVP)<br>Diastolic blood pressure (DIA)<br>Mean arterial blood pressure (MAP)<br>Mean pulmonary artery blood pressure (MPAP)<br>Stroke volume/stroke volume index (SV/SVI)<br>Systemic vascular resistance/systemic vascular resistance index (SVR/SVRI)<br>Stroke volume variation (SVV)<br>Systolic blood pressure (SYS) |
| Indications for<br>Use/Intended Use | The Edwards Pressure Cable when used with an Edwards' pressure monitoring<br>sensor and connected to a compatible monitor, offers continuous assessment of<br>hemodynamic parameters. The Edwards Pressure Cable is indicated for use in<br>critical care patients in which the balance between cardiac function, fluid status,<br>vascular resistance and pressure needs continuous assessment. It may be used for<br>monitoring hemodynamic parameters in conjunction with a perioperative goal<br>directed therapy protocol in a hospital environment. |
{4}------------------------------------------------
| Differences Include: | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • Revision to the intended use and indications for use statement for the<br>Edwards Pressure Cable to allow compatibility with third party monitors.<br>These revisions do not change the inherent intended use and indications for use. | | |
| • Revision to the labeling to remove the specific reference to the<br>HemoSphere Advanced Monitoring System | | |
| Table 1: Device Comparison of the Predicate HemoSphere Pressure Cable with Subject<br>Edwards Pressure Cable | | |
| Features | Predicate Device | Subject Device |
| Indications for Use | The HemoSphere advanced monitor when used<br>with the HemoSphere Pressure Cable is<br>indicated for use in critical care patients in<br>which the balance between cardiac function,<br>fluid status, vascular resistance and pressure<br>needs continuous assessment. It may be used<br>for monitoring hemodynamic parameters in<br>conjunction with a perioperative goal directed<br>therapy protocol in a hospital environment.<br>Refer to the Edwards FloTrac sensor, Acumen<br>IQ sensor, FloTrac IQ sensor, and TruWave<br>DPT indications for use statements for<br>information on target patient population specific<br>to the sensor/transducer being used. | The Edwards Pressure Cable when used<br>with an Edwards' pressure monitoring<br>sensor and connected to a compatible<br>monitor, offers continuous assessment of<br>hemodynamic parameters. The Edwards<br>Pressure Cable is indicated for use in<br>critical care patients in which the balance<br>between cardiac function, fluid status,<br>vascular resistance and pressure needs<br>continuous assessment. It may be used for<br>monitoring hemodynamic parameters in<br>conjunction with a perioperative goal<br>directed therapy protocol in a hospital<br>environment. |
| Intended Use | The HemoSphere advanced monitoring<br>platform is intended to be used by qualified<br>personnel or trained clinicians in a critical<br>care environment in a hospital setting. The<br>HemoSphere advanced monitoring platform<br>is intended for use with compatible<br>Edwards oximetry and Swan-Ganz catheters<br>and pressure monitoring sensors and<br>catheters.<br><br>A comprehensive list of parameters available<br>while monitoring with the HemoSphere<br>advanced monitor and a connected HemoSphere<br>Pressure Cable or connected HemoSphere<br>Pressure Cable and oximetry cable are listed in<br>table 2 and 3. | The Edwards Pressure Cable is intended<br>to be used by qualified personnel or<br>trained clinicians in a critical care<br>environment in a hospital setting. The<br>Edwards Pressure Cable is intended for<br>use with compatible monitors, and<br>Edwards pressure monitoring sensors<br>and compatible catheters. |
| No new testing was performed since the subject and predicate devices are identical<br>and have similar intended use and indications for use. The Edwards Pressure<br>Cable leverages the testing for the predicate HemoSphere Pressure Cable<br>(HemoSphere Advanced Platform, K180881, cleared November 16, 2018). The<br>testing was performed in compliance with the following standards: | | |
| • IEC 60601-1:2005+A1:2012 Medical Electrical Equipment, Part 1:<br>General Requirements for Basic Safety and Essential Performance | | |
| • IEC 60601-1-2 Medical Electrical Equipment, Part 1-2: General<br>Requirements for Safety- Collateral Standard: Electromagnetic<br>Compatibility- Requirements and Tests | | |
| Comparison to<br>Predicate Device<br><br>Performance Data | | |
| | IEC 60601-1-6 Medical Electrical Equipment, Part 1-6: General<br>Requirements for Safety and Essential Performance- Collateral Standard:<br>Usability IEC 60601-1-8 Medical Electrical Equipment, Part 1-8: General<br>Requirements for Basic Safety and Essential Performance- Collateral<br>Standard: General Requirements, Tests, and Guidance for Alarm Systems<br>in Medical Electrical Equipment and Medical Electrical Systems IEC 62304 Medical Device Software- Life Cycle Processes IEC 62366 Medical Devices- Application Usability Engineering to Medical<br>Devices Software Verification:<br>The Edwards Pressure Cable is considered software with a Moderate Level of<br>Concern. Software verification was performed per FDA's Guidance for Industry<br>and FDA Staff, "Guidance for the Content of Premarket Submissions for Software<br>Contained in Medical Devices”. The testing requirements for the predicate<br>HemoSphere Pressure Cable was tested and passed at a sub-system level to ensure<br>the safety of the device. This testing is applicable to the subject Edwards Pressure<br>Cable.<br>All testing for Electrical Safety and Electromagnetic Compatibility (EMC) and<br>Software Verification testing being leveraged for the subject Edwards Pressure<br>Cable passed.<br>All performance testing leveraged for the subject Edwards Pressure Cable was<br>previously submitted in the predicate 510(k) that included the HemoSphere<br>Pressure Cable (HemoSphere Advanced Monitoring Platform), (K180881, cleared<br>November 16, 2018). | |
| Conclusion | The subject Edwards Pressure Cable is identical to the predicate HemoSphere<br>Pressure Cable and has similar intended use and indications for use. Therefore, it<br>is substantially equivalent. | |
{5}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.